Search Results
Found 5 results
510(k) Data Aggregation
(28 days)
The intended use of electrocardiograph is to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart disease. The proposed device has analysis feature, however the ECC with measurements and interpretive statements is offered to clinician on an advisory basis only.
The proposed digital electrocardiograph, iE 101 and iE 300, are designed based on the cleared digital electrocardiographs K141946 (iE 3). The proposed device has same intended use, functional modules design, working modes, analysis function and accessories. the proposed devices are different than the predicate device in that they has new printing channel for iE 101); new layout of user-machine interface; three different physical specification, such as screen size, device dimension, device weight; three different specification, such as battery capacity, noise level and CMRR; and a modified software which is modified to applicable to the proposed devices.
The proposed electrocardiographs, iE 101 and iE 300, follow the same design principle, main components, accessories and software; they acquire ECG electrical signals from patient body surface by ECG electrodes. After been amplified, filtered and transferred, the ECG signal waveforms are displayed on the LCD and recorded on the paper through thermal printer.
They consist of four modules, which are power supply module, signal collection module, amplification module, and control module.
They can acquire ECG signal via twelve leads simultaneously, display or print waveform of ECG signal via one channel (iE 101) or three channel (iE 300).
The provided text describes a 510(k) submission for a Digital Electrocardiograph, models iE101 and iE300. The submission argues for substantial equivalence to a predicate device (K141946).
Here's an analysis of the acceptance criteria and study information:
1. Table of Acceptance Criteria and Reported Device Performance
The submission focuses on establishing substantial equivalence to a predicate device, rather than defining specific performance acceptance criteria for diagnostic accuracy. The "acceptance criteria" here are essentially the compliance with relevant safety and performance standards and the demonstration that the proposed device performs similarly to the predicate.
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Compliance with IEC 60601-1:2005 + A1: 2012 (General safety) | Complies with IEC 60601-1:2005 + A1: 2012 |
| Compliance with IEC 60601-1-2: 2007 (EMC) | Complies with IEC 60601-1-2: 2007 |
| Compliance with IEC 60601-2-25:2011 (Electrocardiographs particular requirements) | Complies with IEC 60601-2-25:2011 |
| Biocompatibility (ISO 10993) | Complies with ISO 10993 |
| Intended Use | Same as predicate device |
| Intended Population | Same as predicate device (Adult and pediatric patients) |
| Intended Environment | Same as predicate device (Hospital, clinics) |
| Configuration | Same as predicate device |
| Lead | Same as predicate device (Standard 12-lead) |
| Acquisition mode | Same as predicate device (Simultaneous 12-lead acquisition) |
| Recording format | Same as predicate device (Automatic / Manual / Rhythm) |
| Analysis mode | Yes (Same as predicate, advisory basis only) |
| Physical/Technical Specifications (e.g., screen size, channels, battery capacity, noise level, CMRR) | Differences noted but deemed not to affect substantial equivalence (e.g., iE101: 5.0" screen, 1 channel; iE300: 5.0" screen, 3 channel vs. predicate's 6.0" screen, 3 channel). Software modified to be applicable to proposed devices. |
2. Sample Size Used for the Test Set and Data Provenance
The document explicitly states: "No clinical study is included in this submission."
Therefore, there is no test set in the context of clinical performance evaluation based on patient data. The "tests" conducted were non-clinical, likely bench testing and engineering verification to demonstrate compliance with standards.
- Sample Size for Test Set: Not applicable as no clinical test set was used.
- Data Provenance: Not applicable as no clinical data was used for a test set. The non-clinical tests were conducted by the manufacturer (Shenzhen Biocare Bio-medical Equipment Co., Ltd.) in China.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
Not applicable, as no clinical study or test set requiring expert-established ground truth was performed for this submission. The device's "analysis feature" is offered on an advisory basis only, meaning its interpretive statements are not intended as definitive diagnoses.
4. Adjudication Method for the Test Set
Not applicable, as no clinical study or test set requiring adjudication was performed.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No. The submission explicitly states "No clinical study is included in this submission." Therefore, no MRMC study was conducted, and no effect size calculation for human readers with or without AI assistance is available.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
While the device has an "analysis feature," the submission does not include any standalone performance study of this algorithm. The analysis feature provides "measurements and interpretive statements" on an "advisory basis only," indicating it's not cleared for diagnostic standalone use.
7. The Type of Ground Truth Used
Not applicable, as no clinical performance study requiring ground truth was conducted. For non-clinical tests (electrical safety, EMC, etc.), the "ground truth" would be the specifications and requirements of the relevant IEC standards.
8. The Sample Size for the Training Set
Not applicable. This device is an electrocardiograph hardware with an accompanying analysis feature. The submission does not describe a machine learning algorithm that underwent training with a specific dataset. The analysis function is likely based on pre-programmed algorithms rather than a trained AI model in the modern sense.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no described training set or AI model training in the submission.
Ask a specific question about this device
(116 days)
Digital Electrocardiograph is intended to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The obtained ECG records can help users to analyze and disease. Digital Electrocardiograph shall be used in healthcare facilities by doctors and/or trained healthcare professionals.
Digital Electrocardiograph, ECG-2000, is intended to acquire ECG signals from adult patients through body surface ECG electrodes. The obtained ECG records can help users to analyze and diagnose heart disease.
It consists of the acquisition box, optional accessories, installation CD, and USB dongle. The acquisition box is connected with ECG electrodes, and intends to acquire ECG signals from patient. The acquired ECG signal is filtered and amplified, then transferred to the working station. The acquisition box is powered by the working station via a USB line. The working station is a Personal Computer installed with ECG-2000 Software, it intends to receive the processed ECG signal from the acquisition box, and further display and record the signals. Optional accessories include lead wires, limb electrodes and chest electrodes, and general-purpose printer (with USB interface, support PCL language, such as HP1010, P2035, and P2055d series). The CD contains Digital Electrocardiograph installer and dongle installation program module.
The provided document is a 510(k) premarket notification for a Digital Electrocardiograph, Model: ECG-2000. It focuses on demonstrating substantial equivalence to predicate devices, rather than presenting a study to prove the device meets acceptance criteria derived from a novel clinical trial.
Here's an analysis based on the provided text:
1. Table of acceptance criteria and the reported device performance
The document does not explicitly define "acceptance criteria" in the context of a clinical performance study with specific metrics like sensitivity, specificity, accuracy, etc. Instead, it demonstrates compliance with technical specifications and safety standards to establish substantial equivalence. The "performance" reported is primarily in terms of meeting these engineering and safety benchmarks.
| Feature | Predicate Device 1 (K092010)Description / Value | Proposed Device (ECG-2000)Reported Performance |
|---|---|---|
| Product Code | DPS | DPS (Same) |
| Regulation Number | 21 CFR 870.2340 | 21 CFR 870.2340 (Same) |
| Class | Class II | Class II (Same) |
| Intended Use | Acquire ECG signals from adult patients to help analyze and diagnose heart disease. Used by healthcare professionals. | Acquire ECG signals from adult and pediatric patients to help analyze and diagnose heart disease. Used by healthcare professionals. (Similar: extended to pediatric patients) |
| Configuration | Chest Electrode and Limb Electrode | Chest Electrode and Limb Electrode (Same) |
| ECG Lead | Standard 12-lead | Standard 12-lead (Same) |
| ECG Gain | (Implied range from "Similar" to proposed device) | 1.25, 2.5, 5, 10, 20, 10-5, 20-10 (mm/mV) (More options than Predicate 1, Same as Predicate 2) |
| Sampling rate | 1000Hz | 1000Hz (Same) |
| Input circuit current | (Implied value from "Similar" to proposed device, complies with IEC 60601-1) | $\leq$ 0.1 μA (Complies with IEC 60601-1) |
| Noise level | (Implied value from "Similar" to proposed device, complies with IEC 60601-1) | $<15\mu$ Vp-p (Complies with IEC 60601-1) |
| Electrical Safety | Comply with IEC 60601-1 | Comply with IEC 60601-1 (Same) |
| EMC | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 (Same) |
| Patient-contact Material | Chest Electrode: Metal, Limb Electrode: ABS | Same |
Note on "Intended Use": The proposed device extends its intended use to include pediatric patients, whereas Predicate Device 1 explicitly mentions "adult patients" in the comparison table. Predicate Device 2, however, is the exact same device, and its previous clearance (K133985) likely covered adult patients, and this submission seeks to add pediatric. The document explicitly states: "This submission is filed for extension of intended use...therefore, only the following standard was tested to verify that the device can meet its additional intended use: IEC 60601-2-25: 2011, Medical electrical equipment - Part 2-25: Particular requirements for the basic safety and essential performance of electrocardiographs."
2. Sample size used for the test set and the data provenance
- Sample Size: Not applicable. The document states "No clinical study is included in this submission." The testing primarily involved product verification to standards (e.g., IEC 60601-2-25) rather than a clinical evaluation with a test set of patient data.
- Data Provenance: Not applicable, as no clinical data set was used for performance evaluation in this submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not applicable. No clinical test set requiring ground truth established by experts was used.
4. Adjudication method for the test set
- Not applicable. No clinical test set requiring adjudication was used.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No. The device is a "Digital Electrocardiograph," which primarily acquires and displays ECG signals for analysis by healthcare professionals. It does not appear to incorporate AI for interpretation or act as an AI-assistive tool for human readers, nor was an MRMC study conducted.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- The document implies the device performs basic signal acquisition, filtering, and amplification, displaying these signals for human interpretation. There is no mention of an "algorithm" performing diagnostic interpretations in a standalone capacity that would require specific performance testing against a ground truth. The device "can help users to analyze and diagnose heart disease," which suggests it provides raw or processed signals for expert review, not an automated diagnosis.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- Not applicable. No clinical performance evaluation requiring a ground truth was conducted. The "ground truth" here is implied by the technical specifications and adherence to international standards for electrocardiographs.
8. The sample size for the training set
- Not applicable. This device is an acquisition and display system for ECG signals. There's no indication it employs machine learning or AI that would require a "training set."
9. How the ground truth for the training set was established
- Not applicable, as there is no training set for an AI/ML algorithm mentioned.
Ask a specific question about this device
(211 days)
The intended use of electrocardiograph is to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart disease. However the ECG with measurements and interpretive statements is offered to clinician on an advisory basis only.
Digital Electrocardiograph, iE 3/ iE 6/ iE 12/ iE 12A/ iE 15/ ECG-1210/ ECG-3010/ ECG-6010, are intended to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart disease. However the ECG with measurements and interpretive statements is offered to clinician on an advisory basis only.
All the proposed models, iE 3/ iE 6/ iE 12/ iE 12A/ iE 15/ ECG1210/ ECG3010/ ECG6010, follow the same design principle, main components, accessories and software; they acquire ECG electrical signals from patient body surface by ECG electrodes. After been amplified, filtered and transferred, the ECG signal waveforms are displayed on the LCD and recorded on the paper through thermal printer.
They consist of four modules, which are power supply module, signal collection module, amplification module, and control module.
They can acquire ECG signal via twelve leads simultaneously, display or print waveform of ECG signal via three channel, six channel, twelve channel or fifteen channel.
The CardioPro analysis program is the analysis program used in this proposed device.
1. Table of acceptance criteria and the reported device performance
The document does not explicitly state acceptance criteria in a quantitative manner for specific performance metrics like accuracy, sensitivity, or specificity regarding the analysis and diagnosis of heart disease. Instead, the submission focuses on demonstrating substantial equivalence to predicate devices and compliance with international safety and performance standards.
The table below summarizes the technical specifications and standards compliance, which implicitly serve as performance requirements for the device's functionality.
| Feature / Standard | Acceptance Criteria (Implied by Predicate & Standards) | Reported Device Performance |
|---|---|---|
| Product Code | DPS | DPS |
| Regulation Number | 21 CFR 870.2340 | 21 CFR 870.2340 |
| Lead Acquisition | Standard 12-lead (or 15-lead for proposed device) | Standard 12-lead or 15 lead |
| Acquisition mode | Simultaneous 12-lead acquisition | Simultaneous 12-lead or 15 lead acquisition |
| Recording format | Automatic / Manual / Rhythm | Automatic / Manual / Rhythm |
| Analysis mode | Yes (presence of analysis program) | Yes (CardioPro analysis program) |
| CMRR | >60dB | >60dB, >100 with AC filter |
| Paper Speed | Multiple levels (e.g., 6.25, 12.5, 25, 50mm/s) | 4-6 levels (e.g., 6.25, 12.5, 25, 50mm/s) |
| Input CIR current | ≤0.1μA | ≤0.1μA |
| Input impedance | >50MΩ | >50MΩ |
| Patient leak current | <10μA | <10μA |
| Frequency response | 0.05~150Hz | 0.05~150Hz |
| Noise level | <15μVp-p | <15μVp-p |
| Electrical Safety | Comply with IEC 60601-1 | Comply with IEC 60601-1 |
| EMC | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 |
| Particular requirements | Comply with IEC 60601-2-25 | Comply with IEC 60601-2-25 |
| Biocompatibility | Comply with ISO 10993 | Comply with ISO 10993 |
2. Sample size used for the test set and the data provenance
The document does not mention a specific "test set" in the context of clinical performance for the interpretation of ECGs or diagnosis of heart disease. The provided information is focused on non-clinical testing to demonstrate compliance with general safety and performance standards (IEC standards) and an assessment of technical characteristics against predicate devices. Therefore, details like data provenance (country of origin, retrospective/prospective) are not applicable or provided for a clinical test set.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
As there is no mention of a clinical test set with ground truth related to ECG interpretation or diagnosis in this document, there is no information about the number or qualifications of experts used for establishing ground truth. The device's interpretive statements are explicitly stated to be "on an advisory basis only," suggesting that human experts remain responsible for final diagnosis.
4. Adjudication method for the test set
Not applicable, as no clinical test set for diagnostic performance is described.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done
No, the document does not indicate that a multi-reader multi-case (MRMC) comparative effectiveness study was performed. The non-clinical tests relate to hardware compliance and functional specifications.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
The document does not describe a standalone performance study for the algorithm's diagnostic capabilities. The non-clinical tests performed were to verify that the device met design specifications and complied with safety and performance standards. The "CardioPro analysis program" is mentioned, and its interpretive statements are stated to be for "advisory basis only," indicating it's not intended for standalone diagnosis.
7. The type of ground truth used
Not applicable, as no clinical test set for diagnostic performance is described. The "ground truth" implicitly referred to in the non-clinical tests would be the established benchmarks and requirements defined by the IEC standards for technical performance (e.g., signal-to-noise ratio, frequency response).
8. The sample size for the training set
The document does not provide any information about a "training set" or explicit details on how the "CardioPro analysis program" was developed or trained.
9. How the ground truth for the training set was established
Not applicable, as no information on a training set or its ground truth establishment is provided.
Ask a specific question about this device
(392 days)
Digital Electrocardiograph is intended to acquire ECG signals from adult patients through body surface ECG electrodes. The obtained ECG records can help users to analyze and diagnose heart disease. Digital Electrocardiograph shall be used in healthcare facilities by doctors and/or trained healthcare professionals.
Digital Electrocardiograph, ECG-2000, is intended to acquire ECG signals from adult patients through body surface ECG electrodes. The obtained ECG records can help users to analyze and diagnose heart disease.
It consists of the acquisition box, optional accessories, installation CD, and USB dongle. The acquisition box is connected with ECG electrodes, and intends to acquire ECG signals from patient. The acquired ECG signal is filtered and amplified, then transferred to the working station. The acquisition box is powered by the working station via a USB line. The working station is a Personal Computer installed with ECG-2000 Software, it intends to receive the processed ECG signal from the acquisition box, and further display and record the signals. Optional accessories include lead wires, limb electrodes and chest electrodes, and general-purpose printer (with USB interface, support PCL language, such as HP1010, P2035, and P2055d series). The CD contains Digital Electrocardiograph installer and dongle installation program module.
The provided text is a 510(k) summary for a Digital Electrocardiograph (ECG-2000) and primarily focuses on demonstrating substantial equivalence to predicate devices through non-clinical testing and comparison of technical specifications. It does not contain information about acceptance criteria for device performance based on clinical studies, nor does it describe such a study.
Therefore, I cannot provide the requested information regarding acceptance criteria, reported device performance from a clinical study, sample sizes, data provenance, expert qualifications, adjudication methods, MRMC studies, standalone performance, or how ground truth was established for a training set.
The document states that non-clinical tests were conducted to verify that the proposed device met all design specifications and complied with relevant IEC and EN standards for electrical safety, EMC, and particular requirements for electrocardiographs. It also mentions "Software Verification Test was performed to verify the software functions against its intended use." However, these are engineering verification tests, not clinical performance studies for diagnostic accuracy.
The closest to "acceptance criteria" presented in the document are the technical specifications compared with predicate devices, such as:
| Item | Acceptance Criteria (from predicate device comparison) | Reported Device Performance (Proposed Device) |
|---|---|---|
| Product Code | DPS | DPS |
| Regulation Number | 21 CFR 870.2340 | 21 CFR 870.2340 |
| Class | Class II | Class II |
| ECG Lead | Standard 12-lead | Standard 12-lead |
| Sampling rate | 1000Hz (at least for one predicate) | 1000Hz |
| Electrical Safety | Comply with IEC 60601-1 | Comply with IEC 60601-1 |
| EMC | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 |
| Input circuit current | ≤0.1 μA (for one predicate) | ≤0.1 μA |
| Noise level | <15μ Vp-p (for one predicate) | <15μ Vp-p |
Summary of missing information:
- A table of acceptance criteria and the reported device performance from a clinical study: Not available. The document refers to engineering specifications and compliance with standards, not diagnostic performance metrics from clinical data.
- Sample sized used for the test set and the data provenance: Not available. No clinical test set is described.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not available. No clinical ground truth establishment is described.
- Adjudication method for the test set: Not available.
- If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not available. This device is a Digital Electrocardiograph, not an AI-assisted diagnostic tool as described by the question.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not available. The device acquires ECG signals; it's not an algorithm that performs diagnosis independently for performance evaluation in this context.
- The type of ground truth used: Not available.
- The sample size for the training set: Not available.
- How the ground truth for the training set was established: Not available.
Ask a specific question about this device
(148 days)
Digital Electrocardiographs are intended to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The obtained ECG records can help users to analyze and diagnose heart disease. Digital Electrocardiographs shall be used in healthcare facilities by doctors and/or trained healthcare professionals.
Digital Electrocardiographs, iE 3, iE 6, are intended to acquire, display and record ECG signals from adult and pediatric patients through body surface by ECG electrodes. ECG data result and patient information could be stored in the memory of the device. The obtained ECG records can help users to analyze and diagnose heart disease.
The provided text is a 510(k) summary for Digital Electrocardiographs (models iE 3, iE 6) and a letter from the FDA determining their substantial equivalence to a predicate device (iE 12). This document does not contain information about specific acceptance criteria, device performance metrics, or details of a study demonstrating the device meets such criteria.
The core of this document is a comparison between the proposed devices and a predicate device to establish "substantial equivalence," meaning that the new devices are as safe and effective as a legally marketed device. This is a regulatory pathway, not a clinical study demonstrating performance against specific benchmarks.
Therefore, many of the requested details cannot be extracted from this document.
Here's what can be inferred or directly stated from the provided text, and what is explicitly missing:
1. Table of Acceptance Criteria and Reported Device Performance
- Acceptance Criteria: Not explicitly stated in terms of performance metrics (e.g., sensitivity, specificity, accuracy for disease detection). The acceptance criteria, in the context of this 510(k), are primarily related to meeting design specifications and demonstrating substantial equivalence to the predicate device, especially regarding safety and intended use.
- Reported Device Performance: No specific numerical performance metrics are reported in this document. The document states a "Non-Clinical Test Conclusion" that "Non clinical tests were conducted to verify that the proposed device met all design specifications as vas Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards: IEC 60601-1, IEC 60601-1-2." This refers to electrical safety and electromagnetic compatibility standards, not clinical performance for diagnosing heart disease.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Not mentioned.
- Data Provenance: Not mentioned.
3. Number of Experts Used to Establish Ground Truth and Qualifications
- Not mentioned, as a clinical study with ground truth established by experts is not detailed here.
4. Adjudication Method
- Not mentioned.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- If done: No, an MRMC comparative effectiveness study is not described or referenced in this document. This submission focuses on establishing substantial equivalence to a predicate device based on technical characteristics and intended use, not on demonstrating improved human reader performance with AI assistance.
- Effect size of human readers improvement: Not applicable, as no such study is presented.
6. Standalone (Algorithm Only) Performance Study
- If done: No, a standalone performance study of an algorithm is not described or referenced. The device is a "Digital Electrocardiograph" which acquires, displays, and records ECG signals for users (doctors/trained professionals) to analyze and diagnose. While it "can help users to analyze and diagnose heart disease," the document does not detail an inherent diagnostic algorithm's performance.
7. Type of Ground Truth Used
- Not mentioned, as no clinical study requiring a ground truth is detailed.
8. Sample Size for the Training Set
- Not mentioned, as this document does not describe the development or training of an AI algorithm.
9. How Ground Truth for the Training Set Was Established
- Not mentioned, as this document does not describe the development or training of an AI algorithm.
Summary of what the document does provide:
- Device Models: iE 3, iE 6
- Intended Use: To acquire ECG signals from adult and pediatric patients through body surface ECG electrodes, with obtained records helping users to analyze and diagnose heart disease. To be used in healthcare facilities by doctors and/or trained healthcare professionals.
- Predicate Device: Digital Electrocardiographs, iE 12 (K122712)
- Regulatory Basis for Acceptance: Substantial Equivalence (SE) to the predicate device, demonstrated through non-clinical tests verifying compliance with design specifications and international safety and EMC standards (IEC 60601-1, IEC 60601-1-2).
- Key Comparison Points for SE: Intended use, technological characteristics, principles of operation, patient contact material, sterile/single use status, energy source, safety standards, and accessories are deemed equivalent to the predicate device.
Ask a specific question about this device
Page 1 of 1