Search Results
Found 11 results
510(k) Data Aggregation
(79 days)
Shanghai AnQing Medical Instrument Co., Ltd.
The Flexible Bronchoscopes have been designed to be used with the video processor, endotherapy accessories and other ancillary equipment for endoscopy within the airways and tracheobronchial tree.
The Flexible Bronchoscope is for use in a hospital environment. The Flexible Bronchoscope is a single-use device designed for use in adults.
The Flexible Bronchoscope (Model: BS27U-12EU, BS27U-12US, BS38U-20EU, BS38U-20US) is intended to be used with the Portable Video Processor (cleared via K243497). The Flexible Bronchoscope is inserted through the airways and tracheobronchial tree during bronchoscopy, and when used with the compatible video processor and monitor, the endoscope system can be operated as intended and indicated. The Flexible Bronchoscope is a single-use endoscope, which consists of a Handle, an Insertion Section, and an Endoscope Connector. The handle includes a deflection lever, a lever lock, a push button for picture taking/video recording, a push button for suction, a connector for suction tubing, a Luer port for insertion of accessory devices and irrigation to the working channel and a LED for illumination. The insertion section contains one working channel, wiring to transmit the image signals to the video processor, and two optical fibers to transmit illumination from the handle to the distal tip. The distal bending section of the insertion section is controlled by the user via the deflection lever on the handle. The distal end of the insertion section contains a CMOS sensor for capturing image and transmitting it to the video processor, optical fibers for transmitting illumination from the LED inside the handle, and the distal opening of the working channel. The Endoscope Connector connects the endoscope handle to the video processor, which provides power and processes video signals from the endoscope. Same as the predicate, the subject device is also provided in 2 deflection versions (US/EU deflection).
This document is an FDA 510(k) clearance letter for a Flexible Bronchoscope, indicating substantial equivalence to a predicate device. However, it does not contain the level of detail requested regarding acceptance criteria and a specific study proving the device meets those criteria, particularly in the context of an AI/algorithm-driven device.
The provided text focuses on the physical and performance characteristics of the bronchoscope itself, its intended use, technological comparison to a predicate device, and various non-clinical tests (electrical safety, photobiological safety, mechanical/optical performance, biocompatibility, sterilization, shelf life, and package validation).
There is no mention of an AI component, an algorithm, or any study involving human readers, ground truth establishment, or performance metrics like sensitivity, specificity, or AUC. The "device performance" in this context refers to the bronchoscope's mechanical and optical functionality, not the diagnostic accuracy of an AI algorithm.
Therefore, most of the requested information cannot be extracted from this document.
Here's an attempt to answer based on the provided text, highlighting what is missing:
1. Table of acceptance criteria and reported device performance
The document does not present a formal table of "acceptance criteria" for a diagnostic algorithm with corresponding "reported device performance" in terms of clinical accuracy metrics (like sensitivity, specificity, AUC). Instead, it states that various non-clinical tests were performed to demonstrate compliance with recognized standards and that the device is "as safe, as effective, and performs as well as the legally marketed device identified above."
Below is a table summarizing the types of tests and the general conclusion, but without specific quantitative acceptance criteria or performance metrics related to diagnostic accuracy.
Test Type | Acceptance Criteria (Implied) | Reported Device Performance (Summary) |
---|---|---|
Electrical Safety | Compliance with ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-18, IEC/TS 60601-4-2 standards. | In compliance with the specified FDA recognized standards. |
Photobiological Safety | Compliance with IEC 62471:2006. | Tested according to IEC 62471:2006. |
Mechanical Performance | Compliance with applicable parts of ISO 8600 (e.g., leakage tightness, bending, deflection endurance, tensile strength). | Tested for leakage tightness, bending, deflection endurance, and tensile strength; implied compliance. |
Optical Performance (Direct) | Compliance with applicable parts of ISO 8600. | Optical measurements performed according to ISO 8600; implied compliance. |
Optical Performance (Comparative) | Image quality parameters (direction of view, field of view, MTF/DOF, color performance, SNR/Dynamic Range, Distortion, Image intensity uniformity) comparable to reference device K173727 and K243857. | Comparative testing performed, results demonstrated subject device is as safe and effective as predicate. |
Biocompatibility | Compliance with ISO 10993-1 and FDA Guidance. Absence of cytotoxicity, sensitization, irritation, pyrogenicity, acute systemic toxicity. | Evaluated for Cytotoxicity, Sensitization, Irritation, Material-mediated pyrogenicity, Acute systemic toxicity; implied compliance. |
Sterilization & Shelf Life | Validation to ISO 11135:2014; EO/ECH residual test per ISO 10993-7:2008; Shelf life determined by stability study per ASTM F1980-21. | Sterilization method validated; EO/ECH residual test performed; shelf life determined through aging test. |
Package Validation | Compliance with ISO 11607-1:2019, ISO 11607-2:2019, ASTM F88/F88M-21, ASTM F1929-15, ASTM D4169-22. | Conducted according to specified standards. |
2. Sample sized used for the test set and the data provenance
- Sample Size for Test Set: Not applicable/Not mentioned. This document describes a medical device (bronchoscope) itself, not an AI algorithm performing a diagnostic task on a dataset. The "tests" mentioned are non-clinical engineering and biological safety tests, not tests on a dataset.
- Data Provenance (e.g. country of origin of the data, retrospective or prospective): Not applicable/Not mentioned.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not applicable. There is no mention of "ground truth" in the context of an AI algorithm's diagnostic performance for which experts would be needed. The "ground truth" for the non-clinical tests would be the established performance specifications and standards for a bronchoscope.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable. As no expert ground truth establishment for a diagnostic AI is mentioned, no adjudication method would be relevant.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No. This document describes the clearance of a flexible bronchoscope, a physical medical device. It does not mention any AI component or a MRMC study.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. There is no mention of an algorithm in this document that would perform as a standalone device.
7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)
- Not applicable. Ground truth in the context of diagnostic AI is not mentioned. For the non-clinical tests, the "ground truth" is compliance with established engineering and safety standards.
8. The sample size for the training set
- Not applicable. There is no mention of an AI model or a training set.
9. How the ground truth for the training set was established
- Not applicable. There is no mention of an AI model or a training set.
Ask a specific question about this device
(198 days)
Shanghai AnQing Medical Instrument Co., Ltd.
The Flexible Bronchoscopes have been designed to be used with the video processor, endotherapy accessories and other ancillary equipment for endoscopy within the airways and tracheobronchial tree.
The Bronchoscope System is for use in a hospital environment. The Flexible Bronchoscope is a single-use device designed for use in adults.
The Bronchoscope System comprises two components: (1) a Portable Video Processor (model: EA101) and (2) a compatible Flexible Bronchoscope (model: BS41H-12EU, BS41H-12US, BS46H-17EU, BS46H-17US, BS50H-20EU, BS50H-20US, BS53H-22EU, BS53H-22US, BS55H-24EU, BS55H-24US, BS59H-28EU, BS59H-28US).
The Flexible Bronchoscope is inserted through the airways and tracheobronchial tree during bronchoscopy. The Video Processor provides power and processes the images for medical electronic endoscope. The Portable Video Processor consists a 13.3" LCD touch screen. It is powered through a Lithium-ion battery or a separate power adaptor.
The Flexible Bronchoscope has following physical and performance characteristics:
- Maneuverable tip controlled by the user
- Flexible insertion cord
- CMOS Camera and LED light source at the distal tip
- Sterilized by Ethylene Oxide
- For single use
Portable Video Processor has following physical and performance characteristics: - Displays the image from the Flexible Bronchoscope on the built-in screen.
- Can be connected to an external monitor.
- Non-sterile Reusable device.
The provided 510(k) clearance letter and summary for K243497 indicate that the device in question is a Flexible Bronchoscope System. This submission appears to be a change to an already cleared flexible bronchoscope (K211169) to make it compatible with a new Portable Video Processor (EA101).
Crucially, there is no mention of any AI or machine learning component in the device description, indications for use, or the non-clinical and/or clinical tests summary. The document describes standard performance testing for medical devices, focusing on electrical safety, EMC, photobiological safety, optical performance, biocompatibility, sterilization, shelf life, and package validation. Software verification and validation are mentioned in the context of IEC 62304, which covers medical device software lifecycle processes, but this does not imply AI.
Therefore, many of the questions related to AI-specific acceptance criteria, study methodologies (like MRMC studies, standalone AI performance), ground truth establishment for AI training/testing sets, and expert involvement for AI adjudication are not applicable to the information provided in this document.
The document does provide information relevant to the overall device's performance validation, which can be presented as acceptance criteria and proof for a non-AI medical device.
Non-AI Device Acceptance Criteria and Study Proof
Since the device described is not an AI/ML device, the concept of "acceptance criteria" and "study that proves the device meets the acceptance criteria" will be interpreted in the context of regulatory clearance for a traditional medical device rather than for AI performance. The studies performed are primarily bench testing and phantom/component level evaluations to demonstrate substantial equivalence to the predicate device and compliance with recognized standards.
1. Table of Acceptance Criteria and Reported Device Performance
For this type of device, "acceptance criteria" are typically defined by compliance with recognized standards and demonstration of equivalent performance to a predicate device. The "reported device performance" is the evidence presented to show this compliance.
Acceptance Criterion | Reported Device Performance (Summary from K243497) |
---|---|
Electrical Safety | In compliance with ANSI AAMI ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] |
Electromagnetic Compatibility (EMC) | In compliance with ANSI AAMI IEC 60601-1-2:2014 Including AMD 1:2021] and IEC 60601-2-18 Edition 3.0 2009-08, IEC/TR 60601-4-2 Edition 1.0 2016-05 |
Photobiological Safety | Tested according to IEC 62471:2006 Medical electrical equipment, Photobiological safety of lamps and lamp systems. |
Optical Performance | Designed to comply with applicable parts of ISO 8600. Optical measurements performed according to applicable parts of ISO 8600 standard. Comparative testing performed for image quality parameters (direction of view, field of view, MTF/DOF, color performance, SNR/Dynamic Range, Distortion, Image intensity uniformity) against predicate device. |
Biocompatibility | Performed according to ISO 10993-1 and FDA Guidance. Tests conducted: Cytotoxicity, Sensitization, Irritation, Material-mediated pyrogenicity, Acute systemic toxicity. |
Sterilization Efficacy | Validated to ISO 11135:2014. EO/ECH residual test per ISO 10993-7:2008. |
Shelf Life | Determined based on stability study including accelerated aging per ASTM F1980-21. |
Package Integrity | Validated per ISO 11607-1:2019, ISO 11607-2:2019, ASTM F88/F88M-21, ASTM F1929-15. Transport and shipping testing per ASTM D4169-22. |
Software Verification & Validation | Conducted according to IEC 62304: 2006 + A1: 2015 and FDA Guidance for Industry and Staff "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". |
Equivalence to Predicate Device (Flexible Bronchoscope) | Flexible Bronchoscopes are the same as cleared in K211169, with changes related to compatibility with new video processor and package configuration. Other performance data from K211169 is applicable. |
Equivalence to Predicate Device (Portable Video Processor) | Demonstrated through comparison of technological characteristics and performance testing to show substantial equivalence. Key differences (IV pole mount, built-in screen, battery power, no USB) were addressed. |
2. Sample Sizes Used for the Test Set and Data Provenance
- Test Set Sample Sizes: Not explicitly stated as "sample sizes" in the manner of a clinical trial. For bench testing (electrical safety, EMC, optical, biocompatibility, sterilization, shelf life, packaging), samples of the device and/or its components were tested to meet specific standard requirements. For example, biocompatibility involves testing material extracts, and sterilization validation involves processing batches of devices. The exact number of units tested for each specific bench test is not detailed in this summary.
- Data Provenance: The document does not specify the country of origin for the non-clinical test data. Given the applicant is in Shanghai, China, it is highly likely the testing was conducted in China. These were non-clinical bench tests, not clinical studies collecting patient data, so "retrospective or prospective" doesn't directly apply in the typical sense.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
- Not applicable in the context of AI/ML ground truth. For traditional device validation, "ground truth" is established by the specifications in recognized consensus standards (e.g., ISO, ASTM, IEC) and the performance characteristics of the predicate device. Expert involvement would be in the form of engineers, test technicians, and quality assurance personnel who perform and verify the tests according to established protocols and standards. Their qualifications are inherent in their professional roles, but not quantified as "number of experts" for ground truth establishment as would be done for clinical image annotation.
4. Adjudication Method for the Test Set
- Not applicable in the context of AI/ML adjudication. For traditional device testing, the "adjudication" is compliance with objective, measurable parameters defined by standards. Test results either meet the specified limits/criteria or they do not. QA processes would review and approve test reports, but there isn't a human "adjudication" process in the way it's used for clinical data interpretation differences.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
- No. This is not an AI/ML device. MRMC studies are typically performed for imaging devices or AI algorithms where human interpretation is involved and needs to be evaluated for improvement with or without AI assistance. This device is a bronchoscope system for visualization and intervention, not for diagnostic image interpretation by humans that would necessitate an MRMC study.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study was done
- No. This is not an AI/ML device. Standalone performance is relevant for AI algorithms. The performance of this device is assessed through its physical, electrical, and optical properties as a medical instrument.
7. The Type of Ground Truth Used
- For this traditional medical device, "ground truth" equates to compliance with recognized industry standards and the specifications / validated performance of the predicate device. For example:
- Electrical Safety: Ground truth is defined by the limits and test methods of IEC 60601-1.
- Biocompatibility: Ground truth is established by the accepted biological responses defined in ISO 10993 series.
- Optical Performance: Ground truth is established by the parameters defined in ISO 8600 and comparative measurements against the predicate device.
8. The Sample Size for the Training Set
- Not applicable. This device does not use an AI/ML model; therefore, there is no "training set."
9. How the Ground Truth for the Training Set Was Established
- Not applicable. As there is no AI/ML model or training set, this question is not relevant.
Ask a specific question about this device
(116 days)
Shanghai AnQing Medical Instrument Co., Ltd.
The Flexible Ureterorenoscope is intended to be used to visualize organs, cavities and canals in the urinary tract (urethra, bladder, ureter, calyces and renal papillae) via transurethral access routes. It can also be used in conjunction with endoscopic accessories to perform various diagnostic and therapeutic procedures in the urinary tract.
The Flexible Ureterorenoscope (Model: US27F-12-EU; US27F-12-US) is intended to be used with the Video Processor (cleared via K211169). The Flexible Ureterorenoscope is inserted through the natural orifice urethra and when used with the compatible Video Processor and monitor, the endoscope system can be operated as intended and indicated. The Flexible Ureterorenoscope is a single-use endoscope, which consists of a Handle, an Insertion Section, and an Endoscope Connector. The handle includes a deflection lever, a lever lock, a push button for picture taking/video recording, a Luer port for insertion of accessory devices and irrigation to the working channel and a LED for illumination. The insertion section contains one working channel, wiring to transmit the image signals to the Video Processor, and two optical fibers to transmit illumination from the handle to the distal tip. The distal bending section of the insertion section is controlled by the user via the deflection lever on the handle. The distal end of the insertion section contains a CMOS sensor for capturing image and transmitting it to the Video Processor, optical fibers for transmitting illumination from the LED inside the Handle, and the distal opening of the working channel. The Endoscope Connector connects the endoscope handle to the video processor, which provides power and processes video signals from the endoscope.
Mechanism of action:
The light emitted from the distal tip of the Flexible Ureterorenoscope is irradiated into the body cavity, and the light reflected from the cavity enters the optical system and is captured by the CMOS image sensor. The CMOS acquisition image is controlled by the CMOS drive circuit, and the RGB video signal is output to the Video Processor via the VI circuit. The Video Processor receives video signals from the endoscope, processes the video signals, and outputs the processed video signal to the attached monitor. The video processor also controls the brightness of the LED on the endoscope.
Flexible Ureterorenoscope has the following physical and performance characteristics:
- Maneuverable tip controlled by the user
- Flexible insertion cord
- Camera at the distal tip
- LED in the handle and transmitted to the distal tip by optical fibers
- Sterilized by Ethylene Oxide
- For single use
The provided FDA 510(k) clearance letter and summary for the Flexible Ureterorenoscope (K243857) does not contain details about a study evaluating AI performance directly or outlining acceptance criteria for such a study.
This document primarily focuses on the substantial equivalence of a medical device (endoscope) based on its physical properties, performance characteristics (mechanical, optical), electrical safety, biocompatibility, sterilization, and packaging. The "non-clinical tests summary & conclusions" explicitly state "Clinical Testing: Not Applicable."
Therefore, I cannot fulfill your request for information regarding acceptance criteria and a study proving a device meets these criteria in the context of AI performance, human reader improvement with AI, or standalone AI performance, as this information is not present in the provided text.
The closest relevant information is about the device's optical performance and image quality, which are foundational for any imaging device, and would implicitly establish a baseline for an AI to process.
Here's what I can provide based on the given information, primarily highlighting the device's tested performance attributes, which serve as foundational acceptance criteria for the hardware itself, not an AI component:
Acceptance Criteria and Device Performance (Flexible Ureterorenoscope - K243857)
Since the provided document is for a medical imaging device (uroscope) and not an AI/software as a medical device (SaMD), the "acceptance criteria" discussed are for the physical and functional performance of the endoscope itself, rather than for an AI algorithm built upon imagery from such a device. There is no mention of AI in this clearance.
1. Table of Acceptance Criteria and Reported Device Performance
Note: The document lists "applicable parts" of standards and general categories of testing rather than specific numerical acceptance thresholds or quantitative performance metrics. The reported performance is adherence to these standards and the successful execution of these tests.
Criteria Category | Acceptance Standard/Test | Reported Device Performance |
---|---|---|
Electrical Safety | ANSI AAMI ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] | In compliance with the standard |
Electromagnetic Compatibility | ANSI AAMI IEC 60601-1-2:2014 [Including AMD 1:2021] | In compliance with the standard |
Particular Requirements for Endoscopic Equipment | IEC 60601-2-18 Edition 3.0 2009-08 | In compliance with the standard |
Electromagnetic Phenomena - Guidance on Basic EMC Aspects | IEC/TR 60601-4-2 Edition 1.0 2016-05 | In compliance with the standard |
Photobiological Safety | IEC 62471:2006 (Medical electrical equipment, Photobiological safety of lamps and lamp systems) | Tested according to the standard and found compliant |
Mechanical Performance | Applicable parts of ISO 8600 | Leakage tightness, bending, deflection endurance, and tensile strength testing performed and met requirements |
Optical Performance | Applicable parts of ISO 8600 | Optical measurements performed and met requirements |
Image Quality Parameters (Comparative to Predicate) | Direction of view, field of view, MTF/DOF, color performance, SNR/Dynamic Range, Resolution, and Image intensity uniformity compared favorably to the predicate to support substantial equivalence. | |
Biocompatibility | ISO 10993-1 and FDA Guidance | Cytotoxicity, Sensitization, Irritation, Material-mediated pyrogenicity, Acute systemic toxicity tests conducted and met requirements |
Sterilization Validation | ISO 11135:2014 | Validated, determined routine control and monitoring parameters |
EO/ECH Residuals | ISO 10993-7:2008 | Test performed and met requirements |
Shelf Life (Accelerated Aging) | ASTM F1980-21 | Determined based on stability study (aging test) |
Package Validation | ISO 11607-1:2019 and ISO 11607-2:2019 | Conducted according to standards |
Package Integrity (Seal Strength) | ASTM F88/F88M-21 | Conducted according to standard |
Package Integrity (Dye Penetration) | ASTM F1929-15 | Conducted according to standard |
Transport & Shipping | ASTM D4169-22 | Testing performed and met requirements |
2. Sample Size and Data Provenance for Test Set
- Sample Size: Not explicitly stated for specific tests (e.g., how many units were tested for mechanical performance). The testing is primarily bench/lab-based.
- Data Provenance: The tests are non-clinical, hardware-focused performance evaluations. Data is generated during these laboratory tests performed by the manufacturer, Shanghai AnQing Medical Instrument Co., Ltd. (China). This is implied to be prospective testing for regulatory submission.
3. Number of Experts and Qualifications for Ground Truth
- Not applicable. The document explicitly states "Clinical Testing: Not Applicable." Ground truth, in this context, would relate to clinical outcomes or expert review of images for diagnostic purposes, which are not part of this 510(k) submission for the device itself. Performance is evaluated against engineering standards and specifications.
4. Adjudication Method for Test Set
- Not applicable. No clinical test set or expert review requiring adjudication is mentioned.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No. The document explicitly states "Clinical Testing: Not Applicable." No MRMC study or AI assistance evaluation is mentioned.
6. Standalone Performance (Algorithm Only)
- Not applicable. This submission is for the endoscope hardware, not a standalone AI algorithm. No algorithm performance is described.
7. Type of Ground Truth Used
- Engineering/Technical Standards: The "ground truth" for this device's performance is adherence to established international and national standards (e.g., IEC, ISO, ASTM, ANSI AAMI) for medical device safety, electrical performance, optical quality, biocompatibility, sterilization, and mechanical integrity. Performance is measured against these technical specifications.
8. Sample Size for Training Set
- Not applicable. There is no mention of an AI component requiring a training set.
9. How Ground Truth for Training Set Was Established
- Not applicable. There is no mention of an AI component or a training set.
Ask a specific question about this device
(181 days)
Shanghai AnQing Medical Instrument Co., Ltd
AnQing Medical Flexible Choledochoscope has been designed to be used with the video processor, monitor, endotherapy accessories and other ancillary equipment for endoscopic diagnosis and treatment within the bile duct. The Flexible Choledochoscope is designed for use in adults.
The Flexible Choledochoscope (Model: CS50H-20EU, CS50H-20US) is a sterile single-use endoscope which is used with the video processor (Model: EOS-H-01. FDA cleared #K211169) produced by AnQing for providing endoscopic imaging within the bile duct for the purpose of diagnosis and treatment. The 2 proposed models are identical except the deflection versions, which is opposite from each other (EU version or US version). The Flexible Choledochoscope is a single-use endoscope, which consists of Handle, Insertion Section, Distal Tip, and Endoscope Connector. The handle includes a deflection lever, a lever lock, an aspiration button, an aspiration connector, a push button for picture taking/video recording and a Luer port for insertion of accessory devices and irrigation to the working channel. The insertion contains one working channel and wiring to transmit the image signals to the Video Processor. The distal bending section of the insertion section is controlled by the user via the deflection lever on the handle. The distal end of the insertion section contains a CMOS sensor for capturing image and transmitting it to the Video Processor, LEDs for illumination, and the distal opening of the working channel. The endoscope connector connects the endoscope handle to the video processor, which provides power and processes video signals from the endoscope.
The provided document is a 510(k) Premarket Notification from the FDA, specifically concerning a Flexible Choledochoscope. This type of document is for a medical device that does not rely on complex algorithms or AI for its primary function. Therefore, the concepts of "AI algorithm," "acceptance criteria for an AI model," "training dataset," "test set," "ground truth establishment," "MRMC studies," and "standalone performance" as they relate to AI are not applicable to this submission.
The acceptance criteria and performance data provided in this document are entirely related to the physical and functional characteristics of the choledochoscope itself, not an AI component.
Here's an analysis of the document in relation to the questions asked, noting where the questions are not applicable due to the nature of the device:
Acceptance Criteria and Device Performance (Non-AI Device)
Since this is a physical medical device (a choledochoscope), the "acceptance criteria" are related to its physical properties, safety standards, and functional performance compared to a predicate device.
1. A table of acceptance criteria and the reported device performance
The document presents a comparison table between the subject device (Flexible Choledochoscope, Models CS50H-20EU, CS50H-20US) and its predicate device (Olympus CHF TYPE V), showing "acceptance criteria" as functional and physical similarities and differences. The "reported device performance" is essentially that the subject device meets or is comparable to the predicate device in these aspects, and where there are differences, non-clinical tests were conducted to demonstrate equivalence without raising new safety/effectiveness concerns.
Feature / Criteria | Subject Device (CS50H-20EU, CS50H-20US) | Predicate Device (CHF TYPE V) (K081456) | Comparison / Performance Outcome |
---|---|---|---|
Indications for Use | Diagnosis & treatment within bile duct, for adults. Used with video processor, monitor, endotherapy accessories. | Diagnosis & treatment within bile duct. Used with Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories. | Equivalent (Note 1: Wording differences do not alter intended use or raise new safety/effectiveness questions). |
Physical Characteristics | |||
Type of Scope | Flexible | Flexible | Same |
Outer diameter (mm) | Max. 5.0 mm | Max. 5.85 mm | Similar (Subject is smaller; Note 2: Differences evaluated via performance testing). |
Inner diameter (mm) | Min. 2.0 mm | Min. 2.0 mm | Same |
Working length | 380 mm | 380 mm | Same |
Deflection | 210° up, 180° down | 160° up, 130° down | Similar (Subject has greater deflection; Note 2: Differences evaluated via performance testing). |
Optical Characteristics | |||
Type of Image sensor | CMOS | Color CCD | Different (Note 2: Differences evaluated via performance testing). |
Field of View | 110° | 120° | Similar (Note 2: Differences evaluated via performance testing). |
Direction of View | 0° | 0° | Same |
Depth of Field | 5mm~100mm | 3mm~50mm | Similar (Note 2: Differences evaluated via performance testing). |
Light Source | Internal LEDs | External light source | Different (Note 2: Differences evaluated via performance testing). |
Patient Contacting Materials | |||
General material type | Compliance with ISO10993-1 | Compliance with ISO10993-1 | Similar (Note 2: Differences evaluated via performance testing). |
Duration and type of contact | "External communication medical device-Tissue" with a contact duration of "Limited ( |
Ask a specific question about this device
(111 days)
Shanghai AnQing Medical Instrument Co., Ltd
AnQing Medical Single Use Flexible Cystoscope is used to provide visualization and operative access during diagnostic and therapeutic endoscopic procedures of the lower urinary tract, including the bladder and urethra.
The Flexible Cystoscope is intended for use in a hospital environment or medical office environment. The Flexible Cystoscope is designed for use in adults.
The Flexible Cystoscope (Model: CY50H-20EU, CY50H-20US, CY55H-24EU, CY55H-24US) is a sterile single-use endoscope which is used with the video processor (Model: EOS-H-01, FDA cleared #K211169) produced by AnQing for providing endoscopic imaging of the lower urinary tract, including the bladder and urethra for the purpose of diagnosis and treatment.
The CY50H-20 and CY55H-24 difference is the size of Distal End Outer Diameter and Working Channel Inner Diameter. The US and EU model are identical except the deflection versions, which is opposite from each other (EU version or US version).
The Flexible Cystoscope is a single-use endoscope, which consists of Handle, Insertion Section, Distal Tip, and Endoscope Connector.
The handle includes a deflection lever, a lever lock, a push button for picture taking/video recording and a Luer port for insertion of accessory devices and irrigation to the working channel. The insertion contains one working channel and wiring to transmit the image signals to the Video Processor. The distal bending section of the insertion section is controlled by the user via the deflection lever on the handle. The distal end of the insertion section contains a CMOS sensor for capturing image and transmitting it to the Video Processor, LEDs for illumination, and the distal opening of the working channel. The endoscope connector connects the endoscope handle to the video processor, which provides power and processes video signals from the endoscope.
The provided document does not describe the acceptance criteria and study proving device meets acceptance criteria in the format requested. Specifically, it does not include a table of acceptance criteria vs. device performance, or details regarding sample sizes, data provenance, ground truth establishment, expert qualifications, adjudication methods, MRMC studies, or standalone algorithm performance.
The document is a 510(k) premarket notification for a Flexible Cystoscope, primarily focusing on demonstrating substantial equivalence to a predicate device (Ambu aScope 4 Cysto) through non-clinical testing.
Here's a breakdown of the information that is available in relation to your request, and what is missing:
1. A table of acceptance criteria and the reported device performance
- Not explicitly provided in the requested format. Instead, the document provides a "Comparison with Predicate Device" table (pages 5-6) which compares various physical, optical, and technological characteristics of the subject device to the predicate device. However, this table lists characteristics and indicates "Same" or "Similar" for comparison, rather than specific acceptance criteria and quantitative performance results for each criterion.
- "Summary of Testing" (page 8-9) lists the types of non-clinical tests performed (Electrical Safety, EMC, Photobiological safety, Mechanical and Optical Performance, Biocompatibility, Sterilization, Shelf life, Package Validation) and the standards they comply with, but does not provide specific acceptance criteria or performance values for the subject device against these criteria.
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
- Not provided. The document describes non-clinical bench testing, but does not specify sample sizes for these tests or the origin of any data (as it's physical device testing, not data collection from patients).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
- Not applicable/Not provided. This type of information is typically relevant for AI/ML-based devices where expert-labeled datasets are used to establish ground truth. The Flexible Cystoscope is a physical medical device, and its performance is evaluated against engineering specifications and standards, not against expert-established ground truth on medical images or diagnoses.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable/Not provided. As above, this is relevant for AI/ML evaluation, not for the direct performance testing of a physical endoscope.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No, not done. The document explicitly states under "Summary of Clinical Tests" (page 9): "The subject of this premarket submission, did not require clinical studies to support substantial equivalence." This means no human-in-the-loop or MRMC studies were conducted.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable/Not done. The device is a physical cystoscope, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- Not applicable/Not provided. The "ground truth" for this device's performance relies on established engineering standards (e.g., ISO 8600 for optical measurements, IEC 62471 for photobiological safety, ISO 10993 for biocompatibility) and direct physical measurements and tests of the device's characteristics (e.g., outer diameter, working length, deflection, depth of field).
8. The sample size for the training set
- Not applicable/Not provided. There is no AI/ML component, so no training set.
9. How the ground truth for the training set was established
- Not applicable/Not provided. There is no AI/ML component, so no training set or ground truth for it.
In summary, the provided document details the non-clinical testing and comparison of a flexible cystoscope to a predicate device to demonstrate substantial equivalence, rather than providing the specific AI/ML-related performance evaluation details requested.
Ask a specific question about this device
(26 days)
Shanghai AnQing Medical Instrument Co., Ltd.
The Flexible Ureterorenoscope is intended to be used to visualize organs, cavities and canals in the urinary tract (urethra, bladder, ureter, calyces and renal papillae) via transurethral access routes. It can also be used in conjunction with endoscopic accessories to perform various diagnostic and therapeutic procedures in the urinary tract.
The Flexible Ureterorenoscope (Model: US31E-12-EU; US31E-12-US) is intended to be used with the Video Processor (cleared in K211169). The Flexible Ureterorenoscope is inserted through the natural orifice urethra and when used with the compatible Video Processor and monitor, the endoscope system can be operated as intended and indicated. The Flexible Ureterorenoscope is a single-use endoscope, which consists of a Handle, an Insertion Section, and an Endoscope Connector. The handle includes a deflection lever, a working channel port for accessory devices and a Luer port for irrigation. The insertion section contains one working channel and wiring to transmit the image signals to the Video Processor. The distal bending section of the insertion is controlled by the user via the deflection lever on the handle. The distal end of the insertion contains a CMOS sensor for capturing image and transmitting it to the Video Processor, LEDs for illumination, and the distal opening of the working channel. The endoscope connects the endoscope handle to the video processor, which provides power and processes video signals from the endoscope. Same as the predicate, the subject device is also provided in 2 deflection versions (US/EU deflection), which is the only difference between the two proposed models.
The provided text describes a 510(k) premarket notification for a medical device called "Flexible Ureterorenoscope." This document is a regulatory submission to the FDA proving substantial equivalence to a predicate device, not a study of an AI/ML algorithm. Therefore, many of the requested details, such as AI-specific acceptance criteria, test set Ground Truth establishment, MRMC studies, or training set information, are not applicable to this kind of device and its regulatory pathway.
The device discussed is a standard medical instrument, an endoscope, not an AI-powered diagnostic or therapeutic device. The "study" mentioned refers to non-clinical performance data and testing to demonstrate that the modified device performs comparably to its predicate and meets safety standards, not a clinical trial involving AI.
Here's an attempt to answer the questions based on the provided text, highlighting where the requested AI-specific information is not applicable:
1. A table of acceptance criteria and the reported device performance
The document states that the device passed all testing in accordance with internal requirements and applicable standards. The acceptance criteria are generally implied to be the successful demonstration that the device's characteristics (mechanical, optical, material, etc.) are equivalent to or better than the predicate device and meet relevant regulatory standards. Specific numerical acceptance criteria were not explicitly provided in the summary, but rather implied by the statement "passed all the testing in accordance with internal requirements and applicable standards to support substantial equivalence of the subject device."
Test Category | Specific Tests | Acceptance Criteria (Implied) | Reported Device Performance |
---|---|---|---|
Mechanical Performance | Tensile strength at joints/connections, coaxiality, deflection endurance, withstand of channel. | Met internal requirements and applicable standards. | Passed all tests. |
Biocompatibility | Cytotoxicity (ISO 10993-5), Irritation (ISO 10993-10), Sensitization (ISO 10993-10), Acute Systemic Toxicity (ISO 10993-11), Material-mediated pyrogenicity (ISO 10993-11). | Met ISO 10993-1 and FDA Guidance requirements for Limited ( |
Ask a specific question about this device
(219 days)
Shanghai AnQing Medical Instrument Co., Ltd.
The Flexible Ureterorenoscope is intended to be used to visualize organs, cavities, and canals in the urinary tract (urethra, bladder, ureter, calyces, and renal papillae) via transurethral access routes. It can also be used in conjunction with endoscopic accessories to perform various diagnostic and therapeutic procedures in the urinary tract.
The Flexible Ureterorenoscope (Model: US31D-12-EU, US31D-12-US) is an endoscope which is used with the Video Processor (Model: EOS-H-01) (FDA cleared K211169) produced by AnQing for providing endoscopic imaging of the ureter and the renal pelvis for the purpose of diagnosis and treatment. The 2 proposed models are identical except the deflection versions, which is opposite from each other (EU version or US version).
The Flexible Ureterorenoscope is a single-use endoscope, which consists of a Handle, an Insertion Section, and an Endoscope Connector. The handle includes a deflection lever, a lever lock, a push button for picture taking/video recording and a Luer port for insertion of accessory devices and irrigation to the working channel. The insertion section contains one working channel and wiring to transmit the image signals to the Video Processor. The distal bending section of the insertion is controlled by the user via the deflection lever on the handle. The distal end of the insertion section contains a CMOS sensor for capturing image and transmitting it to the Video Processor, LEDs for illumination, and the distal opening of the working channel. The endoscope connector connects the endoscope handle to the video processor, which provides power and processes video signals from the endoscope.
Mechanism of action:
The light emitted by the LED cold light source at the distal tip of the disposable Flexible Ureterorenoscope is irradiated into the body cavity, and the light reflected from the cavity enters the optical system and is captured by the CMOS image sensor. The CMOS acquisition image is controlled by the CMOS drive circuit, and the RGB video signal is output to the Video Processor via the VI circuit. The Video Processor receives video signals from the endoscope, processes the video signals, and outputs the processed video signal to the attached monitor. The video processor also controls the brightness of the LEDs on the endoscope.
Flexible Ureterorenoscope has the following physical and performance characteristics:
- Maneuverable tip controlled by the user
- Flexible insertion cord
- Camera and LED light source at the distal tip
- Sterilized by Ethylene Oxide
- For single use
The provided document is a 510(k) premarket notification for a medical device (Flexible Ureterorenoscope). It does not describe an acceptance criteria or report on a study proving the device meets acceptance criteria in the manner requested.
Instead, the document details the device's technical specifications, indications for use, comparison to a predicate device, and a summary of non-clinical bench testing, electrical safety, EMC, photobiological safety, biocompatibility, sterilization, and packaging validation tests performed to demonstrate substantial equivalence to a legally marketed predicate device.
Specifically:
- No acceptance criteria table or reported device performance for a specific study is provided. The document focuses on demonstrating substantial equivalence through various tests, not on meeting predefined performance criteria for a novel clinical function.
- No sample size for a test set, data provenance, number of experts for ground truth, or adjudication method are mentioned. This information would typically be part of a clinical study, which the document explicitly states was not required.
- No multi-reader multi-case (MRMC) comparative effectiveness study was done.
- No standalone (algorithm only) performance study was done. This device is a medical instrument (endoscope), not an AI algorithm.
- No specific type of ground truth is mentioned for performance evaluation. The evaluation relies on engineering and biological safety standards.
- No sample size for a training set or method for establishing ground truth for a training set are provided. This is not relevant for this device type.
The document explicitly states:
- "Summary of Clinical Tests: The subject of this premarket submission did not require clinical studies to support substantial equivalence."
Instead, the document summarizes non-clinical tests to demonstrate substantial equivalence:
- Electrical Safety and Electromagnetic Compatibility: Compliance with ANSI/AAMI ES: 60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012; IEC 60601-2-18:2009; IEC 60601-1-2:2014.
- Photobiological Safety: Compliance with IEC 62471:2006.
- Mechanical and Optical Performance: Designed to comply with applicable parts of ISO 8600. Optical measurements performed according to applicable parts of ISO 8600. Mechanical tests included leakage tightness, bending, deflection endurance, and tensile strength testing. Comparative testing for image quality (color reproduction, resolution, depth of field, image intensity uniformity, distortion, dynamic range) against the predicate device.
- Biocompatibility: Evaluated according to ISO 10993-1 and FDA Guidance, including tests for Cytotoxicity (ISO 10993-5:2009/(R) 2014), Sensitization/Irritation (ISO 10993-10:2010), Material-mediated pyrogenicity (ISO 10993-11:2017), and Acute systemic toxicity (ISO 10993-11:2017).
- Sterilization and Shelf Life Testing: Validation to ISO 11135:2014, EO/ECH residual test to ISO 10993-7:2008. Shelf life (3 years) determined by stability study including accelerated aging (ASTM F1980-16).
- Package Validation: Conducted according to ISO 11607-1:2019 and ISO 11607-2:2019, including visual inspection (ASTM F1886), seal strength (F88/F88M-15), dye penetration (ASTM F1929-15), bubble emission (ASTM D3078), and transport/shipping (ASTM D4169-16).
Ask a specific question about this device
(219 days)
Shanghai AnQing Medical Instrument CO., Ltd.
The flexible bronchoscope have been designed to be used with the video processor, endotherapy accessories and other ancillary equipment for endoscopy within the airways and tracheobronchial tree.
The Bronchoscope System is for use in a hospital environment. The Flexible Bronchoscope is a single-use device designed for use in adults.
The Bronchoscope System consists of Flexible Bronchoscope (twelve models shown in below) to be introduced within the airways or tracheobronchial tree and Video Processor (model: EOS-H-01) for clinical image processing. The Flexible Bronchoscope is inserted through the airways and tracheobronchial tree during Bronchoscopy. The Video Processor provides power and processes the images for medical electronic endoscope.
The Flexible Bronchoscope is a sterile single used flexible bronchoscope. The Video Processor is a reusable monitor.
The light emitted by the LED cold light source at the distal tip of the disposable video bronchoscope is irradiated into the body cavity, and the light reflected from the cavity enters the optical system and is captured by the CMOS image sensor. The CMOS acquisition image is controlled by the CMOS drive circuit, and the RGB video signal is output to the Video Processor via the VI circuit. The Video Processor receives video signals from the endoscope, processes the video signals, and outputs the processed video signal to the attached monitor. The video processor also controls the brightness of the LEDs on the endoscope.
The provided text does not contain information about acceptance criteria and a study proving the device meets those criteria from an AI/ML perspective. The document is a 510(k) summary for a Bronchoscope System, which is a medical device, but it does not describe an AI/ML-driven diagnostic or assistive technology.
The "Performance Data" section details various engineering and biological tests conducted for the Bronchoscope System, such as biocompatibility, sterilization, electrical safety, EMC, software verification, and bench performance testing. These tests ensure the device's physical and functional safety and effectiveness, but they do not involve AI/ML performance metrics like sensitivity, specificity, or AUC against a ground truth.
Specifically, there is no mention of the following in the provided text:
- A table of acceptance criteria and reported device performance for an AI/ML component.
- Sample size, data provenance, number of experts, adjudication method, MRMC study, standalone performance, or ground truth type related to an AI/ML model for a test set.
- Sample size or ground truth establishment for a training set of an AI/ML model.
The document explicitly states under "Clinical Testing": "Based on the similarities of the device specifications, intended use, indications for use between the Bronchoscope System and its predicate device, no clinical studies were needed to support this 510(k) Premarket Notification." This further confirms that no studies involving human interpretation augmented or replaced by AI, or studies evaluating AI performance metrics, were part of this submission.
Therefore, I cannot fulfill your request for information related to AI/ML acceptance criteria and study data from this document.
Ask a specific question about this device
(71 days)
Shanghai AnQing Medical Instrument Co., Ltd.
The Ureterorenoscope System consists of a sterile single-use Flexible Ureteroscope to be introduced within the urinary tract and video processor for clinical image processing. The device is indicated for endoscopic examination in the urinary track and can be used to examine the interior of the kidney, and using additional accessories, to perform various diagnostic and therapeutic procedures.
The Ureterorenoscope System consists of a sterile single-use Flexible Ureteroscope to be introduced within the urinary tract and a non-sterile repeat use Video Processor for clinical image processing.
The provided document is a 510(k) Summary for the AnQing Medical Instrument Co., Ltd. Ureterorenoscope System (K201293). This document declares no clinical performance data was required or conducted to demonstrate substantial equivalence to the predicate device. Therefore, it does not contain the information requested in points 1, 2, 3, 4, 5, 6, 7, 8, and 9 of your prompt regarding acceptance criteria and a study proving the device meets those criteria with specific performance metrics, sample sizes, expert qualifications, or ground truth details.
The document focuses on non-clinical performance data and a comparison to a predicate device (Shanghai AnQing Medical Ureterorenoscope System K180367) to establish substantial equivalence.
Here's a breakdown of the relevant sections from the document that explain why no clinical study or specific performance criteria are described:
- Non-Clinical Performance Data: "The subject and predicate devices have the same fundamental technology, insertion section length, field and direction of view, light source, image display, number of uses and sterilization. The subject ureteroscope differs from the predicate in patient-contacting materials, and size of the primary package. These differences do not raise different questions of safety and effectiveness as compared to the predicate."
- Non-Clinical Performance Data (cont.): "As part of demonstrating substantial equivalence to the predicate, a risk analysis was completed to identify the risks associated with the endoscope material changes and package size change. Verification testing were conducted to evaluate the modifications. The following tests associated with the device modifications were performed on the subject device according to methods and acceptance criteria outlined in the predicate device (K180367). The subject device passed all the testing in accordance with internal requirements and applicable standards to support substantial equivalence of the subject device."
- Clinical Performance Data: "Clinical testing was not required to demonstrate the substantial equivalence to the predicate devices. Non-clinical bench testing was sufficient to establish the substantial equivalence of the modifications."
- Conclusion: "The conclusions drawn from the nonclinical tests demonstrate that the subject device, the Ureterorenoscope System is substantially equivalent to the predicate devices."
Therefore, I cannot provide the requested table or study details because the submission states that no clinical studies were performed for this 510(k) clearance. The device's substantial equivalence was based on non-clinical bench testing and comparison to a predicate device, focusing on material and packaging changes.
Ask a specific question about this device
(332 days)
Shanghai AnQing Medical Instrument CO., Ltd.
The Hysteroscope System is intended to be used to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic and surgical procedures.
Note: Hysteroscopes are used as tools to access the uterine cavity and are not, in and of themselves, a method of surgery.
Generally recognized indications for diagnostic hysteroscopy include:
- Abnormal bleeding
- Infertility and pregnancy wastage
- Evaluation of abnormal hysterosalpingogram
- Intrauterine foreign body
- Amenorrhea
- Pelvic pain
Generally recognized indications for operative hysteroscopy include:
- Directed endometrial biopsy
- Polypectomy
- Submucous myomectomy
- Transection of intrauterine adhesions
- Transection of intrauterine septa
- Endometrial ablation
The Hysteroscope System consists of a sterile single-use disposable Rigid Hysteroscope and video processor for clinical image processing. The Rigid Hysteroscope (available in models GSA01-B, GSB04-B and GSB14-B) are used to permit direct viewing of the cervical canal and the uterine cavity for the purpose of performing diagnostic and surgical procedures. Model GSA01-B is intended for diagnostic hysteroscopy only, and has a rigid shaft without operative channel (injection channel only). Models GSB04-B and GSB14-B are intended for both diagnostic and operative hysteroscopy, and have an operative (insertion) channel. GSB04-B has a rigid shaft, and GSB14-B has a curved tip. The video processor provides power and processes the images from the Rigid Hysteroscopes. Anatomical images are collected via a CMOS chip at the distal end of the hysteroscope and are transmitted through the video processor to a monitor.
The provided text is a 510(k) Summary for a Hysteroscope System. It describes the device, its intended use, and compares it to a predicate device to establish substantial equivalence. However, it does not include information about acceptance criteria and a study proving that the device meets those criteria in the way typically expected for a clinical performance study involving AI or human readers.
The performance data included in this document focuses on non-clinical testing performed to ensure the device's safety and effectiveness in relation to its physical and functional attributes. This is standard for medical devices like hysteroscopes, which are tools for visualization and surgical procedures, rather than AI-driven diagnostic or interpretative devices.
Therefore, the requested information elements related to AI, human readers, effect size, and specific ground truth establishment (like expert consensus, pathology, or outcomes data for complex diagnostic tasks) are not applicable to the content provided.
Here's the information that can be extracted or inferred from the document regarding acceptance criteria and performance data:
Acceptance Criteria and Device Performance for Shanghai AnQing Medical Instrument CO., Ltd. Hysteroscope System (K181545)
The acceptance criteria for the Hysteroscope System are based on engineering and performance standards relevant to a medical device of its type (a hysteroscope). The "study" proving acceptance criteria is a series of non-clinical tests.
1. A table of acceptance criteria and the reported device performance
Acceptance Criteria Category | Reported Device Performance |
---|---|
Biocompatibility | Evaluated per ISO 10993-1:2009 for "Surface – Mucosal Membrane" with "Limited ( |
Ask a specific question about this device
Page 1 of 2