Search Results
Found 4 results
510(k) Data Aggregation
(107 days)
Smartdata Suzhou Co., Ltd
The Vacutore® Air/Water Bottle Tubing (Model GAR025, GAR076) is to connect an air/CO2 source, a sterile water source (water bottle), and an endoscope to supply air/CO2 and water during gastrointestinal endoscopic procedures. It is a 24-hour multi-patient use device.
The Vacutore® Air/Water Bottle Tubing is intended for 24-hour multi-patient use. Air/water bottle tubing is supplied in sterile. The Vacutore® Air/Water Bottle Tubing is manufactured for use in conjunction with sterile water bottle, and together with Olympus 160 and Pentax 90 series endoscopes. The air/water bottle tubing is individually packed in sealed package, sold as a sterile device. The air/water bottle tubing is designed to be attached to the air/water port of the endoscopes to provide irrigation through the air/water channel to the distal end of endoscope.
The provided document is a 510(k) Premarket Notification for a medical device called "Vacutore® Air/Water Bottle Tubing." This type of submission focuses on demonstrating substantial equivalence to a predicate device, rather than proving novel effectiveness through a comprehensive study with specific acceptance criteria related to a diagnostic output or algorithm performance.
Therefore, the typical metrics for acceptance criteria and study descriptions (like sensitivity, specificity, types of ground truth, number of experts, MRMC studies, effect sizes, and separate test/training sets with explicit ground truth establishment) are generally not applicable in the context of this device and submission type.
Instead, the acceptance criteria here relate to the performance, safety, and functional equivalence of the tubing device compared to an existing predicate.
Here's an analysis based on the information provided, addressing the questions as much as possible for this type of device:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state numerical "acceptance criteria" and "reported device performance" in the way one would for an AI diagnostic device (e.g., "sensitivity > 90%"). Instead, it lists tests performed and states that these tests support substantial equivalence. The "performance" is implicitly "met all tests successfully" to be considered substantially equivalent.
Acceptance Criterion (Implicit) | Reported Device Performance (Implicit) |
---|---|
Assembling Integrity | Passed |
Endoscope Compatibility | Passed |
Compatibility with Bottle | Passed |
Two-Way Valve Integrity | Passed |
Flow Clamp Test | Passed |
Water Flow Test | Passed |
Back Flow Performance Test | Passed |
Sterility (SAL of 10⁻⁶) | Passed (Validated EO sterilization) |
EO residuals (ANSI/AAMI/ISO 10993-7) | Below maximum levels |
Shelf Life (3 years) | Passed (Verified by accelerated aging, seal strength, dye penetration, microbial barrier, vacuum leak, sterility tests) |
Biocompatibility (ISO 10993-1) | Passed (Acute Systemic Toxicity, In Vitro Cytotoxicity, Skin sensitization, Irritation tests showed biocompatibility) |
Barrier Integrity & Simulated Use Testing (24-hour multi-patient use) | Passed (Verified back-flow prevention and 24-hour multi-patient use) |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not specify a "sample size" in terms of number of patients or images, as this is a tubing device. The testing was performed on "Smartdata samples from initial production lots." The provenance of these samples would be the manufacturer, Smartdata Suzhou Co., Ltd., in China. These would be considered prospective in the sense they are new production samples being tested.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This is not applicable. The device is a piece of medical tubing, not a diagnostic or AI-driven decision support system that requires expert interpretation for ground truth. The "ground truth" here is physical performance (e.g., does it leak? does it flow water correctly? is it sterile?).
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This is not applicable. Adjudication methods are used to establish ground truth in studies involving human interpretation or clinical outcomes. For bench testing of a physical device, the results are typically objectively measured and do not require expert adjudication in this manner.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No. An MRMC study is relevant for diagnostic or AI-assisted interpretation, where human readers evaluate cases with and without AI assistance. This device is a piece of tubing used in endoscopic procedures, not an AI diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
No. This refers to the performance of an algorithm without human input, which is not relevant for this device. The testing described is "standalone" in the sense that the tubing itself was tested for its physical and functional properties.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for this device's performance is based on objective measurement of physical properties and adherence to established standards. Examples include:
- Physical integrity (no leaks, secure connections)
- Flow rates (water flow test)
- Sterility (Sterility Assurance Level, EO residuals)
- Material compatibility (biocompatibility tests)
- Durability (shelf life accelerated aging)
- Functional performance (back-flow prevention, compatibility with endoscopes/bottles).
8. The sample size for the training set
Not applicable. This device does not involve a "training set" in the context of machine learning or AI models.
9. How the ground truth for the training set was established
Not applicable, as there is no training set.
Ask a specific question about this device
Smartdata Suzhou Co., Ltd
The Andorate Auxiliary Water Connector (Model GAR048) is used in conjunction with irrigation tubing (not supplied), intended to provide irrigation fluids such as sterile water supplied to the Pentax 90 series endoscope during gastrointestinal endoscopic procedures when used in conjunction with an irrigation pump.
The disposable Andorate Valves Set (Model GAR037) consists of one suction valve, one air/water valve and one biopsy valve.
-
The Andorate Disposable Suction Valve is used to control the suction of an endoscope (for Pentax 90 series Endoscope) during a GI Endoscopic procedure.
-
The Andorate Disposable Air/Water valve is used to control the air/water function of an endoscope (for Pentax 90 series Endoscope) during a GI Endoscopic procedure.
-
The Andorate Disposable biopsy valve is used to cover the endoscope biopsy port during an endoscopy procedure. In addition, the valve provides access for endoscopic device passage and exchange, helps maintain insufflation and minimizes leakage of biomaterial from the biopsy port throughout the endoscopic procedure.
The Andorate® Valves Set and Andorate® Auxiliary Water Connector are intended for single-use and are supplied sterile. Single-use valves sets and auxiliary water connector help preventing potential safety risks and eliminate the need for manual cleaning and reprocessing. The valves set and auxiliary water connector are easily incorporated into infection prevention policies as a single use item.
The Andorate® Auxiliary Water Connector is manufactured for use in conjunction with irrigation tubing, and together with Pentax 90 series endoscope. The auxiliary water connector is individually packed in sealed package, sold as a sterile device. The auxiliary water connector is designed to the auxiliary water port of the endoscopes. The auxiliary water connector consists of a backflow valve which prevent the backflow of water or biomaterials from the endoscope to the sterile water bottle.
The Andorate® Valves Set is manufactured for use with Pentax 90 series endoscope. The suction, air/water, and biopsy valves are housed in a single tray and packaged in sealed packed. The valves set is sold as a sterile device.
- The Suction Valve component of the Andorate® Valves Set is designed to o be attached to the suction port of the endoscope and the air/water valve is designed to be attached to the air/ water port of the endoscope. The activation of the suction valve allows the user to aspirate excess fluids or other debris obscuring the endoscopic image, while the activation of the air/ water valve allows the user to control air and water flow to assist in cleansing the lens during procedures.
- The Biopsy Valve component of the Andorate® Valves Set is intended to O cover the endoscope biopsy port during an endoscopy procedure. In addition, the valve provides access for endoscopic device passage and exchange, helps maintain insufflation and minimizes leakage of biomaterial from the biopsy port throughout the endoscopic procedure.
The subject devices in this submission have the same operation and method of action as the predicate devices. In the submission, Andorate® Valves Set and Andorate® Auxiliary Water Connector are the only subject device. This submission does not include irrigation tubing or irrigation pump.
The provided document describes the K191231 submission for the Andorate® Valves Set (GAR037) and Andorate® Auxiliary Water Connector (GAR048). This is a 510(k) summary, which focuses on demonstrating substantial equivalence to predicate devices rather than providing detailed acceptance criteria and performance data for a novel device. Therefore, much of the requested information regarding detailed acceptance criteria, specific device performance metrics against those criteria, and information about expert consensus, training sets, and MRMC studies is not available in this document as it's typically required for software-as-a-medical-device (SaMD) or AI-driven diagnostic devices.
However, based on the information provided, here's what can be extracted:
1. Table of Acceptance Criteria and Reported Device Performance
The document doesn't provide a table of quantitative acceptance criteria with corresponding performance metrics for the proposed devices in the way an AI/SaMD 510(k) summary would. Instead, it lists the types of non-clinical performance tests conducted to demonstrate substantial equivalence to predicate devices, focusing on functional and safety aspects. The reported performance is generally stated as having met the requirements, implying equivalence to the predicates.
Test Category | Acceptance Criteria (Implied) | Reported Device Performance |
---|---|---|
Andorate® Valves Set (GAR037) | ||
Assembling Integrity (Suction Valve, Biopsy Valve) | Maintain structural integrity upon assembly. | Testing performed; implicit compliance. |
Endoscope Compatibility (All Valves) | Compatible with Pentax 90 series Endoscopes. | Testing performed; implicit compliance. |
Depression Force (Suction Valve, Air/Water Valve) | Operate within acceptable force ranges. | Testing performed; implicit compliance. |
Vacuum Leak Test (Suction Valve, Biopsy Valve) | No significant vacuum leakage. | Testing performed; implicit compliance. |
Suction Flow Test (Suction Valve) | Achieve adequate suction flow. | Testing performed; implicit compliance. |
Air Leakage Testing (Air/Water Valve) | No significant air leakage. | Testing performed; implicit compliance. |
Water Flow Test (Air/Water Valve) | Achieve adequate water flow. | Testing performed; implicit compliance. |
Squeegee Leak Test (Biopsy Valve) | No significant leakage during device passage/exchange. | Testing performed; implicit compliance. |
Andorate® Auxiliary Water Connector (GAR048) | ||
Compatibility with Irrigation Tubing | Compatible with standard irrigation tubing. | Testing performed; implicit compliance. |
Compatibility with Endoscope | Compatible with Pentax 90 series Endoscopes. | Testing performed; implicit compliance. |
Water Flow Test | Achieve adequate water flow. | Testing performed; implicit compliance. |
Air Leakage Test | No significant air leakage. | Testing performed; implicit compliance. |
Water Leakage Test | No significant water leakage. | Testing performed; implicit compliance. |
Backflow Performance Test | Prevent backflow of water/biomaterials. (Explicitly mentioned: "prevent the backflow of water or biomaterials from the endoscope to the sterile water bottle.") | Testing performed; implicit compliance. |
Sterilization | SAL of 10^-6 for EO sterilization. EO residuals below ANSI/AAMI/ISO 10993-7. | Achieved SAL of 10^-6. EO residuals below maximum levels. |
Shelf Life | Maintain sterility and intended performance for 3 years. | Demonstrated through accelerated aging, packaging integrity tests. |
Biocompatibility | Meet ISO 10993 guidelines for mucosal membrane contact. | Tests (Systemic Toxicity, Cytotoxicity, Skin Sensitization, Irritation) show biocompatibility. |
2. Sample Size for the Test Set and Data Provenance
- Sample Size of Test Set: The document does not specify exact sample sizes for each non-clinical performance test. It states that "Smartdata performed bench testing to support substantial equivalence. The following testing were performed on Smartdata samples from initial production lots, including sterilization." This implies a representative sample was tested, but no specific number is given.
- Data Provenance: The tests were performed on "Smartdata samples from initial production lots," which would originate from Smartdata Suzhou Co., Ltd. in China. The data is retrospective in the sense that it was collected from manufactured devices for the purpose of the 510(k) submission.
3. Number of Experts Used to Establish Ground Truth and Qualifications
- This information is not applicable and not provided. The devices are physical medical instruments (valves and connectors), not diagnostic or AI-driven devices that would require expert consensus for ground truth. Performance was evaluated through objective bench testing.
4. Adjudication Method for the Test Set
- This is not applicable as the tests performed are objective bench tests, not subjective interpretations requiring adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is typically conducted for diagnostic devices (especially those involving image interpretation by human readers) to assess the impact of AI assistance on reader performance. The devices in this submission are accessories for endoscopes, not diagnostic tools.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
- No, a standalone (algorithm only) study was not done. This is relevant for AI algorithms. The devices are mechanical accessories, not algorithms.
7. Type of Ground Truth Used
- The "ground truth" for the performance tests was defined by established engineering and medical device standards, functional requirements, and comparison to the predicate devices. For example, for a "Vacuum Leak Test," the ground truth is simply the absence of leakage beyond a specified threshold, determined by the test method. For biocompatibility, it's compliance with ISO 10993 standards.
8. Sample Size for the Training Set
- Not applicable. These are physical hardware devices; there is no "training set" in the context of an AI/ML algorithm. Their design and manufacturing processes are developed through traditional engineering methods and validated through bench testing.
9. How the Ground Truth for the Training Set Was Established
- Not applicable as there is no training set for these devices.
Ask a specific question about this device
(169 days)
Smartdata Suzhou Co., Ltd
The Andorate™ Auxiliary Water Connector (Model GAR027) is used in conjunction with irrigation tubing (not supplied), intended to provide irrigation fluids such as sterile water supplied to the Olympus 160 series endoscope during gastrointestinal endoscopic procedures when used in conjunction with an irrigation pump.
The Andorate™ Auxiliary Water Connector is intended for single-use and are supplied sterile. Single-use auxiliary water connector help prevential safety risks and eliminate the need for manual cleaning and reprocessing. The auxiliary water connector easily incorporate into infection prevention policies as a single use item. The Andorate™ Auxiliary Water Connector are manufactured for use in conjunction with irrigation tubing, and together with Olympus 160 series endoscopes. The auxiliary water connector is individually packed in sealed package, sold as a sterile device. In the submission, Andorate™ Auxiliary Water Connector is the only subject device. This submission does not include irrigation tubing or irrigation pump. The auxiliary water connector are designed to the auxiliary water port of the endoscopes. The auxiliary water connector consist of a backflow valve which prevent the backflow of water or biomaterials from the endoscope to the sterile water bottle.
The provided document (K182998) is a 510(k) summary for an auxiliary water connector for endoscopes. This type of device is a physical accessory and not a software algorithm or AI-driven system. Therefore, the questions related to AI/algorithm performance, such as sensitivity, specificity, MRMC studies, ground truth establishment for training/test sets, and sample sizes for training sets, are not applicable to this submission.
The acceptance criteria and proof of device performance (rather than algorithm performance) are based on non-clinical bench testing demonstrating functional equivalence and safety compared to a predicate device.
Here's a summary of the relevant information from the document:
1. Table of Acceptance Criteria and Reported Device Performance
For this medical device (Andorate™ Auxiliary Water Connector), the "acceptance criteria" are typically defined by functional performance requirements identified through risk analysis and comparison to a predicate device, and the "device performance" is demonstrated through bench testing.
Test Conducted | Acceptance Criteria (Implied by equivalence claim & typical device performance standards) | Reported Device Performance (Summary from Section 9) |
---|---|---|
Compatibility with irrigation tubing | Device correctly connects and forms a secure seal with irrigation tubing. | Tested and supported substantial equivalence. |
Compatibility with endoscope | Device correctly connects and forms a secure seal with Olympus 160 series endoscopes. | Tested and supported substantial equivalence. |
Water Flow Test | Adequate water flow through the device for intended irrigation. | Tested and supported substantial equivalence. |
Air Leakage Test | No air leakage when subjected to relevant pressure. | Conducted for verification; results supported equivalence. |
Water Leakage Test | No water leakage when subjected to relevant pressure. | Conducted for verification; results supported equivalence. |
Backflow Performance Test | Effectively prevents backflow of water or biomaterials from the endoscope to the sterile water bottle. | Tested and ensured prevention of backflow. |
Sterility | Sterility Assurance Level (SAL) of 10^-6; EO residuals below defined maximum levels. | Validated EO sterilization cycle achieved SAL of 10^-6; residuals below limits. |
Shelf Life (3 years) | Maintained sterile barrier and intended performance after accelerated aging. | Performance testing after accelerated aging confirmed sterile barrier and performance for at least 3 years. |
Biocompatibility | No adverse biological effects (cytotoxicity, sensitization, irritation). | Tests (Acute Systemic Toxicity, In Vitro Cytotoxicity, Skin Sensitization, Rectal Irritation) showed the device is biocompatible. |
2. Sample Size Used for the Test Set and Data Provenance
The document states, "Smartdata performed bench testing to support substantial equivalence. The following testing was performed on Smartdata samples from initial production lots, including sterilization." It does not specify the exact sample size (number of units tested) for each specific test (e.g., how many connectors were tested for backflow).
- Sample Size: Not explicitly stated as a number of units, but "samples from initial production lots."
- Data Provenance: The tests were conducted by Smartdata Suzhou Co., Ltd. (China) and accredited laboratories (e.g., Sanitation Environment Technology Institute, Soochow University, for packaging tests). The data is prospective in the sense that it was generated specifically for this 510(k) submission on newly manufactured devices.
3. Number of Experts Used to Establish Ground Truth and Qualifications of Experts
This concept is not applicable as this is a medical device (accessory) and not an AI/algorithm-driven diagnostic/screening tool. Ground truth for device performance is established through engineering and performance specifications and validated through bench testing, not expert interpretation of image data.
4. Adjudication Method for the Test Set
Not applicable. This is not an image-based diagnostic study requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
No, an MRMC study was not done because this is a physical medical device (an auxiliary water connector) and not an AI/imaging algorithm. The study focuses on the device's physical and functional performance, not its impact on human reader accuracy.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable. This is a physical device, not an algorithm.
7. The Type of Ground Truth Used
The "ground truth" for this device's performance is established by:
- Engineering Specifications: Design requirements for flow rates, dimensions, connection integrity, and backflow prevention.
- Predicate Device Performance: Benchmarking against the known performance of the legally marketed predicate device (ENDOGATOR FEQ, K092429).
- International Standards: Adherence to relevant standards for sterilization (ANSI/AAMI/ISO 10993-7), biocompatibility (ISO 10993-1), and packaging integrity.
- Risk Analysis: Addressing identified safety concerns, such as fluid leakage and backflow, which were noted in the FDA MAUDE database for similar devices.
8. The Sample Size for the Training Set
Not applicable. This is a physical device, not an AI model that requires a training set.
9. How the Ground Truth for the Training Set was Established
Not applicable, as no training set was used for an AI model.
Ask a specific question about this device
(28 days)
Smartdata Suzhou Co., Ltd
The disposable Endorate™ valves sets consists of one suction valve, one air/water valve and one biopsy valve.
The Endorate™ Disposable Suction valve is used to control the suction function of an endoscope (for Olympus 160 series Endoscope) during a GI Endoscopic procedure.
The Endorate™ Disposable Air/Water valve is used to control the air/water function of an endoscope (for Olympus 160 series Endoscope) during a GI Endoscopic procedure.
The Endorate™ Disposable biopsy valve is used to cover the endoscope biopsy port during an endoscopy procedure. In addition, the valve provides access for endoscopic device passage and exchange, helps maintain insufflation and minimizes leakage of biomaterial from the biopsy port throughout the endoscopic procedure.
The Endorate™ valve set consists of one suction valve, one air/water valve and one biopsy valve. The Endorate™ device is intended for single-use, and is supplied sterile. Sterile, singleuse suction valve, air/water valve and biopsy valve help prevent potential safety risks and eliminate the need for manual cleaning and reprocessing. These valves easily incorporate into infection prevention policies as a single use item. The suction valve is designed to be attached to the suction port of the endoscope and the air/water valve is designed to be attached to the air/water port of the endoscope. The activation of the suction valve allows the user to aspirate excess fluids or other debris obscuring the endoscopic image, while the activation of the air/water valve allows the user to control air and water flow to assist in cleansing the lens during procedures. The biopsy valve is intended to cover the endoscope biopsy port during an endoscopy procedure. In addition, the valve provides access for endoscopic device passage and exchange, helps maintain insufflation and minimizes leakage of biomaterial from the biopsy port throughout the endoscopic procedure. The Endorate™ Disposable Biopsy Valves are manufactured in two configurations - one for use with Olympus®/Fujinon® gastrointestinal endoscopes and the other for use with Pentax® gastrointestinal endoscopes. Both Olympus®/Fujinon® and Pentax® versions are sold as non-sterile device, individually packed.
This document describes the Smartdata Suzhou Co., Ltd. Endorate™ Valves Sets (K181509), which includes a suction valve, air/water valve, and biopsy valve, and its substantial equivalence to predicate devices. The information provided heavily relies on comparisons with existing predicate devices rather than independent performance metrics with acceptance criteria.
1. Table of Acceptance Criteria and Reported Device Performance
The submission does not explicitly state quantitative acceptance criteria in a typical "Pass/Fail" format with specific thresholds. Instead, it demonstrates substantial equivalence through a direct comparison of technological characteristics and performance in bench testing to predicate devices (DEFENDO™ Disposable Suction Valve K102581, DEFENDO™ Disposable Air/Water Valve K102409, and DEFENDO™ Disposable Biopsy Valve K090851). The performance is reported as meeting the functional requirements through "bench testing."
Below is a summary of the non-clinical tests performed and the implied "performance" relative to the predicate, demonstrating that the device functions as intended and is similar to the predicate.
Test Category | Specific Test | Acceptance Criteria (Implied) | Reported Device Performance |
---|---|---|---|
Disposable Suction Valve | Scope Compatibility | Compatible with specified endoscopes (e.g., Olympus 140/160/180/190/240/260 series). | Bench testing performed, demonstrating compatibility. (Implicitly met based on SE claim) |
Valve Depression Force | Functional depression force similar to predicate. | Bench testing performed. (Implicitly met based on SE claim) | |
Leak Test | No leakage during operation. | Bench testing performed. (Implicitly met based on SE claim) | |
Disposable Air/Water Valve | Scope Compatibility | Compatible with specified endoscopes. | Bench testing performed, demonstrating compatibility. (Implicitly met based on SE claim) |
Valve Depression Force | Functional depression force similar to predicate. | Bench testing performed. (Implicitly met based on SE claim) | |
Air Ventilation and Insufflation | Proper air flow and insufflation as intended. | Bench testing performed. (Implicitly met based on SE claim) | |
Water Flow Performance | Proper water flow as intended. | Bench testing performed. (Implicitly met based on SE claim) | |
Biopsy Valve | Scope Compatibility | Compatible with specified endoscopes (Olympus/Fujinon and Pentax GI endoscopes). | Bench testing performed, demonstrating compatibility. (Implicitly met based on SE claim) |
Leak Testing | Minimizes leakage of biomaterial. | Bench testing performed. (Implicitly met based on SE claim) | |
Squeegee Performance | Provides access for device passage and exchange, helps maintain insufflation. | Bench testing performed. (Implicitly met based on SE claim) | |
Sterilization & Shelf Life | Sterilization | Sterility Assurance Level (SAL) of 10⁻⁶, EO residuals below ISO 10993-7. | Validated EO sterilization cycle achieved SAL of 10⁻⁶. EO residuals below maximum levels. |
Shelf Life | Maintain sterile barrier for at least three years. | Three (3) year expiration date supported by accelerated aging, seal strength, dye penetration, and packaging persistence bacteria performance of pouches. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify the exact sample sizes used for the "initial production lots" subjected to bench testing. The data provenance is implied to be from internal testing by Smartdata Suzhou Co., Ltd. (China), and it is a retrospective evaluation against predicate devices rather than a prospective study with human subjects.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. The study did not involve establishing ground truth through expert consensus for a test set, as it was a bench testing comparison to predicate devices, not a diagnostic or AI performance study.
4. Adjudication Method for the Test Set
Not applicable. There was no test set requiring expert adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, an MRMC comparative effectiveness study was not done. The device is an endoscopic accessory, not an AI-assisted diagnostic tool for human readers.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
Yes, in a sense, the entire non-clinical performance evaluation can be considered a "standalone" evaluation of the device's mechanical and functional characteristics. However, this terminology usually applies to AI algorithms. For this medical device (endoscopic valves), the "standalone" performance refers to its ability to meet its functional requirements (e.g., suction, air/water control, biopsy access, leakage prevention) as demonstrated through the bench tests without human intervention in the operation beyond initiating the tests.
7. The Type of Ground Truth Used
The "ground truth" for the non-clinical performance tests was inherent in the design specifications, functional requirements, and established performance of the legally marketed predicate devices. The device was considered to meet "ground truth" if its performance in bench tests was found to be substantially equivalent to the predicate devices and met basic functional expectations. Industry standards (e.g., ISO 10993-7 for sterilization) also served as a form of ground truth for specific parameters.
8. The Sample Size for the Training Set
Not applicable. This is not an AI/machine learning device, so there is no "training set."
9. How the Ground Truth for the Training Set Was Established
Not applicable. There is no training set for this device.
Ask a specific question about this device
Page 1 of 1