Search Results
Found 4 results
510(k) Data Aggregation
(117 days)
The ViSI Mobile Monitoring System is intended for use by clinicians and medically qualified personnel for single or multi-parameter vital signs monitoring of adult patients (18 years or older). It is indicated for ECG (3 or 5 lead-wire), respiration rate (RESP), heart rate (HR), noninvasive blood pressure (NIBP), continuous noninvasive blood pressure (cNIBP), noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate (PR), skin temperature (TEMP), posture tracking and basic arthythmia (Ventricular Tachycardia, Ventricular Fibrillation, Asystole, Atrial Fibrillation) analysis and alarm in hospital-based facilities including general medical-surgical floors, intermediate care floors, and emergency departments.
Continuous non-invasive blood pressure (cNIBP) measurements have not been evaluated on patients during ambulation. The basic arrhythmia analysis feature is intended for use on patients 18 years of age and older; it has not been evaluated on pediatric patients.
The arrhythmia analysis feature is intended for use by healthcare professionals trained in the identification and treatment of arrhythmia events. Automated arrhythmia analysis is an adjunct to clinical assessment; clinician review of the analysis should precede any therapeutic intervention.
The ViSi Mobile Monitoring System may be used as standalone devices or networked to ViSi Mobile Remote Viewers through wireless 802.11 communication.
The Visl Mobile Insight is an optional secondary notification system that communicates alarms directly to an assigned caregiver. It is intended to supplement the primary alarming devices which originate in the ViSI Mobile patient-worn device.
The ViSi Mobile Monitoring System is a patient worn, portable, battery operated, continuous physiological monitoring device intended for the monitoring of ECG (3 or 5 lead-wire), respiration rate (RESP), heart rate (HR), non-invasive blood pressure (NIBP), continuous non-invasive blood pressure (cNIBP), non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate (PR), skin temperature (SKIN TEMP), posture tracking and alarms, basic arrhythmia analysis (ventricular fibrillation, ventricular tachycardia, asystole, atrial fibrillation) and alarms.
The ViSi Mobile Monitoring System consists of the patient worn devices, disposables, backup battery, patient data server and remote viewer.
This document (K180472) is a 510(k) premarket notification for the ViSi Mobile Monitoring System. The focus of the provided text is to demonstrate substantial equivalence to a predicate device, primarily by highlighting changes and comparing technical characteristics and performance, especially concerning arrhythmia detection.
Based on the provided text, here's a detailed breakdown of the acceptance criteria and the study proving the device meets them:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly present a "table of acceptance criteria" in the traditional sense for the acceptance of the device as a whole. Instead, it presents performance metrics against recognized standards (ANSI/AAMI EC57: 2012) for its arrhythmia detection algorithm, comparing the improved "Proposed Device" (Subject Device) with a "Reference Device A" (K142827), which is a previously cleared version of the ViSi Mobile Monitoring System.
Here's a table derived from the provided "C. Atrial Fibrillation Comparison" and "E. ALARM TEST COMPARISON" sections, focusing on the arrhythmia detection performance:
| Metric | Acceptance Criteria (Implicitly, the performance of the Reference Device K142827 or general industry standards like ANSI/AAMI EC57:2012) | Reported Device Performance (Proposed Device K180472) | Commentary (from document) |
|---|---|---|---|
| AF Episode Sensitivity (%) | RA (Reference Device A - K142827): | S (Subject Device - K180472): | |
| NST | 0‡ | - | Null result (test done, but statistic cannot be calculated due to absence of test or reference annotation) |
| MIT-BIH Gross | 90 | 87 | Same or Better |
| MIT-BIH Average | 92 | 87 | |
| AF Episode Positive Predictivity (%) | RA (Reference Device A - K142827): | S (Subject Device - K180472): | |
| NST | 0‡ | - | Null result |
| MIT-BIH Gross | 100 | 65 | Same or Better |
| MIT-BIH Average | 100 | 99 | |
| AF Duration Sensitivity (%) | RA (Reference Device A - K142827): | S (Subject Device - K180472): | |
| NST | ‡ | - | Null result |
| MIT-BIH Gross | 75 | 80 | Same or Better |
| MIT-BIH Average | 71 | 84 | |
| AF Duration Positive Predictivity (%) | RA (Reference Device A - K142827): | S (Subject Device - K180472): | |
| NST | 0‡ | 0 | Same or Better |
| MIT-BIH Gross | 0 | 75 | |
| MIT-BIH Average | 0 | 40 | |
| AF False Positive Report | RA (Reference Device A - K142827): | S (Subject Device - K180472): | |
| NST | 1‡ | 0 | Same or Better |
| MIT-BIH Gross | 0 | 137 | |
| MIT-BIH Average | 0 | ||
| AF False Negative Report | RA (Reference Device A - K142827): | S (Subject Device - K180472): | |
| NST | 0‡ | 0 | Same or Better |
| MT-BIH Gross | 0 | 10 | |
| MIT-BIH Average | 0 | ||
| AF Time to Detection (sec) | RA (Reference Device A - K142827): | S (Subject Device - K180472): | |
| NST | - | - | Null result |
| MIT-BIH Gross | 00:20.94 | 0:06.4 | Same or Better |
| MIT-BIH Average | |||
| VF/VT Sensitivity (%) | Reference Device A (K142827) | Subject Device (K180472) | |
| AHA | 90 | 91 | |
| MIT | 100 | 100 | |
| CU | 94 | 97 | |
| VF/VT Positive Predictivity (%) | Reference Device A (K142827) | Subject Device (K180472) | |
| AHA | 60 | 100 | |
| MIT | 67 | 100 | |
| CU | 46 | 94 | |
| AFIB Sensitivity (%) | Reference Device A (K142827) | Subject Device (K180472) | |
| MIT | 97 | 90 | |
| AFIB Positive Predictivity (%) | Reference Device A (K142827) | Subject Device (K180472) | |
| MIT | 65 | 100 |
The study that proves the device meets the acceptance criteria is primarily non-clinical, using established databases.
2. Sample size used for the test set and the data provenance
The test set utilizes publicly available, established databases:
- AHA (American Heart Association)
- MIT-BIH (Massachusetts Institute of Technology - Beth Israel Hospital Arrhythmia Database)
- CU (Creighton University Ventricular Tachyarrhythmia Database)
- NST
The specific sample sizes (e.g., number of ECG recordings or patients) from these databases are not explicitly stated in the provided text. The provenance is from these known databases, which are widely accepted for arrhythmia algorithm testing. The data within these databases is retrospective.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
The document states: "The arrhythmia analysis is based on the same technology as the currently cleared Monebo Technologies, Inc Arrhythmia library (K062282), which has been adapted for real time analysis in the ViSi System. The ViSi Monitoring System arrhythmia analysis has been validated by comparison to the AHA, MIT-BIH, CU, and NST databases as prescribed in ANSI/AAMI EC57: 2012."
The ground truth for these databases (AHA, MIT-BIH, CU, NST) was established by experts, typically cardiologists or electrophysiologists, during the creation of the databases. The specific number and qualifications of these experts are not detailed within this 510(k) document but are inherent to the accepted nature of these standard databases.
4. Adjudication method for the test set
The document does not describe an adjudication method for the test set data itself. The ground truth for the test sets (AHA, MIT-BIH, CU, NST databases) was established independently by the creators of those databases.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No. A multi-reader multi-case (MRMC) comparative effectiveness study involving human readers is not described in this document. The study focuses purely on the algorithm's performance against established gold-standard databases, comparing an updated algorithm to a previous version and recognized standards. The device is not presented as an AI-assistance tool for human readers but rather as a standalone monitoring system with an integrated arrhythmia detection algorithm.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Yes. The performance evaluation presented is a standalone (algorithm only) performance assessment. The tables show the algorithm's sensitivity and positive predictivity for various arrhythmia events (VF/VT, AFIB) when processing the specified ECG databases. The context further reinforces this, stating "Automated arrhythmia analysis is an adjunct to clinical assessment; clinician review of the analysis should precede any therapeutic intervention," indicating it acts as an aid rather than requiring human modification of its output for the test.
7. The type of ground truth used
The ground truth used for the performance testing is expert consensus from established, recognized ECG databases (AHA, MIT-BIH, CU, NST). These databases contain expert-annotated ECG waveforms, which serve as the gold standard for evaluating arrhythmia detection algorithms.
8. The sample size for the training set
The document does not explicitly state the sample size of the training set used to develop or refine the arrhythmia detection algorithm. It mentions that "The arrhythmia analysis is based on the same technology as the currently cleared Monebo Technologies, Inc Arrhythmia library (K062282)," which implies that the training might have occurred prior to this submission (or the core technology was pre-trained).
9. How the ground truth for the training set was established
The document does not explicitly detail how the ground truth for the training set was established. Given that the algorithm is based on (or an adaptation of) the Monebo Technologies, Inc Arrhythmia library, it's highly probable that the training also utilized similar expert-annotated ECG datasets or proprietary clinical data. The document does state that the arrhythmia algorithm was "Rewritten to improve specificity and detection metrics" and later "New arrhythmia detection algorithm to improve arrhythmia detection accuracy and positive predictivity," implying development and testing that would have involved ground truth data.
Ask a specific question about this device
(293 days)
The ViSi Mobile Monitoring System is intended for use by clinicians and medically qualified personnel for single or multi-parameter vital signs monitoring of adult patients (18 years or older). It is indicated for ECG (3 or 5 lead-wire), respiration rate, noninvasive blood pressure (NIBP), continuous non-invasive blood pressure (cNIBP), non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, basic arrhythmia analysis (Ventricular Tachycardia, Ventricular Fibrillation, Asystole, Atrial Flutter) and alarm in hospital-based facilities; including general medical-surgical floors, intermediate care floors, and emergency departments.
Continuous non-invasive blood pressure (cNIBP) measurements have not been evaluated on patients during ambulation.
The basic arrhythmia analysis feature is intended for use on patients 18 years of age and older. It has not been evaluated on pediativ patients or neonates.
The arrhythmia analysis feature is intended for use by healthcation and treatification and treatment of arrhythmia events. Automated arrhythmia analysis is an adjunct to clinician review of the analysis should precede any therapeutic intervention.
The ViSi Mobile Monitoring System may be used as standalone devices or networked to a ViSi Mobile Remote Viewer through wireless 802.11 communication.
The ViSi Mobile Monitoring System is a lightweight, body-worn vital signs monitor featuring a high resolution, full color touch screen display, with visual and audible alarms and alerts. The ViSi Mobile Monitor is designed to continuously non-invasively measure ECG, basic arrhythmias [including ventricular tachycardia, ventricular fibrillation, asystole, and atrial fibrillation/atrial flutter], heart rate, SpO2, blood pressure, pulse rate, respiration rate, and temperature. The ECG, Sp02, and Respiration waveforms are viewable on demand. The ViSi Mobile Monitoring System is capable of one-time and continuous NIBP measurements.
The ViSi Mobile Monitoring System for basic arrhythmia analysis (Ventricular Tachycardia, Ventricular Fibrillation, Asystole, Atrial Fibrillation/Atrial Flutter) includes the following acceptance criteria and study information:
1. Table of Acceptance Criteria and Reported Device Performance
| Parameter | Acceptance Criteria (ANSI/AAMI EC57: 2012) | Reported Device Performance (ViSi Mobile Monitoring System) |
|---|---|---|
| QRS Detection | (Not explicitly defined in the provided text, but implied as being comparable to predicate) | Comparable to predicate device |
| Arrhythmia Analysis | Based on ANSI/AAMI EC57: 2012 | Validated by comparison to AHA, MIT-BIH, CU, and NST databases |
| False Alarms | Minimized by artifact and unclassified rhythm detection | Artifact detection prevents noisy signals from being detected as beats; Unclassified rhythm for unclassifiable rhythms (>30s triggers technical alarm) |
Note: The document specifies that the basic arrhythmia analysis is based on the same technology as the currently cleared Monebo Technologies, Inc Arrhythmia library (K062282), which was adapted for real-time analysis. It also states the device has been validated by comparison to specific databases as prescribed by ANSI/AAMI EC57: 2012.
2. Sample Size Used for the Test Set and Data Provenance
The test set included data from the following databases:
- AHA (American Heart Association)
- MIT-BIH Arrhythmia Database
- CU (Creighton University Arrhythmia Database)
- NST (Northrop-Swales Transient Database)
The document does not detail the specific sample size (number of recordings or patients) used from each database, nor does it explicitly state the country of origin or whether the data was retrospective or prospective. However, these are established, publicly available retrospective databases commonly used for arrhythmia algorithm validation.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
The ground truth for the AHA, MIT-BIH, CU, and NST databases was established by expert consensus during the creation of these databases. The specific number and qualifications of experts involved in the original annotation of these databases are not provided in this document but are well-documented within the respective database literature.
4. Adjudication Method for the Test Set
The document does not explicitly describe an adjudication method for the test set beyond relying on the established ground truth of the referenced databases (AHA, MIT-BIH, CU, NST). These databases typically feature annotations derived from multiple expert reviews.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No Multi-Reader Multi-Case (MRMC) comparative effectiveness study is mentioned in the provided text, nor is there any effect size of human readers improving with AI vs without AI assistance. The study described focuses on standalone algorithm performance against established databases.
6. Standalone Performance Study
Yes, a standalone (algorithm only without human-in-the-loop performance) study was done. The ViSi Mobile Monitoring System's basic arrhythmia analysis was validated by comparison to the AHA, MIT-BIH, CU, and NST databases as prescribed in ANSI/AAMI EC57: 2012. The QRS detection sensitivity of the ViSi System was reported as comparable to the predicate device.
7. Type of Ground Truth Used
The type of ground truth used was expert consensus based on the established annotations within the AHA, MIT-BIH, CU, and NST databases.
8. Sample Size for the Training Set
The document does not provide specific details on the sample size used for the training set. It states that the basic arrhythmia analysis is based on the same technology as a previously cleared device (Monebo Technologies, Inc Arrhythmia library, K062282). This implies that a substantial training set was likely used during the development of the original algorithm, and possibly for adaptive training, but no specific numbers are given for the ViSi Mobile's training.
9. How the Ground Truth for the Training Set Was Established
Similarly, the document does not explicitly state how the ground truth for the training set was established. Given that the algorithm is based on a previously cleared library, it is highly probable that the training data and its ground truth were established through similar expert-annotated ECG datasets as those used for validation (e.g., subsets or similar types of data to the AHA, MIT-BIH databases).
Ask a specific question about this device
(77 days)
The ViSi Mobile Monitoring System is intended for use by clinicians and medically qualified personnel for single or multi-parameter vital signs monitoring of adult patients (18 years or older). It is indicated for ECG (3 or 5 lead-wire), respiration rate (RESP), heart rate (HR), noninvasive blood pressure (NIBP), continuous noninvasive blood pressure (cNIBP), noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate (PR), and skin temperature (TEMP) in hospital-based facilities; including, general medical-surgical floors, intermediate care floors, and emergency departments.
The ViSi Mobile Monitoring System may be used as standalone devices or networked to ViSi Mobile Remote Viewers through wireless 802.11 communication.
The ViSi Mobile Monitoring System is a lightweight, body-worn vital signs monitor featuring a high resolution, full color touch screen display, with visual and audible alarms and alerts. The ViSi Mobile Monitor is designed to continuously non-invasively measure ECG, heart rate, SpO2, blood pressure, pulse rate, respiration rate, and temperature. The ECG, SpO2, and Respiration waveforms are viewable on demand. The ViSi Mobile Monitoring System is capable of one-time and continuous NIBP measurements.
This document describes the ViSi Mobile Monitoring System, a vital signs monitor. However, the provided text does not contain the detailed study information required to fully answer all aspects of your request regarding acceptance criteria and performance study specifics. Specifically, it lacks quantitative performance data, sample sizes for test sets, details on expert panels, adjudication methods, or MRMC study results.
Based on the available information, here's what can be extracted and inferred:
1. A table of acceptance criteria and the reported device performance
The document mentions that acceptance criteria were met, but it does not provide specific numerical criteria or reported device performance metrics in a table format. It states: "The results demonstrated that all acceptance criteria were met, and therefore conforms to expected device performance and intended use."
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not specify the sample size used for the test set, nor does it provide information on data provenance (country of origin, retrospective or prospective).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
The document does not mention the number or qualifications of experts used to establish ground truth for the test set.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
The document does not describe any adjudication method for the test set.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
The document does not mention a multi-reader multi-case (MRMC) comparative effectiveness study. This device is a vital signs monitor, not an AI-assisted diagnostic tool for human readers, so such a study would likely not be applicable in this context. The modifications are related to posture tracking and associated alarms.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
The document implies that standalone performance testing of the algorithms was conducted, specifically:
- "SVT-000061 - Walking Algorithm Software Verification"
- "SVT-000062 Undesirable Posture, Immobility and Fall Detection . Software verification"
- "TP-670 – Undesirable Posture Alarms Validation"
- "TP-671 Immobility Alarms Validation"
- "TP-672 Patient Fall Alarm Validation"
- ". TP-673 - Posture Definitions and Walking Validation"
These tests suggest evaluating the algorithms' accuracy in detecting posture, immobility, and falls. However, the specific metrics or acceptance criteria for these standalone tests are not provided in the summary.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The document does not explicitly state the type of ground truth used for the posture, immobility, and fall detection features. Given the nature of these features, ground truth would likely involve:
- Direct observation: Human observers physically verifying the patient's posture, movement, or fall event.
- Reference sensors/equipment: Other validated sensors (e.g., accelerometers, video analysis) used as a gold standard to establish true states.
8. The sample size for the training set
The document does not provide any information regarding the sample size for the training set.
9. How the ground truth for the training set was established
The document does not provide any information on how the ground truth for the training set was established.
Summary of available information regarding acceptance criteria and performance study:
The device under review is the ViSi Mobile Monitoring System, which added new posture alarm features (undesirable posture, immobility, patient fall) and the ability to display if a patient is walking.
Acceptance Criteria and Reported Performance:
- Acceptance Criteria for new features: The document generally states that "all acceptance criteria were met." However, the specific, quantitative acceptance criteria for each new feature (undesirable posture alarm, immobility alarm, patient fall alarm, walking detection) are not detailed.
- Reported Device Performance: No specific performance metrics (e.g., sensitivity, specificity, accuracy, precision) are provided for these features within this document. The statement is qualitative: "The results demonstrated that all acceptance criteria were met, and therefore conforms to expected device performance and intended use."
Study Information (Inferred/Partial):
- Type of Study: Non-clinical performance (verification and validation) testing.
- Test Set Sample Size: Not specified.
- Data Provenance: Not specified.
- Experts for Ground Truth: Not specified.
- Adjudication Method: Not specified.
- MRMC Study: Not applicable/not performed for this type of device modification.
- Standalone Performance: Yes, algorithm-only performance was evaluated through various software verification and validation tests related to walking, undesirable posture, immobility, and fall detection.
- Ground Truth Type: Not explicitly stated but likely involved direct observation or comparison with reference sensors for posture and movement.
- Training Set Sample Size: Not specified.
- Training Set Ground Truth Establishment: Not specified.
In conclusion, while the document confirms that verification and validation testing was conducted and acceptance criteria were met for the new posture features, it lacks the granular detail about the studies themselves, including specific performance metrics, sample sizes, and ground truth methodologies.
Ask a specific question about this device
(23 days)
The ViSi Mobile Monitoring System is intended for use by clinicians and medically qualified personnel for single or multi-parameter vital signs monitoring of adult patients (18 years or older). It is indicated for ECG (3 or 5 lead-wire), respiration rate (RESP), heart rate (HR), non-invasive blood pressure (NIBP), continuous non-invasive blood pressure (cNIBP), non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate (PR), and skin temperature (TEMP) in hospital-based facilities; including, general medical-surgical floors, intermediate care floors, and emergency departments.
The ViSi Mobile Monitoring System may be used as standalone devices or networked to ViSi Mobile Remote Viewers through wireless 802.11 communication.
The ViSi Mobile Monitoring System is a lightweight, body-worn vital signs monitor featuring a high resolution, full color touch screen display, with visual and audible alarms and alerts. The ViSi Mobile Monitor is designed to continuously non-invasively measure ECG, heart rate, SpO2, blood pressure, pulse rate, respiration rate, and temperature. The ECG, SpO2, and Respiration waveforms are viewable on demand. The ViSi Mobile Monitoring System is capable of one-time and continuous NIBP measurements.
This document is a 510(k) summary for the ViSi Mobile Monitoring System. It describes the device, its intended use, and provides a summary of non-clinical performance testing conducted to demonstrate substantial equivalence to previously cleared predicate devices.
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state formal acceptance criteria in a quantitative manner (e.g., "sensitivity must be > X%"). Instead, it describes performance in terms of improvements or conformance to standards.
| Metric / Standard | Acceptance Criteria (Implicit) | Reported Device Performance |
|---|---|---|
| QRS Detection | Equal to or better performance than prior beat-picker | MIT Database: Equal to or better performance in gross Q sensitivity. Negligible reduction (no specific value given) in gross Q positive predictivity. |
| AHA Database: Equal to or better performance in gross Q sensitivity and gross Q positive predictivity. | ||
| Overall improvement in performance compared to existing algorithm | Determined to be an improvement in performance over the existing algorithm. | |
| IEC 60601-2-27 | Conformance to the standard | Conformance to IEC 60601-2-27 was demonstrated. |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
- Sample Size for Test Set: The document mentions the use of the "MIT and AHA databases." These are well-known, publicly available benchmark databases for ECG analysis. The specific number of records or patients from these databases used for testing is not explicitly stated in this summary.
- Data Provenance: The MIT and AHA databases are standard, established datasets. The country of origin and whether the data is retrospective or prospective is not specified in this document but is inherent to the nature of these established benchmark databases (typically retrospective and collected over time from various sources).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
The document does not provide information on the number or qualifications of experts used to establish the ground truth for the MIT and AHA databases. For these widely recognized benchmark databases, the ground truth is typically meticulously annotated by multiple qualified experts over many years.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
The document does not describe the adjudication method used for establishing the ground truth of the MIT and AHA databases. This information is typically detailed in the documentation accompanying those specific databases.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
There is no mention of a multi-reader multi-case (MRMC) comparative effectiveness study, nor any evaluation of human reader improvement with or without AI assistance. The study described focuses on the standalone performance of the QRS detection algorithm against benchmark databases.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, a standalone study was done. The performance testing described, which involved the "new ECG beat-picker" against the MIT and AHA databases, is a standalone evaluation of the algorithm's performance without human-in-the-loop involvement.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The ground truth for the MIT and AHA databases, which were used to evaluate QRS detection, is based on expert consensus annotations of the ECG waveforms.
8. The sample size for the training set
The document does not provide information regarding the sample size of the training set used for the development or training of the new ECG beat-picker algorithm.
9. How the ground truth for the training set was established
The document does not provide information on how the ground truth for the training set was established.
Ask a specific question about this device
Page 1 of 1