Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K193197
    Device Name
    QuickClear Mechanical Thrombectomy System
    Manufacturer
    Volcano AtheroMed Inc.
    Date Cleared
    2020-04-20

    (152 days)

    Product Code
    QEZ, DXE
    Regulation Number
    870.5150
    Why did this record match?
    Applicant Name (Manufacturer) :

    Volcano AtheroMed Inc.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
    Intended Use
    The QuickClear Mechanical Thrombectomy System is intended for removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems.
    Device Description
    The QuickClear Mechanical Thrombectomy System is an aspiration thrombectomy system designed for removal of fresh, soft emboli and thrombi from vessels of the peripheral arterial and venous systems. The QuickClear Mechanical Thrombectomy System removes thrombus from the peripheral vasculature using continuous vacuum aspiration, through a single lumen aspiration catheter. The system components are sterile, single-use devices that are sterilized using Ethylene Oxide (EO). The QuickClear Mechanical Thrombectomy System comprises the following devices and accessories: - QuickClear Aspiration Pump - A 60cc Syringe - QuickClear Aspiration Catheters: Catheters are available in the following sizes: 6F straight tip, 8F straight tip, 8F shaped tip and 10F shaped tip. - An Obturator is supplied with the 8F and 10F catheters. - A Waste Collection Bag
    Ask a Question
    K Number
    K182972
    Device Name
    Phoenix Atherectomy System
    Manufacturer
    Volcano AtheroMed Inc.
    Date Cleared
    2018-12-20

    (55 days)

    Product Code
    MCW
    Regulation Number
    870.4875
    Why did this record match?
    Applicant Name (Manufacturer) :

    Volcano AtheroMed Inc.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
    Intended Use
    The Phoenix Atherectomy System is intended for use in atherectomy of the peripheral vasculature. It is not intended for use in coronary, carotid, iliac or renal vasculature.
    Device Description
    The Phoenix Atherectomy System is a sterile, single-use device designed for atherectomy of the peripheral vasculature. The Phoenix Atherectomy System has two main components: the Phoenix Catheter and the Phoenix Handle. The Phoenix Catheter is a flexible, over-the-wire (OTW), front-cutting Catheter that continuously captures and clears debulked plaque proximally through the Catheter and Handle into a collection reservoir that resides outside the patient. For use, the Phoenix Catheter is inserted into the Phoenix Handle. The Handle incorporates a self-contained battery-powered motor designed to drive and rotate the cutter of the Phoenix Atherectomy Catheter at its specified rotational speed. The device is activated by an ON/OFF slider switch on the top of the Handle. A Wire Support Clip is used to hold a guidewire in a fixed position relative to the Handle and prevent guidewire rotation during the procedure. The Catheter, Handle, and Wire Support Clip are packaged as sterile, single-use components of the Phoenix Atherectomy System.
    Ask a Question
    K Number
    K181877
    Device Name
    Phoenix 2.4mm Atherectomy Plus System
    Manufacturer
    Volcano AtheroMed Inc.
    Date Cleared
    2018-12-17

    (157 days)

    Product Code
    MCW
    Regulation Number
    870.4875
    Why did this record match?
    Applicant Name (Manufacturer) :

    Volcano AtheroMed Inc.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
    Intended Use
    The Phoenix Atherectomy System is intended for use in atherectomy of the peripheral vasculature. The system is not intended for use in the coronary, carotid, iliac, pulmonary, or renal vasculature. Phoenix Atherectomy Plus System: When used with the Phoenix Aspiration Pump as the vacuum source, the Phoenix 2.4mm Deflecting Atherectomy System is indicated for the removal of thrombus from vessels of the peripheral arterial vasculature.
    Device Description
    The Phoenix 2.4mm Atherectomy Plus System comprises the currently cleared Phoenix 2.4mm Deflecting Atherectomy System (K172386) and a new Phoenix Aspiration Pump. which when used in conjunction with each other are designed for removal of thrombus from vessels of the peripheral arterial vasculature. The Phoenix 2.4mm Deflecting Atherectomy System and the Phoenix Aspiration Pump are sterile, single-use devices that are sterilized using Ethylene Oxide (EO). The Phoenix 2.4mm Deflecting Atherectomy System comprises the Phoenix 2.4mm Deflecting Atherectomy Catheter and the Phoenix Atherectomy Handle with Wire Support Clip. The Phoenix Catheter is an over-the-wire (OTW), multi-lumen Catheter with a cutter at the distal tip that continuously captures and clears debulked (excised) material proximally through the Catheter and Handle into a collection reservoir that resides outside the patient. For use, the Phoenix Catheter is inserted into the Phoenix Handle incorporates a self-contained battery-powered motor designed to drive and rotate the cutter of the Phoenix Atherectorny Catheter at its specified rotational speed, and is activated by an ON/OFF slider switch on the top of the Handle. A Wire Support Clip is used to hold a guidewire in a fixed position relative to the Phoenix Handle and prevent guidewire rotation during the procedure. The Catheter, Handle, and Wire Support Clip are packaged as sterile, single-use components of the Phoenix Atherectomy System. The Phoenix Catheter is compatible with commercially available 0.014" exchange length (260 cm or greater) guidewires. For the purpose of this 510(k), the design of the Phoenix 2.4mm Deflecting Atherectomy System remains unchanged. The Phoenix Aspiration Pump connects to the disposal outlet of the Phoenix Catheter via the pump's aspiration tubing. The exit end of the aspiration tubing is connected to a waste disposal bag to collect aspirated material. A 60 ml syringe is provided as part of the Aspiration Pump Assembly to assist in priming the pump aspiration tubing and purging the aspiration system of air. The Aspiration Pump is battery-operated and is operated by an ON/OFF switch. It serves as a vacuum source for the Aspiration System for aspirating the thrombus from the target vessel out of the Phoenix Catheter via the pump aspiration tubing and into the disposal bag. The Aspiration Pump is non-patient contacting.