Search Results
Found 4 results
510(k) Data Aggregation
(123 days)
Well Brain International Ltd.
The Gymform® TOTAL ABS (Model: WB-245) is intended to stimulate healthy muscles in order to improve or facilitate muscle performance.
The large belt is intended for use on the muscles in abdomen. The small belt for working the leg and arm muscles, intended for use on the muscles in arms, legs (lower extremities) and thighs areas separately.
The Gymform® TOTAL ABS( Model: WB-245) consists of a control unit, large belt, small belt and electrode pads. The control unit enclosed in ABS plastic case. The device is design based on Microcurrent technology (Electro Muscle Stimulation-EMS) which uses a gentle current to condition, tone and strengthen muscles.
The Gymform® TOTAL ABS( Model: WB-245) is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. The TOTAL ABS may be considered a technique or method for muscle training. It is a battery operated muscle stimulation system specifically designed to training muscles.
The large belt is intended for use on the muscles in abdomen. The small belt for working the leg and arm muscles, intended for use on the muscles in arms, legs (lower extremities) and thighs areas separately.
The stimulator sends gentle electrical current to targeted muscle group through the electrodes placed on the skin. The parameters of the unit are controlled by the buttons. Its intensity level can be adjustable by user.
There are 6 modes for output stimulation, and for each mode, there are 10 levels of output intensity can be choose. Default time is 10minutes for output stimulation.
Power is derived from 3 pieces of AAA batteries located in a compartment protected by a removable battery cover for the Fitness Belt.
The provided document, a 510(k) Summary for the Gymform® TOTAL ABS, Model: WB-245, describes the device's technical specifications and claims substantial equivalence to predicate devices. However, it does not detail specific acceptance criteria for a study proving the device meets those criteria in clinical performance metrics. Instead, the document focuses on demonstrating safety and performance through compliance with recognized standards and a usability study, and by comparing its technical specifications to legally marketed predicate devices.
Here's an analysis of the provided information, structured around your request, with an emphasis on what is available and what is not explicitly stated:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state "acceptance criteria" for clinical performance. Instead, it relies on demonstrating compliance with recognized electrical safety, EMC, and biocompatibility standards, and demonstrating substantial equivalence to predicate devices, which generally implies that if the predicate device is safe and effective, and the new device is sufficiently similar and meets safety standards, it is also safe and effective.
The performance data summarized focuses on technical benchmarks and compliance:
Category | Acceptance Criteria (Implied) | Reported Device Performance |
---|---|---|
Electrical Safety | Compliance with IEC 60601-1, IEC 60601-2-10 | Comply with IEC 60601-1 and IEC 60601-2-10 |
EMC | Compliance with IEC 60601-1-2 | Comply with IEC 60601-1-2 |
Biocompatibility | Compliance with ISO 10993-5, ISO 10993-10 | Non cytotoxic, No skin sensitization, No skin irritation |
Software Integrity | Compliance with FDA Guidance for Software in Medical Devices (Moderate concern) | Software verification and validation testing conducted and documented |
Waveform Parameters | Verified to demonstrate substantial equivalence to predicate devices | Output voltage, output current, Output waveform frequency, pulse width, maximum current and power density verified |
Usability (OTC) | Lay user can: 1) Self-select, 2) Apply safely, 3) Understand indications/contraindications/warnings | Usability testing completed in 24 subjects; tests demonstrate device meets requirements |
Max Output Voltage | Comparable to predicate devices, within safe limits | 81.6V @ 500Ω, 96V @ 2kΩ, 99.2V @ 10kΩ. Lower than primary predicate, covered by range of predicate and reference devices. |
Max Output Current | Comparable to predicate devices, within safe limits | 163mA @ 500Ω, 48mA @ 2kΩ, 9.92mA @ 10kΩ. Lower than primary predicate, covered by range of predicate and reference devices. |
Frequency Range | Comparable to predicate devices, within safe limits | 50, 60, 70 Hz. Within predicate devices' ranges and IEC 60601-2-10/FDA guidance. |
Pulse Width Range | Comparable to predicate devices, within safe limits | 200 us. Within predicate devices' ranges and IEC 60601-2-10/FDA guidance. |
Max Current Density | Compliance with IEC 60601-2-10 and FDA guidance | 0.041 mA/cm² @ 500Ω. Compliant with standards. |
Max Power Density | Compliance with IEC 60601-2-10 and FDA guidance, below 0.25 W/cm² | 28.81 µW/cm². Compliant with standards and below 0.25 W/cm². |
2. Sample Size for the Test Set and Data Provenance
- Usability Study Test Set: 24 subjects
- Data Provenance: Not specified (e.g., country of origin, retrospective/prospective). The study is described as "Usability testing (OTC Study)," which is typically prospective.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts
Not applicable for a usability study. The "ground truth" for a usability study is typically the direct observation of user interaction and self-reported comprehension, not expert consensus on a medical diagnosis.
4. Adjudication Method for the Test Set
Not applicable. Usability studies do not typically involve adjudication in the same way clinical diagnostic studies do. The assessment is usually based on predefined tasks and criteria for successful completion and comprehension by the "lay user."
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC comparative effectiveness study was mentioned. This device is a muscle stimulator, not an AI-powered diagnostic tool for interpretation by human readers.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done
This device is not an algorithm for diagnosis. Its performance is inherent in its electrical stimulation output and how users interact with it. The stated performance evaluations (electrical, EMC, biocompatibility, waveform, usability) are effectively "standalone" in that they assess the device itself without requiring a human medical expert to interpret its output for diagnostic purposes.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- For Biocompatibility: Laboratory testing results based on established ISO standards.
- For Electrical Safety/EMC/Waveform: Measurement against recognized engineering standards (IEC, FDA guidance).
- For Usability: User performance and comprehension against specified usability objectives for "lay users." There isn't a "ground truth" in the diagnostic sense, but rather a determination of whether users can operate the device safely and effectively according to instructions.
8. The sample size for the training set
Not applicable. This document pertains to the evaluation of a physical medical device (muscle stimulator) rather than an AI/ML algorithm that requires a training set.
9. How the ground truth for the training set was established
Not applicable, as no training set for an AI/ML algorithm was used.
Ask a specific question about this device
(286 days)
WELL BRAIN INTERNATIONAL, LTD.
GYMFORM®ABS& CORE is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. The ABS& CORE may be considered a technique or method for muscle training. 2-area belt is intended for use on the muscles in abdomen or lower back separately. Mini belt is intended for use on the muscles in arms, legs, thighs or buttocks areas separately.
GYMFORM® ABS&CORE is a two channels battery operated muscle stimulation system specifically designed to stimulation the muscles.
The 2-area belt is intended for use on the muscles in abdomen or lower back separately.
Mini belt is intended for use on the muscles in arms, legs, thighs and buttocks areas separately.
It is comprised of a console for signal generation, two belts (2-area belt and mini belt) for fixation, and electrode pads for signal connection to skin. The electrode pads are replaceable.
Power is derived from 2 "AAA" batteries located in a compartment protected by a removable battery cover. There is no current passed from side to side. The user cannot access the wiring or connectors within the belt.
The stimulator sends gentle electrical current to targeted muscle group through the electrodes placed on the skin. The parameters of the device are controlled by the buttons. Its intensity level can be adjustable by user.
The provided text describes the 510(k) submission for the GYMFORM® ABS&CORE, Model: VDPGYCSET0042, a powered muscle stimulator. The purpose of a 510(k) submission is to demonstrate that the new device is substantially equivalent to a legally marketed predicate device, meaning it is as safe and effective. It does not typically involve detailed studies proving novel device performance against specific clinical acceptance criteria in the way a new drug or advanced diagnostic AI might.
Instead, the "acceptance criteria" here are compliance with recognized safety and performance standards, and the "study" is a series of laboratory bench tests comparing the new device's specifications and performance against these standards and the predicate devices.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for the GYMFORM® ABS&CORE are largely based on compliance with international standards for medical electrical equipment and a comparison to predicate devices, demonstrating substantial equivalence in terms of safety and effectiveness.
Acceptance Criterion (Standard/Comparison) | Reported Device Performance |
---|---|
Safety and Effectiveness (General) | Substantially Equivalent to predicate devices (K130074, K111476) |
Electrical Safety (IEC 60601-1 and IEC 60601-2-10 standards) | Complies with IEC 60601-1 and IEC 60601-2-10 |
Electromagnetic Compatibility (EMC) (IEC 60601-1-2 standard) | Complies with IEC 60601-1-2 |
Software Verification and Validation (FDA Guidance for Pre-Market Submissions and for Software Contained in Medical Devices) | Verified and Validated |
Waveform Output Specifications (IEC 60601-2-10 and Guidance for Powered Muscle Stimulator) | Verified specifications within acceptable ranges compared to predicates |
- Maximum Output Voltage (Vpeak-to-peak) | 132V @ 500Ω; 138V @ 2kΩ; 140V @ 10kΩ (vs. 108-126V for predicates) |
- Maximum Current Density (Ipeak-to-peak) | 264mA @ 500Ω; 69mA @ 2kΩ; 14mA @ 10kΩ (vs. 216-12.6mA for predicates) |
- Frequency range | 2 Hz, 10 Hz, 50 Hz, 90 Hz, 120 Hz (vs. 1-120 Hz for predicates) |
- Pulse width range | 108μs / 124μs (vs. 100-120 μs for predicates) |
- Pulse duration | Varies by mode, e.g., 500ms (Mode 1), 11.1ms (Mode 2) (similar to Predicate I) |
- Net Charge | 19.2µC @ 500Ω (vs. 15.7µC for Predicate I) |
- Maximum Phase Charge | 16.4µC @ 500Ω (vs. 13.0µC for Predicate I) |
- Maximum Current Density (mA/cm²) | 0.082 mA/cm² @ 500Ω (vs. 0.057-0.55 mA/cm² for predicates) |
- Maximum Power Density | 94.8 μW/cm² @ 500Ω (vs. 53.8 μW/cm² for Predicate I) |
- ON time / OFF time | 0.5s / 0.5s (vs. 1s / 1s for Predicate I) |
- Duty Cycle | 0.02%~1.28% (vs. 0.02%~1.19% for Predicate I) |
Dispersion and Shelf Life (ASTM F 1980-07, Guidance: Shelf Life of Medical Device) | Tested |
Adhesion (Section 5.4 of AAMI EC 12_2000_(R) 2010 Disposable ECG electrodes) | Tested |
Biocompatibility (ISO10993-5 and ISO10993-10 requirements) | All user contacting materials comply |
Functional Specifications (Power sources, leakage current, channels, modes, intensity, timer, weights, dimensions) | Compared to predicate devices; minor differences noted but compliant with IEC standards and guidance |
2. Sample Size Used for the Test Set and Data Provenance
This document describes a 510(k) submission, which relies primarily on bench testing and a comparison to predicate devices, not human clinical trials or retrospective/prospective medical data sets in the typical sense of AI/diagnostic device studies.
- Test Set Sample Size: Not applicable in the context of human subjects or medical images. The "test set" here refers to the device itself being subjected to various laboratory tests.
- Data Provenance: The data comes from internal laboratory bench testing conducted by the manufacturer (Well Brain International Ltd.) and comparisons against specifications of legally marketed predicate devices. The country of origin for the data is implied to be Hong Kong, China (where the manufacturer is located), though specific test labs are not detailed. It is retrospective in the sense that the new device's characteristics are compared against existing standards and predicate devices' known specifications.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not provided and is generally not relevant for this type of 510(k) submission. "Ground truth" in this context is defined by international safety and performance standards (e.g., IEC standards, ASTM standards, AAMI standards) and the specifications of the predicate devices. Compliance with these established standards and equivalence to the predicates serves as the "ground truth" for regulatory clearance.
4. Adjudication Method for the Test Set
This information is not applicable. Adjudication methods (like 2+1, 3+1) are typically used for establishing ground truth in clinical studies involving expert interpretation of medical data (e.g., radiology reads). For a powered muscle stimulator demonstrating substantial equivalence through bench testing, the "adjudication" is essentially the determination by the manufacturer (and subsequently, the FDA) of whether the device meets the referenced standards and is substantially equivalent to the predicates.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No, an MRMC comparative effectiveness study was not done. This type of study is specifically relevant for AI-powered diagnostic devices that assist human readers (e.g., radiologists interpreting images). The GYMFORM® ABS&CORE is a powered muscle stimulator, and its submission focuses on electrical safety, performance, and functional equivalence, not on improving human reader performance with AI.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
No, a standalone algorithm performance study was not done or is not applicable. The device is hardware-based with integral software/firmware for control, but it's not an "algorithm only" device in the diagnostic AI sense. Its function is direct muscle stimulation, not data analysis or diagnosis.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for this submission is established through:
- Compliance with recognized international standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, FDA Guidance documents (e.g., for software, powered muscle stimulators), ASTM F 1980-07, AAMI EC 12_2000_(R) 2010 electrodes, ISO10993-5, ISO10993-10.
- Specifications and performance of legally marketed predicate devices: Gymform® ABS-A-ROUND (K130074) and Contour Technology Muscle Stimulator (K111476).
The ultimate ground truth is the determination of substantial equivalence to these predicates, which implies equivalent safety and effectiveness.
8. The Sample Size for the Training Set
This information is not applicable in the context of this 510(k) submission. A "training set" refers to data used to train machine learning models. This device is a traditional medical device (powered muscle stimulator) and does not involve AI or machine learning that would require a training set.
9. How the Ground Truth for the Training Set Was Established
As no training set (for AI/ML) is mentioned or relevant to this device, the question of how its ground truth was established is not applicable.
Ask a specific question about this device
(213 days)
WELL BRAIN INTERNATIONAL, LTD.
The GYMFORM® ABS-A-ROUND is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. The ABS-A-ROUND may be considered a technique or method for muscle training. The 3-area belt is intended for use on the muscles in abdomen, left waist and right waist alternately. The Mini belt accessory is intended for use on the muscles in arms, legs (lower extremities), thighs and buttocks areas separately.
The ABS-A-ROUND is a three-channel battery operated muscle stimulation system specifically designed to training muscles. The 3-area belt is intended for use on the muscles in abdomen, left waist and right waist alternately. The stimulation of each channel is active atternately for 2 cycles with its set output intensity level and mode program in below default activation sequence: F channel is active for 180 seconds solely, stop for 10 seconds as idle interval; then L channel is active for 180 seconds solely, stop for 10 seconds as idle interval; then R channel is active for 180 seconds solely, stop for 10 seconds as idle interval; repeat one cycle with above activation sequence. The Mini belt accessory is intended for use on the muscles in arms, legs (lower extremities), thighs and buttocks areas separately. It is comprised of an electronic stimulator module for signal generation, two belts for fixation, and electrodes for signal connection to skin. The electrodes are replaceable. Power is derived from 3 batteries located in a compartment protected by a removable battery cover for the Fitness Belt. There is no current passed from side to side. The stimulator sends gentle electrical current to targeted muscle group through the electrodes placed on the skin. The parameters of the unit are controlled by the buttons. Its intensity level can be adjustable by user.
Here's an analysis of the provided text regarding the GYMFORM® ABS-A-ROUND device, focusing on acceptance criteria and supporting studies.
Important Note: The provided document is a 510(k) summary for a medical device. 510(k) submissions primarily focus on demonstrating "substantial equivalence" to a legally marketed predicate device, rather than proving efficacy through clinical studies that would typically define "acceptance criteria" and "device performance" in the way one might expect for a novel device or a high-risk device. The "performance" discussed here largely refers to compliance with electrical safety and EMC standards, and a comparison of technical specifications to predicate devices. There is no mention of a formal clinical performance study in the provided text that would establish typical diagnostic or therapeutic performance metrics like sensitivity, specificity, accuracy, or effect sizes for human improvement.
Acceptance Criteria and Reported Device Performance
Given the nature of a 510(k) summary, the "acceptance criteria" are primarily related to safety, electrical performance, and substantial equivalence to predicate devices, rather than clinical efficacy. The reported "performance" aligns with these compliance aspects.
Acceptance Criterion (Implied/Explicit) | Reported Device Performance |
---|---|
Electrical Safety | Complies with IEC 60601-1 and IEC 60601-2-10 standards. |
Electromagnetic Compatibility (EMC) | Complies with IEC 60601-1-2 standard. |
Waveform Parameters | Waveform test report provided to verify device parameters. |
Biocompatibility | All user-contacting materials comply with ISO10993-5 and ISO10993-10. |
Substantial Equivalence | Considered substantially equivalent to predicate devices; differences do not raise new safety or effectiveness issues. |
Intended Use | Stimulate healthy muscles to improve/facilitate muscle performance; considered a technique/method for muscle training. |
Functional Specifications | Different from predicate devices in channels, programs, phase variations, output, frequency range, pulse width, power, operating/storage environment, electrode size, and maximum current density, but all comply with relevant IEC standards and FDA guidance, thus not raising safety/effectiveness issues. |
Study Details (Based on available information)
-
Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- No specific test set or clinical study data is mentioned for evaluating muscle performance or clinical outcomes. The "testing" mentioned refers to compliance with electrical safety and EMC standards. These are typically bench tests or simulations, not human subject studies.
- The manufacturer is Well Brain International Ltd., located in Hong Kong SAR, China.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable. There is no mention of a study requiring expert-established ground truth for a test set in the context of clinical performance. The documentation focuses on engineering and regulatory compliance.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable. No clinical test set or adjudication method is described.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This device is a powered muscle stimulator, not an imaging analysis or diagnostic AI device. There is no mention of an MRMC study or AI assistance.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This device is a direct-use muscle stimulator, not an algorithm.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Not applicable in the clinical sense. The "truth" or reference used in this submission relates to established electrical safety standards (IEC 60601-1, IEC 60601-2-10, IEC 60601-1-2) and biocompatibility standards (ISO10993-5, ISO10993-10), as well as FDA guidance for EMS devices. These are engineering and regulatory benchmarks, not clinical ground truth.
-
The sample size for the training set
- Not applicable. This device does not use machine learning or require a training set in the conventional sense.
-
How the ground truth for the training set was established
- Not applicable. As above, no training set or associated ground truth establishment process is mentioned.
Summary of the 510(k) Submission's Approach:
The Well Brain International Ltd. 510(k) submission for the GYMFORM® ABS-A-ROUND is a typical example of demonstrating substantial equivalence for a Class II medical device. The "acceptance criteria" are primarily regulatory and engineering compliance requirements. The "study" mentioned consists of testing the device against recognized electrical safety, EMC, and biocompatibility standards, and then demonstrating that its technical specifications and intended use are similar enough to existing, legally marketed predicate devices such that any differences do not raise new concerns about safety or effectiveness. There is no clinical performance study data presented in this summary to establish efficacy or diagnostic accuracy, which is common for devices of this type seeking a 510(k) clearance based on substantial equivalence.
Ask a specific question about this device
(383 days)
WELL BRAIN INTERNATIONAL, LTD.
The electrical muscle stimulation (EMS) operation mode is indicated for the improvement of abdominal muscle tone, and for the development of a firmer abdomen.
The therapeutic vibrator (Vibration) operation mode is indicated for the promotion of muscle relaxation.
The new device with its proprietary name/model number of Gymform Dual Flex Belt/WB-162 is a single device featured with two modes of operation, independent each other - the electrical muscle stimulation for abdominal muscles conditioning and the therapeutic vibrator for abdominal muscles relaxation.
Upon selected the electrical muscle stimulator (EMS) mode, the flex belt will function as an electrical muscle stimulator capable to achieve toning and firming of muscles in the abdominal region via using transcutaneous electrical muscle stimulation which is a technology for muscle conditioning by sending a stream of electric impulses to the motor nerve of muscle via the pair of single-patient/multiple-application biocompatible cutaneous electrodes on the inner surface of the flex belt, the muscle responds to each electric impulse by producing a contractile twitch, just as it would create voluntary muscle contractions during normal training & strengthening practice, and continues to contract until the electric impulse is over the muscle then returns to its relax state. The level of muscle contraction and relaxation is determined by the electric impulse intensity and the duration applied to the motor nerve of muscle, in which is user selectable via the flex belt control panel.
Upon selected the therapeutic vibrator (Vibration) mode, the Gymform Dual Flex Belt (new device) will function as a therapeutic vibrator to perform muscles relaxation. The therapeutic vibrator (Vibration) mode employs the method of electric muscle massage that simulates natural massage movement of a massage therapist, using oscillating movement of the eccentric motor that installed in the instrument casing to produce linear pulling vibrations to relaxing the abdominal muscle region of a patient.
The provided text is a 510(k) Summary of Safety and Effectiveness for the Gymform Dual Flex Belt. It compares the device to a predicate device, the Slendertone FLEX Abdominal Training System Type 515 (K030708), to establish substantial equivalence.
Based on the document, here's an analysis of the acceptance criteria and study information:
1. Acceptance Criteria and Reported Device Performance
The document does not explicitly state "acceptance criteria" in the typical sense of numerical thresholds for performance metrics. Instead, it relies on demonstrating substantial equivalence to a predicate device. The primary "performance" is compliance with safety standards and showing "same effectiveness" as the predicate device.
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Safety: Compliance with recognized consensus safety standards | The device successfully passed testing for conformity with: |
- IEC 60601-1 (General Requirements for Safety)
- IEC 60601-1-2 (Electromagnetic Compatibility)
- IEC 60601-2-10 (Particular requirements for the safety of nerve and muscle stimulators)
- ISO 10993-5 (In Vitro cytotoxicity)
- ISO 10993-10 (Tests for irritation and delayed-type hypersensitivity)
- Complies with FDA safety requirements and IEC 60601-2-10 |
| Effectiveness: Similar to predicate device | "effectiveness test report of Gymform Dual Flex Belt demonstrates that the new device has same effectiveness of predicate device." (Specific metrics for effectiveness are not detailed in this summary.) |
| Technological Characteristics: Similar to predicate device | The EMS mode is "designed, developed and built on the identical technology of electrical muscle stimulation." While specific output parameters differ (e.g., frequency ranges, max intensity levels), the overall mode of action is identical. The document states "verification and validation tests demonstrated the Gymform Dual Flex Belt maintains same safety and effectiveness as that of the cleared predicate device." |
| Intended Use/Indications: Identical to predicate device | The EMS mode of the Gymform Dual Flex Belt is indicated for "improvement of abdominal muscle tone, and for the development of a firmer abdomen," which is stated as "essentially identical to the predicate device." |
2. Sample Size Used for the Test Set and Data Provenance
The document mentions an "effectiveness test report" but does not provide any details regarding the sample size used, the characteristics of the test participants, or the data provenance (e.g., country of origin, retrospective or prospective nature of the study). It only states that the report demonstrates "same effectiveness" as the predicate device.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
The document does not provide information about experts or ground truth establishment for an effectiveness study. The assessment of effectiveness appears to be based on an internal report that concluded similar effectiveness to the predicate device, but the methodology of that assessment is not detailed.
4. Adjudication Method for the Test Set
The document does not mention any adjudication method.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, and Effect Size
No MRMC comparative effectiveness study is mentioned or implied. The device is a direct-to-consumer electrical muscle stimulator, and comparative effectiveness studies of this nature are generally not required for 510(k) clearance in this context. The comparison is primarily against the performance characteristics of the predicate device for substantial equivalence.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done
This is not applicable as the device is a physical muscle stimulator, not an algorithm or AI system. Its performance is inherent in its electrical output and mechanical function.
7. The Type of Ground Truth Used
The document does not specify the type of "ground truth" used for effectiveness testing. Given the nature of a muscle stimulator, effectiveness might be assessed through metrics such as muscle contraction force, muscle circumference changes, or perceived muscle tone/firmness, but these details are not provided. The primary ground for clearance is "comparison of the electric muscles stimulation (EMS) mode" against a predicate.
8. The Sample Size for the Training Set
This question is not applicable as the Gymform Dual Flex Belt is a physical medical device, not a machine learning model. There is no concept of a "training set" in the context of this 510(k) submission.
9. How the Ground Truth for the Training Set Was Established
This question is not applicable for the same reason as point 8.
Ask a specific question about this device
Page 1 of 1