510(k) Data Aggregation

    K Number
    K151170
    Manufacturer
    Date Cleared
    2015-07-02

    (62 days)

    Product Code
    Regulation Number
    884.5470
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Tosama 100% Organic Cotton Menstrual Tampon is intended for insertion into the vagina for the absorption of menstrual or other vaginal discharge.

    Device Description

    The device will be offered as a traditional unscented menstrual 100% organic cotton tampon consisting of an absorbent pledget, a withdrawal cord, and an applicator. The pledget is of the traditional cylindrical, bullet-like shape and the applicator has a standard rounded tip to ease insertion. Each tampon is individually wrapped and packaged in multi-unit containers for retail sale. It will be offered in three absorbencies: Regular, Super, and Super Plus.

    AI/ML Overview

    This document is a 510(k) premarket notification for a menstrual tampon. It is not a study about an AI-powered medical device. Therefore, it does not contain the information required to answer your request regarding acceptance criteria and a study proving a device meets them for an AI device.

    The document discusses the substantial equivalence of the "Tosama 100% Organic Cotton Menstrual Tampon" to a predicate device (Naturalena 100% Cotton Tampon). It outlines the general safety and effectiveness of the traditional medical device through non-clinical testing.

    Here's why the requested information cannot be provided from this document:

    1. Nature of the Device: The device described is a physical menstrual tampon, not an AI-powered software or system.
    2. Type of Study: The document refers to "non-clinical data" and "testing" (chemical, microbiological, biocompatibility, efficacy) for the tampon, which are standard for physical medical devices to demonstrate safety and effectiveness. It does not describe a clinical study of an AI algorithm's performance.
    3. Missing AI-Specific Information: There is no mention of algorithms, machine learning, AI assistance, human readers, ground truth establishment by experts, or training/test sets in the context of AI.

    The requested information is specific to the development and validation of AI/ML-based medical devices, which are not covered by this submission for a menstrual tampon.

    Ask a Question

    Ask a specific question about this device

    K Number
    K150897
    Manufacturer
    Date Cleared
    2015-04-20

    (18 days)

    Product Code
    Regulation Number
    884.5470
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Tosama 100% Cotton Tampon is intended for insertion into the vagina for absorption of menstrual or other vaginal discharge.

    Device Description

    The device will be offered as a traditional unscented menstrual 100% cotton tampon consisting of an absorbent pledget, a withdrawal cord, and an applicator. The pledget is of the traditional cylindrical, bullet-like shape and the applicator has a standard rounded tip to ease insertion. Each tampon is individually wrapped and packaged in multi-unit containers for retail sale. It will be offered in three absorbencies: Regular, Super, and Super Plus.

    AI/ML Overview

    The provided document describes a medical device (Tosama 100% Cotton Menstrual Tampon) and its substantial equivalence to predicate devices, rather than an AI-powered diagnostic tool. Therefore, many of the requested categories for acceptance criteria and study design (such as Number of experts used to establish the ground truth, Adjudication method, MRMC comparative effectiveness study, Standalone performance, Type of ground truth, and Sample size for the training set and Ground truth for training set) are not applicable.

    However, I can extract information related to the device's performance, the tests conducted, and the basis for its equivalence.

    Here's the information derived from the provided text, adapted to the closest relevant categories:

    1. A table of acceptance criteria and the reported device performance

    The document doesn't explicitly state "acceptance criteria" but lists various physical and functional specifications for the tampon, along with reported values for the Tosama 100% Cotton Menstrual Tampon and the predicate devices. The study objective was to demonstrate substantial equivalence, meaning the Tosama device performs within the same established ranges as the predicates.

    Feature / Acceptance Criteria (Implied)Reported Device Performance (Tosama 100% Cotton Menstrual Tampon)Predicate Performance (Naturalena / Cottons 100% Natural Cotton Tampon)
    Indications for UseIntended for insertion into the vagina for absorption of menstrual or other vaginal discharge.Same
    DesignTampon with cylindrical shape and bullet-like tip. Applicator with smooth, rounded tip.Same
    Pledget Material100% Cotton100% Cotton
    Withdrawal Cord Material100% Cotton100% Cotton
    Applicator MaterialPolyethylenePolyethylene
    Individual Wrapping MaterialPolypropylenePolypropylene (Naturalena), Cellophane (Cottons)
    Additives & Finishing AgentsHydrophobic coating on withdrawal cord.Hydrophobic coating on withdrawal cord (Naturalena), None (Cottons)
    Digital Regular Tampon
    Length43.0mm – 46.0mm43.0mm – 46.0mm
    Diameter11.0mm – 12.0mm11.0mm – 12.0mm
    Mass2.30g – 2.70g2.30g – 2.70g
    Withdrawal Cord Length130mm – 160mm130mm – 160mm
    Syngina Absorbency6.0g - 9.0g6.0g - 9.0g
    Digital Super Tampon
    Length48.0mm – 51.0mm48.0mm – 51.0mm
    Diameter12.0mm – 13.0mm12.0mm – 13.0mm
    Mass2.70g – 3.10g2.70g – 3.10g
    Withdrawal Cord Length130mm - 160mm130mm - 160mm
    Syngina Absorbency9.0g - 12.0g9.0g - 12.0g
    Digital Super Plus Tampon
    Length48.0mm - 51.0mm48.0mm - 51.0mm
    Diameter14.0mm - 15.0mm14.0mm - 15.0mm
    Mass3.30g – 3.80g3.30g – 3.80g
    Withdrawal Cord Length130mm - 160mm130mm - 160mm
    Syngina Absorbency12.0g – 15.0g12.0g – 15.0g
    Applicator Regular Tampon
    Tampon Length43.0mm – 46.0mm43.0mm - 46.0mm
    Applicator Length120mm ± 5%120mm ± 5%
    Tampon Diameter11.0mm – 12.0mm11.0mm - 12.0mm
    Applicator Diameter13mm13mm
    Tampon Mass2.00g - 2.30g2.00g - 2.30g
    Applicator Mass3.40g - 3.60g3.40g - 3.60g
    Total Mass5.40g - 5.90g5.40g - 5.90g
    Withdrawal Cord Length130mm - 160mm130mm - 160mm
    Syngina Absorbency6.0g - 9.0g6.0g - 9.0g
    Applicator Super Tampon
    Tampon Length43.0mm - 46.0mm43.0mm - 46.0mm
    Applicator Length120mm ± 5%120mm ± 5%
    Tampon Diameter12.0mm - 13.0mm12.0mm - 13.0mm
    Applicator Diameter16mm16mm
    Tampon Mass2.60g - 3.00g2.60g - 3.00g
    Applicator Mass4.10g - 4.30g4.10g - 4.30g
    Total Mass6.70g - 7.30g6.70g - 7.30g
    Withdrawal Cord Length130mm - 160mm130mm - 160mm
    Syngina Absorbency9.0g - 12.0g9.0g - 12.0g
    Applicator Super Plus Tampon
    Tampon Length43.0mm - 46.0mm43.0mm - 46.0mm
    Applicator Length120mm ± 5%120mm ± 5%
    Tampon Diameter14.0mm - 15.0mm14.0mm - 15.0mm
    Applicator Diameter18mm18mm
    Tampon Mass3.00g - 3.50g3.00g - 3.50g
    Applicator Mass4.20g - 4.40g4.20g - 4.40g
    Total Mass7.20g - 7.90g7.20g - 7.90g
    Withdrawal Cord Length130mm - 160mm130mm - 160mm
    Syngina Absorbency12.0g - 15.0g12.0g - 15.0g

    The "study that proves the device meets the acceptance criteria" is described as a non-clinical data summary. This non-clinical data was used for substantial equivalence determination to predicate devices already on the market.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: Not explicitly stated for each test, but it is implied that representative samples of the Tosama 100% Cotton Menstrual Tampon and its components were tested.
    • Data Provenance: The manufacturer is TOSAMA, d.o.o. from Slovenia, EU. The testing appears to have been conducted by or for the manufacturer. The data is likely prospective, as it relates to testing the specific device for regulatory submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This is a physical and chemical testing study for a medical device, not an AI-powered diagnostic tool requiring expert ground truth for interpretation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is a physical and chemical testing study, not based on human review/adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI device. The device itself is a standalone medical product.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    The "ground truth" in this context refers to established test methodologies and results for similar predicate devices, and the physical, chemical, and microbiological properties of the materials and finished product. Testing followed "recommended requirements."

    • Chemical Analysis: Tests for pesticide, dioxin, heavy metal, phthalate, and PCB residues.
    • Microbiological Testing: Enumeration per USP <61>/<62>, Toxic Shock Syndrome Toxin-1 production, vaginal microflora growth alteration, and S. aureus growth enhancement.
    • Biocompatibility Testing: In vitro cytotoxicity, intracutaneous and vaginal irritation, sensitization, and acute systemic toxicity.
    • Efficacy Testing: Syngina absorbency, fiber loss, withdrawal cord pull-out resistance, and tampon integrity.
    • Physical Characteristics: Length, diameter, mass, withdrawal cord length, applicator dimensions.

    8. The sample size for the training set

    Not applicable. This is not an AI device, so there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable. There is no training set as it is not an AI device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K120481
    Manufacturer
    Date Cleared
    2012-07-13

    (148 days)

    Product Code
    Regulation Number
    884.5470
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Lil-lets tampons (both types) are unscented tampons for:
    • Women's personal hygiene with respect to intra vaginal absorption of menstrual or other vaginal discharge.
    • The plastic applicator is for easing the placement of the tampon correctly into the vagina (only the Lil-lets Silk Comfort Compact Plastic Applicator Tampon).

    Device Description

    Lil-lets tampons are used to absorb menstrual fluid. Lil-lets series tampons come with a plastic applicator and without a plastic applicator in sizes: Lite, Regular, Super, Super plus and Ultra. Lil-lets tampons are made of commercial cotton and rayon, a polyethylene/polyester cover, and cotton or rayon string.

    AI/ML Overview

    The provided 510(k) summary (K120481) for the Lil-lets Silk Comfort Compact Plastic Applicator Tampon & Lil-lets Non-Applicator Tampon describes pre-clinical testing to demonstrate substantial equivalence to a predicate device, MAXIM Plastic Applicator and Non Applicator Tampons (K080775). However, it does not contain a comprehensive study proving that the device meets specific acceptance criteria in the manner typically described for clinical studies of new medical devices with defined performance metrics, patient populations, and statistical analysis.

    Instead, the summary focuses on non-clinical tests to ensure safety and equivalence based on existing standards and knowledge of the predicate device. The "acceptance criteria" are implied by adherence to these standards and by demonstrating that the Lil-lets tampons do not present new questions of safety or effectiveness.

    Here's a breakdown of the information based on the provided text, using the requested format where applicable:


    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicsAcceptance Criteria (Standard)Reported Device Performance (Summary)
    BiocompatibilitySubstantial equivalence to predicate's materials (MAXIM Plastic Applicator and Non Applicator Tampons, K080775)."No additional testing was necessary because the materials of Lil-lets tampons are substantially equivalent to MAXIM Plastic Applicator and Non Applicator Tampons, K080775."
    Microbiology Tests- Enhance the growth of Staphylococcus aureus (Zone of inhibition)- Alter the growth of normal vaginal microflora (Zone of inhibition)- Increase the production of Toxic Shock Syndrome Toxin-1 (TSST-1)(Not explicitly stated as "met" or specific results, but the submission implies these were assessed and deemed acceptable for substantial equivalence. The overall conclusion states substantial equivalence.)
    Syngyna TestingAdherence to 21 CFR 801.430(f)(2) to verify absorbency ranges as specified in the regulation. Tampons must be labeled in accordance with required ranges."Syngyna testing was conducted in accordance to 21 CFR 801.430(f)(2) to verify that the subject tampons met absorbency ranges as specified in the regulation. The tampons are labeled in accordance to required ranges." (This directly states the criteria were met.)
    Physical and Chemical TestsInternal methods for:- Fibre loss test- Stability of digital tampons- Withdrawal cord attachment test- Expulsion force of applicator tampons(Not explicitly stated as "met" or specific results, but the submission implies these were assessed and deemed acceptable for substantial equivalence. The overall conclusion states substantial equivalence.)
    Technological CharacteristicsNo differences from the technical characteristics of the substantially equivalent predicate devices (MAXIM tampons under K080775), differing only in dimensions across sizes (Lite, Regular, Super, Super plus, Ultra). The tampons contain commercial cotton and rayon, a polyethylene/polyester cover, and cotton or rayon string."There are no differences between the technical characteristics of the Lil-lets tampons and the predicate characteristics of the substantially equivalent devices MAXIM tampons under K080775. All sizes... have the same characteristics; they differ only in dimensions."

    Detailed Study Information:

    This document is a 510(k) summary focused on demonstrating substantial equivalence for tampons, not a report of a traditional clinical performance study for a novel device. Therefore, many of the requested points do not apply or are not explicitly detailed in the provided text, as the regulatory pathway chosen (510(k)) relies heavily on comparing technical and safety characteristics to a legally marketed predicate device.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • The document does not specify general "test set" sample sizes or data provenance in the context of a clinical study.
    • For Syngyna testing, it states "Syngyna testing was conducted," but does not provide the number of tampons tested. This is a bench test performed in a lab, not on human subjects.
    • For Microbiology and Physical/Chemical tests, no sample sizes are provided for individual tests.
    • Data provenance for these non-clinical tests would likely be the manufacturing site or a contract lab performing the tests, which is in Slovenia, given the submitter's address.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • Not applicable in this context. "Ground truth" in the sense of expert consensus for diagnostic performance is not relevant to non-clinical tests of tampon safety and absorbency. Expertise is implied by adherence to regulatory standards and testing methodologies by competent personnel.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. This concept is for clinical studies involving human interpretation or challenging outcomes.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is a tampon, not an AI-assisted diagnostic tool or an imaging device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This device is a physical product (tampon), not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • The "ground truth" for these non-clinical tests would be the established regulatory standards and validated test methodologies themselves (e.g., 21 CFR 801.430(f)(2) for absorbency, specific microbiological assays, internal physical testing protocols).
    • For biocompatibility, the ground truth is the accepted safety profile of the predicate device's materials.

    8. The sample size for the training set:

    • Not applicable. There is no "training set" in the context of such a device and its regulatory submission.

    9. How the ground truth for the training set was established:

    • Not applicable.