Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K142206
    Device Name
    Dental Unit
    Date Cleared
    2015-03-13

    (214 days)

    Product Code
    Regulation Number
    872.6640
    Panel
    Dental
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Zhuhai Siger Medical Equipment Co., Ltd

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Dental Unit is intended to supply power to and serve as a base for dental devices; and accessories. It is intended for use in the dental clinic /office environment and used by trained dentists and/or dental technicians and assistants. This product is attached with a dental chair.

    Device Description

    The proposed devices Dental Unit are well equipped dental unit, which are intended to supply power to and serve as a base for dental devices and accessories. They are intended for use in the dental clinic/office environment and used by trained dentists and/or dental technicians and assistants. The products are attached with dental chair.

    The proposed devices include two models. U300 and U500. Both of the two models mainly consist of instruments, instrument arm tray, cabinet group, operation light arm, assistant position, pedal switch, and patient chair.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical device called "Dental Unit." It is a premarket notification to the FDA to demonstrate substantial equivalence to a legally marketed predicate device. This type of document typically focuses on comparing the new device to existing ones based on technical specifications and safety standards, rather than clinical performance or diagnostic accuracy studies.

    Therefore, the information regarding acceptance criteria and a study to prove the device meets these criteria is limited to non-clinical tests and compliance with safety and performance standards. There is no mention of studies involving human subjects, AI assistance, expert ground truth, or MRMC comparative effectiveness studies.

    Here's the information extracted from the document:

    1. A table of acceptance criteria and the reported device performance:

    The document presents a comparison table between the proposed device (Dental Unit U300, U500) and the predicate device (Dental Unit with Chair S2310) across several technical characteristics and standards. The "acceptance criteria" can be inferred as compliance with the listed standards and achieving similar technical specifications to the predicate device.

    ItemAcceptance Criteria (Predicate)Reported Device Performance (Proposed)
    Product CodeEIAEIA
    Regulation No.21 CFR 872.664021 CFR 872.6640
    Class11
    Intended UseSupply power to and serve as a base for dental devices; and accessories. Intended for use in dental clinic/office environment by trained dentists/technicians/assistants. Attached with a dental chair.Same as predicate.
    Operating LightHalogenLED
    Connection JointComply with ISO9168Comply with ISO9168
    Water HeatingYesNo
    Operation MethodControl Panel / Assistant Control Panel / Foot ControllerControl Panel / Assistant Control Panel / Foot Controller
    Power Supply110V110V
    Frequency50/60Hz50/60Hz
    Power (with dental chair)900VA400VA
    Pressure of Water Supply200 kPa ~ 400 kPa0.2MPa-0.4MPa (200 kPa - 400 kPa)
    Pressure of Air Supply≥550 kPa0.55MPa (550 kPa)
    Dental Chair Loading Capacity135Kg200kg
    Dental Chair Movement Range (Chair)420-820 MM390mm-740mm
    Dental Chair Movement Range (Backrest)0°~80°1°~70°
    Dental Chair Movement Range (Headrest)200MM150mm
    Accessories can be attached to the deviceHandpiece / Scaler / Curing Light / SyringeHandpiece / Scaler / Curing Light / Three-way-Syringe
    Performance StandardsComply with ISO7494-1, ISO7494-2 and ISO6875Comply with ISO7494-1, ISO7494-2 and ISO6875
    Rate of Water Suction (Suction)≥ 1L/min1 L/min
    Rate of Water Suction (Silva Ejector)> 750mL/min750ml/min
    Electrical SafetyComply with IEC 60601-1Comply with IEC 60601-1
    EMCComply with IEC 60601-1-2Comply with IEC 60601-1-2
    Patient Contact MaterialPU, PVC, Polyamide, ABS, Stainless steelArmrest: PU, Leather for patient chair: PVC, Syringe: Stainless steel, Tubes: TPU

    2. Sample size used for the test set and the data provenance:

    The document describes non-clinical tests conducted to verify that the proposed device met design specifications and was substantially equivalent. It lists several international standards to which the device was tested for compliance (e.g., IEC 60601-1, IEC 60601-1-2, ISO7494-1, ISO7494-2, ISO 6875, ISO 10993-5, ISO 10993-10).

    • Sample size: Not applicable in the context of clinical test sets as no clinical study is mentioned. The tests were likely performed on a limited number of device units (e.g., prototypes or production samples).
    • Data provenance: Not explicitly stated as it's not a clinical data set. The tests were conducted by the manufacturer or accredited testing laboratories as part of verifying compliance with standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    Not applicable. The "ground truth" for the non-clinical tests is established by the specified requirements of the international safety and performance standards themselves. There is no mention of experts establishing ground truth in a clinical sense.

    4. Adjudication method for the test set:

    Not applicable. This concept applies to clinical studies where different expert opinions on a case need to be resolved. For non-clinical device testing, the outcome is typically a pass/fail against a specific standard or specification.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No, an MRMC comparative effectiveness study was not done. The device is a "Dental Unit," which is a physical piece of equipment for dental procedures, not an AI-powered diagnostic or assistive tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    No, a standalone performance study in the context of an algorithm or AI was not done. The device is a traditional dental unit.

    7. The type of ground truth used:

    For the non-clinical tests, the "ground truth" is defined by the technical specifications outlined in the international standards (e.g., IEC 60601-1 for electrical safety, ISO7494-1 for general dental unit requirements, ISO 10993 for biocompatibility). The device either complies with these standards or it does not.

    8. The sample size for the training set:

    Not applicable. This device is not an AI/ML algorithm that requires a training set.

    9. How the ground truth for the training set was established:

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1