K Number
K220999Device Name
Hipro Glycosylated Hemoglobin (HbA1c) Test System
Manufacturer
Shijiazhuang Hipro Biotechnology Co., Ltd.
Date Cleared
2024-09-12
(892 days)
Product Code
PDJ, LCP
Regulation Number
862.1373Why did this record match?
Applicant Name (Manufacturer) :
Shijiazhuang Hipro Biotechnology Co., Ltd.
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Hipro® Glycosylated Hemoglobin (HbA1c) Test System comprised of the Hipro Glycosylated hemoglobin (HbA1c) test kit and the HP-AFS/1 automatic immunoassay analyzer is used as an aid in diagnosis of diabetes mellitus, as an aid to identify patients who may be at risk for developing diabetes mellitus, and for the monitoring of long-term blood glucose control in individuals with diabetes mellitus. It is an in vitro diagnostics reagent system intended for quantitative determination of % hemoglobin A1c (DCCT/NGSP) in venous whole blood.
Device Description
The Hipro® Glycosylated hemoglobin (HbA1c) test system is intended for quantitative determination of % hemoglobin A 1c (DCCT/NGSP) in venous whole blood. It is composed of Glycosylated hemoglobin (HbA1c) test kit and automatic immunoassay analyzer. Glycosylated hemoglobin (HbA1c) test kit consists of two reagents R1 and R2, which are liquid and ready to use. Reagent R1 contains glycine buffer and latex, and reagent R2 contains glycine buffer and two types of antibodies, Goat anti-mouse IgG, mouse anti-human HbAlc monoclonal antibody.
HP-AFS/1 Automatic Immunoassay Analyzer is made up of light absorption test module, scattered light test module, fluorescence test module, press components, data transmission interface and printer. And the absorption test module is made up of absorption light path unit, and lightabsorption sensor part. The scattered light test module is made up of scattered light path unit and scattered light sensor part.