Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medical Diode Laser Hair Removal Device (PZ-606VI, PZ-BDT68-01, PZ-BDT68-02, PZ-BDT68-03, PZ-BDT68-04) is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin.

    Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

    Device Description

    The proposed device, Medical Diode Laser Hair Removal Device (PZ-606VI、PZ-BDT68-01、PZ-BDT68-02、PZ-BDT68-03、PZ-BDT68-04), is a surgical device, which is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI);

    The device incorporates a diode laser that emits invisible infrared laser radiation centered on a wavelength of 808 nm. According to the selective light absorption theory, the laser can be preferentially absorbed by the melanin in the hair follicle, thus achieving the purpose of permanent hair removal. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. Hair growth must remain stable for a certain amount of time and must exceed the entire growth cycle of the hair follicle. Permanent hair removal does not mean that all hair in the treated area is completely lost. Special cooling technology is applied simultaneously during treatment to cool the skin and protect skin.

    There are 5 models included, PZ-606VI, PZ-BDT68-01, PZ-BDT68-02, PZ-BDT68-03 and PZ-BDT68-04, the five models have the same intended use, mechanism of action and principle, only the minimum values of energy density and pulse duration are different.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and synopsis are for a Medical Diode Laser Hair Removal Device. This document is a regulatory submission for a physical medical device, not an AI/software as a medical device (SaMD). Therefore, the information requested in your prompt regarding acceptance criteria and studies for AI/SaMD (e.g., sample size for test/training sets, data provenance, expert adjudication, MRMC studies, standalone performance, ground truth establishment) is not applicable to this device submission.

    The document describes the device, its intended use, a comparison to a predicate device, and the non-clinical tests performed to demonstrate substantial equivalence for a hardware device. It explicitly states: "No clinical study is included in this submission."

    Therefore, I cannot provide a table of AI acceptance criteria or details about AI-specific studies as requested, because this information is not present in the provided document and is not relevant to the type of device being cleared.

    However, I can extract the acceptance criteria and performance information that is relevant to this type of physical device, based on the non-clinical testing mentioned:

    Acceptance Criteria and Reported Device Performance (Non-Clinical for a Physical Device)

    Acceptance Criteria CategorySpecific Test/StandardAcceptance Criteria (Implied by Compliance)Reported Device Performance
    Electrical SafetyIEC 60601-1:2005+A1:2012+A2:2020Medical electrical equipment - Part 1: General requirements for basic safety and essential performance met."Complies with IEC 60601-1, IEC 60601-2-22"
    Electromagnetic Compatibility (EMC)IEC 60601-1-2:2014+A1:2020Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic - Requirements and tests met."Complies with IEC 60601-1-2"
    Laser SafetyIEC 60601-2-22:2019, IEC 60825-1:2014Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment met; Equipment classification and requirements for laser safety met."Complies with IEC 60601-2-22, IEC 60825"
    BiocompatibilityISO 10993-5:2009 (Cytotoxicity)No cytotoxic effects."No Cytotoxicity"
    ISO 10993-10:2021 (Skin Sensitization)No evidence of sensitization."No evidence of sensitization"
    ISO 10993-23:2021 (Irritation)No evidence of irritation."No evidence of irritation"
    Performance AccuracyN/A (General Performance Testing)Spot size accuracy within specifications."Performance Testing for Spot Size Accuracy...met"
    N/A (General Performance Testing)Energy output accuracy within specifications."Performance Testing for...Energy Output Accuracy...met"
    Mechanical PerformanceIEC 60601-1 (Implied)Mechanical performance accepted despite dimension/weight differences."the mechanical performance of the proposed device is determined to be accepted"

    Regarding the AI/SaMD specific questions from your prompt:

    1. Sample size used for the test set and the data provenance: Not applicable. No AI/SaMD test set described.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. No AI/SaMD ground truth described.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable.
    7. The sample size for the training set: Not applicable.
    8. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K250782
    Date Cleared
    2025-06-12

    (90 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    • CO2 LASER Part:
      Fractional mode is indicated only for ablative skin resurfacing.

    Non-fractional mode (Impulse mode) is indicated for incision, excision, ablation, vaporization and coagulation of body soft tissues including intraoral tissues, in medical specialties including aesthetic (dermatology and plastic surgery), otorhinolaryngology (ENT), gynecology, neurosurgery, dental and oral surgery and genitourinary surgery.

    Device Description

    The Fractional CO2 Laser Machine uses a CO2 laser with the emission wavelength of 10600nm to emit focal spots through a high focus lens. The target is water in skin tissue. A certain amount of thermal stripping, more thermal coagulation and obvious thermal effect are produced at the part directly penetrated by the focal spot.

    The Fractional CO2 Laser Machine has two operational modes, fractional mode and impulse mode.

    • Fractional mode includes three modes: normal, random and midsplit;
    • Impulse mode includes three modes: single, continuous and impulse.

    The Fractional CO2 Laser Machine has 5 modules: Control display panel module, Main control power module, Light guide arm module, Fan cooling module, Laser module.

    AI/ML Overview

    The provided FDA 510(k) clearance letter describes a "Fractional CO2 Laser Machine" (K250782) and its substantial equivalence to a predicate device, the "Ilooda Fraxis CO2 Laser" (K172096).

    However, the document primarily focuses on demonstrating substantial equivalence through non-clinical performance and safety comparisons rather than detailing specific acceptance criteria for a study that proves the device meets those criteria, particularly in the context of device performance outcomes for intended use. The clearance is based on the device being as safe and effective as its predicate through these comparisons and compliance with relevant standards.

    Therefore, many of the requested sections regarding the study that proves acceptance criteria (especially clinical study details, ground truth, expert involvement, and reader studies) cannot be extracted from this document as such studies are explicitly stated to be not included.

    Here's an analysis of what can and cannot be extracted:

    Acceptance Criteria and Device Performance (Based on Substantial Equivalence and Standards Compliance)

    The acceptance criteria are implicitly met by demonstrating substantial equivalence to the predicate device and compliance with recognized safety and performance standards. The "reported device performance" is presented through direct comparison with the predicate.

    Table of Acceptance Criteria and Reported Device Performance

    Criteria Category / ParameterAcceptance Criteria (Implicitly based on Predicate & Standards)Reported Device Performance (Proposed Device)Comparison / Remark
    General
    Device NameEquivalent to predicate (Ilooda Fraxis CO2 Laser)Fractional CO2 Laser MachineEquivalent
    Classification Regulation21 CFR 878.4810 (SAME as Predicate)21 CFR 878.4810SAME
    Classification PanelGeneral & Plastic Surgery (SAME as Predicate)General & Plastic SurgerySAME
    ClassII (SAME as Predicate)IISAME
    Product CodeGEX, ONG (SAME as Predicate)GEX, ONGSAME
    Common NamePowered Laser Surgical Instrument, Powered Laser Surgical Instrument With Microbeam\Fractional Output (SAME as Predicate)Powered Laser Surgical Instrument, Powered Laser Surgical Instrument With Microbeam\Fractional OutputSAME
    Indications for UseAblative skin resurfacing (fractional mode); incision, excision, ablation, vaporization, coagulation of body soft tissues (non-fractional mode) (SAME as Predicate)Same as predicate (listed in document)SAME
    Prescription UsePrescription use (SAME as Predicate)Prescription useSAME
    Performance (Technical Specifications)
    Laser TypeCO2 (SAME as Predicate)CO2SAME
    Laser Wavelength10.6μm (SAME as Predicate)10.6μmSAME
    Output power30W (SAME as Predicate)30WSAME
    Pulse Duration20-5,000μs (Predicate range)100-5,000μsSAME (within range)
    Fractional Pulse energyMax 150mJ (SAME as Predicate)Max 150mJSAME
    Repetition rate1,000Hz (SAME as Predicate)1,000HzSAME
    Scan area20x20mm (SAME as Predicate)20x20mmSAME
    Spot size100-200μm (Fractional), Max 1.3mm (Non-fractional) (SAME as Predicate)100-200μm (Fractional), Max 1.3mm (Non-fractional)SAME
    Number of microbeams per surface area (fractional)Max 289 spot/cm2 (SAME as Predicate)Max 289 spot/cm2SAME
    Energy per microbeam (fractional)150mJ (SAME as Predicate)150mJSAME
    Total power per surfaced area (fractional)Max 30W (SAME as Predicate)Max 30WSAME
    Treatment Time10-15 min (SAME as Predicate)10-15 minSAME
    Pulse rate (non-fractional)1Hz – 1,000Hz (SAME as Predicate)1Hz – 1,000HzSAME
    Pulse width (non-fractional)20μs –5000μs (Predicate range)100μs –5000μsSAME (within range)
    Operational modeFractional mode, Normal mode (CW, Pulse, Single Pulse) (Predicate)Fractional mode, Impulse mode (single/continuous/impulse)SAME
    Aiming beamDiode laser (Red), Max 4mW (SAME as Predicate)Diode laser (Red), Max 4mWSAME
    CoolingAir cooling (SAME as Predicate)Air coolingSAME
    User InterfaceLCD touch screen (SAME as Predicate)LCD touch screenSAME
    Optical guideArticulated arm (SAME as Predicate)Articulate systemSAME
    Electrical Requirements100-240VAC, 50-60 Hz, 6.3 A (Predicate)100-120V~, 50-60HzAnalysis
    Energy flux per μ beams(4.810^5~1.910^6) mJ/cm2 (SAME as Predicate)(4.810^5~1.910^6) mJ/cm2SAME
    Power flux per μbeam(9.510^7~3.810^8)mW/cm2 (SAME as Predicate)(9.510^7~3.810^8)mW/cm2SAME
    Inter-beam spacingDetermined by micro beam diameter & density of 289 spot/cm2 (Predicate)0.6mm-2.6mm (Derived)Analysis
    Safety
    Electrical SafetyComply with IEC 60601-1 (SAME as Predicate)Comply with IEC 60601-1SAME
    EMCComply with IEC 60601-1-2 (SAME as Predicate)Comply with IEC 60601-1-2SAME
    Laser SafetyComply with IEC 60601-2-22, IEC 60825-1 (SAME as Predicate)Comply with IEC 60601-2-22, IEC 60825-1SAME
    Biocompatibility: CytotoxicityNo Cytotoxicity (SAME as Predicate & ISO 10993-5)No CytotoxicitySAME
    Biocompatibility: IrritationNo evidence of irritation (SAME as Predicate & ISO 10993-23)No evidence of irritationSAME
    Biocompatibility: SensitizationNo evidence of sensitization (SAME as Predicate & ISO 10993-10)No evidence of sensitizationSAME

    Study Information (Based on the provided document)

    As per Section 10, "No clinical study is included in this submission," and Section 9 "Non Clinical Test Conclusion," the device relies on non-clinical testing and substantial equivalence to a predicate. Therefore, the following sections cannot be fully addressed with the provided text for a clinical study proving acceptance criteria.

    1. Sample size used for the test set and the data provenance:

      • Test Set Sample Size: Not applicable (N/A) for clinical outcomes given no clinical study was performed. For non-clinical tests, specific sample sizes (e.g., number of units tested for electrical safety, number of samples for biocompatibility) are not detailed in this summary, but are typically defined by the referenced standards.
      • Data Provenance: N/A for clinical data. For non-clinical compliance testing, the provenance is from manufacturer-conducted tests or accredited labs demonstrating adherence to international standards (e.g., IEC, ISO).
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • N/A. No clinical study for which ground truth would be established by experts is described.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • N/A. No clinical study with adjudicated outcomes is described.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • N/A. Such a study would be relevant for devices involving image interpretation or AI assistance, which this laser device is not described as having.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • N/A. This is not an AI/algorithm-driven diagnostic device.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • N/A. For non-clinical tests, the "ground truth" is defined by the passing criteria of the referenced international standards (e.g., specific thresholds for electromagnetic compatibility, absence of cytotoxicity).
    7. The sample size for the training set:

      • N/A. No algorithm or AI component requiring a training set is mentioned.
    8. How the ground truth for the training set was established:

      • N/A. No algorithm or AI component requiring a training set is mentioned.
    Ask a Question

    Ask a specific question about this device

    K Number
    K250038
    Date Cleared
    2025-05-19

    (131 days)

    Product Code
    Regulation Number
    890.5850
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Muscle Stimulator Device is indicated to be used for:

    • Improvement of abdominal tone, strengthening of the abdominal muscles, development of firmer abdomen.
    • Strengthening, Toning and Firming of buttocks, thighs and calves.
    • Improvement of muscle tone and firmness, for strengthening muscles in arms.
    Device Description

    The Muscle Stimulator Device (model PZ-100) is a non-invasive therapeutic device. The device produces electromagnetic field that interacts with the tissues of the human body. The device has four outputs, and two applicators perform simultaneous treatment.

    The Muscle Stimulator Device is equipped with a large color touch screen with wide view angle that significantly facilitates the use of the device. The on-screen information guides the user step-by-step through the entire therapy procedure. The therapeutic parameters are easily set using the touch screen of the device. During the therapy the device keeps information about the applied therapy type, remaining therapy time and main therapy parameters on the screen.

    The Muscle Stimulator Device is composed of control display panel module, main control module, temperature module, treatment handle module, and fan cooling module.

    The proposed device can simultaneously apply multiple treatment handles to patients, and the handles are held in place by bandages.

    AI/ML Overview

    This document is a 510(k) clearance letter for the Muscle Stimulator Device (PZ-100). It focuses heavily on demonstrating substantial equivalence to a predicate device through non-clinical testing and comparison of specifications.

    Crucially, there is ABSOLUTELY NO MENTION of acceptance criteria related to a study proving performance, sample sizes, expert involvement, ground truth, or MRMC studies.

    The document explicitly states:

    • "No clinical study is included in this submission." This means there was no study to prove the device met acceptance criteria in a clinical setting with human subjects.
    • The clearance is based on "non clinical tests" to verify that the proposed device met design specifications and was substantially equivalent to the predicate device. These typically involve engineering tests, electrical safety, EMC, and biocompatibility, as listed in the "Non-Clinical Test Conclusion" section.

    Therefore, I cannot provide the requested information regarding acceptance criteria and study details for proving the device meets those criteria because this information is not present in the provided document.

    The document describes regulatory compliance and equivalency based on technical specifications and non-clinical testing, not on a clinical performance study with human subjects, which would typically involve the criteria you asked about (sample size, expert ground truth, MRMC, etc.).

    In summary, based on the provided text, the device was NOT proven to meet acceptance criteria through a clinical performance study. Its clearance is based on demonstrating substantial equivalence to a predicate device through non-clinical testing.

    Ask a Question

    Ask a specific question about this device

    K Number
    K250041
    Date Cleared
    2025-05-14

    (126 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medical Picosecond ND:YAG Laser System is intended for use in surgical and aesthetic application in the medical dermatology and general and plastic surgery as follows:

    1064nm wavelength:

    • Removal of tattoos on all skin type (Fitzpatrick skin types I-VI) with the following tattoo colors: black, brown, green, blue and purple.
    • Treatment of benign pigmented lesions on Fitzpatrick skin types I-IV.

    532nm wavelength:

    • Removal of tattoos on Fitzpatrick skin types I-III with the following tattoo colors: red, yellow and orange.
    • Treatment of benign pigmented lesions on Fitzpatrick skin types I-IV.
    Device Description

    The Medical Picosecond ND: YAG Laser System mainly consists of a host, a light guide arm, a treatment handle, a foot switch, a power cord, and other auxiliary treatment accessories. The host mainly includes laser power supply, laser, water circulation cooling system, control board, and display screen. Accessories include a foot switch, laser protective glasses, power cords, and a water filling funnel.

    The Medical Picosecond ND: YAG Laser System is a multi-wavelength, pulsed laser system designed for the treatment of benign pigmented lesions. A key feature of the device is its ability to produce multiple laser wavelengths (1064 nm and 532 nm).

    AI/ML Overview

    I am sorry, but the provided FDA 510(k) clearance letter for the Medical Picosecond ND: YAG Laser System (PZ-6) does not contain the information requested regarding acceptance criteria and a study proving the device meets those criteria.

    Specifically, the document states:

    • "No clinical study is included in this submission." (Page 11, Section 8.0 Clinical Test Conclusion)
    • The "Non clinical tests were conducted to verify that the subject devices met all design specifications as were Substantially Equivalent (SE) to the predicate device." (Page 11, Section 7.0 Non-Clinical Test Conclusion) These non-clinical tests relate to compliance with various IEC and ISO standards for medical electrical equipment, laser safety, and biocompatibility, as well as software validation. These are not performance acceptance criteria for the intended use (tattoo and lesion removal) in the way you are asking.

    Therefore, I cannot provide the detailed information requested in your prompt as it is not present in the provided text.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1