Search Filters

Search Results

Found 75 results

510(k) Data Aggregation

    Why did this record match?
    Applicant Name (Manufacturer) :

    Shanghai SIGMA High-tech Co., Ltd.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    UV phototherapy device is intended to be used for the treatment of psoriasis, vitiligo and atopic dermatitis (eczema). It is to be used on intact skin only. Patients shall use it at home under the guidance of the physician.

    Device Description

    UV phototherapy device is intended to be used for the treatment of psoriasis, vitiligo, atopic dermatitis, and eczema. It is to be used on intact skin only. Patients shall use it at home under the quidance of the doctor.

    Users can set UV irradiation dose through the controller, and confirm the light emission by pressing the start button. The controller will control the light emitting time of the set dose. The light emission will stop automatically after the UV irradiation dose reaches the set value.

    AI/ML Overview

    The provided text is a 510(k) Summary for a UV Phototherapy Device. It describes the device, its intended use, and a comparison to predicate devices, but it does not contain information about acceptance criteria or specific study results that prove the device meets acceptance criteria.

    The document states that "Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device." It then lists several international standards (IEC and ISO) with which the device complies. However, it does not provide:

    • A table of actual acceptance criteria (e.g., minimum efficacy rates, safety thresholds) for the device's clinical performance.
    • The reported device performance against such criteria.
    • Details about specific studies (clinical or non-clinical) that "prove" the device meets these criteria in terms of clinical outcomes for psoriasis, vitiligo, and atopic dermatitis.
    • Information on sample sizes for test sets or training sets, data provenance, expert qualifications, adjudication methods, MRMC studies, standalone performance, or ground truth establishment.

    The document focuses on substantiating equivalence to predicate devices primarily through technological comparison and compliance with electrical safety, EMC, and biocompatibility standards, rather than direct evidence of clinical efficacy meeting predefined acceptance criteria for the treatment of dermatologic conditions.

    Therefore, I cannot fulfill your request for a table of acceptance criteria and reported device performance, or details about the studies and ground truth, as this information is not present in the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223912
    Manufacturer
    Date Cleared
    2023-08-17

    (231 days)

    Product Code
    Regulation Number
    872.3930
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    SigmaGraft Inc.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    InterCollagen® Guide alone or in combination with suitable augmentation materials (like autogenous bone, allogeneic, xenogeneic or alloplastic bone replacement materials) is immediate or delayed guided tissue and bone regeneration.

    • · in the context of a treatment of fenestration defects
    • · in case of dehiscence defects
    • · after apicoectomy and resection of retained teeth
    • · in extraction sockets after tooth extractions
    • · in case of immediate or delayed augmentation around implants in extraction sockets
    Device Description

    InterCollagen® Guide is a resorbable collagen membrane, derived from porcine pericardium. InterCollagen® Guide is intended for periodontal and/or dental surgical procedures as a barrier membrane restricting the entry of rapidly proliferating non-osteogenic cells within the bone defect while allowing the ingrowth of slow-growing bone forming cells. The membrane is a bioresorbable barrier which eventually is remodeled and/or incorporated by the host tissue. InterCollagen® Guide is substantially resorbed within 15 weeks after implantation. It is adaptable and easy to handle. It can be trimmed to the desired size and conforms easily when hydrated. The product is terminally sterilized via gamma irradiation.

    AI/ML Overview

    This FDA 510(k) summary describes the InterCollagen® Guide, a resorbable collagen membrane. The submission seeks to prove substantial equivalence to a predicate device, the Straumann Jason Membrane, rather than demonstrate strict acceptance criteria and performance against those criteria. Therefore, the response will be structured to extract relevant performance claims and the study details provided, even if they don't explicitly fit a "table of acceptance criteria" format with numerical targets.

    1. Table of Acceptance Criteria and Reported Device Performance

    As this is a 510(k) submission seeking substantial equivalence, explicit numerical "acceptance criteria" with defined thresholds are not typically presented in the same way as a de novo or PMA submission. Instead, the focus is on demonstrating similar performance to a legally marketed predicate device. The following table summarizes the key performance aspects that were evaluated and the results presented, which essentially serve as the "reported device performance" against the implicitly accepted standard of the predicate device's performance.

    Performance AspectAcceptance Standard (Implicitly based on Predicate Device)Reported Device Performance (InterCollagen® Guide)
    BiocompatibilityMeets ISO 10993 requirements for medical devices.Non-mutagenic (Mouse Lymphoma, Ames Test)
    (General)Non-sensitizer (Kligman Maximization)
    Non-irritant (Intracutaneous irritation)
    Non-cytotoxic (Cytotoxicity L929 Neutral Red uptake)
    Non-toxic (Acute Systemic Toxicity, 90-day Subchronic Toxicity)
    Non-pyrogenic (Pyrogenicity)
    Minimum tissue reaction at 2, 11, and 15 weeks of implantation; no adverse tissue reaction to the host (Local effects after Implantation)
    In Vivo PerformancePerformance substantially equivalent to the predicate device (Jason membrane) in a canine model, with similar histological, histomorphometric, and micro-CT outcomes.Performed in a manner substantially equivalent to the cleared predicate device across pathology, histology, histomorphology, and micro-CT endpoints.
    Resorption TimeSubstantially resorbed by 12 weeks (Predicate device: Straumann Jason Membrane)Substantially resorbed by 15 weeks
    Sterility AssuranceSterility Assurance Level (SAL) of 10^-6 (Predicate device also achieves this)Achieves a Sterility Assurance Level of 10^-6 (via Irradiation)
    Viral InactivationMeets ISO 22442-3Viral inactivation studies performed in accordance with ISO 22442-3 to ensure viral safety.
    Pyrogenicity (Batch)Non-pyrogenic (release test)Each batch tested for endotoxin (LAL test, USP , USP ) as finished product release test.
    Shelf Life/StabilityPerformance testing of packaging system; Selection, qualification, and validation of packaging.Product and packaging stability determined using real-time aging data; Packaging tested per ASTM D4169, ISO 11607.

    2. Sample size used for the test set and the data provenance

    The primary "test set" for performance evaluation was the canine two-wall intrabony defect model.

    • Sample Size: Not explicitly stated in terms of the number of animals or defects. The document mentions "End points for pathology, histology, histomorphology and micro-CT were taken after 2, 6, and 13 weeks," implying a longitudinal study.
    • Data Provenance: Conducted as an animal study (canine model). The country of origin is not specified in the provided text. The study design was a comparison between the test article (InterCollagen® Guide), a predicate device (Jason membrane), and an empty control. This is a prospective study design for evaluating the device's performance post-implantation.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not explicitly state the number or qualifications of experts used to establish ground truth for the animal study. Histology, histomorphology, and pathology evaluations are typically conducted by trained veterinary pathologists or researchers specialized in bone regeneration, but specific details are not provided.

    4. Adjudication method for the test set

    The document does not specify any adjudication method for establishing ground truth in the animal study. Evaluations of pathology, histology, histomorphology, and micro-CT are usually performed by experts, but whether independent reviews or consensus methods were used is not mentioned.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A multi-reader multi-case (MRMC) comparative effectiveness study was not performed. This device is a resorbable collagen membrane for guided bone and tissue regeneration, not an AI-assisted diagnostic or interpretative tool. Therefore, the concept of "human readers improve with AI" is not applicable here.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    A standalone performance study of an algorithm was not performed. This is a medical device (collagen membrane), not a software or AI/ML algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the animal study:

    • Pathology: Evaluation of pathological changes in tissues.
    • Histology: Microscopic examination of tissue samples.
    • Histomorphology: Quantitative analysis of tissue morphology (e.g., bone formation).
    • Micro-CT: Three-dimensional imaging for bone volume and structure analysis.

    These are considered objective biological and imaging endpoints, typically interpreted by experts, to establish the "ground truth" regarding the device's in-vivo performance and tissue response.

    8. The sample size for the training set

    This is not applicable. This device is a physical medical implant, not an AI/ML algorithm that requires a "training set."

    9. How the ground truth for the training set was established

    This is not applicable as there is no training set for this type of medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221808
    Manufacturer
    Date Cleared
    2023-03-18

    (269 days)

    Product Code
    Regulation Number
    872.3930
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    SigmaGraft Inc.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    InterOss® Collagen is indicated for filling of extraction sockets to enhance preservation of the alveolar ridge. InterOss Collagen is recommended for:

    Filling of extraction sockets to enhance preservation of the alveolar ridge

    Filling of periodontal defects in extraction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR)

    Device Description

    InterOss Collagen is a combination of InterOss®, an anorganic hydroxyapatite bone substitute, and collagen fibers for use in periodontal, oral and maxillofacial surgery. This product is a composite of 90% InterOss® (granules of size 0.25-1mm) and 10% porcine collagen fibers. InterOss®, which is already a cleared device by the FDA (K151209), is a hydroxyapatite material derived from Australian bovine bone. The osteoconductive mineral structure is produced from bone through a multi-step purification process. Following placement in bony voids or gaps InterOss® acts as an osteoconductive scaffold for the ingrowth of adjacent viable bone. InterOss® gradually resorbs and is replaced with bone during the healing process. The collagen component facilitates the adaptation of Inter Osse to the allowing easier handling. The product is non-pyrogenic, single use only, and terminally sterilized via gamma-irradiation.

    The product is available in the following shapes and sizes:

    | Type | Weight
    (mg) | Dimension
    (mm) | Ref# |
    |-------|----------------|-------------------|----------|
    | Block | 50 | 6 x 6 x 3 | IOC-50 |
    | Block | 100 | 6 x 6 x 6 | IOC-100 |
    | Block | 250 | 7 x 9 x 8 | IOC-250 |
    | Block | 350 | 8 x 10 x 9 | IOC-350 |
    | Block | 500 | 10 x 12 x 10 | IOC-500 |
    | Plug | 150 | 6 x 10 | IOC-P150 |
    | Plug | 250 | 8 x 10 | IOC-P250 |
    | Plug | 400 | 11 x 9 | IOC-P400 |
    | Plug | 450 | 10 x 12 | IOC-P450 |

    AI/ML Overview

    The provided document is a 510(k) premarket notification for the InterOss® Collagen bone grafting material. It describes the device, its intended use, and a comparison to predicate devices, along with non-clinical testing performed to demonstrate substantial equivalence.

    However, the document does not contain information about a study proving the device meets acceptance criteria in the context of an Artificial Intelligence (AI) or machine learning device. The "acceptance criteria" and "device performance" in the tables are for the physical and chemical properties and biocompatibility of the bone grafting material itself, not for an AI-powered system evaluating its efficacy or providing diagnostic information.

    Therefore, I cannot provide the requested information regarding:

    • A table of acceptance criteria and reported device performance for an AI device.
    • Sample size used for the test set and data provenance.
    • Number of experts and their qualifications for establishing ground truth.
    • Adjudication method for the test set.
    • Multi-Reader Multi-Case (MRMC) comparative effectiveness study.
    • Standalone performance (algorithm only).
    • Type of ground truth used.
    • Sample size for the training set.
    • How ground truth for the training set was established.

    This document is for a traditional medical device (bone grafting material), not an AI/ML medical device, and thus the requested AI-specific information is not present.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221065
    Device Name
    MediLab
    Date Cleared
    2022-06-10

    (59 days)

    Product Code
    Regulation Number
    892.2050
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Sigma Scientific Services LLC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MediLab is intended for use as a diagnostic and analysis tool for diagnostic images for hospitals, imaging centers, radiologists, reading practices and any user who requires and is granted access to patient image, demographic and report information. MediLab displays and manages diagnostic quality DICOM images. MediLab is not intended for diagnostic use with mammography images. Usage for mammography is for referral only. MediLab is not intended for diagnostic use on mobile devices.

    Contraindications: MediLab is not intended for the acquisition of mammographic image data and is meant to be used by qualified medical personnel.

    Device Description

    The MediLab is a DICOM medical image viewer that allows downloading, reviewing, manipulating, visualizing and printing medical multi-modality image data in DICOM format, from a client machine. MediLab is a server-based solution that connects to any PACS and displays DICOM images within the hospital, securely from remote locations, or as an integrated part of an EHR or portal. MediLab enables health professionals to access, manipulate, measure DICOM images and collaborate real-time over full quality medical images using any web-browser without installing client software.

    AI/ML Overview

    The provided text describes the MediLab device, its intended use, and compares it to predicate devices. However, it does not explicitly state acceptance criteria in a quantitative manner or detail a specific study proving the device meets said criteria for diagnostic performance.

    The document primarily focuses on demonstrating substantial equivalence to predicate devices (ZeeroMED View, IMPLANTER DENTAL PLANNING SOFTWARE, and aPROMISE X) based on intended use, technical characteristics, and functionalities as a medical image management and processing system.

    Here's a breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not provide specific quantitative acceptance criteria or diagnostic performance metrics for the MediLab device. The comparison focuses on feature similarity with predicate devices. The closest to "performance" mentioned is "Performance Testing (Measurement Accuracy) was conducted on the MediLab system to determine measurement accuracy when performing the various distance and area measurements." However, the results of this testing, or any specific acceptance criteria for these measurements, are not provided.

    Acceptance CriteriaReported Device Performance
    Not explicitly stated in terms of diagnostic performance metrics."Performance Testing (Measurement Accuracy) was conducted on the MediLab system to determine measurement accuracy when performing the various distance and area measurements."
    Functionality (e.g., DICOM image loading, display operations, user authentication, report generation)"Subject device has similar features and functionality as the predicate and reference devices."
    Safety and Effectiveness"The non-clinical performance test data and software verification and validation demonstrate that the MediLab system performs comparably to and it is as safe and effective as the predicate device."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the text. The document mentions "Non-clinical product evaluation to demonstrate safety and effectiveness was conducted" and "Performance Testing (Measurement Accuracy) was conducted," but it does not specify the details of the test data (sample size, origin, retrospective/prospective nature).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided in the text. The document does not mention any expert review or ground truth establishment process for diagnostic performance validation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided in the text, as no details of a diagnostic performance study are given.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no indication in the text that an MRMC comparative effectiveness study was conducted. The MediLab device is described as a medical image management and processing system and a diagnostic/analysis tool, not specifically an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    There is no indication in the text that a standalone algorithm performance study was conducted. The device is presented as a medical image viewer and management system for use by qualified medical personnel.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    This information is not provided in the text, as no diagnostic performance study details are given.

    8. The sample size for the training set

    The document describes "Software Verification and Validation" and mentions "software requirements and specifications, design architecture, risk analysis and software validation and verification." However, it does not mention a "training set" or "training data" in the context of an AI/ML device. The MediLab is described as a DICOM image viewer and processor, not as a device utilizing machine learning for diagnostic interpretations.

    9. How the ground truth for the training set was established

    As there is no mention of a training set in the context of an AI/ML algorithm, this information is not applicable and therefore not provided in the text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K202462
    Date Cleared
    2021-04-23

    (239 days)

    Product Code
    Regulation Number
    880.6850
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Sigma Medical Supplies Corporation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SIGMA sterilization pouch and roll are intended to provide health care workers with an effective method to enclose devices intended for sterilization in the Steam Sterilizer and via Ethylene Oxide (EO). The recommended sterilization cycles are as follows:

    • · Gravity steam at 121°C (250°F) for 30 minutes; Drying time of 30 minutes.
    • · Gravity steam at 132°C (270°F) for 15 minutes; Drying time of 30 minutes.
    • · Gravity steam at 135°C (275°F) for 10 minutes; Drying time of 30 minutes.
    • · Pre-vacuum steam at 132°C (270°F) for 4 minutes; Drying time of 20 minutes.
    • · Pre-vacuum steam at 135°C (275°F) for 3 minutes; Drying time of 16 minutes.
      · EO sterilization cycle is 4 hours at 55°C (131°F) with a relative humidity between 50%-85% and a sterilant concentration of 600 mg/L.
      Furthermore, the sterilization pouch and roll maintains the enclosed devices up until 3 years post EO gas sterilization and maintains the enclosed devices up until 6 months post Steam sterilization. Lastly, the pouch's external chemical ink indicators are designed to indicate to the pouch has undergone either a steam or EO sterilization process.
    Device Description

    The SIGMA sterilization pouches and rolls are inserted into the Pouch/Roll, sealed, and then sterilized in the Sterilization System. After completion of the sterilization process, the Pouch/Roll maintain sterility of the enclosed medical devices until the seal is opened. The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed devices until used up to 3 years post EO gas sterilization and maintains the enclosed devices up until 6 months post Steam sterilization.
    The Self-seal pouch permits sealing of the pouch without need of heat- sealing equipment, while the heat sealed pouches and rolls are heat sealed prior to processing in the steam/or EO Sterilization cycles.
    The chemical indicators ink printed on the medical grade paper will exhibit a color change after the pouch is exposed to steam or ethylene oxide gas. The color of Chemical Indicator changes from Blue to Greenish Black, when exposed to Steam. And the color changes from red to yellow, when exposed to EO gas.
    The Chemical Indicator offers an addition way to verity processing in the sterilization cycle. The Chemical Indicator should be used in addition to, not in place of, the biological indicator. The Chemical Indicators do not signify sterilization; they only indicate that the indicator has been exposed to Steam/or EO gas.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for the SIGMA Sterilization Pouch and Roll. It does not describe an AI/ML medical device, but rather a sterilization packaging product. Therefore, the questions related to AI/ML device performance metrics such as sample size for test set, data provenance, expert ground truth, adjudication methods, MRMC studies, standalone performance, and training set details are not applicable.

    The document focuses on demonstrating substantial equivalence to a predicate device (K180661) through non-clinical performance testing.

    Here's an analysis of the acceptance criteria and study as presented for this non-AI/ML medical device:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria and reported device performance are extensively detailed across multiple tables, primarily Table 2 ("Summary of the Proposed and Predicate Devices Technological Characteristics") and Table 3 ("Summary of Non-Clinical Testing"). I will extract the key information from these tables.

    FeatureTest/StandardAcceptance CriteriaReported Device Performance (Proposed Device)Result
    Sterilant Penetration / Drying Time / Aeration
    Steam Sterilization - SALANSI/AAMI/ISO 17665-1:2006/(R)2013, ANSI/AAMI/ISO TIR 17665-2:2009 (R2016)Meet the requirement of SAL 10-6, test BI (Steam processed): No bacterial growthUsing half-cycle and full-cycle analysis. Test BI: No bacterial growthPass
    Steam Sterilization - Drying TimeANSI/AAMI/ISO 17665-1:2006/(R)2013, ANSI/AAMI/ISO TIR 17665-2:2009 (R2016)Weight difference before sterilization and after drying shall not exceed 0%Weight difference = 0%Pass
    Steam Sterilization - Visual Drying(Standard not explicitly listed for this specific bullet, but falls under the general standards above)(Implied visual dryness)Visual are drying.Pass
    EO Gas Sterilization - SALAAMI / ANSI / ISO 11135:2014Meet the requirement of SAL 10-6, test BI (EO processed): No bacterial growthUsing half-cycle analysis. Test BI: No bacterial growthPass
    EO Gas Sterilization - ResidualsISO 10993-7:2008 (R) 2012EO 400 (kgf/mm2), MD > 450 (kgf/mm2)CD = 458, MD = 462 (min. value after Steam Sterilization)Pass
    Tear Resistance Test (Plastic Film)ASTM D 1004 -13CD > 300 (kgf/mm2), MD > 350 (kgf/mm2)CD = 443, MD = 453 (min. value after Steam Sterilization)Pass
    Thickness VariationASTM F 2251-13±0.02±0.003Pass
    Internal Pressurization Failure ResistanceASTM F1140/F1140M-13Pressurize to 80% of burst value and hold for 30s. Burst value > 5.0 (kPa)Average Burst pressure = 5.8 (kPa) (after Accelerated Aging)Pass
    Dye Penetration Test (Seal Leaks)ASTM F1929-15No leakage across the seal width of sterile barrier systemNo Channels identified on packagePass
    Seal StrengthASTM F88/F88M-15Strength > 356 (gf /25.4mm)Site 1= 588, Site 2= 598, Site 3= 909, Site 4= 1016 (min. value of test results)Pass
    Accelerated Aging (Steam)ASTM F 1980-07Incubated for 22.5 days under controlled conditions (Temp: 55°C / RH: 50%) simulating 6 months post-steam sterilization storage.a. Test articles (no product included and steam sterilized 135°C/15min) were accelerated aged at 55°C and 50% R.H to simulate the real time aging of 6 months. b. Test articles (product included and steam sterilized 135°C/15min) were accelerated aged at 55°C and 50% R.H to simulate the real time aging of 6 months. (Implied that package integrity was maintained after aging, as it "Pass"ed)Pass
    Accelerated Aging (EO Gas, Sterility)ASTM F 1980-07Incubated for 137 days under controlled conditions (Temp: 55°C / RH: 50%) simulating 3 years post-EO gas sterilization storage.a. Test articles (no product included) were accelerated aged at 55°C and 50% R.H to simulate the real time aging of 3 years. b. Test articles (product included and EO sterilized) were accelerated aged at 55°C and 50% R.H to simulate the real time aging of 3 years. (Implied that package integrity was maintained after aging, as it "Pass"ed)Pass
    Accelerated Aging (General Shelf Life)ASTM F 1980-07Incubated for 137 days under controlled conditions (Temp: 55°C / RH: 50%) simulating 3 years storage.a. Test articles (no product included) were accelerated aged at 55°C, and 75% RH to simulate the real time aging of 3 years. b. Test articles (product included) were accelerated aged at 55°C, and 75% RH to simulate the real time aging of 3 years. (Implied that package integrity was maintained after aging, as it "Pass"ed)Pass
    Bubble Emission (Gross Leaks)ASTM D3078-02(R2013)No LeakageNo LeakagePass
    Internal Pressurization (Gross Leaks)ASTM F2096-11No LeakageNo LeakagePass
    Microbial Barrier TestDIN 58953-6CFU 1.7Average LRV = 3.31Pass
    Chemical Indicator Efficacy Testing (Type 1 Process Indicators)
    Steam CI Color Change (Low Exposure)AAMI/ANSI/ISO 11140-1:2014121°C / 3.0 min & 134°C / 0.5 min: Unacceptable result (Color should not change significantly)121°C / 3.0 min & 134°C / 0.5 min: the result of color is Blue (No change)Pass
    Steam CI Color Change (High Exposure)AAMI/ANSI/ISO 11140-1:2014121°C / 10.0 min & 134°C / 2.0 min: Acceptable result (Color should change to endpoint color)121°C / 10.0 min & 134°C / 2.0 min: the result of color is from Blue to Greenish Black (Endpoint color)Pass
    Steam CI Color Change (Dry Heat Control)AAMI/ANSI/ISO 11140-1:2014Dry heat 140°C / 30 min: Unacceptable result (Color should not change significantly)Dry heat 140°C / 30 min: the result of color is Blue (No change)Pass
    EO Gas CI Color Change (Absence of EO)AAMI/ANSI/ISO 11140-1:2014Absence of EO gas / 90 min: Unacceptable result (Color should not change significantly)Absence of EO gas / 90 min: the result of color is Red (No change)Pass
    EO Gas CI Color Change (Low Exposure)AAMI/ANSI/ISO 11140-1:2014EO gas Teat / 2 min: Unacceptable result (Color should not change significantly)EO gas Teat / 2 min: the result of color is Red (No change)Pass
    EO Gas CI Color Change (High Exposure)AAMI/ANSI/ISO 11140-1:2014EO gas Teat / 20 min: Acceptable result (Color should change to endpoint color)EO gas Teat / 20 min: the result of color is Yellow (Endpoint color)Pass
    Chemical Indicator Shelf LifeAAMI/ANSI/ISO 11140-1:2014All performance attributes should maintain the original color: 3 years shelf lifeThe real-time test was carried out from October 15, 2007 to December 15, 2010. Test group exposed to Steam maintained Dark Green; test group exposed to EO maintained Yellow; Control group maintained original color.Pass

    Regarding the sections that are not applicable to non-AI/ML devices (points 2-9):

    1. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective): Not applicable, as this is a physical medical device. The testing involves laboratory evaluations of material properties and sterilization efficacy, not interpretation of data. The document does not specify general "sample sizes" in terms of number of patients/cases, but rather the number of units tested for specific physical and chemical properties. The provenance is likely from the manufacturer's testing facilities in Taiwan.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable. "Ground truth" in the context of AI/ML refers to expert-labeled data. For this device, "ground truth" is established by adherence to recognized international standards (e.g., ASTM, ISO, AAMI) and their specified pass/fail criteria for physical and chemical properties.

    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable. Adjudication methods are typically for resolving discrepancies in expert labeling of AI/ML data. The "acceptance criteria" here are objective measurements against defined standards.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is a standalone physical product, not an AI-assisted diagnostic or therapeutic device.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. There is no AI algorithm. The device's performance is inherently standalone in its function as a sterilization pouch.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable to an AI device. For this physical device, the "ground truth" for its performance is determined by adherence to established international and national standards for sterilization packaging, material science, and chemical indicators (e.g., microbial barrier, tensile strength, peel strength, chemical indicator color change).

    7. The sample size for the training set: Not applicable. This is a physical medical device, not an AI/ML model for which a training set would be required.

    8. How the ground truth for the training set was established: Not applicable. Same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180672
    Date Cleared
    2018-06-06

    (84 days)

    Product Code
    Regulation Number
    880.6850
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Sigma Medical Supplies Corporation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sterilization Pouch/Roll Made with Tyvek® are intended to provide health care workers with an effective method to enclose devices intended for sterilization in the STERRAD® 100S Sterilizer. The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed devices until used up to 3 years post sterilization.

    The pouches and rolls are printed with a chemical indicator bar which is a process indicator (ISO 11140-1:2005) that changes from red to blue (or lighter) when exposed to hydrogen peroxide vapor during processing in the STERRAD® 100S Sterilizer.

    The Sterilization Pouch/Roll Made with Tyvek® is offered in the follow 5 types:

    • Self-sealing sterilization pouches
    • Sterilization pouches, Flat
    • Sterilization pouches, Gusseted
    • Sterilization rolls, Flat
    • Sterilization rolls, Gusseted
    Device Description

    Sterilization Pouch/Roll Made with Tyvek® is intended to contain medical devices to be terminally sterilized in the STERRAD® 100S Sterilization System. The medical devices are inserted into the Pouch/Roll, sealed, and then sterilized in the STERRAD® 100S Sterilization System (The recommended cycle parameter is 6 minutes (Injection volume: 2880µL) at 50°C.). After completion of the sterilization process, the Pouch/Roll maintains sterility of the enclosed medical devices until the seal is opened. The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed devices until used up to 3 years post sterilization.

    The Pouch/Roll is printed with a chemical indicator bar that changes from red to blue (or lighter) when exposed to hydrogen peroxide vapor during processing in the STERRAD® 100S Sterilization System.

    The pouches are constructed from Tyvek®/plastic films, with H2O2 Chemical Indicator printed onto the Tyvek® film. The Self-seal pouch permits sealing of the pouch without need of heat- sealing equipment, while the heat sealed pouches and rolls are heat sealed prior to processing in the STERRAD® Sterilization Systems.

    The H2O2 Chemical Indicator offers an additional way to verify processing in the sterilization cycle. The Chemical Indicator should be used in addition to, not in place of, the biological indicator. H2O2 Chemical Indicators do not signify sterilization; they only indicate that the indicator has been exposed to the hydrogen peroxide. The color of the Chemical Indicator changes from red to blue (or lighter) when exposed to hydrogen peroxide.

    The Sterilization Pouch/Roll Made with Tyvek® is offered in the follow 5 types:

    • Self-sealing sterilization pouches
    • Sterilization pouches, Flat
    • Sterilization pouches, Gusseted
    • Sterilization rolls, Flat
    • Sterilization rolls, Gusseted
    AI/ML Overview

    The provided text describes the regulatory submission for a medical device (Sterilization Pouch/Roll Made with Tyvek®) seeking 510(k) clearance from the FDA. This document focuses on demonstrating substantial equivalence to a previously cleared predicate device, rather than proving the safety and effectiveness of a novel AI/software-as-a-medical-device (SaMD).

    Therefore, the information typically requested regarding acceptance criteria and study design for AI/SaMD devices (like sample size for test/training sets, data provenance, expert
    qualifications, ground truth establishment, MRMC studies, or standalone algorithm performance) is not applicable to this submission. This device is a physical product, and its performance is evaluated through material science, sterility, and packaging tests.

    However, I can extract the acceptance criteria and reported performance for the physical device based on the non-clinical testing summaries.

    1. A table of acceptance criteria and the reported device performance:

    Test CategoryTest Method / Standard (Acceptance Criteria)Reported Device Performance
    Sterilant PenetrationAAMI / ANSI / ISO 14937:2009, "Sterilization of Health Care Products - General Requirements for Characterization of a Sterilizing Agent and the Development, Validation, and Routine Control of a Sterilization Process for Medical Devices" (Acceptance: Able to penetrate and sustain direct contact with the medical instrument inside)Confirmed that the sterilant is able to penetrate the device and sustain direct contact with the medical instrument inside.
    Sterilant PenetrationHalf-Cycle Efficacy (Acceptance: 6 log reduction of Geobacillus stearothermophilus)Showed a 6 log reduction of Geobacillus stearothermophilus.
    BiocompatibilityISO 10993-10:2010(E), "Biological Evaluation of Medical Devices - Part 10: Tests for irritation and skin sensitization" (Acceptance: Meets requirements, non-irritating)Showed "negative reaction" (non-irritating) and met requirements.
    Package IntegrityASTM F2251-03, "Standard Test Method for Thickness Measurement of Flexible Packaging Material" (Acceptance: Passed)Passed
    Package IntegrityASTM D882, "Standard Test Method for Tensile Properties of Thin Plastic Sheeting" (Acceptance: Passed)Passed
    Package IntegrityASTM D 5035, "Standard Test Method for Breaking Force and Elongation of Textile Fabrics (Strip Method)" (Acceptance: Passed)Passed
    Package IntegrityASTM D1922, "Standard Test Method for Propagation Tear Resistance of Plastic Film and Thin Sheeting by Pendulum Method" (Acceptance: Passed)Passed
    Package IntegrityASTM F1140-07, "Standard Test Methods for Internal Pressurization Failure Resistance of Unrestrained Packages" (Acceptance: Passed)Passed
    Package IntegrityASTM F88/F88M; ISO 11607-1, "Standard Test Method for Seal Strength of Flexible Barrier Materials" (Acceptance: Passed)Passed
    Package IntegrityASTM F1929-98 (04); ISO 11607-1, "Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration" (Acceptance: Passed)Passed
    Material CompatibilityAAMI / ANSI / ISO 11607-1:2006//A1:2014, "Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems" (Acceptance: Met criteria)Met material compatibility acceptance criteria.
    Sterility MaintenanceDIN 58953-6:2010-05, "Sterilization - Sterile supply - Part 6: Microbial barrier testing of packaging materials for medical devices which are to be sterilized" (Acceptance: Passed)Passed
    Sterility MaintenanceAAMI / ANSI / ISO 11607-1:2006//A1:2014 (Acceptance: Met criteria for sterility maintenance after aging)Met sterility maintenance acceptance criteria.
    Sterility MaintenanceDurability: Accelerated Aging Test (ASTM F 1980-2007; ISO 11607-1) (Acceptance: Passed for 3 years accelerated aging, maintaining seal strength)Passed (3 years accelerated aging, seal strength maintained).
    Chemical Indicator EfficacyAAMI / ANSI / ISO 11140-1:2014, "Sterilization of Health Care Products - Chemical Indicators - Part 1: General Requirements" (Acceptance: Sterilant penetrated through pouch, CI color changed to endpoint color, color stability for 3 years, completeness/uniformity of color change)The sterilant penetrated through the pouch, the CI color changed to the endpoint, indicator dye remained stable for 3 years real-time aging, and the color change was effective.

    For the following points, as explained, they are not applicable because this is a 510(k) submission for a physical medical device (sterilization pouches) based on substantial equivalence, not an AI/SaMD (Software as a Medical Device). The evaluation focuses on material properties and performance under sterilization conditions rather than algorithmic performance on complex data (like images or medical records).

    2. Sample sizes used for the test set and the data provenance: Not applicable. The tests performed are engineering and material-science based, using samples of the device and specific test setups, not patient-derived data. The document does not specify a "test set" in the context of an AI algorithm, nor "data provenance" (e.g., country of origin) relevant to patient data. The non-clinical tests were likely conducted in a controlled laboratory environment.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable. Ground truth as typically defined for AI/SaMD (e.g., disease diagnosis from medical images) is not relevant here. The "ground truth" for these tests are objective measurements against established engineering and biological standards (e.g., quantitative measures of tensile strength, observable microbial growth, or chemical indicator color change).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. Adjudication methods are used to resolve disagreements among human annotators/experts for ground truth establishment in AI/SaMD; this is not relevant for materials testing.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI-assisted device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The "ground truth" for the performance evaluations are based on established industry standards and scientific principles for material properties (e.g., ASTM standards for mechanical properties), biological efficacy (e.g., 6-log reduction for sterilization), and chemical reactivity (e.g., color change of indicators). No human expert consensus or outcomes data in a clinical context is used as "ground truth" for this device's acceptance.

    8. The sample size for the training set: Not applicable. There is no AI training set involved.

    9. How the ground truth for the training set was established: Not applicable. There is no AI training set involved.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180661
    Date Cleared
    2018-06-05

    (83 days)

    Product Code
    Regulation Number
    880.6850
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Sigma Medical Supplies Corp.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SIGMA sterilization pouch and roll are intended to provide health care workers with an effective method to enclose devices intended for sterilization in steam auto claves and via Ethylene Oxide (EO). The recommended steam sterilization cycle parameters are 30 minutes at 121°C. The recommended EO sterilization cycle is 4 hours at 55°C with a relative humidity between 50%-85% and a sterilant concentration of 600 mg/L. Furthermore, the sterilization pouch and roll maintains the enclosed devices up until 3 years post EO gas sterilization and maintains the enclosed devices up until 6 months post Steam sterilization. Lastly, the pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EO sterilization process.

    Device Description

    The SIGMA sterilization pouch and roll are intended to provide health care workers with an effective method to enclose devices intended for sterilization in steam auto claves and via Ethylene Oxide (EO). The recommended steam sterilization cycle parameters are 30 minutes at 121°C. The recommended EO sterilization cycle is 4 hours at 55°C with a relative humidity between 50%-85% and a sterilant concentration of 600 mg/L. The medical devices are inserted into the Pouch/Roll, sealed, and then sterilized in the Sterilization System. After completion of the sterilization process, the Pouch/Roll maintain sterility of the enclosed medical devices until the seal is opened. The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed devices until used up to 3 years post EO gas sterilization and maintains the enclosed devices up until 6 months post Steam sterilization.

    The Self-seal pouch permits sealing of the pouch without need of heat- sealing equipment, while the heat sealed pouches and rolls are heat sealed prior to processing in the steam/or EO Sterilization cycles.

    The chemical indicators ink printed on the medical grade paper will exhibit a color change after the pouch is exposed to steam or ethylene oxide gas. The color of Chemical Indicator changes from Blue to Greenish Black, when exposed to Steam. And the color changes from red to yellow, when exposed to EO gas.

    The Chemical Indicator offers an addition way to verity processing in the sterilization cycle. The Chemical Indicator should be used in addition to, not in place of, the biological indicator. The Chemical Indicators do not signify sterilization; they only indicate that the indicator has been exposed to Steam/or EO gas.

    AI/ML Overview

    This document is a 510(k) Premarket Notification from Sigma Medical Supplies Corp. to the FDA regarding their SIGMA Sterilization Pouch and Roll. It seeks to prove substantial equivalence to a previously cleared device (K102158).

    Here's an analysis of the acceptance criteria and the study proving the device meets them, based solely on the provided text:

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria here are derived from the comparison of the proposed device to the predicate device and the requirement to meet various ISO and ASTM standards. The study's "performance" is reported as "Passed" or "Meets requirements."

    Table of Acceptance Criteria and Reported Device Performance

    Feature/TestAcceptance Criteria (Expected Outcome)Reported Device Performance
    Material CompositionSame as predicate (Medical Grade Paper, CPP, PET, PU adhesive, EO and Steam Process Indicator Print Ink)Same as predicate (Medical Grade Paper, CPP, PET, PU adhesive, EO and Steam Process Indicator Print Ink)
    Steam Sterilization Cycle Parameters30 minutes at 121°CSame (30 minutes at 121°C)
    EO Gas Sterilization Cycle Parameters4 hours at 55°C, 50%-85% RH, 600 mg/L sterilant concentrationSame (4 hours at 55°C, 50%-85% RH, 600 mg/L sterilant concentration)
    Pouch TypesSame as predicate (Self-sealing, Flat, Gusseted pouches; Flat, Gusseted rolls)Same as predicate
    Sterilant Penetration (Half-Cycle Efficacy)Meet requirement of SAL 10^-6^The test meets the requirement of SAL 10^-6^
    Chemical Indicator (CI) Functionality & EndpointSterilant penetrates and causes CI color change to endpoint colorThe sterilant penetrated and affected the CI color change to the endpoint color
    Package Integrity (Physical Properties)Passed (based on various ASTM/ISO standards)Passed for all listed tests (Thickness Variations, Tensile Strength of plastic film, Tensile Strength of paper, Air Permeance of paper, Tear Resistance Test, Burst Strength, Bubble Leak Test, Seal Peel Test, Dye Penetration Test, Microbial Barrier Test)
    Toxicological Properties (Biocompatibility Test)Passed (based on ANSI/AAMI/ISO 10993-10)Passed. Showed "negative reaction". Meets ISO 10993-10:2010(E). EO sterilization residuals meet AAMI/ANSI/ISO 10993-7:2008 (R) 2012.
    Durability: Accelerated Aging TestPassed (based on ASTM F 1980; ISO 11607-1)Passed
    End Point / Post Processing Color Stability (Steam)6 months6 months
    End Point / Post Processing Color Stability (EO)3 Years3 Years
    Shelf Life (Chemical Indicator Functionality)3 Years3 Years
    Shelf Life (Accelerated aging test - Seal Strength)3 Years3 Years
    Class 1 Process Indicators for STEAM (Pouch & CI)Remain stable before use, change color in presence of sterilant, maintain endpoint stability, meet ISO 11140-1:2014 requirementsDemonstrates ability to remain stable, change color, maintain endpoint stability. Meets ISO 11140-1:2014.
    Class 1 Process Indicators for EO Gas (Pouch & CI)Remain stable before use, change color in presence of sterilant, maintain endpoint stability, meet ISO 11140-1:2014 requirementsDemonstrates ability to remain stable, change color, maintain endpoint stability. Meets ISO 11140-1:2014.
    Maintenance of Sterility (Steam)Up to 6 months post Steam sterilizationMaintains up until 6 months post Steam sterilization
    Maintenance of Sterility (EO)Up to 3 years post EO gas sterilizationMaintains up until 3 years post EO gas sterilization

    Study Details:

    1. Sample Size Used for the Test Set and Data Provenance:

      • The document does not specify the exact sample sizes used for each test. It states that "The testing demonstrates the ability of the Sigma Sterilization Pouch and Roll" and "The testing results demonstrate the ability of the Sigma Sterilization Pouch and Roll."
      • Data Provenance: The manufacturer is Sigma Medical Supplies Corp. located in Taiwan (R.O.C.). The testing (non-clinical) would likely have been conducted by or for this company. The document does not specify if the data is retrospective or prospective, but given it's a premarket notification for a new device (albeit similar to a predicate), it would be prospective data generated specifically for this submission.
    2. Number of Experts Used to Establish Ground Truth and Qualifications:

      • This document describes performance testing of a medical device (sterilization pouches and rolls) against established standards (ISO, ASTM). It does not involve human interpretation of images or other data for which "ground truth" would be established by experts in a clinical sense (e.g., radiologists for diagnostic AI).
      • The "ground truth" in this context is defined by the technical specifications and performance requirements of the relevant industry standards (e.g., attainment of SAL 10^-6^ for sterilization, specified physical properties, color change of chemical indicators).
      • Therefore, the concept of "ground truth" established by a panel of human experts in the way it's used for AI/clinical diagnostic studies is not directly applicable here. The validity of the tests themselves is based on adherence to the referenced standards.
    3. Adjudication Method for the Test Set:

      • Not applicable in this context. Adjudication is typically used for reconciling disagreements among human readers or evaluators in clinical studies. Here, the tests are objective measurements against defined standards.
    4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

      • No, an MRMC study was not done. This type of study is relevant for assessing the impact of AI on human reader performance in diagnostic tasks, which is not the purpose of this device or its reported testing.
    5. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Yes, implicitly. All the described tests (Sterilant Penetration, Package Integrity, Biocompatibility, Chemical Indicator Efficacy, etc.) are evaluating the physical and functional properties of the device itself, independent of human interaction during its use (beyond following instructions). These are "algorithm only" in the sense that the device's performance is measured against objective criteria, not as part of a human-AI system.
    6. The Type of Ground Truth Used:

      • The ground truth is based on industry-recognized technical standards (e.g., ISO, ASTM) for sterilant efficacy, package integrity, biocompatibility, and chemical indicator performance. For example, "SAL 10^-6^" or physical property thresholds defined by the standards.
    7. The Sample Size for the Training Set:

      • This document describes premarket notification for a physical medical device, not an AI/ML algorithm. Therefore, there is no "training set" in the context of machine learning. The device is manufactured based on design specifications and then tested.
    8. How the Ground Truth for the Training Set was Established:

      • Not applicable, as there is no "training set" for this type of device. The design and manufacturing processes are likely informed by prior engineering knowledge and adherence to quality systems for medical device production.
    Ask a Question

    Ask a specific question about this device

    K Number
    K151209
    Device Name
    InterOss
    Manufacturer
    Date Cleared
    2015-11-06

    (184 days)

    Product Code
    Regulation Number
    872.3930
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    SigmaGraft, Inc

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    InterOss® small granules are recommended for:

    • Augmentation or reconstructive treatment of the alveolar ridge .
    • . Filling of infrabony periodontal defects
    • . Filling of defects after root resection, apicocectomy, and cystectomy
    • . Filling of extraction sockets to enhance preservation of the alveolar ridge
    • . Elevation of the maxillary sinus floor
    • . Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR)
    • . Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR)

    InterOss® large granules are recommended for:

    • . Augmentation or reconstructive treatment of the alveolar ridge
    • . Elevation of the maxillary sinus floor
    • . Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR)
    Device Description

    InterOss® is a hydroxyapatite material derived from Australian bovine bone. The osteoconductive mineral structure is produced from bone through a multi-step purification process. Following placement in bony voids or gaps, InterOss acts as an osteoconductive scaffold for the ingrowth of adjacent viable bone. InterOss® gradually resorbs and is replaced with bone during the healing process.

    InterOss® is available in granule form and is packaged in vials or a syringe-like applicator.

    InterOss®, in the vial form, will be available to the United States market in 8 versions: Filled with 0.25g, 0.5g, 1.0g, 2.0g, or 5.0g of small granules (0.25 - 1.0mm) or filled with 0.5g, 1.0g, or 2.0g of large granules (1.0 - 2.0mm).

    InterOss , in the syringe-like applicator form, will be available to the United States market in 6 versions: Filled with 0.25cc, 0.5cc, or 1.0cc of small granules (0.25 - 1.0mm) or filled with 0.5cc, 1.0cc, or 1.5cc of large granules (1.0mm - 2.0mm).

    The syringe-like applicator was designed to deliver InterOss® granules more precisely to the intended treatment site without having to use other sterile instruments. The InterOss® granules can be wetted with either the patient's blood or sterile physiological saline solution back and then pressing down on the plunger. A removable filter cap prevents granules from falling out of the syringe-like applicator during storage and wetting.

    During the manufacturing process of InterOss®, the granules are placed into a polymer syringe-like applicator or a glass vial, the vial or syringe is then capped with a rubber cap (the vial is also sealed with an aluminum cap), packaged into a polyethylene terephthalate tray, covered with a Tyvek lid, sealed, and then sterilized by gamma irradiation. The sterilized device is place in a protective package (outer box) along with its Instructions for Use and doctors notes.

    All InterOss® products are supplied sterile and are intended for single use only.

    AI/ML Overview

    The provided text describes the regulatory clearance of a bone grafting material called InterOss® and does not contain information about a medical device that relies on AI or offers performance metrics that would typically be described with acceptance criteria like sensitivity, specificity, or AUC, as commonly seen in AI/ML device descriptions.

    Instead, the document focuses on demonstrating substantial equivalence to predicate devices through various tests and comparisons. The "acceptance criteria" here are implied by the standards for bone grafting materials and the equivalence to the predicate devices.

    Here's an attempt to structure the information based on the request, reinterpreting "acceptance criteria" as the properties and test outcomes required for substantial equivalence in this context:

    1. Table of Acceptance Criteria and Reported Device Performance

    Parameter / Acceptance Criteria (Implied by Equivalence to Predicate & Standards)InterOss® Reported Performance
    Physical/Chemical Properties
    Crystallinity ratio106%
    Phase purity100% hydroxyapatite
    Ca/P ratio1.67
    MorphologyGranular and porous
    Average granule size (small)568 µm (0.25 - 1.0mm range)
    Average granule size (large)1625 µm (1.0 - 2.0mm range)
    Pore interconnectivityPositive
    Absence of heavy metalsConfirmed
    Soluble Ca elution (small granules)0.533 ± 0.016 (3 days), 0.322 ± 0.012 (7 days), 0.366 ± 0.024 (14 days)
    Soluble P elution (small granules)3.723 ± 0.004 (3 days), 6.277 ± 0.079 (7 days), 5.865 ± 0.019 (14 days)
    Soluble Ca elution (large granules)0.513 ± 0.018 (3 days), 0.315 ± 0.009 (7 days), 0.351 ± 0.021 (14 days)
    Soluble P elution (large granules)3.640 ± 0.085 (3 days), 6.241 ± 0.041 (7 days), 5.847 ± 0.019 (14 days)
    Pore size (small granules)149.1660 Å
    Pore size (large granules)47.3236 Å
    Surface area (small granules)114.4452 m²/g
    Surface area (large granules)109.0581 m²/g
    Solubility (small granules)74.274 ± 13.599 µg/mm³
    Solubility (large granules)70.247 ± 12.612 µg/mm³
    Average pH (small granules)7.86 ± 0.05
    Average pH (large granules)7.83 ± 0.05
    Compressive strength (small granules)1.00 MPa
    Compressive strength (large granules)1.17 MPa
    Physiochemical properties (comparison to Bio-Oss®)Comparable (pore structure, microstructure, phase structure, chemical composition, residual organic substance)
    Biocompatibility
    Skin SensitizationNon-sensitizing (Stimulation Index = 1.03±0.26)
    MutagenicityNon-mutagenic (Micronucleus and AMES test)
    Intracutaneous ReactivityNo skin reactions
    PyrogenicityNon-pyrogenic
    HemolysisNo hemolytic reaction
    Systemic Toxicity / CytotoxicityNon-toxic and non-cytotoxic (Acute Systemic Injection Test and Cytotoxicity Test)
    Systemic Toxicity (leachables)Negative
    SterilityPassed (SAL 1 X10⁻⁶)
    Bacterial Endotoxin (LAL)Passed (
    Ask a Question

    Ask a specific question about this device

    K Number
    K133838
    Date Cleared
    2015-03-11

    (448 days)

    Product Code
    Regulation Number
    880.6850
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    SIGMA MEDICAL SUPPLIES CORP.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sterilization Pouch/Roll Made with Tyvek® are intended to provide health care workers with an effective method to enclose devices intended for sterilization in the STERRAD® 100S Sterilizer. The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed devices until used up to 3 years post sterilization.

    The pouches and rolls are printed with a chemical indicator bar which is a process indicator (ISO 11140-1:2005) that changes from red to blue (or lighter) when exposed to hydrogen peroxide vapor during processing in the STERRAD® Sterilization System.

    Device Description

    Sterilization Pouch/Roll Made with Tyvek® is intended to contain medical devices to be terminally sterilized in the STERRAD® Sterilization Systems. The recommended Gas Plasma sterilization cycle parameter is 6 minutes (Injection volume: 2880µ L) at 50°C. The medical devices are inserted into the Pouch/Roll, sealed, and then sterilized in the STERRAD® Sterilization System. After completion of the sterilization process, the Pouch/Roll maintain sterility of the enclosed medical devices until the seal is opened. The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed devices until used up to 3 years post sterilization.

    The Pouch/Roll is printed with a chemical indicator bar that changes from red to blue (or lighter) when exposed to hydrogen peroxide vapor during processing in the STERRAD® Sterilization System.

    The proposed pouches are constructed from Tyvek® plastic films, with H2O2 Chemical Indicator printed onto the Tyvek" film. The Self-seal pouch permits sealing of the pouch without need of heat- sealing equipment, while the heat sealed pouches and rolls are heat sealed prior to processing in the STERRAD® Sterilization Systems.

    The H2O2 Chemical Indicator offers an addition way to verity processing in the sterilization cycle. The Chemical Indicator should be used in addition to, not in place of, the biological indicator. H2O2 Chemical Indicators do not signify sterilization; they only indicate that the indicator has been exposed to the hydrogen peroxide. The color of the Chemical Indicator changes from red to blue (or lighter) when exposed to hydrogen peroxide.

    The Sterilization Pouch/Roll Made with Tyvek® is offered in the follow 5 types:

    • Self-sealing sterilization pouches .
    • . Sterilization pouches, Flat
    • . Sterilization pouches, Gusseted
    • Sterilization rolls, Flat
    • Sterilization rolls, Gusseted
    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device called "Sterilization Pouch/Roll Made with Tyvek®". It details the device's characteristics, intended use, and a comparison to a predicate device to establish substantial equivalence, which is the basis for its FDA clearance.

    Here's a breakdown of the requested information:

    1. A table of acceptance criteria and the reported device performance

    The document presents performance data primarily in a comparative table (Table 5-2, continued on subsequent pages in the original document) comparing the new device against its predicate. It also lists the standards used for testing throughout the "Effectiveness and Safety" section.

    Performance MetricAcceptance Criteria (Standard Reference and/or Implied Criterion)Reported Device Performance
    Sterilant PenetrationMeet the requirement of SAL 10-6 (ANSI/AAMI/ISO 14937:2009)The test met the requirement of SAL 10-6
    Package Integrity (Physical Properties)(Tested before and after Gas Plasma Sterilization)
    Thickness Variations (mm)(ASTM F 2251-03) No specific numerical acceptance criteria provided, but "Small" values for both new and predicate are similar.Small: 0.145 (before/after); Large: 0.146 (before/after)
    Tear Resistance (g)(ASTM D1922) No specific numerical acceptance criteria provided, but similar to predicate.CD: 259 (before), 258 (after); MD: 282 (before), 280 (after)
    Tensile strength of plastic film (kgf/mm2)(ASTM D882) No specific numerical acceptance criteria provided, but similar to predicate.CD: 575 (before), 531 (after); MD: 577 (before), 531 (after)
    Tensile strength of Tyvek® (N/2.54cm)(ASTM D 5035) No specific numerical acceptance criteria provided, but similar to predicate.CD: 174 (before), 170 (after); MD: 163 (before), 162 (after)
    Burst Strength (kPa)(ASTM F1140-07) No specific numerical acceptance criteria provided, but similar to predicate.Small: 21.4 (before), 17.95 (after); Large: 4.49 (before), 2.03 (after)
    Seal Peel Test (g/15mm)Pass (ASTM F88/F88M-09; ISO 11607-1) - The predicate also reports "Pass".Upper: 340.3 (before), 493.6 (after); Down: 506.7 (before), 709.9 (after); Left: 345.7 (before), 436.0 (after); Right: 316.5 (before), 518.5 (after). Result: Pass
    Dye penetration TestNo Channels identified on package (ASTM F 1929-12; ISO 11607-1)No Channels identified on package
    Microbial Barrier Test
    Ask a Question

    Ask a specific question about this device

    K Number
    K102158
    Date Cleared
    2011-08-02

    (368 days)

    Product Code
    Regulation Number
    880.6850
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    SIGMA MEDICAL SUPPLIES CORP.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SIGMA sterilization pouch and roll are intended to provide health care workers with an effective method to enclose devices intended for sterilization in steam auto claves and via Ethylene Oxide (EO). The recommended steam sterilization cycle parameters are 30 minutes at 121 ℃. The recommended EO sterilization cycle is 4 hours at 55 C with a relative humidity between 50%-85% and a sterilant concentration of 600 mg/L. Furthermore, the sterilization pouch and roll maintains the enclosed devices up until 3 years post sterilization. Lastly, the pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EO sterilization process.

    The SIGMA sterilization pouch and roll is offered in the following 5 types:

    • Self-sealing sterilization pouches .
    • . Sterilization pouches, Flat
    • . Sterilization pouches, Gusseted
    • . Sterilization rolls, Flat
    • . Sterilization rolls, Gusseted
    Device Description

    The SIGMA sterilization pouch and roll are intended to provide health care workers with an effective method to enclose devices intended for sterilization in steam auto claves and via Ethylene Oxide (EO). The recommended steam sterilization cycle parameters are 30 minutes at 121 ℃. The recommended EO sterilization cycle is 4 hours at 55 C with a relative humidity between 50%-85% and a sterilant concentration of 600 mg/L. Furthermore, the sterilization pouch and roll maintains the enclosed devices up until 3 years post sterilization. Lastly, the pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EO sterilization process.

    The SIGMA sterilization pouch and roll is offered in the following 5 types:

    • Self-sealing sterilization pouches: These pouches are made from a medical grade plastic film that . 0 is heat sealed on three sides. The forth side has an adhesive strip that is used to seal the pouch. Release paper used in the pouch is a laminated sheet with composing structure of PE/paper/PE. It is a strip to cover the adhesive area and is released before seal the pouch. The medical grade paper conforms to recognized material standards and can be sterilized by steam or ethylene oxide gas. The Process Indicators Ink printed on the medical grade paper will exhibit a color change after the pouch is exposed to steam or ethylene oxide gas.
    • Sterilization pouches, Flat: These pouches has the same components with the Self-sealing . sterilization pouches, except the forth side is left opened instead of an adhesive strip and will be heat-sealed when using.
    • Sterilization pouches, Gusseted: These pouches are the same with the Sterilization pouches, flat, . except that the plastic film is folded on both longest sides instead of flat. This design is convenient to enclose the medical devices with certain height.
    • Sterilization rolls, Flat: These rolls are made from a medical grade paper and plastic film that are heat sealed on opposite two sides. It will be cut into the suitable length and the opened sides will be heat-sealed. The indicators printed on the medical grade paper are the same with the self-sealing sterilization pouches.
    • Sterilization rolls, Gusseted: These rolls are the same with the flat sterilization roll, except that the plastic film is folded on both longest sides instead of flat. This design is convenient to enclose the medical devices with certain height.
    AI/ML Overview

    The SIGMA Sterilization Pouch and Roll underwent numerous validation studies to demonstrate its effectiveness and safety, confirming its substantial equivalence to predicate devices. These studies adhered to recommended practices, standards, and guidelines from organizations such as AAMI, ISO, and ASTM.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Test CategoryAcceptance Criteria (Standard)Reported Device Performance
    Sterilant PenetrationAAMI / ANSI / ISO 17665-1:2006, ISO TS 17665-2:2009, AAMI / ANSI / ISO 11135-1:2007Effectively adequate sterilant penetration to most difficult areas. Confirmed sterilant able to penetrate and sustain direct contact with instruments.
    Drying TimeAAMI / ANSI / ISO 17665-1:2006, ISO TS 17665-2:2009, AAMI / ANSI / ISO 11135-1:2007Testing performed as described. Results confirm effective performance (specific quantitative result not detailed, but implied by meeting standard).
    AerationAAMI / ANSI / ISO 10993-7:2008Aeration time validation test meets requiremeAAMI / ANSI / ISO 10993-7:2008.
    BiocompatibilityISO 10993-10:2010(E), ISO 10993-1:2009/Cor. 1:2010(E), ISO 10993-12:2007(E), ISO/IEC 17025:2005Showed "negative reaction." Meets requirements of ISO 10993-10:2010(E).
    Package IntegrityAAMI / ANSI / ISO 11607-1:2006, ISO 1924-2, ISO 5636-3, ASTM D 3078-02, ASTM D882, ASTM F 2251-03, ASTM D 1004, ASTM E 1140-07, ASTM E 1929-98 (04), ASTM E 88-2007, ASTM F 1980-2007, ASTM F 1608-00Effectively adequate Package Integrity.
    Material CompatibilityAAMI / ANSI / ISO 11607-1:2006, ISO 1924-2, ISO 5636-3, ASTM D 3078-02, ASTM D882, ASTM F 2251-03, ASTM D 1004, ASTM E 1140-07, ASTM E 1929-98 (04), ASTM E 88-2007, ASTM F 1980-2007, ASTM F 1608-00Testing performed as described. Results confirm effective performance (specific quantitative result not detailed, but implied by meeting standard).
    Sterility MaintenanceAAMI / ANSI / ISO 11607-1:2006, ISO 1924-2, ISO 5636-3, ASTM D 3078-02, ASTM D882, ASTM F 2251-03, ASTM D 1004, ASTM E 1140-07, ASTM E 1929-98 (04), ASTM E 88-2007, ASTM F 1980-2007, ASTM F 1608-00Testing performed as described. Results confirm effective performance (specific quantitative result not detailed, but implied by meeting standard).
    Chemical Indicator EfficacyAAMI / ANSI / ISO 11140-1:2005Demonstrates effective stability of ink before use, lasting quality (color stability), completeness and uniformity of color change, and all-or-none change (unless color standard provided). Meets requirements of ISO 11140-1:2005.

    2. Sample Size Used for the Test Set and the Data Provenance:
    The document does not explicitly state the sample sizes used for each specific test within the validation studies. The studies reference adherence to international standards (AAMI, ISO, ASTM), which typically define appropriate sample sizes for such tests.
    The data provenance can be inferred as originating from testing laboratories in connection with Sigma Medical Supplies Corp., which is based in Taiwan. The studies appear to be prospective, as they were conducted specifically for the validation of the SIGMA Sterilization Pouch and Roll.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:
    This information is not provided in the document. The studies rely on established international standards and laboratory testing rather than expert-based ground truth for the performance criteria of sterilization pouches and chemical indicators.

    4. Adjudication Method for the Test Set:
    This information is not provided in the document. The validation methods described are primarily objective physical and chemical tests, rather than subjective assessments requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:
    No, an MRMC comparative effectiveness study was not done. This type of study is typically used for diagnostic devices involving human interpretation of images or clinical data. The SIGMA Sterilization Pouch and Roll is a Class II device for sterilization wraps and indicators, and its performance is evaluated through laboratory testing against established physical and chemical standards.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:
    Yes, the studies described are standalone performance evaluations of the device itself (the pouches and their integrated chemical indicators). This is inherent to the nature of testing sterilization packaging and chemical indicators, which are designed to function independently based on physical and chemical principles, without human interpretation of "algorithm-only" performance in the context of AI.

    7. The Type of Ground Truth Used:
    The ground truth used for these studies is based on established scientific and engineering principles codified in international consensus standards (AAMI, ISO, ASTM). These standards define the acceptable performance parameters for sterilant penetration, drying time, aeration, biocompatibility, package integrity, material compatibility, sterility maintenance, and chemical indicator efficacy. The "ground truth" is therefore the successful compliance with these predefined objective criteria.

    8. The Sample Size for the Training Set:
    This is not applicable. The device is a physical product (sterilization pouch and roll) and does not involve AI or machine learning algorithms that require a training set. The validation studies focused on testing the manufactured product's compliance with performance standards.

    9. How the Ground Truth for the Training Set Was Established:
    This is not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 8