Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K150549
    Date Cleared
    2015-06-01

    (90 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sharp Medical Reactor™ Trocar and Sleeve has indications in the creation of a port of entry into the thoracic cavity and allows the user to create and maintain limited access into the intention of chest tube placement.

    Device Description

    The Reactor™ Trocar consists of an obturator with a flat tip at the distal end which encloses a circular scalpel shaped knife blade. The blade extends approximately 1mm and immediately retracts when the trigger is squeezed. The clear sleeve with printing accommodates instruments up to 36 Fr. or 12 mm.

    AI/ML Overview

    This document describes a 510(k) premarket notification for a medical device called the "Reactor™ Trocar and Sleeve". This is a regulatory submission for a device to demonstrate substantial equivalence to a legally marketed predicate device, rather than a study designed to establish acceptance criteria and prove performance against them in the traditional sense of a clinical or analytical validation study for an AI/ML device.

    Therefore, the specific information requested in the prompt (sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, training sets) is not applicable to this type of document, which focuses on demonstrating equivalence to an existing device through comparison of features and performance testing.

    However, I can extract the information that is present concerning performance.

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present acceptance criteria in a quantitative table with corresponding reported performance metrics as one might see in a clinical trial report for a novel medical device. Instead, the "Performance Data" section describes the types of tests conducted to demonstrate equivalence to the predicate device. The implied "acceptance criteria" for these tests would be that the Reactor™ Trocar and Sleeve performs at least as well as, or comparably to, the predicate device in the specified tests to establish substantial equivalence.

    Performance Test TypeImplied Acceptance CriteriaReported Device Performance
    Dry Fire TestComparable performance to predicate deviceBench tests confirm the performance of the Reactor™ Trocar and Sleeve is substantially equivalent to the predicate device.
    Tissue Insertion TestComparable performance to predicate deviceBench tests confirm the performance of the Reactor™ Trocar and Sleeve is substantially equivalent to the predicate device and validate the Reactor™ Trocar and Sleeve perform as intended.
    Chest Tube Placement TestSuccess in chest tube placement comparable to predicate deviceBench tests confirm the performance of the Reactor™ Trocar and Sleeve is substantially equivalent to the predicate device and validate the Reactor™ Trocar and Sleeve perform as intended. Success in chest tube placement was demonstrated in the TraumaMan surgical simulator.
    Shaft Rotation TestComparable performance to predicate deviceBench tests confirm the performance of the Reactor™ Trocar and Sleeve is substantially equivalent to the predicate device.
    Torque TestComparable performance to predicate deviceBench tests confirm the performance of the Reactor™ Trocar and Sleeve is substantially equivalent to the predicate device.
    Surgical Imitation Tests (TraumaMan simulator)Successful and intended operation within the simulated surgical environment, comparable to predicate deviceThese tests demonstrate the device performs as intended in a simulated surgical environment for chest tube placement. "Bench tests confirm the performance... is substantially equivalent... and validate the Reactor™ Trocar and Sleeve perform as intended."
    Biocompatibility (ISO 10993)Compliance with ISO 10993 (Biological Evaluation of Medical Devices)The materials are biocompatible and comply with ISO 10993 (Biological Evaluation of Medical Devices).
    Sterilization (ISO 11137)Compliance with ISO 11137 (Sterilization of Healthcare Products)The device is a single use, sterile product compliant with ISO 11137 (Sterilization of Healthcare Products).

    Note: The phrase "substantially equivalent" is the overarching "acceptance criterion" for a 510(k) submission, meaning the new device is as safe and effective as a legally marketed device. The performance tests are conducted to support this claim.

    The following information is not present in the provided document:

    1. Sample size used for the test set and the data provenance: The document mentions "bench tests" and "surgical imitation tests in the TraumaMan surgical simulator." It does not specify the number of devices tested, the number of simulated cases, or any details about the origin of the test "data" (e.g., retrospective/prospective).
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: This is not applicable. The device is a mechanical surgical instrument, not an AI/ML diagnostic tool requiring expert ground truth for interpretation. Performance is assessed by direct physical testing.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable for the type of bench testing described.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: This is not applicable as the device is a surgical instrument, not an AI-assisted diagnostic tool.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For mechanical devices like this, the "ground truth" for performance is typically objective physical measurements or successful completion of a task, not expert interpretation or pathological findings. In this case, successful performance in the described tests (e.g., successful tissue insertion, chest tube placement in a simulator) serves as the basis for performance evaluation.
    7. The sample size for the training set: Not applicable as this is not an AI/ML device.
    8. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1