510(k) Data Aggregation

    K Number
    K210359
    Device Name
    Hexanium TLIF
    Manufacturer
    Date Cleared
    2021-03-10

    (30 days)

    Product Code
    Regulation Number
    888.3080
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    SpineVision SAS

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Hexanium® TLIF (Transforaminal Lumbar Interbody Fusion) system is an intervertebral body fusion device indicated for use with autogenous bone graft in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two continuous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). Patients should have received at least 6 months of non-operative treatment prior to treatment with Hexanium® TLIF system. This device has to be filled with autogenous bone graft material. This device is implanted via transforaminal approach. Hexanium® TLIF system must be used in combination with supplemental internal spinal fixation which has been cleared by the FDA for use in the lumbar spine.

    Device Description

    The Hexanium® TLIF is a titanium alloy (Ti6Al4V ELI) interbody cage manufactured via an Additive Manufacturing method. The honeycomb structure allows for bone through-growth through the structure of the device as well as providing lateral and vertical bone graft windows in the body of the cage.
    Hexanium® TLIF is available in a straight and curved shape in heights of 7mm to 16mm in 1mm increments in a Small footprint (28mm x 10mm) and a Medium footprint (32mm x 10mm) for the straight cage and a Small footprint (28mm x 9mm) and a Medium footprint (32mm x 10.5mm) for the curved cage. Hexanium® TLIF has a 5° Lordosis angle.
    Hexanium® TLIF devices are provided sterile.

    AI/ML Overview

    This document describes the regulatory clearance for the Hexanium® TLIF intervertebral body fusion device and focuses on its substantial equivalence to a predicate device rather than its performance against specific acceptance criteria for an AI/ML algorithm. Therefore, many of the requested categories regarding AI/ML study design, such as MRMC studies, expert ground truth establishment, and training/test set details, are not applicable or cannot be extracted from this document.

    The acceptance criteria described here pertain to the device's mechanical properties, materials, and overall safety and efficacy as an implant, demonstrated through testing and comparison to a legally marketed predicate device.

    Here's an attempt to populate the table and answer the questions based on the provided text, indicating "Not Applicable" (NA) or "Not Provided" (NP) where the information isn't present in the context of an AI/ML study.

    Acceptance Criteria and Device Performance (for a Medical Device, not AI/ML)

    Acceptance Criteria CategorySpecific Criterion (if stated)Reported Device Performance
    Mechanical Testing (Fatigue, Compression, etc.)Compliance with ASTM 2267-04 and ASTM F2077-14 standards, specifically demonstrating equivalent or superior mechanical properties to the predicate for:
    • Axial compression
    • Compression shear
    • Torsion
    • Expulsion
    • Bone fusion (in terms of surface area and volume for grafting) | The mechanical testing performed for the predicate device (Hexanium TLIF system, K180437) was submitted and found to support the subject Hexanium TLIF cages.
      The study showed that the predicate Hexanium TLIF cage remains the worst-case construct compared to the subject device for:
    • Highest von Mises stress: in axial compression, compression shear, torsion, and expulsion. This implies the subject device experiences less stress under similar loads.
    • Smallest surface area between bone graft and endplates: This means the predicate is the worst-case for bone fusion, implying the subject device offers better or equivalent bone fusion potential.
    • Lowest bone graft volume: Reinforcing the predicate as worst-case for bone fusion. |
      | Material Biocompatibility | Use of biocompatible materials. | Manufactured from titanium alloy (Ti6Al4V ELI). (Implicitly accepted as a common, biocompatible material for implants). |
      | Sterilization & Cleaning | Demonstrates ease of cleaning and effective sterilization. | The predicate Hexanium TLIF (K180437) is the worst-case for cleaning and sterilization as it is the most difficult-to-clean, has the largest surface area and volume, and has the heaviest weight. This implies the subject device is easier to clean and sterilize. |
      | Pyrogenicity | Meets pyrogen limit specifications. | Bacterial Endotoxins Test was performed in accordance with USP to demonstrate that the device meets pyrogen limit specifications. |
      | Design Consistency | Equivalent in design, mechanical properties, and function to predicate. | Hexanium® TLIF is substantially equivalent to the primary predicate device in terms of intended use, design, mechanical properties, and function. The specific design difference noted is the honeycomb structure, which is an additive manufacturing feature, but the overall form, lordosis angle (5°), and size ranges are similar to the predicate, and critically, the new design is shown not to be mechanically inferior. |
      | Regulatory Compliance | Conforms to Class II Special Controls Guidance Document. | The Hexanium® TLIF device conforms to the Class II Special Controls Guidance Document: Intervertebral Body Fusion Device Document issued on June 12, 2007. |

    Study Details (Relevant to Device Clearance, Not AI/ML Performance)

    1. Sample size used for the test set and the data provenance:

      • Sample Size: Not applicable in the context of an AI/ML test set. For mechanical testing, devices were tested according to ASTM standards, which would imply specific sample sizes for destructive and non-destructive testing, but these exact numbers are not provided in the summary. The study relies on demonstrating that predefined worst-case predicate constructs were mechanically and cleaning/sterilization-wise more challenging than the proposed device.
      • Data Provenance: Not specified for a "test set" in the AI/ML sense. The "data" comes from mechanical testing of the Hexanium TLIF system (K180437) and the subject devices, which would have been conducted in a lab setting. Geographical origin of this specific testing data is not provided. The study is retrospective in the sense that it leverages prior testing data from the predicate device and compares it to the new device via analysis (e.g., FEA) rather than new large-scale clinical trials.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable (NA). This document does not pertain to an AI/ML system requiring expert-established ground truth from images or clinical data. Device performance is established via engineering principles, material science, and mechanical testing against recognized standards.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • NA. No human adjudication of clinical or image data for an AI/ML test set was performed.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • NA. This is a hardware medical device (an implant), not an AI/ML diagnostic or assistive software. No human reader studies were conducted or are relevant for this type of clearance.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • NA. This is not an AI/ML algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" here is based on established engineering standards (ASTM), material properties, and physical performance under controlled mechanical testing conditions. In some cases, it's a comparison to a "worst-case" scenario of an existing cleared device.
    7. The sample size for the training set:

      • NA. No AI/ML training set was involved.
    8. How the ground truth for the training set was established:

      • NA. No AI/ML training set was involved.

    Summary of Device Acceptance:

    The Hexanium® TLIF device was accepted through a 510(k) pathway, demonstrating substantial equivalence to a previously cleared predicate device (Hexanium® TLIF, K180437). This was primarily achieved by:

    • Mechanical Testing and Analysis: Leveraging existing mechanical testing data from the predicate and performing further analysis (e.g., Finite Element Analysis - FEA) to show that the new device meets or exceeds the performance of the predicate. Crucially, the predicate device was shown to represent the "worst-case" scenario in terms of mechanical stress response (axial compression, shear, torsion, expulsion) and properties affecting bone fusion (surface area, volume) and cleaning/sterilization. This means if the predicate passed, the subject device, being "less worst-case" or "better," would also pass.
    • Material Equivalence: Using the same established biocompatible material (Ti6Al4V ELI).
    • Sterilization and Pyrogenicity: Demonstrating compliance with specific standards (USP) for bacterial endotoxins and showing the predicate was the worst-case for cleaning/sterilization.
    • Compliance with Special Controls: Adhering to the specific guidance for intervertebral body fusion devices.

    No clinical data was presented or required for this 510(k) clearance, as functional and safety equivalence was demonstrated through non-clinical testing and comparison to the predicate.

    Ask a Question

    Ask a specific question about this device

    K Number
    K193000
    Device Name
    Hexanium® ACIF
    Manufacturer
    Date Cleared
    2020-10-20

    (358 days)

    Product Code
    Regulation Number
    888.3080
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    SpineVision SAS

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Hexanium® ACIF (Anterior Cervical Interbody Fusion) system is intervertebral body fusion device indicated for use with autogenous bone graft in skeletally mature patients with Degenerative Disc Disease (DDD) at one level from C3-T1. DDD is defined as discogenic neck pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). Patients should have received at least 6 weeks of non-operative treatment prior to treatment with Hexanium® ACIF system. This device has to be filled with autogenous bone graft material. This device is implanted via an anterior approach.

    Device Description

    The Hexanium® ACIF is a titanium alloy (Ti6Al4V ELI) interbody cage manufactured via an Additive Manufacturing method. The honeycomb structure allows for bone through-growth through the structure of the device as well as providing lateral and vertical bone graft windows in the body of the cage. Hexanium® ACIF is available in 2 sagittal profiles (lordotic 6° and convex 6°), heights of 5-12 mm, and 3 footprints (15x12mm, 17x14mm, and 19x15mm). The Hexanium® ACIF screws are self-drilling and available in lengths of 10, 12, 14, and 16mm and diameters of 3.50 and 3.80mm. Hexanium® ACIF is provided sterile.

    AI/ML Overview

    This is a 510(k) premarket notification for a medical device (Hexanium® ACIF, an intervertebral body fusion device) seeking substantial equivalence to a predicate device, not an AI/ML medical device. Therefore, the information requested about acceptance criteria, study design, ground truth, sample sizes for training/test sets, expert adjudication, and MRMC studies is not applicable in this context.

    The document states:

    • "No clinical data has been presented." This means no human study was performed to assess the performance of the device in patients.
    • "The following non-clinical tests were conducted: static and dynamic axial compression, static and dynamic shear compression, static and dynamic torsion testing according to ASTM F2077, subsidence testing according to ASTM F2267, and expulsion testing. Results demonstrate comparable mechanical properties to the predicate device." This indicates the performance data relates to mechanical properties tested in a lab setting, not clinical performance or AI/ML algorithm performance.

    The acceptance criteria for this device would be based on these mechanical tests demonstrating equivalence to the predicate device, generally through meeting established ASTM standards. The "performance data" section states that the device demonstrated "comparable mechanical properties to the predicate device," which implies it met the acceptance criteria for these physical tests.

    Since this is a traditional medical device submission and not an AI/ML device, the detailed requirements for AI/ML performance studies (like MRMC, standalone performance, ground truth establishment, expert qualifications, etc.) are not relevant here.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180437
    Device Name
    Hexanium® TLIF
    Manufacturer
    Date Cleared
    2018-05-04

    (73 days)

    Product Code
    Regulation Number
    888.3080
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    SpineVision, SAS

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Hexanium® TLIF (Transformational Lumbar Interbody Fusion) system is an intervertebral body fusion device indicated for use with autogenous bone graft in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two continuous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). Patients should have received at least 6 months of non-operative treatment prior to treatment with Hexanium® TLIF system. This device has to be filled with autogenous bone graft material. This device is implanted via transforaminal approach. Hexanium® TLIF system must be used in combination with supplemental internal spinal fixation which has been cleared by the FDA for use in the lumbar spine.

    Device Description

    The Hexanium® TLIF device is a titanium alloy (Ti6Al4V ELI) interbody cage manufactured via an Additive Manufacturing method. The honeycomb structure allows for bone through-growth through the structure of the device as well as providing lateral and vertical bone graft windows in the body of the cage. Hexanium® TLIF is available in heights of 7mm to 16mm in 1mm increments in a Small footprint (28mm X 9mm) or a Medium footprint (32mm X 10.5mm). Hexanium® TLIF has a 5° Lordosis angle. Hexanium® TLIF devices are provided sterile.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Hexanium® TLIF device:

    This document is a 510(k) Premarket Notification from the FDA for the Hexanium® TLIF device. It primarily focuses on demonstrating substantial equivalence to a predicate device rather than providing detailed acceptance criteria and a study report in the way a clinical trial or algorithm validation study would.

    Based on the provided text, the "acceptance criteria" and "study" are described in terms of mechanical and biological performance, aiming for equivalence to a predicate device, as typically required for a 510(k) submission for an intervertebral body fusion device.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria Category/TestAcceptance Criteria (Implicit from Device Guidance & Predicate Comparison)Reported Device Performance
    Mechanical Performance
    Static CompressionConforms to ASTM F2077-14; Comparable to predicate device.Performed according to ASTM F2077-14; Results demonstrate comparable mechanical properties to the predicate device.
    Dynamic CompressionConforms to ASTM F2077-14; Comparable to predicate device.Performed according to ASTM F2077-14; Results demonstrate comparable mechanical properties to the predicate device.
    Static Compression ShearConforms to ASTM F2077-14; Comparable to predicate device.Performed according to ASTM F2077-14; Results demonstrate comparable mechanical properties to the predicate device.
    Dynamic Compression ShearConforms to ASTM F2077-14; Comparable to predicate device.Performed according to ASTM F2077-14; Results demonstrate comparable mechanical properties to the predicate device.
    SubsidenceConforms to ASTM F2267-04; Comparable to predicate device.Performed according to ASTM F2267-04; Results demonstrate comparable mechanical properties to the predicate device.
    Expulsion TestingComparable to predicate device.Performed; Results demonstrate comparable mechanical properties to the predicate device.
    Biological Performance
    Bacterial Endotoxins TestMeets pyrogen limit specifications (in accordance with USP).Performed in accordance with USP; Device meets pyrogen limit specifications.
    General ConformanceConforms to "Class II Special Controls Guidance Document: Intervertebral Body Fusion Device Document issued on June 12, 2007."Device conforms to the specified guidance document.

    Explanation of the "Study":

    The "study" described here is a series of benchtop mechanical and biological tests designed to demonstrate that the Hexanium® TLIF device performs equivalently to its predicate device (SpaceVision® Interbody Fusion Cages, K153783) and meets relevant industry standards and FDA guidance for intervertebral body fusion devices.

    Missing Information (based on the provided text):

    The provided text is a 510(k) summary, which is a regulatory document and not a detailed study report. Therefore, several of the requested points are not applicable (N/A) or not explicitly stated in this type of document, especially those related to AI algorithm validation, human-reader studies, or clinical trials.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size for Test Set: Not specified in terms of number of devices/units tested for each mechanical test. Typically, these benchtop tests involve a small, statistically valid number of samples (e.g., 5-10 per test condition), but the exact number is not provided.
    • Data Provenance: The tests are described as being performed to ASTM standards and USP. This implies laboratory testing. The manufacturing entity, SpineVision SAS, is based in France, so the testing could have been conducted in France or at a qualified testing facility elsewhere. The document doesn't specify a "country of origin for data" in the context of patient data, as this is a benchtop study. The testing is prospective in the sense that it was conducted specifically to support this 510(k) submission, not retrospectively analyzing existing data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    • N/A. This is a mechanical/biological benchtop study, not a study involving interpretation of medical images or clinical outcomes by human experts to establish ground truth. The "ground truth" for these tests is defined by the ASTM standards and USP specifications.

    4. Adjudication Method for the Test Set:

    • N/A. As it's a benchtop physical and biological test comparing to standards and a predicate, there's no "adjudication method" in the sense of expert consensus for interpretations. The results are quantitative measurements against defined criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    • No. An MRMC comparative effectiveness study is typically performed for diagnostic devices, especially those involving image interpretation or AI algorithms, to assess inter-reader variability and the impact of the device on human performance. This is a spinal implant for fusion, and no such study is mentioned or required for this type of device in a 510(k) context.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • No. This device is a physical medical implant, not an AI algorithm. Therefore, "standalone algorithm performance" is not applicable.

    7. The Type of Ground Truth Used:

    • For mechanical testing: Industry Standards (ASTM F2077-14, ASTM F2267-04) and Predicate Device Performance (side-by-side comparison).
    • For biological testing: Pharmacopoeial Standards (USP for Bacterial Endotoxins Test).

    8. The Sample Size for the Training Set:

    • N/A. This study does not involve a "training set" as it is not an AI/machine learning model.

    9. How the Ground Truth for the Training Set was Established:

    • N/A. This study does not involve a "training set" or "ground truth for a training set."

    In summary, the provided document details the regulatory pathway for a conventional medical device (spinal implant) through a 510(k) submission. The "acceptance criteria" and "study" revolve around demonstrating mechanical and biological equivalence to a previously cleared predicate device and compliance with established industry standards through benchtop testing, rather than complex clinical or AI-based performance studies.