510(k) Data Aggregation
K Number
K063873Device Name
ICARE TONOMETER TA01I
Manufacturer
TIOLAT OY
Date Cleared
2007-05-23
(145 days)
Product Code
HKY, HKX
Regulation Number
886.1930Why did this record match?
Applicant Name (Manufacturer) :
TIOLAT OY
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The iCare Tonometer TA01i is intended to be used for the measurement of intraocular pressure of the human eye.
Device Description
The iCare Tonometer TA01i is a portable, handheld, battery operated tonometer, which measures intraocular pressure (IOP). To measure IOP, a very light (26.5 mg) and a slow moving (0.25-0.30 m/s) probe makes momentary contact with the cornea of the eye. The device measures the speed of the probe during the time probe is in contact with cornea. Resulting IOP is displayed on a liquid crystal display (LCD). The force used for the measurement which is directed onto the eye is extremely small, because of the low weight and slow speed of the probe.