Search Results
Found 9 results
510(k) Data Aggregation
(261 days)
Summit Medical Inc.
The InstruSafe® Sterilization Container is a reusable sterilization container system intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. It is intended to allow sterilization of the enclosed device and also maintain sterility for 180 days. The InstruSafe® Sterilization Container includes accessories such as trays, baskets, filters, data cards and tamper proof locks.
The InstruSafe® Sterilization Container consists of various models within the validated product line. All models are available with perforated lids and either perforated or non-perforated bottoms. All models are intended to be used with single use paper filter.
The container is a reusable device designed to be used with the following sterilization cycle parameters:
Pre-Vacuum Steam Sterilization Cycle:
4 minutes
132°C (270 °F)
Drying Time Minimum: 30 minutes
The InstruSafe® Sterilization Container is a reusable sterilization container system intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. They consist of various sizes of bottoms and lids that provide an effective packaging method for sterilization, storage and transportation of surgical instruments by healthcare providers. These containers are suitable for use in pre-vacuum Steam Sterilizers with a tray or basket accessory. Metal (Stainless Steel) surgical instruments like forceps, scissors, clamps, etc. including lumened/cannulated instruments are placed in the trays or baskets for sterilization.
Note: The maximum number of Lumens/Lumen Configuration is as follow:
2 lumens with ≥ 1mm I.D. x ≤ 200mm and a second lumen ≥ 5mm I.D. x ≤ 400mm
The containers are available in different models such as full size (1/1), three-quarter size (3/4), half size (1/2) and XL.
InstruSafe® Sterilization Container is manufactured using aluminum alloy, stainless steel, silicone and the surfaces are protected by a layer of anodized aluminum oxide to prevent corrosion.
The containers must only be used with single use paper filters.
This document describes the regulatory submission for the InstruSafe® Sterilization Container (K180528) and demonstrates its substantial equivalence to predicate devices. The information provided heavily relies on non-clinical performance data rather than a study with human readers or AI algorithms. As such, many of the requested data points related to MRMC studies, training sets, and expert consensus for ground truth are not applicable to this type of device and submission.
Here's the breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:
Acceptance Criteria and Reported Device Performance
The acceptance criteria for this device are demonstrated through various non-clinical performance tests, ensuring its ability to sterilize enclosed medical devices and maintain sterility.
Acceptance Criteria Category | Specific Test/Attribute | Reported Device Performance |
---|---|---|
Device Functionality | Sterilization Efficacy and Thermal Profile (Pre-Vacuum Steam) | Successful; met AAMI TIR12:2010, ANSI/AAMI ST77:2013, and ANSI/AAMI ST79:2013/A4:2013 standards. Tested with lumened/cannulated instruments. |
Dry Time | Successful; met AAMI TIR12:2010, ANSI/AAMI ST77:2013, and ANSI/AAMI ST79:2013/A4:2013 standards. | |
Sterility Maintenance | Maintenance of Sterility (Shelf Life) | Maintained sterility for 180 days after sterilization efficacy testing. Met AAMI TIR12:2010, ANSI/AAMI ST77:2013, and ANSI/AAMI ST79:2010/A3:2012 standards. |
Microbial Aerosol Challenge | Successful; met AAMI TIR12:2010, ANSI/AAMI ST77:2013, and ANSI/AAMI ST79:2010/A3:2012 standards. | |
Biocompatibility | Biocompatibility | Meets requirements of ISO 10993-1, ANSI/AAMI/ISO 10993-5:2009, and ANSI/AAMI/ISO 10993-12:2012. |
Device Performance Under Load | Load Weight Capability | Successfully tested with load weights up to 25 lbs (as claimed in indications for use) during sterilization efficacy and dry time testing. |
Reusable Device Performance | Cleaning/Reusability (Mechanical Cleaning) | Successful; met AAMI TIR12:2010, AAMI TIR30:2011, and ANSI/AAMI ST81:2004(R)2010 standards. |
Mechanical Integrity | Handle Testing | Successful; met ANSI/AAMI ST77:2013 standards. |
Gasket Testing | Successful; met ANSI/AAMI ST77:2013 standards. | |
Lid/Base Compatibility Testing | Successful; met ANSI/AAMI ST77:2013 standards. | |
Locking Tag/Label Compatibility Testing | Successful; met ANSI/AAMI ST77:2013 standards. | |
Design Feature Justification | Absence of security lid to prevent damage/contamination of filter unit during transport (as present in some predicates) | The sterilization containers were tested without a security lid during sterilization efficacy, dry time, microbial aerosol challenge, and shelf life testing and were found to have no failures that were caused by damage or contamination of the filter unit. The subject device functions as intended. The lack of this feature does not compromise safety or effectiveness. |
Material Composition | Materials (Anodized Aluminum Alloy, Stainless Steel, Silicone) | Confirmed to be within acceptable and known safe specifications, comparable to predicate devices. |
Compatibility with Instruments | Compatibility with Stainless Steel Medical Devices | Verified. |
Compatibility with Lumened/Cannulated Instruments | Verified through sterilization efficacy testing. |
Study Proving Device Meets Acceptance Criteria
The studies detailed for the InstruSafe® Sterilization Container are non-clinical, bench-validation studies, rather than clinical trials or AI performance evaluations.
-
Sample Size Used for the Test Set and Data Provenance:
- The document does not specify numerical sample sizes for the non-clinical tests (e.g., number of containers tested, number of sterilization cycles performed, or number of microbial challenges). It states "The sterilization containers were tested..." which implies a sufficient number for validation, but no precise count is given.
- Data Provenance: The studies were non-clinical bench testing. The country of origin for the data is not explicitly stated but can be inferred to be in the USA, given the FDA submission. The studies were designed to be prospective validation tests according to established medical device testing standards.
-
Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
- This is not applicable as the studies are non-clinical, bench-level performance validations for a sterilization container. Ground truth is established by objective measurements against predefined industry standards (e.g., AAMI, ISO) rather than subjective expert interpretation of medical images or data.
-
Adjudication Method for the Test Set:
- Not applicable. Adjudication methods like "2+1" or "3+1" are typical for clinical studies involving human readers and subjective assessments (e.g., radiology reads). For non-clinical, objective performance testing, results are determined by adherence to pre-defined scientific and engineering standards and pass/fail criteria.
-
If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This device is a reusable sterilization container and does not involve AI or human "readers" in the context of medical image interpretation or diagnosis. Therefore, an MRMC comparative effectiveness study was not performed.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Not applicable. This device is a physical medical device (sterilization container), not an AI algorithm.
-
The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.):
- The "ground truth" for this device is based on objective, verifiable measurements against established medical device sterilization, biocompatibility, and mechanical performance standards, such as those published by AAMI (Association for the Advancement of Medical Instrumentation) and ISO (International Organization for Standardization). For example:
- Sterilization Efficacy: Measured by indicator organisms (bioburden reduction) after a sterilization cycle.
- Dry Time: Measured objectively following a sterilization cycle.
- Sterility Maintenance/Microbial Aerosol Challenge: Measured by preventing microbial ingress over time.
- Biocompatibility: In vitro and in vivo tests against ISO 10993 standards.
- Mechanical Integrity: Measured by stress tests, weight limits, and functional checks.
- The "ground truth" for this device is based on objective, verifiable measurements against established medical device sterilization, biocompatibility, and mechanical performance standards, such as those published by AAMI (Association for the Advancement of Medical Instrumentation) and ISO (International Organization for Standardization). For example:
-
The Sample Size for the Training Set:
- Not applicable. This device is hardware, not a machine learning model, so there is no "training set."
-
How the Ground Truth for the Training Set was Established:
- Not applicable. As there is no training set for a machine learning model, this question is irrelevant to the described device.
Ask a specific question about this device
(260 days)
SUMMIT MEDICAL, INC.
The MINNE TIES™ MMF Suture System is an adjustable flexible plastic band that wraps between and around a tooth to create an anchorage point for maxilla-mandibular fixation and immobilization. MINNE TIES™ MMF Suture System is suitable for:
- Pre-operative fixation
- Per-operative fixation
- Short-term (up to 3 weeks) fixation for minimally displaced fractures
- Splintage post jaw dislocation
The MINNE TIES™ MMF Suture System contains:
- Four (4) 1.0mm sized sutures
- Seven (7) 0.7mm sized sutures
- One (1) 0.5mm Sized suture
and are used to establish maxilla-mandibular fixation (MMF) and immobilization by applying independently balanced series of sutures bilaterally between and/or around the teeth in the interdental space (i.e. triangular apical embrasure). The sutures provide a stabilizing occlusive force allowing for short term (up to 3 weeks) fixation for minimally displaced fractures for pre-operative and per-operative fixation and splintage post jaw dislocation. An Introducer is used to advance sutures between and/or around the teeth. A cheek retractor is included with the system to organize the ties during the procedure.
The MINNE TIES™ MMF Sutures are made of polypropylene, polyester and stainless steel. The cheek retractor is made of polypropylene.
The provided text describes a 510(k) premarket notification for the "MINNE TIES™ MMF Suture System". This document indicates the device's substantial equivalence to a legally marketed predicate device rather than a ground-up study proving its performance against acceptance criteria for a novel AI/ML device.
Therefore, the information required to answer the prompt for an AI/ML device (e.g., acceptance criteria for AI performance, sample sizes for AI training/test sets, expert ground truth establishment, MRMC studies, standalone performance, etc.) is not present in the provided document.
The document focuses on:
- Device Description: The components and general use of the MINNE TIES™ MMF Suture System.
- Intended Use/Indications for Use: What the device is intended for (maxilla-mandibular fixation).
- Predicate Device Comparison: A detailed comparison to a previously cleared device (Rapid IMF™ - K030605) to demonstrate substantial equivalence.
- Non-Clinical Testing: Physical and material tests (tensile strength, fatigue, biocompatibility, sterilization, packaging, shelf life, corrosion resistance, system handling/cadaver test) to show that the new device performs similarly and safely to the predicate.
- Clinical Testing: Explicitly states that clinical data was not required for this device, which further indicates the absence of human-in-the-loop performance studies or deep learning-based performance evaluations.
In summary, this document is for a medical device that does not involve artificial intelligence or machine learning, and thus the specific metrics and study methodologies requested in the prompt related to AI/ML device acceptance criteria are not applicable or provided.
Ask a specific question about this device
(109 days)
Summit Medical Inc.
Instru-Safe® Instrument Protection System cassettes are used to organize and protect other medical devices that are sterlized by a healthcare provider. Instrument Protection System cassettes are intended to allow sterilization of the enclosed medical devices during a Sterilization Cycle. The Instru-Safe Instrument Protection System cassettes are intended to be used in conjunction with a legally marketed wrap or Aesculap rigid container. The Instru-Safe Instrument Protection System cassettes are not intended on their own to maintain sterility. A full list of device models is provided in table 1.
Sterilization methods and configurations - Sterrad NX Standard Sterilization Cycle
Lumen claims for Sterrad NX Standard Sterilization Cycle
Summit Cassette Model | Minimum Inside Diameter | Maximum Length | Number of Lumens |
---|---|---|---|
IN-2681 | 1 mm | 65 mm | 1 |
IN-2681 | 3 mm | 200 mm | 1 |
IN-8987-S | 1 mm | 500 mm | 5 |
IN-8615 | 2.3 mm | 210 mm | 5 |
IN-6105 | 4 mm | 235 mm | 1 |
The worst case validated load by vent-to-volume calculation is the IN-2681 tray.
Summit Medical Inc. InstruSafe Instrument Protection System are cassettes / trays used to enclose and hold surgical instruments and instrument accessories in an organized manner during the sterilization process and subsequent storage and transportation. The cassettes / trays do not have direct patient contact. The cassettes / trays by themselves do not maintain sterility.
The cassettes / trays are different sizes of the same basic configuration: a rectangular base with latchable cover. The cassettes / trays have perforations to allow sterilant penetration. The cassettes / trays contain silicone inserts in the base and/or cover to hold, organize and protect the surgical instruments within the cassette / tray during the sterilization process and subsequent storage and transportation.
InstruSafe Instrument Protection System cassettes include
- perforated base
- perforated cover
- silicone inserts (hold-it / hold down)
- Handles
- Latches
- Feet
- Posts (optional)
- Divider (optional)
- Shelf (optional)
The cassette contains components made of anodized aluminum, stainless steel, blue silicone, black silicone, polyester, ultem™ 1000
The provided document is a 510(k) premarket notification for the InstruSafe® Instrument Protection System. It describes the device, its intended use, and a comparison to a predicate device, focusing on sterilization methods.
However, the document does not contain information typically associated with acceptance criteria and a study proving a device meets those criteria in the context of an AI/ML medical device. This document is for a non-AI medical device and primarily focuses on demonstrating substantial equivalence to a predicate device, particularly regarding a new sterilization cycle. Therefore, I cannot extract the requested information as it is not present.
Specifically, the document does NOT contain information on:
- A table of acceptance criteria and reported device performance for an AI/ML algorithm.
- Sample sizes for test sets, data provenance, number of experts, adjudication methods, MRMC studies, standalone performance, type of ground truth, or training set sample size related to an AI/ML device.
The information provided pertains to the physical and functional characteristics of instrument protection cassettes and their compatibility with a specific sterilization cycle. The "performance data" mentioned refers to sterilization validation testing, not AI algorithm performance.
Ask a specific question about this device
(111 days)
SUMMIT MEDICAL, INC.
Instru-Safe® Instrument Protection System cassettes are used to organize and protect other medical devices that are sterilized by a healthcare provider. Instru-Safe Instrument Protection System cassettes are intended to allow sterilization of the enclosed medical devices during Amsco V-PRO Low Temperature Sterilization Cycles. The Instru-Safe Instrument Protection System cassettes are intended to be used in conjunction with a legally marketed wrap or Aesculap rigid container. The Instru-Safe Instrument Protection System cassettes are not intended on their own to maintain sterility. A full list of device models is provided in table 1.
Summit Medical Inc. Instru-Safe Instrument Protection System are cassettes / trays used to enclose and hold surgical instruments and accessories in an organized manner during the sterilization process and subsequent storage and transportation. The cassettes / trays by themselves do not maintain sterility. The cassettes / trays are different sizes of the same basic configuration: a rectangular base with latchable cover. The cassettes / trays have perforations to allow sterilant penetration. The cassettes / trays contain silicone inserts in the base and/or cover to hold, organize and protect the surgical instruments within the cassette / tray.
The provided document is a 510(k) premarket notification for a medical device (Instru-Safe® Instrument Protection System), which focuses on demonstrating substantial equivalence to a predicate device, rather than proving that the device meets specific acceptance criteria through a clinical study. The document primarily details the device's intended use and compares its characteristics and performance to a previously cleared predicate device.
Therefore, many of the requested elements for describing an acceptance criteria study, such as sample size for test sets, ground truth establishment for test and training sets, the number and qualifications of experts, adjudication methods, or MRMC studies, are not explicitly mentioned or applicable in this type of regulatory submission.
However, based on the information provided, here's an attempt to address the request as much as possible:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implied by the intended use and the predicate device's performance. The "Performance Data" section in the summary directly addresses this.
Acceptance Criterion (Implied) | Reported Device Performance |
---|---|
Sterilant Penetration in Amsco V-PRO Low Temperature Sterilization Systems (Standard, Lumen, Non Lumen Cycles) | Sterilization validation testing was performed to demonstrate Instru-Safe Instrument Protection System compatibility when used in an Amsco V-PRO Low Temperature Sterilization Systems with a legally marketed wrap or Aesculap rigid container. |
Material Compatibility | No changes from predicate device (K133015). |
Toxicological Properties (Biocompatibility/Cytotoxicity) | MEM Elution Cytotoxicity (ISO 10993-5): Test samples met USP and ISO 10993-5 requirements; considered non-cytotoxic under test conditions. |
Shelf Life | No Change (Reusable, 5-year accelerated shelf life study for predicate device). |
2. Sample size used for the test set and the data provenance
The document does not specify a separate "test set" in the context of an AI/algorithm study. For the sterilization validation, the specific number of units tested (cassettes/trays, instruments, sterilization cycles) is not provided. The data provenance would be from internal testing conducted by Summit Medical Inc. as part of their regulatory submission. The study is prospective in nature for a new device attempting to be cleared for the specified sterilization cycles.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This document does not describe the establishment of ground truth by experts for a test set in the context of an AI/algorithm. The "ground truth" for sterilization effectiveness would be determined by standard microbiological and physical indicator tests, which do not typically involve human expert interpretation in the same way as, for example, image analysis.
4. Adjudication method for the test set
Not applicable. There is no mention of adjudication for a test set as this is not an AI/algorithm performance study.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI-assisted device, and no MRMC study was performed or described.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This device is a physical instrument protection system, not a software algorithm.
7. The type of ground truth used
For the sterilization validation, the ground truth would be established through sterilization effectiveness testing using biological and chemical indicators, which are standard methods to confirm the complete inactivation of microorganisms. For the biocompatibility claim, the ground truth was established by laboratory testing against ISO 10993-5 requirements.
8. The sample size for the training set
Not applicable. This is not an AI/algorithm, so there is no training set.
9. How the ground truth for the training set was established
Not applicable. As there is no training set mentioned.
Ask a specific question about this device
(246 days)
SUMMIT MEDICAL, INC.
Instru-Safe® Instrument Protection System cassettes are used to organize and protect other medical devices that are sterilized by a healthcare provider. Instrument Protection System cassettes are intended to allow sterilization of the enclosed medical devices during a Sterrad 100NX Flex Sterilization Cycle. The Instru-Safe Instrument Protection System cassettes are intended to be used in conjunction with a legally marketed wrap or Aesculap rigid container. The Instru-Safe Instrument Protection System cassettes are not intended on their own to maintain sterility. A full list of device models is provided in table 1.
Summit Medical Inc. Instru-Safe Instrument Protection System are cassettes / trays used to enclose and hold surgical instruments and accessories in an organized manner during the sterilization process and subsequent storage and transportation. The cassettes / trays by themselves do not maintain sterility. The cassettes / trays are different sizes of the same basic configuration: a rectangular base with latchable cover. The cassettes / trays have perforations to allow sterilant penetration. The cassettes / trays contain silicone inserts in the base and/or cover to hold, organize and protect the surgical instruments within the cassette / tray.
The provided document is a 510(k) premarket notification for a medical device (Instru-Safe® Instrument Protection System) and describes the device's intended use and performance testing related to sterilization. It is not an AI/ML device, and therefore, many of the requested categories for AI/ML studies are not applicable.
Here's a breakdown based on the provided text, while noting the irrelevance of AI-specific questions:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria (Implied from Performance Data Section) | Reported Device Performance (from "Performance Data" and "Performance" sections) |
---|---|
Sterilization compatibility (allowing sterilization of enclosed medical devices during a Sterrad 100NX Flex Sterilization Cycle with legally marketed wrap or Aesculap rigid container) | Sterilization validation testing was performed to demonstrate Instru-Safe Instrument Protection System compatibility when used in a Sterrad 100NX Flex Sterilization Cycle with a legally marketed wrap or Aesculap rigid container. This new sterilization cycle does not affect safety and effectiveness of the Instru-Safe Instrument Protection System. The device's "Performance" section explicitly states "Sterrad 100NX Flex Sterilization Cycle" under "Sterilant Penetration." |
Material Biocompatibility (cytotoxicity) | MEM Elution Cytotoxicity (ISO 10993-5) test samples met USP and ISO 10993-5 requirements, passed, and are considered non-toxic. |
Shelf Life (reusability) | Reusable (5-year accelerated shelf life study). "No Change" (from predicate device's shelf life). |
2. Sample size used for the test set and the data provenance
The document does not specify a numerical sample size for the sterilization validation testing. It mentions "Sterilization validation testing was performed" for compatibility with the Sterrad 100NX Flex Sterilization Cycle. The data provenance is not explicitly stated in terms of country of origin, nor whether it was retrospective or prospective. Given the nature of validation for a physical medical device, it would most likely be prospective testing conducted in a laboratory or clinical setting.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. The "ground truth" for this device involves demonstrating the physical property of sterilization compatibility and material biocompatibility, which are typically assessed through standardized laboratory tests, not expert interpretation of diagnostic data.
4. Adjudication method for the test set
Not applicable. This is not a study involving human interpretation of data requiring adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is not an AI/ML device.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. This is not an AI/ML device.
7. The type of ground truth used
The ground truth used would be based on the validated efficacy of the Sterrad 100NX Flex Sterilization Cycle to sterilize medical devices and the results of ISO 10993-5 cytotoxicity testing. This is objective, measurable data from physical and biological tests, not expert consensus, pathology, or outcomes data in a diagnostic sense.
8. The sample size for the training set
Not applicable. This is not an AI/ML device.
9. How the ground truth for the training set was established
Not applicable. This is not an AI/ML device.
Ask a specific question about this device
(97 days)
Summit Medical Inc.
Instru-Safe® Instrument Protection System cassettes are used to organize and protect other medical devices that are sterilized by a healthcare provider. Instru-Safe Instrument Protection System cassettes are intended to allow sterilization of the enclosed medical devices during a pre-vacuum steam sterilization cycle. The Instru-Safe Instrument Protection System cassettes are intended to be used in conjunction with a legally marketed Genesis rigid containers. The Instru-Safe Instrument Protection System cassettes are not intended on their own to maintain sterility.
Summit Medical Inc. Instru-Safe Instrument Protection System are cassettes / trays used to enclose and hold surgical instruments and accessories in an organized manner during the sterilization process and subsequent storage and transportation. The cassettes / trays by themselves do not maintain sterility. The cassettes / trays are different sizes of the same basic configuration: a rectangular base with latchable cover. The cassettes / trays have perforations to allow sterilant penetration. The cassettes / trays contain silicone inserts in the base and/or cover to hold, organize and protect the surgical instruments within the cassette / tray.
The provided document is a 510(k) summary for the Instru-Safe® Instrument Protection System, which is a sterilization tray. This document focuses on demonstrating substantial equivalence to a predicate device (K133015) rather than on presenting a clinical study or performance data of a new AI device. Therefore, many of the requested elements for describing an AI device's acceptance criteria and study proving its performance (such as sample size for test/training sets, data provenance, number/qualifications of experts, adjudication methods, MRMC studies, standalone performance, or type of ground truth) are not applicable or cannot be extracted from this document.
The document discusses the performance data in terms of sterilization validation testing for the device's ability to allow sterilization of enclosed medical devices.
Acceptance Criteria and Reported Device Performance
The acceptance criteria are implied by the "Indications for Use" and the "Sterilization methods and configurations" which specify the conditions under which the device is intended to function effectively for sterilization. The reported performance is that the device has passed "Sterilization validation testing" under these conditions.
Acceptance Criteria Category | Specific Criteria | Reported Device Performance |
---|---|---|
Intended Use | Organize and protect other medical devices sterilized by a healthcare provider. Allow sterilization of enclosed medical devices during a pre-vacuum steam sterilization cycle. Used in conjunction with legally marketed Genesis rigid containers. Not intended on their own to maintain sterility. | The device's indications for use align with this, and its performance is validated through sterilization testing. |
Sterilization Parameters | Cycle: Pre-Vacuum | Sterilization validation testing was performed to demonstrate compatibility with a pre-vacuum steam sterilization cycle. |
Temperature: 270°F (132°C) | Sterilization validation testing was performed at 270°F (132°C). | |
Minimum Exposure Time: 4 minutes | Sterilization validation testing was performed with a 4-minute exposure time. | |
Minimum Dry Time: 30 minutes | Sterilization validation testing was performed with a 30-minute dry time. | |
Lumen Claims | Specific dimensions for minimum inside diameter, maximum length, and number of lumens for various Summit Cassette Models (e.g., IN-2681: 3mm ID, 200mm length, 1 lumen; IN-0000: 1mm ID, 400mm length, 5 lumens). | The device's performance is demonstrated for the specified lumen claims in the table. |
Worst Case Load | IN-2681 tray identified as the worst case validated load by vent-to-volume calculation. | The system is validated for the worst-case load. |
Total Weight Limit | Total weight of the container system (container, tray, instrument load) must not exceed 25 lbs. | Not explicitly stated as "met," but implied by the successful sterilization validation. The table for "Estimated Weight - Tray w/ instruments" for various part numbers shows individual tray weights well below 25 lbs, so the cumulative system weight would need to adhere to this. |
Material Composition | No changes from predicate device; components made of anodized aluminum, stainless steel, blue silicone, black silicone, polyester, ultem™ 1000. | Confirmed as "No changes from predicate device." |
Physical Properties | Includes perforated base, perforated cover, silicone inserts, handles, latches, feet, and optional posts, dividers, and shelves. | Confirmed as identical to the predicate device. |
Toxicological Properties | No change from predicate device; biocompatibility, including sterilant residue limits. | Confirmed as "No change." |
Shelf Life | Reusable (5-year accelerated shelf life study). | Confirmed as "No change." |
Since the provided text is for a traditional medical device (sterilization tray) and not an AI/ML powered device, the following points are largely not applicable. However, I will answer them based on what can be inferred or explicitly stated, noting the limitations.
-
Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size for Test Set: The document does not specify a distinct "test set" sample size in terms of cases or patients, as this is a device validation, not a clinical trial evaluating diagnostic or predictive accuracy. The "test" in this context refers to the sterilization validation testing of the physical device models and lumen configurations. For lumen claims, specific models like IN-2681, IN-0000, and IN-6105 with various lumen dimensions were tested. The "worst case validated load" (IN-2681 tray) suggests testing was performed to cover the most challenging scenarios.
- Data Provenance: Not applicable in the context of clinical data. The performance data is derived from laboratory-based sterilization validation testing of the physical devices.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable. The "ground truth" for a sterilization device is whether it effectively sterilizes, which is determined by microbiological and physical testing methods, not expert human interpretation of medical images or pathologies.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable. Adjudication methods are relevant for subjective interpretations, whereas sterilization effectiveness is assessed objectively through validated protocols (e.g., biological indicators, chemical indicators, physical parameters).
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. MRMC studies are used for evaluating diagnostic performance of imaging modalities or AI systems that assist human readers. This device is a physical sterilization tray.
-
If a standalone (i.e. algorithm only without human-in-the loop performance) was done
- Not applicable. This is not an algorithm or AI system. Its performance is inherent in its design and its ability to withstand and facilitate sterilization processes.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- The "ground truth" for this device's performance is the successful completion of sterilization validation testing, typically involving:
- Biological Indicators: Confirmation of microbial kill (e.g., Geobacillus stearothermophilus spores).
- Chemical Indicators: Verification of exposure to sterilant.
- Physical Monitoring: Verification of temperature, pressure, and exposure time within the sterilizer.
- Dryness Assessment: Ensuring instruments are dry after the cycle.
- The document states "Sterilization validation testing was performed to demonstrate Instru-Safe Instrument Protection System compatibility when used in a Four (4) minute pre-vacuum steam sterilization cycle within a Genesis rigid container." This implies the use of standard sterilization validation protocols to confirm efficacy.
- The "ground truth" for this device's performance is the successful completion of sterilization validation testing, typically involving:
-
The sample size for the training set
- Not applicable, as this is not an AI/ML device that requires a training set.
-
How the ground truth for the training set was established
- Not applicable, as there is no training set for this device.
Ask a specific question about this device
(189 days)
SUMMIT MEDICAL, INC.
Instru-Safe® Instrument Protection System cassettes are used to organize and protect other medical devices that are sterilized by a healthcare provider. Instrument Protection System cassettes are intended to allow sterilization of the enclosed medical devices during a Sterrad® 100NX Express Cycle. The Instru-Safe Instrument Protection System cassettes are intended to be used in conjunction with a legally marketed wrap. The Instru-Safe Instrument Protection System cassettes are not intended on their own to maintain sterility. The Instru-Safe Instrument Protection System has no lumen claims for the Sterrad 100NX Express Cycle.
Instru-Safe® Instrument Protection System cassettes are used to organize and protect other medical devices that are sterilized by a healthcare provider.
The provided text is a 510(k) premarket notification for a medical device called the "Instru-Safe® Instrument Protection System." This document is a regulatory filing with the FDA and does not contain information about the acceptance criteria or a study proving the device meets those criteria in the context of an AI/ML medical device.
The device described, an "Instrument Protection System," is a sterilization tray used to organize and protect other medical devices during sterilization. It is a traditional medical device, not an AI/ML diagnostic or therapeutic system. Therefore, the questions related to AI/ML device performance metrics (such as effect size of human readers with AI assistance, standalone algorithm performance, training set details, or ground truth establishment for AI models) are not applicable to this document.
The document lists "Indications for Use" which specify how the device is intended to be used, but these are not "acceptance criteria" in the sense of performance metrics for an AI/ML algorithm. The table on page 4/6 provides technical specifications for different parts of the "Instru-Safe® Instrument Protection System" (e.g., maximum number of instruments, estimated weight), which are characteristics of the physical device rather than performance metrics of a diagnostic algorithm.
Therefore, I cannot extract the requested information regarding acceptance criteria and a study proving device performance as it pertains to an AI/ML medical device from the provided document.
Ask a specific question about this device
(142 days)
SUMMIT MEDICAL, INC.
Instru-Safe ® Instrument Protection System cassettes are used to organize and protect other medical devices that are sterilized by a healthcare provider. Instrument Protection System cassettes are intended to allow sterilization of the enclosed medical devices during a Sterrad 100S Standard and Sterrad 100NX Standard sterilization cycles. The Instrument Protection System cassettes are intended to be used in conjunction with legally marketed wrap or Aesculap rigid container. The Instrument Protection System cassettes are not intended on their own to maintain sterility. A full list of device models is provided in table 1.
Instru-Safe® Instrument Protection System cassettes are used to organize and protect other medical devices that are sterilized by a healthcare provider.
The provided text is a 510(k) premarket notification letter from the FDA regarding a medical device, the "Instru-Safe Instrument Protection System." This document describes the device, its intended use, and lists various models. However, it does not contain information about acceptance criteria, device performance studies, sample sizes, expert qualifications, or ground truth establishment.
Based on the provided text, I can only provide the following information:
Device Description:
- Trade/Device Name: Instru-Safe Instrument Protection System
- Regulation Number: 21 CFR 880.6850
- Regulation Name: Sterilization Tray
- Regulatory Class: II
- Product Code: KCT
- Manufacturer: Summit Medical Incorporated
Intended Use/Indications for Use:
"Instru-Safe® Instrument Protection System cassettes are used to organize and protect other medical devices that are sterilized by a healthcare provider. Instrument Protection System cassettes are intended to allow sterilization of the enclosed medical devices during a Sterrad 100S Standard and Sterrad 100NX Standard sterilization cycles. The Instrument Protection System cassettes are intended to be used in conjunction with legally marketed wrap or Aesculap rigid container. The Instrument Protection System cassettes are not intended on their own to maintain sterility."
Sterilization Validation:
The device models were validated with specified rigid containers for:
- Sterrad 100S Standard Cycle
- Sterrad 100NX Standard Cycle
Lumen Claims for Sterrad 100S Standard and Sterrad 100NX Standard Cycles:
- 4 Stainless steel lumens with 3mm inner diameter × 400mm length (All appropriate sized models listed in Table 1 with the exception of IN-2681)
- 2 Stainless steel lumens with 3mm inner diameter × 200mm length
- 1 Stainless steel lumens with 3mm inner diameter × 200mm length (IN-2681)
- 1 Stainless steel lumens with 1mm inner diameter × 65mm length
Missing Information:
The document does not contain any of the following information explicitly related to acceptance criteria or a study proving the device meets them:
- A table of acceptance criteria and the reported device performance.
- Sample size used for the test set and the data provenance.
- Number of experts used to establish the ground truth for the test set and their qualifications.
- Adjudication method for the test set.
- If a multi-reader multi-case (MRMC) comparative effectiveness study was done, or the effect size of human readers improvement with AI vs. without AI assistance. (This device is a physical sterilization tray, not an AI software, so this section is not applicable).
- If a standalone (i.e. algorithm only, without human-in-the-loop performance) was done. (Not applicable for a physical device).
- The type of ground truth used (expert consensus, pathology, outcomes data, etc.).
- The sample size for the training set. (Not applicable for a physical device, as there is no "training set" in the AI/ML sense).
- How the ground truth for the training set was established. (Not applicable).
To obtain the missing information, one would need to review the actual submission documents (e.g., test reports, validation studies) that were provided to the FDA by Summit Medical Incorporated for K140289. The letter provided here is merely the FDA's decision letter based on that submission.
Ask a specific question about this device
(169 days)
SUMMIT MEDICAL, INC.
Instru-Safe® Instrument Protection System cassettes are used to organize and protect other medical devices that are sterilized by a healthcare provider. Instrument Protection System cassettes are intended to allow sterilization of the enclosed medical devices during a pre-vacuum steam sterilization cycles. The Instru-Safe Instrument Protection System cassettes are intended to be used in conjunction with a legally marketed wrap or Aesculap rigid container. The Instru-Safe Instrument Protection System cassettes are not intended on their own to maintain sterility.
Not Found
The provided text is a 510(k) premarket notification from the FDA regarding the "Instru-Safe® Instrument Protection System." This document is primarily concerned with regulatory approval and does not contain information about the acceptance criteria or a study proving the device meets said criteria in the context of a "study" related to AI device performance.
The device in question, the "Instru-Safe® Instrument Protection System," is classified as a "Sterilization Wrap" (21 CFR 880.6850). This is a physical medical device used to organize and protect other medical devices during pre-vacuum steam sterilization cycles.
Therefore, the requested information for an AI-related study (such as sample size for test/training sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, and effect size with AI assistance) is not applicable to this document.
The document states the Intended Use of the device, which acts as the de facto acceptance criteria in terms of functionality. These are:
- To organize and protect other medical devices.
- To allow sterilization of the enclosed medical devices during pre-vacuum steam sterilization cycles.
- To be used in conjunction with a legally marketed wrap or Aesculap rigid container.
- Not intended on their own to maintain sterility.
The document also specifies the validated sterilization parameters:
- Cycle: Pre-vacuum
- Temperature: 270°F (132°C)
- Exposure Time: 4 minutes
- Minimum Dry Time: 30 minutes
And for certain models used with Aesculap containers:
- Sterilization: Steam prevacuum sterilizers ONLY operating at 270°F (132°C) for 4 minutes exposure time.
And specific lumen size validations:
- Minimum Inside Diameter: 1mm, 3mm, 5mm
- Maximum Length: 65mm, 76mm, 177mm, 200mm, 241mm, 400mm
- Number of Lumens: 1, 2, 4, 17
This information pertains to the physical and functional characteristics of the sterilization cassettes, ensuring they are compatible with the specified sterilization processes and can hold certain instruments. It does not involve any AI or software validation.
In summary, none of the requested AI-related study information can be extracted from this document because the device described is a physical medical device (sterilization wrap/cassette), not an AI/software as a medical device.
Ask a specific question about this device
Page 1 of 1