Search Results
Found 4 results
510(k) Data Aggregation
(101 days)
TOP-RANK Health Care Co., Ltd.
Iontophoresis Electrodes are intended to be used to introduce soluble salts and other drugs into the body as an alternative to hypodermic injection.
The proposed device is indicated for the administration of soluble salts or other drugs into the body for medical purposes. The Iontophoresis Electrodes utilize adhesive electrodes have both anode and cathode electrodes that adhere to the skin. The electrodes may either be connected by leads (wires) to a separate controller Forming a current loop. The portion of the active electrode that contacts the skin also has a drug/solution reservoir. End users saturate the reservoir with the ionic solution to be delivered. The reservoir contacts the electrode and the drug/solution is delivered to the patient transdermally.
Five models (Small, Medium, Large, Butterfly and Return Electrode) are included in this series products, and they have the same intended purposes, working theories and specifications.
The provided document describes the acceptance criteria and the study that proves the device meets those criteria for the Iontophoresis Electrodes (K232020).
Here's the breakdown of the requested information:
1. A table of acceptance criteria and the reported device performance
The document presents a comparison to a predicate device (Iontophoresis Drug Delivery Electrodes, K040495) to demonstrate substantial equivalence, rather than setting explicit numerical acceptance criteria for a new device's performance. The "acceptance criteria" are implied by the predicate device's specifications and the non-clinical test conclusions.
Acceptance Criteria (Implied by Predicate & Standards Compliance) | Reported Device Performance (Proposed Device) | Remark |
---|---|---|
Product Code: EGJ | EGJ | Same |
Regulation Number: 21 CFR 890.5525 | 21 CFR 890.5525 | Same |
Indications for Use: Introduce soluble salts and other drugs into the body as an alternative to hypodermic injection. | Introduce soluble salts and other drugs into the body as an alternative to hypodermic injection. | Same |
Patient Population: Patients requiring iontophoresis treatment | Patients requiring iontophoresis treatment | Same |
Usage Environment: Hospitals, medical clinics | Hospitals, medical clinics | Same |
Design (Electrode Size): Comparable to predicate sizes (e.g., Small, Medium, Large, Butterfly, Return Electrode) | ION-T01 (69x69 mm), ION-T02 (82x82 mm), ION-T03 (88x88 mm), ION-T04 (84x94 mm), ION-T05 (64x64 mm) | Same (Note 1 - differences in design requirements, not affecting safety/performance) |
Main Materials: Silver/Silver Chloride (Ag/AgCl), Buffering Agent, PU Conductive carbon film | Silver/Silver Chloride (Ag/AgCl), Buffering Agent, PU Conductive carbon film | Same |
Anatomical Sites: For body surface skin | For body surface skin | Same |
Working mode: Must be used as a complete set, consisting of an active drug delivery electrode and a passive return electrode. | Must be used as a complete set, consisting of an active drug delivery electrode and a passive return electrode. | Same |
Compatibility with other devices: Designed for use with iontophoresis devices only. | Designed for use with iontophoresis devices only. | Same |
Impedance: 2MΩ ~ 10MΩ | 2MΩ ~ 10MΩ | Same |
System current: 4.0mA | 4.0mA | Same |
Maximum Dosage: 80 mA-min | 80 mA-min | Same |
Fill Volume: Comparable to predicate fill volumes (e.g., 1.5cc, 2.5cc, 4.0cc, 2.0cc for specific models) | 1.5cc (ION-T01), 2.5cc (ION-T02), 4.0cc (ION-T03), 2.0cc (ION-T04) | Same |
Biocompatibility (Cytotoxicity): No Cytotoxicity | No Cytotoxicity | Same |
Biocompatibility (Skin sensitization): No evidence of sensitization | No evidence of sensitization | Same |
Biocompatibility (Irritation): No evidence of irritation | No evidence of irritation | Same |
EMC, Electrical Safety (Electrical safety): Comply with IEC 60601-1 | Comply with IEC 60601-1 | Same |
EMC, Electrical Safety (EMC): Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | Same |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
The document explicitly states: "No clinical study is included in this submission."
Therefore, there is no sample size for a clinical test set or information on its provenance. The testing was non-clinical.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable, as no clinical study was conducted that would require expert ground truth. The evaluation focused on non-clinical performance and safety based on recognized standards.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable, as no clinical study was conducted.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is an Iontophoresis Electrode, which is a physical device for drug delivery, not an AI or imaging diagnostic tool that would involve human readers or AI assistance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is a physical medical device, not an algorithm or software.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
Not applicable, as no clinical study was conducted. The "ground truth" for the non-clinical tests was established by compliance with international standards (IEC 60601-1, IEC 60601-1-2, ISO 10993-5, ISO 10993-10, ISO 10993-23) and comparison to a legally marketed predicate device.
8. The sample size for the training set
Not applicable. This is a physical medical device, not a machine learning model.
9. How the ground truth for the training set was established
Not applicable. This is a physical medical device, not a machine learning model.
Ask a specific question about this device
(90 days)
Top-Rank Health Care Co.,Ltd.
Intended for use with electric stimulation therapy devices, such as TENS and EMS. Conductive Gel is used with external electrodes to reduce the impedance of the contact between the electrode and the skin.
Conductive Gel can be used with any of dozens of stimulating devices that are legally sold, to reduce the impedance between the skin and the stimulating device. It consists of Purified water: 98.25%, Carbopol: 0.5%, Glycerin: 1%, Sodium hydroxide: 0.2% and Triclosan: 0.05%. And the Purified water used as the solvent, the Carbopol as a gel forming material, the Glycerin as a Moisturizing, the Sodium hydroxide as a Buffering and the Triclosan as a preservative. The Conductive Gel is used on intact skin surfaces. The entire surface of Conductive Gel is very conductive, smeared evenly on the stimulation device, so that the current is evenly distributed. The gel is to be generously applied to the area where an electrode will be used. The gel can be washed off the skin after use.
This document describes the validation of a "Conductive Gel" medical device. Here's a breakdown of the acceptance criteria and the study conducted to prove it:
1. Table of Acceptance Criteria and Reported Device Performance
Test Item | Purpose of the Test | Reference Standard | Acceptance Criteria | Test Results |
---|---|---|---|---|
In vitro Cytotoxicity Test | Determine if the target device extract is cytotoxic | ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity | Non-cytotoxicity to L929 cells | Pass |
Skin Sensitization Test | Evaluate skin sensitization after topical applications | ISO 10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization | Frequency of positive challenge results in sample extract and Negative control animals are 0%, Positive control is 100% | Pass |
Skin Irritation Test | Evaluate skin irritation after single topical applications | ISO 10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization | Negligible irritation in a rabbit skin single exposure test | Pass |
Usability Study | Study how usability will be performed | IEC 62366-1 Edition 1.0 2015-02, Medical devices - Part 1: Application of usability engineering to medical devices [Including CORRIGENDUM 1 (2016)] | The subject device can meet the usability goal of IEC 62366-1 standards | Pass |
Shelf Life Test | Study if performance meets 2-year shelf life requirements | ASTM F1980-16: Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices; Guidance document for the "Shelf Life of Medical Devices" issued in April 1991 | All items (Exterior, Package, pH, Impedance and conductivity) tested on both before and after aging samples meet the performance required | Pass |
Additional Performance Data from Comparison to Predicate Device:
Item | Acceptance Criteria (Predicate) | Reported Device Performance (Subject Device) |
---|---|---|
Impedance (at 1 MHz) | ≤ 500 Ω | ≤ 500 Ω |
pH | 7.0 - 7.5 | 6.5 - 7.5 |
Percent Concentration of Ingredients | Purified water: 98.25%, Carbopol: 0.5%, Glycerin: 1%, Sodium hydroxide: 0.2%, Triclosan: 0.05% | Purified water: 98.25%, Carbopol: 0.5%, Glycerin: 1%, Sodium hydroxide: 0.2%, Triclosan: 0.05% |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the exact sample sizes for each non-clinical test (cytotoxicity, sensitization, irritation, usability, shelf life). However, it refers to standard biological evaluation and aging test methods (ISO 10993-5, ISO 10993-10, IEC 62366-1, ASTM F1980-16), which inherently define sample sizes for their respective procedures, typically involving animal models for biological tests or specific numbers of units for usability/shelf life.
- Data Provenance: The studies were identified as "lab bench testing." The location of these labs or the country of origin for the data is not specified beyond the submitter's and correspondent's addresses (China). The tests are non-clinical, meaning they were not conducted on human subjects.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not provided as the validation is based on non-clinical (lab bench) testing against established standards, not on expert consensus or human reader performance.
4. Adjudication Method for the Test Set
This information is not applicable/provided as the validation is based on objective lab bench testing against predefined acceptance criteria, not subjective interpretations requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a conductive gel, which is a physical product, not an AI or software device that would involve human readers or comparative effectiveness studies with AI assistance.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. As mentioned, this is a physical medical device (conductive gel), not an algorithm or AI.
7. The Type of Ground Truth Used
The "ground truth" for the non-clinical tests is based on established scientific and regulatory standards and their defined criteria.
- For biological tests (cytotoxicity, sensitization, irritation), the ground truth is whether the device elicits a toxic or adverse biological response as determined by standardized in-vitro and in-vivo (animal) models.
- For usability, the ground truth is compliance with IEC 62366-1 usability engineering principles.
- For shelf life, the ground truth is the maintenance of specified physical and chemical properties (like pH, impedance, conductivity, exterior, package integrity) over the declared shelf life based on accelerated aging models.
- For material composition, the ground truth is the precise percentage concentration of each ingredient.
- For impedance and pH, the ground truth is the measured value compared to the specified range or limit.
8. The Sample Size for the Training Set
Not applicable. This is a physical device, and there is no "training set" in the context of machine learning or AI.
9. How the Ground Truth for the Training Set was Established
Not applicable. As above, there is no training set for this type of device.
Ask a specific question about this device
(321 days)
Top-Rank Health Care Co.,Ltd.
The TOPTENS-01 can be used for the Symptomatic relief of chronic intractable pain, Post traumatic pain adjunctive treatment, Post-surgical pain adjunctive treatment.
The proposed device is transcutaneous electrical nerve stimulator for pain relief. this is a device used to apply an electrical current to electrodes on a patient's skin to treat pain. It only has one model: TOPTENS-01. TOPTENS-01 is mainly composed: host, built-in power supply, output lead and electrode. TOPTENS-01 power supply is DC4.5V, using built-in removable 3 AAA batteries. TOPTENS-01 comes with 2 independent output channels (A&B). Users can be adjust independently the two output channels Intensity by pressing the Left and right sides Intensity adjustment key. LCD screen display the remaining stimulation time in the top left corner. The default stimulation time is 30 minutes. TOPTENS-01 has 9 types stimulation modes: Mode1, Mode2, Mode 3, Mode 4, Mode 5, Mode6, Mode 7, Mode 8, Mode 9.
This is a 510(k) premarket notification for a medical device called the Transcutaneous Electrical Nerve Stimulator (TOPTENS-01) manufactured by Top-Rank Health Care Co., Ltd. The document aims to demonstrate that the TOPTENS-01 device is substantially equivalent to a legally marketed predicate device (K152374, also a Transcutaneous Electrical Nerve Stimulator).
This document does not contain acceptance criteria or a study proving the device meets those criteria in the context of clinical performance or diagnostic accuracy. Instead, it demonstrates substantial equivalence based on a comparison to a predicate device, focusing on technical specifications and safety standards.
Therefore, many of the requested sections (acceptance criteria related to clinical performance, sample sizes for test/training sets, data provenance, expert ground truth, MRMC studies, standalone performance, and ground truth types) are not applicable or cannot be extracted from this specific 510(k) summary, as these types of studies were not conducted or reported for this submission.
Here's an analysis based on the provided document:
1. A table of acceptance criteria and the reported device performance
This document does not establish performance acceptance criteria in terms of clinical outcomes or accuracy. Instead, it demonstrates substantial equivalence based on compliance with harmonized electrical safety and EMC standards, and a comparison of the device's technical specifications to a predicate device. The "performance" being evaluated is its adherence to these standards and the similarity of its operational parameters to the predicate.
Acceptance Criteria Category (from 510(k) Summary) | Description of Criteria / Standard | Reported Device Performance and Compliance |
---|---|---|
Electrical Safety | Compliance with IEC 60601-1: Medical electrical equipment Part 1: General Requirements For Basic Safety And Essential Performance; and IEC 60601-2-10: Medical electrical equipment Part 2-10: Particular requirements for the safety of nerve and muscle stimulators. | The device complies with IEC 60601-1 and IEC 60601-2-10. The 510(k) states: "The new device passed Safety testing, so these differences will not cause any safety and effectiveness issues." (Difference 4). "The pulse parameters meet the maximum current density |
Ask a specific question about this device
(192 days)
TOP-RANK Health Care Co., Ltd
Electrosurgical disposable grounding pads are to conduct electrosurgical energy from target tissue of a patient back to an electrosurgical unit (ESU), or generator.
Electrosurgical disposable grounding pads construct of a layer of conductive aluminum foil with a foam as the supportive base of the pad. The aluminum foil is coated with adhesive hydro-gel layer. There is another film placed above the adhesive hydro-gel layer, which is used to protect the gel during the transportation and storage and will be removed prior to use.
They are available in different combinations of configurations, types, intended populations and conductive area.
They are not provided in sterile conditions and not required to be sterilized by the end user prior to operation. However, they are for single use only.
The provided text is a 510(k) summary for Electrosurgical Disposable Grounding Pads. This document is a regulatory submission to the FDA for a medical device and not a study proving the device meets acceptance criteria for an AI/ML powered device.
Therefore, many of the requested criteria related to AI/ML model performance evaluation (e.g., sample size for test/training sets, expert ground truth, MRMC study, human reader improvement, standalone performance, etc.) are not applicable to this document as it describes a physical, non-software medical device.
The document focuses on demonstrating substantial equivalence to a predicate device through non-clinical testing (bench testing) and biocompatibility, not through AI/ML performance metrics.
Here's a breakdown based on the provided document, addressing the applicable points and explaining why others are not applicable:
Acceptance Criteria and Study for Electrosurgical Disposable Grounding Pads
1. Table of Acceptance Criteria and Reported Device Performance
For this physical device, "acceptance criteria" are derived from compliance with recognized medical device standards and demonstrating substantial equivalence to a predicate device. The "reported device performance" is its successful compliance with these standards.
Acceptance Criteria Category | Specific Criteria (Standard/Test) | Reported Device Performance |
---|---|---|
Electrical Safety | IEC 60601-1:2005 | Complies (results demonstrated device met design specifications and SE) |
IEC 60601-2-2:2009 | Complies (results demonstrated device met design specifications and SE) | |
Electromagnetic Compatibility | IEC 60601-1-2:2007 | Complies (results demonstrated device met design specifications and SE) |
Biocompatibility | ISO 10993-5:2009/(R) 2014 (Cytotoxicity) | Complies (tested with ISO 10993 Series Standards) |
ISO 10993-10:2010 (Irritation & Sensitization) | Complies (tested with ISO 10993 Series Standards) | |
Functional/Design Equivalence | Conduct electrosurgical energy from target tissue to ESU/generator | Functionally equivalent to predicate device (K091672) |
Configuration: Single and Split aluminum film; Adult and pediatric; Single Use | Configurations are equivalent to predicate device | |
Material Equivalence | Adhesive hydro-gel, Aluminum foil, Polyester (PET), Foam (Polyester Fibre) | Materials are equivalent to predicate device |
2. Sample size used for the test set and the data provenance
- Sample Size: Not specified in the document. For non-clinical (bench) testing of physical devices, sample sizes are typically determined by engineering and statistical principles to ensure robust testing, but the explicit number of units tested is not provided in this regulatory summary.
- Data Provenance: The tests were non-clinical tests conducted to verify compliance with international standards (IEC, ISO). The location of the testing is not explicitly stated, but the company is based in China. The data would be prospective, as it was generated specifically for this submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not Applicable. This is a physical electrosurgical device, not an AI/ML model for which "ground truth" would be established by experts reviewing data (e.g., medical images). Ground truth for this device is based on objective measurements and compliance with engineering and safety standards.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not Applicable. This concept is relevant to expert review of data (e.g., medical images) to establish ground truth for AI/ML validation, not for the non-clinical testing of a physical medical device.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not Applicable. This device is a physical electrosurgical pad, not an AI/ML system, so no MRMC study or assessment of human reader improvement with AI assistance was performed.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not Applicable. This device is a physical electrosurgical pad, not an AI/ML algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- For this physical device, "ground truth" is established through objective measurements and compliance with recognized international standards (e.g., specified resistance limits, temperature rise limits, material compatibility, etc.) during non-clinical (bench) testing. There is no human interpretation of data for ground truth establishment as would be the case for an AI/ML diagnostic device.
8. The sample size for the training set
- Not Applicable. This device is a physical medical device, not an AI/ML model that requires a training set.
9. How the ground truth for the training set was established
- Not Applicable. As no AI/ML model was developed, there was no training set and thus no ground truth established for it.
Ask a specific question about this device
Page 1 of 1