Search Results
Found 5 results
510(k) Data Aggregation
(50 days)
Ask a specific question about this device
(34 days)
Ask a specific question about this device
(42 days)
A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment.
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Classified by FDA's General and Plastic Surgery Device panel as Class I, 21 CFR 880.6250, Powder-Free (yellow) Vinyl Patient Examination Glove, 80LYZ, and meets all requirements of ASTM Standard D5250-92.
This is a 510(k) summary for a medical device that is a Shijiazhuang Great Eagle Plastics Products Co., Ltd. Synthetic Powder-Free (Yellow) Vinyl Patient Examination Glove. This submission is for a Class I medical device, which generally means it poses a lower risk to patients and users and often relies on established performance standards rather than extensive clinical studies for approval.
The document describes non-clinical tests performed to demonstrate substantial equivalence to predicate devices, rather than a study proving the device meets acceptance criteria in the way one might describe for an AI/ML digital health product or more complex diagnostic device. For this type of glove device, acceptance criteria are tied to manufacturing standards and physical properties.
Here's the breakdown of the information requested, tailored to what's available for this specific medical device submission:
| Information Requested | Details for Shijiazhuang Great Eagle Plastics Products Co., Ltd. Synthetic Powder-Free (Yellow) Vinyl Patient Examination Gloves |
|---|---|
| 1. Acceptance Criteria and Reported Device Performance | |
| Acceptance Criteria | Reported Device Performance |
| ASTM-D-5250-92 Standard Compliance (Physical and Dimensions Testing) | All testing meets requirements for Physical and Dimensions Testing conducted on gloves, Inspection Level S-2, AQL 4.0. |
| FDA 1000 ml. Water Fill Test (Pinhole integrity) | Samplings of AQL 2.5, Inspection Level S-4, met requirements. |
| Biocompatibility (Primary Skin Irritation and Skin Sensitization) | Results showing no primary skin irritant or sensitization reactions. |
| Powder-Free Claim (USP Iodine Test for Starch) | Revealed passing results, adhering to USP Iodine test methodology. |
| 2. Sample size for test set and data provenance | The document does not specify a distinct "test set" sample size in the context of an AI/ML study. For the physical and performance tests, "samplings" were taken from production, with "Inspection Level S-2, AQL 4.0" for physical and dimensions, and "AQL 2.5, Inspection Level S-4" for the water fill test. This refers to statistical sampling plans used in quality control. The data provenance is implied to be from the manufacturer's internal testing as part of their production and quality assurance processes (China). The study is retrospective in the sense that these are manufacturing quality control tests conducted on finished products. |
| 3. Number of experts used to establish ground truth & qualifications | For this device (patient examination gloves), "ground truth" is established through standardized laboratory tests and adherence to established material and performance specifications (ASTM standards, FDA water fill test, USP tests). There isn't a team of human "experts" establishing a diagnostic ground truth in the way described for AI/ML. The "experts" are the laboratory technicians and quality control personnel performing the defined tests and interpreting the results against the established standards. No specific number or qualifications are provided beyond the implicit expertise in conducting these standard tests. |
| 4. Adjudication method for the test set | Not applicable in the context of diagnostic performance. Quality control for manufacturing relies on standardized test methods with pass/fail criteria. There isn't an "adjudication method" involving multiple human readers as in clinical image interpretation. |
| 5. Multi-Reader Multi-Case (MRMC) comparative effectiveness study | Not applicable. This is for a physical medical device (examination gloves), not an AI/ML system or diagnostic tool where human readers' performance would be evaluated. |
| 6. Standalone (algorithm only) performance study | Not applicable. This is for a physical medical device (examination gloves). There is no algorithm. |
| 7. Type of ground truth used | The ground truth is based on established industry standards and regulatory testing protocols: - ASTM Standard D5250-92 (for physical and dimensional properties) - FDA 1000 ml. Water Fill Test (for pinhole integrity) - Biocompatibility testing (Primary Skin Irritation and Skin Sensitization tests) - USP Iodine Test (for powder-free claim verification). |
| 8. Sample size for the training set | Not applicable. This is for a physical medical device, not an AI/ML system requiring a training set. |
| 9. How the ground truth for the training set was established | Not applicable. This is for a physical medical device. |
Ask a specific question about this device
(49 days)
A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare similar personnel to prevent contamination between and healthcare personnel and the patient's body, fluids, waste or environment.
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Classified by FDA's General and Plastic Surgery Device Panel 21 CFR 880.6250, Powdered Vinyl Patient Class I, as Examination Glove, 80LYZ, Powdered with Absorbable Dusting Powder, USP, Class III and meets all requirements of ASTM Standard D5250-92.
The provided text is for a 510(k) summary for medical gloves, which are low-risk devices and do not typically involve the types of studies and acceptance criteria that are relevant to AI/ML-powered medical devices. Therefore, much of the requested information, such as multi-reader multi-case studies, standalone algorithm performance, and expert ground truth establishment, is not applicable.
Here's the information that can be extracted or deduced from the provided document, formatted to best fit the request for an AI/ML device, with explanations where the information is not directly applicable:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Physical and Dimensions Testing: | Shijiazhuang Great Eagle Plastic Products Co., Ltd. glove production standards are based on ASTM-D-5250-92. All testing meets requirements for Physical and Dimensions Testing conducted on gloves, Inspection Level S-2, AQL 4.0. |
| Pinhole Detection (Water Fill Test): | The FDA 1000 ml. Water Fill Test was conducted with samplings of AQL 2.5, Inspection Level S-4, meeting these requirements. |
| Biocompatibility (Primary Skin Irritation): | Primary Skin Irritation testing was conducted with results showing no primary skin irritant reactions. |
| Biocompatibility (Skin Sensitization): | Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing no sensitization reactions. |
| Overall Conformance to Standards: | Conform fully to ASTM-D-5250-92 standards as well as applicable CFR 21 pinhole references, and meets FDA requirements, biocompatibility requirements and labeling claims as shown by data in Section 7. There are no safety/efficacy issues or new claims from the "substantial equivalence" products cited. |
2. Sample Size Used for the Test Set and Data Provenance
The document specifies "samplings of AQL 2.5, Inspection Level S-4" for the Water Fill Test and "Inspection Level S-2, AQL 4.0" for Physical and Dimensions Testing. These are statistical sampling plans from ASTM standards, not a fixed sample size in the context of typical AI/ML test sets. The exact number of gloves tested for each lot would depend on the size of the manufacturing lot.
Data Provenance: The document does not specify the country of origin of the data beyond the manufacturer being in China. The data is from non-clinical testing performed by the manufacturer to demonstrate compliance with standards. It is implicitly retrospective in the sense that the tests were performed on manufactured batches.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. The ground truth for medical gloves is established by objective physical and chemical tests (e.g., water leak tests, measurements of dimensions, chemical assays for irritation). There are no "experts" in the sense of clinical reviewers establishing ground truth for individual cases.
4. Adjudication Method for the Test Set
Not applicable. Adjudication methods like 2+1 or 3+1 are used for human review of images or data to establish ground truth in diagnostic studies. For glove testing, the results are typically quantitative measurements or pass/fail criteria against established standards.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a medical glove, not an AI/ML diagnostic tool. Therefore, MRMC studies are not relevant.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable. This device is a physical product and does not involve an algorithm.
7. The Type of Ground Truth Used
The ground truth used is based on objective measurement against established industry and regulatory standards. Specifically:
- Physical and Dimensional Characteristics: Defined by ASTM-D-5250-92.
- Barrier Integrity (Pinhole): Defined by FDA 1000 ml. Water Fill Test (referencing CFR 21 pinhole references) and ASTM-D-5250-92.
- Biocompatibility: Established through Primary Skin Irritation and Skin Sensitization testing, performed to determine the absence of adverse biological reactions.
8. The Sample Size for the Training Set
Not applicable. This device is a physical product, not an AI/ML model that requires a training set. The "training" for glove manufacturing would involve process control and quality assurance to meet the specified standards across all production.
9. How the Ground Truth for the Training Set was Established
Not applicable. As there is no AI/ML training set, there is no ground truth to be established in this context. The "ground truth" equivalent for manufacturing is the adherence to the ASTM-D-5250-92 standard and internal quality specifications during the production process.
Ask a specific question about this device
(46 days)
A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare similar personnel to prevent contamination between and healthcare personnel and the patient's body, fluids, waste or environment.
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Classified by FDA's General and Plastic Surgery Device Panel Class I, 21 CFR 880.6250, Powder-Free Vinyl Patient as Examination Glove, 80LYZ and meets all requirements of ASTM Standard D5250-92.
The provided document describes the 510(k) summary for the Shijiazhuang Great Eagle Plastic Products Co., Ltd. Synthetic Powder-Free Vinyl Patient Examination Gloves. It details the device, its intended use, comparison to predicate devices, and non-clinical tests performed to demonstrate substantial equivalence.
Here's an analysis of the acceptance criteria and the study that proves the device meets them:
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Physical and Dimensions Testing (ASTM D5250-92) | All testing meets requirements for Physical and Dimensions Testing conducted on gloves, Inspection Level S-2, AQL 4.0. |
| FDA 1000 ml. Water Fill Test | Conducted with samplings of AQL 2.5, Inspection Level S-4, meeting these requirements. |
| Primary Skin Irritation | Results showing no primary skin irritant reactions. |
| Skin Sensitization (allergic contact dermatitis) | Results showing no sensitization reactions. |
| USP Iodine Test for Starch (Powder-Free Claim) | Adheres to all USP Iodine Test methodology and testing conducted revealed passing results, insuring the "powder-free" claim. |
| Biocompatibility requirements | Met (as stated in the conclusion). |
| Labeling claims (no hypoallergenic claim) | Met (no special labeling claims, no hypoallergenic claim). |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not explicitly state the specific sample sizes for each test conducted. It mentions "samplings of AQL 2.5, Inspection Level S-4" for the Water Fill Test and "Inspection Level S-2, AQL 4.0" for Physical and Dimensions Testing, which are statistical sampling plans but don't provide a direct sample count. The tests were performed on gloves produced by Shijiazhuang Great Eagle Plastic Products Co., Ltd. in Shijiazhuang, He Bei Province, P.R. China. The data would be prospective, as it's generated during the production and testing of new devices.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This section is Not Applicable. The device is a medical glove, and its performance is evaluated against established technical standards (ASTM, FDA water fill, USP iodine test) and biocompatibility tests. These tests have objective criteria and do not rely on expert interpretation to establish "ground truth" in the way a diagnostic imaging device might.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This section is Not Applicable. As explained above, the tests performed on the gloves are objective and rely on direct measurements and observations against predefined pass/fail criteria, not on subjective expert consensus or adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This section is Not Applicable. The device is a patient examination glove, not an AI-powered diagnostic tool. Therefore, an MRMC study or an assessment of human reader improvement with AI assistance is not relevant.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This section is Not Applicable. The device is a physical medical glove and does not involve any algorithms or AI for standalone performance evaluation.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for the device, a medical glove, is defined by objective, measurable compliance with established international and regulatory standards and test methods:
- ASTM Standard D5250-92: This standard defines physical properties and dimensions for vinyl examination gloves. The "ground truth" is whether the glove's measured properties (e.g., tensile strength, elongation, dimensions) fall within the specified ranges.
- FDA 1000 ml. Water Fill Test: This test determines the watertight integrity (pinhole defects) of the gloves. The "ground truth" is a pass/fail outcome based on whether water leaks through the glove under specific conditions.
- Primary Skin Irritation and Skin Sensitization Testing: These are biological tests conducted to assess the biocompatibility of the glove material. The "ground truth" is the absence of observed irritation or sensitization reactions in test subjects/models.
- USP Iodine Test for Starch: This test objectively determines the presence or absence of starch, which is critical for a "powder-free" claim. The "ground truth" is a passing result indicating minimal or no starch.
8. The sample size for the training set
This section is Not Applicable. The device is a medical glove, not a machine learning model. Therefore, there is no "training set" in the context of AI. The described tests are for product verification and validation.
9. How the ground truth for the training set was established
This section is Not Applicable. As stated above, there is no training set for this type of device.
Ask a specific question about this device
Page 1 of 1