Search Results
Found 1 results
510(k) Data Aggregation
K Number
K212905Device Name
Sotair Device
Manufacturer
Safe BVM Corporation
Date Cleared
2022-08-18
(339 days)
Product Code
BTM
Regulation Number
868.5915Why did this record match?
Applicant Name (Manufacturer) :
Safe BVM Corporation
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Sotair™ device is intended for use with a manual resuscitator with air-tight connections, in non-breathing patients that require flow controlled ventilation using ambient air or a supplemental oxygen source. The Sotair device is a single use, disposable device added to a manual resuscitator and can be used for in-hospital, emergency, and transport care. The adult Sotair device comprises a flow-limiting valve that limits the inspiratory flow enabling providers to ventilate approximately 55 liters-per-minute (LPM). The flow-limiting valve is intended to minimize gastric inflation during manual ventilation. The Sotair device can be disabled by removing the manual resuscitator to its conventional operation. The Sotair device is intended for adult use only.
Device Description
The Sotair™ device is a universal safety accessory device that is compatible with all manual resuscitators and airway interfaces that have standard ISO 5356-1:2015 15/22 mm connections. The Sotair device's flow-limiting function has been tested with the following bag resuscitators: [List of compatible resuscitators provided in the document]. The Sotair device is not compatible with manual resuscitators that have a built-in flow limiting device. The Sotair device uses a flapper valve with a blade design to limit the flow rate.
Ask a Question
Page 1 of 1