Search Results
Found 2 results
510(k) Data Aggregation
(155 days)
The device is intended for ultrasound imaging, measurement, display and analysis of the human body and fluid.
The operating modes supported by the device are B, M.PWD CWD, Tissue Doppler, Color Doppler, Power Doppler, Tissue Velocity Imaging, Harmonic Imaging, 3D/4D, Combine modes.
This device is indicated for Abdominal; Fetal/Obstetrics; Gynecology; Transvagmal; Urology(including prostate); Trans rectal; Cardiac(adult and child); Peripheral Vascular; Small Organs/Parts(thyroid, breast, testicle), Musculoskeletal(Conventional and Superficial); Pediatrics(including neonatal cephalic); interventional(nerve block and vascular access); Intraoperative (abdominal, brain) and Adult Cephalic diagnostic Ultrasound applications.
This device is intended to use by, or by the order of, and under the supervision of an appropriately-trained healtheare professional qualified to direct the use of the device in hospitals or clinics.
VINNO G86,VINNO G86E, VINNO M86E, VINNO G65, VINNO G65, VINNO G65E, VINNO G65D, VINNO G65D, VINNO G90, VINNO G90E ultrasound devices are professional digital color ultrasonic apparatus.It transmits ultrasound waves into the body tissues and displays the echo images of the tissues and blood flow accordingly.
The provided document is a 510(k) summary for an ultrasound system, not a document detailing the performance study of an AI/ML-based medical device. Therefore, it does not contain the specific information requested regarding acceptance criteria and performance study details for an AI-powered device.
The document discusses the substantial equivalence of the VINNO G86, G86E, M86, M86E, G65, G65P, G65E, G65D, G90, and G90E ultrasound devices to a predicate device (Resona 7 series). It focuses on safety and performance characteristics of generalized ultrasound systems, not a specific AI/ML diagnostic or assistive tool.
Key sections that would typically contain the requested information for an AI/ML device are missing or not applicable in this document:
- Acceptance Criteria for an AI device: This document lists technical and safety standards (e.g., IEC 60601-1, NEMA UD 2) rather than diagnostic performance metrics (sensitivity, specificity, AUROC) for an AI output.
- Study Proving Acceptance Criteria: The performance data section mentions biocompatibility, electrical safety, EMC, software V&V, and acoustic output testing, which are standard for all medical devices. It does not describe a clinical performance study for an AI algorithm's diagnostic accuracy.
- Sample size for test set and data provenance: Not mentioned as there's no AI performance study described.
- Number of experts and qualifications for ground truth: Not mentioned.
- Adjudication method: Not mentioned.
- MRMC comparative effectiveness study: Not mentioned.
- Standalone (algorithm only) performance: Not mentioned.
- Type of ground truth: Not mentioned.
- Training set sample size and ground truth establishment: Not mentioned.
In summary, based on the provided text, it is not possible to answer the questions about acceptance criteria and the study proving the device meets them for an AI/ML powered medical device. The document describes a standard ultrasound system clearance, not one with an integral AI diagnostic or assistive component that would require such performance evaluations.
Ask a specific question about this device
(364 days)
The device is general purpose diagnostic ultrasound system for use by qualified healthcare professionals. It is applicable for adults, pregnant women, pediatric patients and neonates.
The device is intended for ultrasound imaging, measurement and analysis of human body and fluid for multiple clinical applications including: abdominal (GYN and Urology), Thoracic/Pleural, Fetal/Ob, small organ (including breast, thyroid, testes), peripheral vessel, neonatal cephalic, pediatric, musculo-skeletal (conventional, superficial), transrectal, trans-vaginal, cardiac adult, cardiatric, magnetic Needle guidance and imaging guidance of interventional procedures (e.g. biopsy).
This device is intended to use by, or by the order of, and under the supervision of a licensed physician qualified to direct the use of the device.
The device is used in hospital, clinics and clinical point-of-care for diagnosis of patients.
The VINNO 8, VINNO 6, VINNO 5 ultrasound devices are laptop digital color ultrasonic diagnostic devices which transmit ultrasound waves into the body tissues and display the echo images of the tissues and blood flow accordingly. The devices are capable of digital acquisition, processing and display and operate from an integrated battery or separate power supply/charger.
The systems also provide for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.
The VGuide NGS function detects the position and orientation of magnetized needles in the presence of the probe and displays this information relative to the ultrasound image. Spatial positioning of the needle, with respect to the ultrasound image, is then updated in real time. This guides the operator to better visualize the needle in the ultrasound image during ultrasound guided needling procedures.
The provided text describes a 510(k) premarket notification for the VINNO 8, VINNO 6, and VINNO 5 ultrasound devices. This document focuses on demonstrating substantial equivalence to predicate devices rather than proving a novel device's performance against specific clinical acceptance criteria through a clinical study.
Therefore, the response will focus on the information relevant to the substantial equivalence determination and the general performance testing mentioned, noting where specific information requested in the prompt (e.g., sample size for test/training sets, number of experts for ground truth, MRMC study, effect size) is not applicable or not provided within this type of regulatory submission.
Here's a breakdown based on the provided text:
1. A table of acceptance criteria and the reported device performance
For a 510(k) submission intending to demonstrate substantial equivalence, the "acceptance criteria" are generally that the new device performs "as safely and as effectively" as the legally marketed predicate device. This is primarily shown through a comparison of technological characteristics and compliance with recognized standards, rather than specific clinical performance metrics for a novel AI algorithm.
| Acceptance Criteria (Demonstrated through Equivalence) | Reported Device Performance (Summary) |
|---|---|
| Technological Characteristics Equivalence | The VINNO 8, VINNO 6, VINNO 5 ultrasound devices have the same technological characteristics and fundamental design as the predicate devices (GE Healthcare LOGIQ e and Philips CX50 Diagnostic Ultrasound System). Differences do not alter suitability for intended use. Features like indications for use, user qualification, physical design, patient contact materials, operating modes, operating controls, measurements, probe types, display monitor, acoustic output, conformity standards, and peripherals are compared and found to be similar or equivalent. |
| Biocompatibility | Evaluated in accordance with ISO 10993-1:2009. All evaluation acceptance criteria were met. |
| Electrical Safety and Electromagnetic Compatibility (EMC) | Complies with IEC 60601-1 (safety), IEC 60601-2-37 (safety for ultrasonic equipment), and IEC 60601-1-2 (EMC). |
| Software Verification and Validation | Conducted and documented as recommended by FDA guidance ("Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices"). The software was considered a "moderate" level of concern. |
| Acoustic Output | Performed according to NEMA UD2 and IEC60601-2-37. Complies with Track 3 limits: Ispta.3 ≤ 720mW/cm² and MI ≤ 1.9. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This document describes a submission for a general diagnostic ultrasound system, not an AI/ML device that requires a test set of patient data to evaluate algorithmic performance. The "performance data" provided relates to engineering and regulatory compliance testing.
Therefore, information on "sample size used for the test set" and "data provenance" (country of origin, retrospective/prospective) for a performance study of an AI algorithm is not applicable and not provided in this submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Since there is no "test set" of patient data for evaluating an AI algorithm's performance described, information on experts for establishing ground truth is not applicable and not provided.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Similarly, as no patient-data-based test set evaluation is described, adjudication methods are not applicable and not provided.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This document describes a general diagnostic ultrasound system. It does not appear to incorporate an AI component for interpretation or assistance to human readers that would necessitate an MRMC study. Therefore, an MRMC comparative effectiveness study is not applicable and not mentioned.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is for a diagnostic ultrasound machine, not a standalone AI algorithm. Therefore, "standalone algorithm performance" is not applicable.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
As this submission is for an ultrasound system itself, not an AI diagnostic algorithm, the concept of "ground truth" concerning patient diagnoses is not applicable to the performance data presented (which covers electrical safety, biocompatibility, software verification, and acoustic output).
8. The sample size for the training set
This document does not describe the development or testing of an AI/ML algorithm that would involve a training set. Therefore, "sample size for the training set" is not applicable and not provided.
9. How the ground truth for the training set was established
As there is no AI/ML training set mentioned, this information is not applicable and not provided.
In summary: The provided 510(k) submission demonstrates the substantial equivalence of the VINNO ultrasound systems to predicate devices through technical specifications, compliance with relevant industry standards (e.g., IEC, ISO, NEMA), and standard safety and performance testing for general medical imaging equipment. It does not involve the clinical study methodologies (e.g., test sets, training sets, expert review, MRMC studies) typically associated with the evaluation of AI/ML-driven diagnostic devices.
Ask a specific question about this device
Page 1 of 1