Search Results
Found 8 results
510(k) Data Aggregation
(112 days)
VITALCARE GROUP, INC.
This device is a safety hypodermic syringe for intramuscular and subcutaneous injection of medication to patients. This device aids in prevention of needlestick injuries. EO sterilized, non-pyrogenic, and latex free.
The OMI Retractable Safety Syringe works like this: STEP 1 After cap removal, fluid draw-up and aspiration is unchanged from conventional syringes. STEP 2 As the plunger rod is depressed and the plunger stopper makes contact with the needle hub, the elastic membrane conforms to the conical shape of the needle hub. The plunger stopper membrane stretches until the plunger rod is fully depressed and maximum fluid is expelled from the svringe. STEP 3 At the point of full plunger rod depression, the application of the injection force overcomes the strength of the web joining the inner and outer portions of the needle hub. As this web fractures and the outer portion of the needle hub moves forward, the plunger stopper membrane is simultaneously pierced by the conical end of the hub, releasing the inner portion of the hub carrying the needle to be retracted into the hollow stem of the plunger rod by the coiled spring. The activation force is controlled. The extra pressure required at the end of the injection stroke to retract the needle is minimal. The entire procedure designed for singlehanded use, with no change to common habits. STEP 4 The fracturing of the needle hub and the piercing of the plunger stopper allow the needle to be forced into the plunger rod by the coiled spring. The speed of this retraction and the minimal dead space avoids blood or fluid spatter. Retraction is accompanied by an audible "click" and the highly visible separation of the needle hub. The syringe is EO sterilized, non-pyrogenic, and latex free.
The provided text describes a 510(k) premarket notification for the OMI Ret retractable Safety Syringe. It focuses on demonstrating substantial equivalence to a predicate device, the SEZ Safety Syringe (K031163), rather than presenting a detailed study with specific acceptance criteria and performance metrics in the format of AI device evaluation.
However, I can extract the relevant information and present it in a manner that aligns with your request where possible, while noting limitations due to the nature of the legal filing.
Here's an interpretation based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
Criterion (Type of Test/Characteristic) | Acceptance Criteria (Predicate Device K031163) | Reported Device Performance (OMI Retractable Safety Syringe) |
---|---|---|
Intended Use | Safety hypodermic syringe for intramuscular and subcutaneous injection, aids in prevention of needlestick injuries. EO sterilized, non-pyrogenic, latex free. | Identical: "This device is a safety hypodermic syringe for intramuscular and subcutaneous injection of medication to patients. This device aids in prevention of needlestick injuries. EO sterilized, non-pyrogenic, and latex free." |
Principle of Operation (Safety Feature) | Activation in two steps: Disassemble needle assembly, then retract into barrel and confine by pushing plunger. | Activation in a single step: "After injecting the medicine, the plunger is pushed in just a bit harder causing the retraction mechanism to pull the needle permanently inside the barrel." (This is presented as an improvement/difference but still meeting the safety intent.) |
Volume (ml/cc) | 3 and 5 ml | 1, 3, 5 and 10 ml sizes (Broader range, but implies individual sizes meet performance) |
Nozzle Type | Female conical lock fitting with rotatable internally threaded neck. | Needle and hub are integral to the syringe, not separable. (Structural difference, but functionally equivalent for intended use). |
Barrel Marking | Conforms to ISO7886-1:1993(E) | Identical |
Reuse | Non-reusable | Identical |
Biocompatibility | Conforms to ISO 10993-1 | Identical |
Materials | Plastic parts: polypropylene; Gasket: thermoplastic rubber; Packing film: Medipeel film; Packing paper: Ethypel paper. | Identical |
Sterility | Sterilized by ethylene oxide gas (SAL = 10⁻⁶) | Identical |
Clinical Acceptability | (Implicitly acceptable as a legally marketed device) | "The results of the investigation showed that the OMI Ret retractable Safety Syringe is clinically acceptable." |
Study Proving Device Meets Acceptance Criteria:
The document refers to a "clinical investigation" and "test for the comparison between OMI Ret retractable Safety Syringe and the legally marketed predicate device." This study was performed "in accordance with 'Supplementary Guidance on Premarket Notifications for Medical Devices with Sharps Injury Prevention Features; Guidance for Industry and FDA'."
2. Sample Size for the Test Set and Data Provenance
- The document does not explicitly state the sample size used for the clinical investigation or comparison test.
- The data provenance (country of origin, retrospective/prospective) is not specified. However, the study was conducted to meet FDA regulations for a US-marketed device.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications
- The document does not provide details on the number of experts, their qualifications, or how ground truth was established for the clinical investigation. For a physical device like a syringe, "ground truth" would likely refer to objective measurements of performance (e.g., retraction force, completeness of retraction, prevention of needlesticks in simulated or actual use).
4. Adjudication Method for the Test Set
- The document does not specify any adjudication method.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No, an MRMC comparative effectiveness study was not performed. This type of study is typically relevant for interpretative devices (like AI in radiology) where human readers make a judgment. The OMI Ret retractable Safety Syringe is a physical medical device.
6. Standalone Performance Study
- Yes, a standalone study (referred to as a "clinical investigation" and "test for the comparison") was performed to demonstrate that the OMI Ret retractable Safety Syringe itself ("algorithm only" in AI terms) met the necessary safety and performance criteria, including comparison to the predicate device. The goal was to show "clinical acceptability."
7. Type of Ground Truth Used
- Based on the context of a safety syringe, the "ground truth" would likely involve objective measurements related to:
- Mechanical performance: Retraction force, completeness of needle retraction, prevention of re-exposure.
- Usability/Human factors: Ease of use, activation force, audible/visible confirmation of retraction, potential for needlestick incidents in simulated use.
- Biocompatibility: In vitro and in vivo testing demonstrating material safety.
- Sterility assurance.
- The document states that the findings were based on "biocompatibility testing, clinical testing, compliance with voluntary standards, and comparison to the previous device."
8. Sample Size for the Training Set
- This is not applicable as the OMI Ret retractable Safety Syringe is a physical device, not an AI/software device that would require machine learning training data.
9. How the Ground Truth for the Training Set Was Established
- Not applicable (no training set for a physical device).
Ask a specific question about this device
(134 days)
VITALCARE GROUP, INC.
The Intravascular Administration Set is a single use, sterile device that provides access for the administration of fluids from a container to the patients vascular system through the administration set's needle or catheter, which is inserted into a vein.
The 3 in 1 IV Administration Set is a single use, sterile, device sterilized with Ethylene Oxide gas. The 3 in 1 set is used to administer fluids from an IV container or syringe to a patient's vascular system through a needle or catheter. A choice can be easily made between 10, 15, or 60, drops/cc volume without breaking the line. This feature allows the healthcare provider the flexibility to treat the patient condition by turning the selector. The device may include a flow regulator, a drip chamber, backflow valve, fluid delivery tubing, connectors between parts of the set, needleless injection site, Y port, extension set, and a hollow spike to penetrate and connect the tubing to an IV bag or other infusion fluid container. VitalCare Group Inc. will offer standard sets and custom sets to meet customer requirements and specifications.
The provided text describes a medical device, the "3 In 1" I.V. Administration Set, and its 510(k) summary for clearance. However, it does not detail a study proving the device meets specific acceptance criteria in the way a clinical trial or performance study for an AI/ML device would.
Instead, the submission focuses on demonstrating substantial equivalence to a predicate device (K925465). This is a different type of validation process.
Here's an analysis based on the information provided, addressing your requested points where applicable, and noting where the information is not present for this type of medical device submission:
1. Table of Acceptance Criteria and Reported Device Performance
For this type of device (an IV administration set), the "acceptance criteria" are generally tied to demonstrating that the device performs as intended and is as safe and effective as a legally marketed predicate device. The performance data presented are primarily engineering bench tests to ensure the physical integrity and functionality of the set.
Acceptance Criteria Category (Implied by Submission) | Reported Device Performance (from text) |
---|---|
Intended Use Equivalence | Identical to predicate device (K925465) |
Component Similarity | Similar components and materials to predicate device |
Biocompatibility | Performed (implies met relevant standards, though specific results not detailed) |
Bench Testing - Physical Integrity | Pull testing, pressure testing, joint strength testing, and drop testing performed (implies met internal specifications, though exact criteria/results not detailed) |
Sterilization/Residuals | Performed (Sterile via EO, implies met standards, though specifics not detailed) |
Labeling Similarity | Similar to predicate device |
Packaging Similarity | Similar to predicate device |
Sterilization Method | Identical to predicate device (Sterile via EO) |
Materials Used | Materials are used in legally marketed devices under comparable conditions (implies generally accepted and safe) |
2. Sample size used for the test set and the data provenance
- Sample Size: Not specified. For bench testing of physical properties, specific sample sizes are typically used for each test (e.g., n=30 for tensile strength). This information is not provided in the 510(k) summary.
- Data Provenance: The tests were performed internally by VitalCare Group, Inc., or by a contracted lab. The country of origin of the data is implicitly the USA (where the company is based and the submission was made). The testing is prospective in the sense that it was conducted specifically for this 510(k) submission to demonstrate device performance and safety.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. For a device like an IV administration set, "ground truth" as it relates to expert consensus in image analysis or clinical diagnosis is not relevant. The "ground truth" for performance is established by engineering specifications, material standards, and industry best practices. The "experts" would be the engineers and quality control personnel designing and testing the device, but their specific numbers and qualifications are not detailed in this summary.
4. Adjudication method for the test set
Not applicable. Adjudication methods (like 2+1, 3+1) are used for resolving disagreements among human reviewers (e.g., radiologists, pathologists) when establishing ground truth for diagnostic or AI performance studies. This is not a human-in-the-loop or diagnostic device.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done
No. This type of study is for evaluating human performance, often with and without AI assistance, typically in diagnostic or screening contexts. It is not relevant for an IV administration set.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
No. This device does not have an "algorithm" or AI component. It is a physical medical device.
7. The type of ground truth used
The "ground truth" for this device's performance is based on engineering specifications, material standards, and functional requirements for IV administration sets. This would include:
- Physical specifications (e.g., dimensions, flow rates)
- Material properties (e.g., tensile strength, chemical compatibility)
- Biocompatibility standards (e.g., ISO 10993)
- Sterility assurance (e.g., ISO 11135 for EO sterilization)
- Leakage and connection integrity standards.
The 510(k) summary indicates that "Pull testing, pressure testing, joint strength testing, and drop testing has been performed," which are standard engineering tests to verify these aspects.
8. The sample size for the training set
Not applicable. This device does not use machine learning or AI, so there is no training set.
9. How the ground truth for the training set was established
Not applicable, as there is no training set.
Ask a specific question about this device
(391 days)
VITALCARE GROUP, INC.
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiners hand or finger to prevent contamination between the patient and examiner.
VitalCare Powder Free Synthetic Vinyl Examination Glove. A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare professional and similar personnel to prevent contamination between the healthcare personnel and the patient's body, fluids, waste or environment.
This document details the acceptance criteria for the "VitalCare Powder Free Synthetic Vinyl Examination Glove" and the studies performed to demonstrate its compliance.
1. Table of Acceptance Criteria and Reported Device Performance
Feature/Claim | Acceptance Criteria | Reported Device Performance |
---|---|---|
Physical and Dimensional Testing | Meets all requirements for ASTM Standard D-5250-00 | VitalCare Powder Free Vinyl Examination Glove meets all requirements for ASTM Standard D-5250-00 physical and dimensional testing. |
Powder-Free Claim (Starch Content) | No more than 2mg powder per glove (ASTM D6124-01) | VitalCare Powder Free Vinyl Examination Glove meets ASTM D6124-01 for starch to determine the gloves meet the powder-free claim, no more than 2mg powder per glove. |
Water Leak Test | FDA 1000ml Water Leak Test | VitalCare Powder Free Vinyl Examination Glove meets FDA 1000ml Water Leak Test. |
Primary Skin Irritation | Demonstrates no skin irritation | Primary Skin Irritation tests demonstrate no skin irritation. |
Skin Sensitization | Demonstrates no skin sensitization | Skin Sensitization tests demonstrate no skin sensitization. |
Intended Use | A disposable device intended for medical purposes that is worn on the hand of healthcare professional and similar personnel to prevent contamination between the healthcare personnel and the patient's body, fluids, waste or environment. | The intended use is clearly stated as: "A patient examination glove is a disposable device intended for medical purposes that is worn on the examiners hand or finger to prevent contamination between the patient and examiner." This aligns with the acceptance criteria. |
Materials | Similar to predicate device (Vinyl, Synthetic, Plastic Bag, Corrugated) | The materials used (Vinyl, Synthetic, Plastic Bag, Corrugated) are stated as similar to the predicate device. |
Labeling | Similar to predicate device (Reorder Number, Size, Quantity, Non-Sterile, pouch and shipping case, Manufacturer Address) with the addition of "Powder Free" and absence of "Powdered". | Labeling information provided adheres to the required details, specifically mentioning "VitalCare Powder Free Synthetic Vinyl Examination Glove," and other details similar to the predicate device, excluding "Powdered." |
Packaging | Similar to predicate device (One plastic bag, 100 pouches per case) | Packaging is stated as "One plastic bag, 100 pouches per case," which is similar to the predicate device. |
Sterilization Information | Bulk Non-Sterile | Device is described as "Bulk Non-Sterile." |
Comparative Claims | No comparative claims (e.g., hypoallergenic, comparison to other devices) | "No comparative claims are made for the VitalCare Powder Free Synthetic Vinyl Examination glove. The glove is not claimed as hypoallergenic. It will not be compared in labeling or advertising to other devices." |
Unique Designs | Not unique | "The design of the VitalCare Powder Free Vinyl Examination Glove is not unique." |
2. Sample Size Used for the Test Set and the Data Provenance
The document does not explicitly state the sample sizes used for each specific test (e.g., physical, dimensional, water leak, skin irritation, skin sensitization). It only refers to the tests performed to meet the standards. The data provenance (country of origin of the data, retrospective or prospective) is not specified.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
The document does not mention the number of experts or their qualifications for establishing ground truth for any of the performance tests. The "ground truth" for these tests is defined by the objective metrics and standards themselves (e.g., ASTM D-5250-00, ASTM D-6124-01, FDA 1000ml Water Leak Test, and standard dermatological tests for irritation/sensitization).
4. Adjudication Method for the Test Set
No adjudication method is described. The performance data seems to be based on direct measurement and adherence to established standard specifications rather than subjective expert review requiring adjudication.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done
No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not performed. The device is an examination glove, and its performance is evaluated against physical and chemical standards, not through human reader diagnostic accuracy.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done
Yes, the performance tests described (ASTM standards, FDA water leak test, skin irritation/sensitization) are inherently standalone evaluations of the device's physical and biological properties. They do not involve human-in-the-loop performance in the way an AI diagnostic algorithm would.
7. The Type of Ground Truth Used
The ground truth used for performance evaluation is based on objective standards and test methods:
- ASTM Standard D-5250-00: For physical and dimensional testing.
- ASTM D-6124-01: For starch content to quantitatively determine the "powder-free" claim.
- FDA 1000ml Water Leak Test: For assessing barrier integrity.
- Primary Skin Irritation and Skin Sensitization tests: Standard toxicological/dermatological testing to assess biocompatibility.
8. The Sample Size for the Training Set
This information is not applicable. The device is a physical product (an examination glove), not an AI algorithm requiring a "training set."
9. How the Ground Truth for the Training Set Was Established
This information is not applicable for the same reason as point 8.
Ask a specific question about this device
(83 days)
VITALCARE GROUP, INC.
VitalCare Sterile Water and 0.9% Sodium Chloride. The10cc/30cc syringes are used for balloon inflation. Devices are not intended for wound irrigation, injection or IV Administration.
Sterile Water and 0.9% Sodium Chloride are products that have been used in the medical community for decades, the only ingredient in the two solutions other than water is Sodium Chloride; there are no preservatives or stabilizers. The syringes are manufactured from 100% medical grade polypropylene and contain no color or preservatives or additives.
The containers are filled with distilled water and distilled saline solution, sealed, capped and gamma irradiated for single use only.
This 510(k) summary (K031748) is for the VitalCare Sterile Water and 0.9% Saline 10cc, 30cc pre-filled inflation syringes. The document describes the device, its intended use, and compares it to a predicate device (Orion Life Systems Sterile Water and 0.9% 10cc pre-filled inflation syringes) to establish substantial equivalence.
The provided document does not contain information about acceptance criteria or a study that proves the device meets specific performance criteria in the way typically expected for a diagnostic or AI-driven medical device submission.
Instead, this 510(k) emphasizes substantial equivalence to an already legally marketed predicate device. This means the manufacturer is asserting their device is as safe and effective as the predicate device based on similar technological characteristics and intended use, rather than conducting a new, detailed performance study against specific acceptance criteria.
Therefore, many of the questions regarding acceptance criteria, study design, sample sizes, ground truth, and expert involvement are not applicable to this type of submission.
However, I can extract the relevant information from the provided text based on the 510(k) submission's structure:
1. A table of acceptance criteria and the reported device performance:
This 510(k) submission, being for a pre-filled syringe (a manufacturing and material equivalence claim), does not present specific "acceptance criteria" and "reported device performance" in the way a clinical study would for a diagnostic or AI device. Instead, the "performance" is based on the equivalence of features and contents to the predicate device.
Feature / Claim | Acceptance Criteria (Implied by Predicate Equivalence) | Reported Device Performance (VitalCare Device) |
---|---|---|
Intended Use | Same as predicate device (catheter balloon inflation; not for wound irrigation or IV administration). | "The intended use is the same." "Used for catheter balloon inflation. The devices are not intended to be used for wound irrigation or IV administration." |
Contents | Sterile water (for sterile water syringes) and 0.9% Sodium Chloride (for saline syringes); no preservatives or stabilizers. | "Sterile Water\0.9% Sodium Chloride." "The only ingredient in the two solutions other than water is Sodium Chloride; there are no preservatives or stabilizers." "The containers are filled with distilled water and distilled saline solution." |
Labeling | Equivalent to predicate device regarding contents, volume, intended use (balloon inflation), and prohibitions (not for injection, no antimicrobial/other substances). | "Sterile Water\Saline 10cc\30cc syringes. For Balloon Inflation. Not for injection. No antimicrobial or other substances added." (Stated as "Similar" to predicate) |
Materials | Medical grade materials, specifically polypropylene for syringes and Santoprene rubber. | "Polypropylene, Santoprene rubber." "Manufactured from 100% medical grade polypropylene and contain no color or preservatives or additives." (Stated as "Similar" to predicate) |
Sterilization | Sterilized for single use. | "Sealed, capped and gamma irradiated for single use only." (Implied to be equivalent to predicate in safety and effectiveness) |
Performance Attributes | Equivalent to predicate device. | "The performance attributes are the same." |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
Not applicable. This is a 510(k) submission for substantial equivalence of a manufacturing product (pre-filled syringes), not a clinical trial or performance study involving a patient test set or data originating from a specific country. The "test" for this type of device is primarily a comparison of its physical and chemical properties, manufacturing processes, and intended use against a legally marketed predicate device.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
Not applicable. There was no "ground truth" to establish in a clinical sense for a test set. The validation for this type of device involves scientific and engineering assessment of material properties, sterilization methods, and product design compared to the predicate device, typically performed by the manufacturer's internal quality and R&D teams and reviewed by the FDA.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
Not applicable. No test set requiring expert adjudication was conducted for this 510(k) submission.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This device is a pre-filled syringe, not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not applicable. This device is a pre-filled syringe, not an algorithm or software device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
Not applicable. The concept of "ground truth" as used in diagnostic or AI studies is not relevant here. The "truth" for this submission revolves around demonstrating that the device's physical and chemical attributes, along with its manufacturing process, result in a product that is safe and effective and functions equivalently to the predicate device.
8. The sample size for the training set:
Not applicable. There is no "training set" for this product.
9. How the ground truth for the training set was established:
Not applicable. There is no "ground truth" or "training set" for this product.
Ask a specific question about this device
(28 days)
VITALCARE GROUP, INC.
A patient Vinyl Exam Glove, is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Vinyl, Patient Exam Gloves, Powdered
This document is a 510(k) clearance letter from the FDA for a Vinyl Powdered Examination Glove. It is primarily a regulatory communication and does not contain information about acceptance criteria or a study proving the device meets specific performance criteria.
The 510(k) clearance process focuses on demonstrating substantial equivalence to a predicate device, not necessarily on detailed performance studies with acceptance criteria in the same way a PMA (Premarket Approval) or de novo submission might.
Therefore, I cannot provide the requested information from the given text. The document does not describe:
- A table of acceptance criteria and reported device performance.
- Sample size used for the test set or data provenance.
- Number of experts and their qualifications.
- Adjudication method.
- Multi-reader multi-case (MRMC) comparative effectiveness study.
- Standalone (algorithm-only) performance.
- Type of ground truth used.
- Sample size for the training set.
- How ground truth for the training set was established.
The document states the device is "substantially equivalent" to pre-amendments devices and focuses on regulatory compliance (GMP, labeling, etc.). For examination gloves, the performance criteria typically relate to things like barrier integrity (e.g., freedom from holes), tensile strength, and dimensions, which are usually tested according to industry standards, but these details are not present in this regulatory letter.
Ask a specific question about this device
(168 days)
VITALCARE GROUP, INC.
Wound cover for patient care.
Gauze, Sponges Various Sizes, Non-Sterile and Sterile
The provided text is a 510(k) premarket notification letter from the FDA regarding "Gauze Sponges." This document is a regulatory approval letter and does not contain any information about acceptance criteria, device performance studies, sample sizes, expert qualifications, or ground truth methodologies typically associated with AI/ML device evaluations.
Therefore, I cannot fulfill your request for the specific information regarding a study, as this information is not present in the provided text. The document is simply an FDA clearance for a medical device (gauze sponges) that was deemed substantially equivalent to existing devices, without the need for the types of performance studies you are asking about for AI/ML products.
Ask a specific question about this device
(63 days)
VITALCARE GROUP, INC.
A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina.
Disposable Nonmetal Vaginal Speculum
The provided document is a 510(k) clearance letter from the FDA for a VitalCare Disposable Nonmetal Vaginal Speculum, along with its Indications for Use. This type of document is for medical devices and does not contain the kind of information requested about acceptance criteria and studies for AI/software-as-a-medical-device (SaMD) performance.
Therefore, I cannot extract the requested information from this document. The questions you've asked are specific to the validation of AI algorithms or software, which involves performance metrics like sensitivity, specificity, sample sizes for training/test sets, expert adjudication, and comparative effectiveness studies. This 1998 FDA clearance document for a physical medical device does not provide any of that information.
Ask a specific question about this device
(126 days)
VITALCARE GROUP, INC.
A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
VitalCare Exam Gloves (Latex), Paude
I am sorry, but based on the provided document, I cannot answer the question regarding acceptance criteria and the study proving the device meets them. The document is an FDA 510(k) clearance letter for VitalCare Latex Exam Gloves, Powdered, and it primarily focuses on the regulatory clearance process and the device's substantial equivalence to legally marketed predicate devices.
The letter does not include:
- A table of acceptance criteria and reported device performance.
- Details about a specific study, including sample sizes, data provenance, number or qualifications of experts, adjudication methods, or ground truth establishment.
- Information on Multi-Reader Multi-Case (MRMC) comparative effectiveness studies or standalone algorithm performance.
The document confirms the device's classification and allows its marketing, but it does not provide the detailed study results or acceptance criteria you are asking for.
Ask a specific question about this device
Page 1 of 1