Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K092809
    Device Name
    HRV ACQUIRE
    Date Cleared
    2010-10-19

    (403 days)

    Product Code
    Regulation Number
    870.2300
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    WR MEDICAL ELECTRONICS CO.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Heart Rate Variability (HRV) Acquire device is intended to record and indicate the following parameters during autonomic testing maneuvers:

    • Expiratory pressure recording and display
    • Respiratory effort
    • Breathing cue metronome
    • Heart rate via electrocardiography (ECG)
    • Non-invasive beat-to-beat blood pressure from optional external device

    The HRV Acquire does not make a diagnosis or indicate by itself any disease state exists. The HRV Acquire is not designed for vital signs monitoring or self monitoring of patients.

    Device Description

    The HRV Acquire (HRV) Device with TestWorks software can be used as a standalone cardiovagal device to support heart rate variability testing and expiratory pressure evaluation during valsalva breathing testing. The HRV device supports measurement of heart rate variability, respiratory effort from commercially available circumferential chest bellows, display of a valsalva breathing metronome breathing cue and display of non-invasive beat to beat blood pressure from an external device. An example of an external blood pressure device is the commercially available FMS Finometer, cleared to market by 510(k) K880572. Other commercially available, FDA cleared to market external beat to beat blood pressure devices with an analog blood pressure signal output may be applied.

    The HRV device user interface is integrated into the existing WR Medical Electronics CO TestWorks software. The Test Works software was cleared to market with the Q-Sweat device, 510(K) K992874. The Test Works software is modified to support an interface to the HRV device.

    The HRV device may be mounted to a standard IV Pole.

    The HRV device has two modes of operation:

      1. Valsalva Mode supporting valsalva breathing cue to the subject under test,
      1. Heart Rate Deep Breathing (HRDB) Mode.

    In both modes, the HRV device acquires and the TestWorks software displays heart rate and blood pressure. The HRV device eliminates the need for a separate ECG acquisition device when these types of studies are performed. The HRV device acquires a Lead II ECG signal.

    The HRV device and TestWorks software do not supply any alarms. All interpretation of acquired and displayed data is the responsibility of the physician/clinician supervising the test.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study information for the WR Medical Electronics CO. Heart Rate Variability Device (HRV Acquire), based on the provided 510(k) Summary K092809.

    This document describes a medical device rather than an AI/ML device. Therefore, many of the typical AI/ML study components (like expert ground truth for a test set, MRMC studies, standalone algorithm performance, and training set details) are not applicable or detailed in this submission. The "study" described here focuses on demonstrating the device's functionality, safety, and equivalence to predicate devices, which is standard for traditional medical device clearances.


    Acceptance Criteria and Reported Device Performance

    The acceptance criteria for this device are largely based on the technical specifications and comparison to predicate devices, rather than clinical performance metrics typical of diagnostic algorithms. The "performance" is reported as compliance with these specifications and standards.

    Acceptance Criterion (Feature/Specification)Reported Device Performance
    ECG Specifications
    CMRR90 dB
    Lead SelectionLead II
    Ground Isolation4 kV rms, 5.5 kV peak
    Input Impedance20 MOhm at 10 Hz
    Frequency Response0.2 to 100 Hz
    R-Wave Detection
    Range15 to 300 bpm
    Accuracy± 2%
    Resolution1 bpm
    Sensitivity300 μV peak
    Tall T Wave RejectionRejects T waves
    Ask a Question

    Ask a specific question about this device

    K Number
    K063560
    Date Cleared
    2007-01-05

    (38 days)

    Product Code
    Regulation Number
    874.1820
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    WR MEDICAL ELECTRONICS CO.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Hilger Dual-Stim Nerve Stimulator provides electrical stimulation to nerves during diagnostic and surgical procedures.

    Device Description

    The Hilger Dual-Stim Nerve Stimulator is a self contained, battery powered device. There is an option to provide a line powered option in the future. Two modes of nerve stimulation are provided, a Clinical Mode and a Surgical Mode.

    AI/ML Overview

    The provided document describes the Hilger Dual-Stim Nerve Stimulator, a device used for electrical nerve stimulation during diagnostic and surgical procedures. The 510(k) summary focuses on demonstrating substantial equivalence to predicate devices rather than conducting a de novo clinical study with specific acceptance criteria based on performance metrics. Therefore, the requested information regarding acceptance criteria, study details, and expert involvement for ground truth cannot be fully provided in the format typically associated with AI/ML device evaluations.

    However, based on the provided text, the device's "acceptance criteria" are implied through its compliance with safety standards and a design control process, and its equivalence to predicate devices in terms of intended use and technical characteristics.

    Here’s a breakdown of the information that can be extracted, and where limitations exist:

    1. Table of Acceptance Criteria and Reported Device Performance

    Feature/CriterionAcceptance Criteria (Implied)Reported Device Performance
    Safety and EffectivenessDemonstrated through compliance with design control process and safety standards."The safety and effectiveness of the Hilger Dual-Stim Nerve Stimulator was demonstrated by testing in compliance with the Design Control process."
    Intended Use EquivalenceSame as predicate devices."The intended use... of the Hilger Dual-Stim Nerve Stimulator is the same as the predicate devices."
    Technological EquivalenceEquivalent to predicate devices."The technical characteristics of the Hilger Dual-Stim Nerve Stimulator are equivalent to those of the predicate devices."
    Safety Standards ComplianceMeet specified international safety standards (EN/IEC 60601-1, -1-2, -2-10, -2-40).The device "met" or was tested "in compliance with" these standards.
    Software Verification & ValidationSuccessfully completed."Software verification and validation" was performed.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not applicable. This document describes a traditional 510(k) submission for a non-AI/ML device where a "test set" in the context of data evaluation (like for AI performance) is not relevant. The device underwent engineering testing (software V&V, safety standard compliance) rather than performance evaluation against a labeled dataset.
    • Data Provenance: Not applicable in the context of a dataset for AI evaluation. The "data" here refers to test results from design verification and validation activities.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    • Number of Experts & Qualifications: Not applicable. Ground truth, in the context of expert consensus for labeling data, is not mentioned because this is not an AI/ML device submission. The "ground truth" for this device's performance relies on engineering specifications and compliance with safety standards as verified by qualified engineers and testers.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. There is no mention of an adjudication process for a "test set" as understood in AI/ML performance evaluation. The compliance was determined through internal design control processes and testing against engineering requirements and safety standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and Effect Size

    • MRMC Study: No, an MRMC comparative effectiveness study was not done. This type of study is relevant for evaluating human reader performance with and without AI assistance, which is not applicable to this nerve stimulator device.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Standalone Performance: Not applicable. This device is an electrical nerve stimulator; it does not rely on an algorithm to interpret data or make diagnostic decisions in a standalone capacity. Its function is to provide electrical stimulation for human interpretation of nerve response.

    7. The Type of Ground Truth Used

    • Type of Ground Truth: The "ground truth" in this context is based on engineering specifications, international safety standards (e.g., EN/IEC 60601 series), and the established performance of predicate devices. The device's electrical output parameters (current, pulse width, frequency) are measured and compared against these defined specifications and those of the predicates.

    8. The Sample Size for the Training Set

    • Sample Size for Training Set: Not applicable. This is not an AI/ML device that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth for Training Set: Not applicable. This is not an AI/ML device that requires a training set or ground truth for such a set.

    Summary of the Study that Proves the Device Meets Acceptance Criteria:

    The "study" referenced in the document is the internal design control process and associated testing, which included:

    • Software verification and validation: Ensuring the device's software functions as intended and meets specifications.
    • Declaration of safety standard compliance: Demonstrating that the device adheres to relevant international safety standards (EN/IEC 60601-1, EN/IEC 60601-1-2, EN/IEC 60601-2-10, EN/IEC 60601-2-40). This would involve electrical safety testing, EMC testing, and other relevant tests depending on the specific standard.

    The conclusion states that these activities were sufficient to demonstrate the safety and effectiveness of the Hilger Dual-Stim Nerve Stimulator and to show its substantial equivalence to predicate devices, thus raising "No new questions of safety or effectiveness."

    Ask a Question

    Ask a specific question about this device

    K Number
    K000574
    Date Cleared
    2000-08-09

    (169 days)

    Product Code
    Regulation Number
    890.5525
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    WR MEDICAL ELECTRONICS CO.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K992874
    Date Cleared
    1999-11-18

    (84 days)

    Product Code
    Regulation Number
    890.5525
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    WR MEDICAL ELECTRONICS CO.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Q-Sweat™ Quantitative Sweat Measurement System is designed to measure the sweat output of the skin of humans. This device does not make a diagnosis or indicate by itself that any disease state exists; it simply documents sweat output. This device is to be used in scientific studies of the anatomy, physiology, and biochemistry of the skin & associated structures.

    Device Description

    The O-Sweat device is measurement-only device which is designed to measure the rate & volume of sweating by capturing a sample of sweat (water) inside a small measuring chamber which is affixed to the skin. It does not measure any other parameters of the sweat sample. The measurement made is simply a calculation of moisture given off by the skin.

    AI/ML Overview

    Here's an analysis of the provided text regarding the Q-Sweat Quantitative Sweat Measurement System, focusing on acceptance criteria and study details.

    Based on the provided text, there is no specific study described that proves the device meets explicitly stated acceptance criteria. The document is a 510(k) summary, which focuses on demonstrating substantial equivalence to predicate devices rather than providing detailed performance study results against predefined metrics.

    The document primarily states that the device is substantially equivalent to other sweat measuring devices. This is a claim made by the applicant, and the FDA's letter indicates their agreement with this claim for market clearance, not necessarily a detailed performance validation against specific criteria.

    Therefore, many of the requested sections (1, 2, 3, 4, 5, 6, 7, 8, 9) cannot be fully populated as the information is not present in the provided text.

    However, I can extract the relevant available information:


    1. Table of acceptance criteria and the reported device performance

    Acceptance Criteria (Explicitly stated)Reported Device Performance
    Not explicitly stated in terms of quantitative metrics or thresholds.The device operates in "nearly exact same fashion" as predicate devices. The use of the device does not affect the body differently or raise new questions of safety or effectiveness compared to predicate devices.

    2. Sample sized used for the test set and the data provenance

    • Sample Size (Test Set): Not specified in the provided text.
    • Data Provenance: Not specified in the provided text.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • This information is not provided in the text as no specific performance study is detailed with an independent ground truth establishment.

    4. Adjudication method for the test set

    • This information is not provided in the text as no specific performance study is detailed.

    5. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done. If so, what was the effect size of how much human readers improve with AI vs without AI assistance.

    • No information about an MRMC study or AI assistance is present. This device is a measurement-only system and does not appear to involve AI or human interpretation in the context of diagnostic decision-making that would necessitate an MRMC study.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done.

    • The document implies standalone performance comparison to predicate devices, but no dedicated "standalone study" with detailed metrics is described. The device is a direct measurement tool, not an algorithm.

    7. The type of ground truth used

    • This information is not provided in the text as no specific performance study is detailed with an independent ground truth. The basis of equivalence is implied to be through comparison with the operational principles and intended use of existing predicate devices.

    8. The sample size for the training set

    • Not applicable/Not provided. The text does not describe a machine learning model or a training set.

    9. How the ground truth for the training set was established

    • Not applicable/Not provided. The text does not describe a machine learning model or a training set.

    Summary of what is provided:

    The document is a 510(k) summary for the Q-Sweat Quantitative Sweat Measurement System. Its primary goal is to establish substantial equivalence to existing predicate devices.

    • Intended Use: "To measure the sweat output of the skin of humans. This device does not make a diagnosis or indicate by itself that any disease state exists; it simply documents sweat output. This device is to be used in scientific studies of the anatomy, physiology, and biochemistry of the skin & associated structures."
    • Comparison to Predicate Devices: The applicant claims the Q-Sweat operates in "nearly exact same fashion" as predicate devices and does not raise new questions of safety or effectiveness. The FDA's letter confirms substantial equivalence, allowing the device to be marketed.
    • Device Type: A "measurement-only device" designed to measure the rate and volume of sweating. It captures a sample of sweat inside a small measuring chamber affixed to the skin and calculates moisture given off by the skin.

    The absence of detailed performance study data, acceptance criteria, and ground truth establishment methods is typical for 510(k) applications that rely heavily on demonstrating substantial equivalence to already cleared devices, especially for devices where the technology and intended use are well-established. The equivalence claim focuses on the device's operational similarity and lack of new safety/effectiveness concerns, rather than a quantifiable performance comparison against specific metrics.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1