Search Results
Found 2 results
510(k) Data Aggregation
(48 days)
Shenzhen Baoan Medical Supplies Co., Ltd
Sterile High-pressure Angiographic Syringes for Single-use are intended for the injection of contrast media or saline; they shall be used with an US legally marketed angiographic injectors.
The proposed devices include 2 models of plastic, disposable syringes and 5 connection tubes of different specifications. They are intended to be used with an U.S. legally marketed angiography injector, compatibilities are shown in Table 1. The connection tubes are used to connect the syringe and the catheter. The tubes are also available in three configurations, which are straight tube (used with single shot syringe), type Y and type T tube (used with dual shot syringe), they can withstand maximum pressure of 400 psi. The proposed devices are used with the J shape tube / spike cleared in K151960.
The provided FDA 510(k) summary (K182289) describes the substantial equivalence of Sterile High-pressure Angiographic Syringes for Single-use to a predicate device, focusing on mechanical and material properties rather than AI/algorithm performance. Therefore, most of the requested information regarding AI acceptance criteria, study design (MRMC, standalone), ground truth establishment, and training/test set details for AI models is not applicable to this document.
The document primarily addresses the physical and functional characteristics of a medical device (syringes and connection tubes) and its compliance with relevant international standards.
However, based on the non-clinical tests described, we can infer a form of acceptance criteria and the "study" (non-clinical testing) that proves the device meets these criteria.
Here's an interpretation based on the provided document:
Acceptance Criteria and Device Performance (Non-AI Medical Device)
Acceptance Criteria (Inferred from Standards & Claims) | Reported Device Performance (Non-Clinical Test Conclusion) |
---|---|
Resistant Liquid Leak Pressure: Must withstand a certain pressure without leakage. (Implicitly, the proposed device seeks to demonstrate a higher resistance than the predicate) | Syringe: 400 psi (Tested) |
Connection Tube: 400 psi (Tested) | |
Material & Design Compatibility: Conformance to specified luer taper standards for secure connections. | ISO 594-1:1986 (Conical fittings, Part 1: General requirements) - Complies |
ISO 594-2:1998 (Conical fittings, Part 2: Lock fittings) - Complies | |
Syringe Functionality (Manual Use): Conformance to general requirements for single-use hypodermic syringes for manual use. | ISO 7886-1:2017 (Sterile hypodermic syringes for single use, Part 1: Syringes for manual use) - Complies |
Syringe Functionality (Power-Driven Use): Conformance to requirements for single-use hypodermic syringes for use with power-driven pumps. | ISO 7886-2:1996 (Sterile hypodermic syringes for single use, Part 2: Syringes for use with power-driven syringe pumps) - Complies |
Sterility: EO sterilized. | EO Sterilized |
Single Use: Designed for single use. | Yes |
Biocompatibility: Conformance to standards for biological evaluation of medical devices. | ISO 10993 series Standards - Conforms to requirements |
Explanation of "Study" Proving Acceptance Criteria (Non-Clinical Testing):
The "study" here refers to non-clinical bench testing conducted to verify the device's technical specifications and compliance with relevant standards.
-
Sample Size Used for the Test Set and Data Provenance:
- Sample Size: Not explicitly stated as a number of units tested. The document mentions "The test samples of syringe 100104 and 100114 with accessories are tested for liquid leakage and flow characteristics." This implies a series of physical tests on multiple units of the specified models. Regulatory submissions for devices of this nature typically involve testing a statistically representative sample size, but the exact number isn't provided in this summary.
- Data Provenance: The tests were conducted internally by the manufacturer, Shenzhen Baoan Medical Supplies Co., Ltd. (China). The data is retrospective in the sense that it was collected as part of product development and validation prior to submission.
-
Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
- Not Applicable. For a medical device like a syringe, "ground truth" is established through engineering specifications, material science, and performance standards, not through expert consensus on medical images or diagnoses. The "truth" is whether the device meets the physical and functional requirements (e.g., does it leak at 400 psi? Does it conform to luer taper dimensions?). These are objective measurements not requiring expert clinical adjudication.
-
Adjudication Method for the Test Set:
- Not Applicable. As mentioned above, the tests are objective, measurable physical and mechanical tests. There is no subjective interpretation requiring adjudication by multiple readers or experts.
-
If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
- No. An MRMC study is a clinical study design typically used for diagnostic devices (like imaging AI) to assess the impact of a new technology on human reader performance. This document concerns a physical device (syringe), not a diagnostic tool or AI.
-
If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- No. This is not an AI/algorithm. This question is irrelevant to the device described.
-
The Type of Ground Truth Used:
- Engineering Specifications and International Standards: The ground truth for this device's performance is defined by adherence to established international standards (ISO 594-1, ISO 594-2, ISO 7886-1, ISO 7886-2, ISO 10993 series) and the manufacturer's own design specifications (e.g., target resistant liquid leak pressure). This is a physical/mechanical "ground truth."
-
The Sample Size for the Training Set:
- Not Applicable. There is no "training set" as this is not an AI/machine learning device. The design and manufacturing processes are based on engineering principles and established device standards, not trained data.
-
How the Ground Truth for the Training Set was Established:
- Not Applicable. Again, this is not an AI/machine learning device. "Ground truth" for manufacturing and design would be derived from engineering drawings, material specifications, and regulatory requirements rather than labeled training data.
Ask a specific question about this device
(175 days)
SHENZHEN BAOAN MEDICAL SUPPLIES CO., LTD
Sterile High-pressure Angiographic Syringes for Single-use are intended for the injection of contrast media or saline; they shall be used with an US legally marketed angiographic injectors.
The proposed devices are available in packs, which may include different configurations of syringes and accessories. The syringes are plastic, disposable syringes, which are available in various models and configurations. They are intended to be used with an U.S. legally marketed angiography injector, compatibilities are shown in Table 1.
The provided document is a 510(k) summary for the Shenzhen Baoan Medical Supplies Co., Ltd. Sterile High-pressure Angiographic Syringes for Single-use. This type of submission is for medical devices, and the "acceptance criteria" and "device performance" primarily refer to compliance with recognized standards and biocompatibility, rather than clinical efficacy or diagnostic accuracy.
Here's an analysis based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance:
The document lists various international and national standards that the device complies with, which serve as the acceptance criteria for its non-clinical performance and safety. The reported device performance is that it "complies with" these standards.
Acceptance Criteria (Standards Complied With) | Reported Device Performance |
---|---|
ISO 10993-7:2008 Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals | Complies |
ASTM F 88/F88M-09 Standard test method for seal strength of flexible barrier materials | Complies |
ASTM F1140/F1140M-13 Standard test methods for internal pressurization failure resistance of unrestrained packages | Complies |
ASTM F1929-12 Standard Test Method for Detecting Seal Leaks in Porous Medical Package by Dye Penetration | Complies |
USP37-NF32 Bacterial Endotoxins Limit | Complies |
ISO 11737-2:2009 Sterilization of medical devices - Microbiological methods - Part 2: Test of sterility | Complies |
ISO 7886-1:1993 Sterile hypodermic syringes for single use - Part 1: Syringes for manual use | Complies |
ISO 7886-2:1996 Sterile hypodermic syringes for single use - Part 2: Syringes for use with power-driven Syringe Pumps | Complies |
ISO594:1986 Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 1 | Complies |
ISO594:1998 Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 2 | Complies |
ISO 10993-4:2002 AMD 2006 Biological evaluation of medical devices -- Part 4: Selection of tests for interactions with blood | Complies |
ISO 10993-5:2009 Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity | Complies |
ISO 10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization | Complies |
ISO 10993-11:2006 Biological evaluation of medical devices -- Part 11: Tests for systemic toxicity | Complies |
ASTM F 756-08, Standard practice for assessment of hemolytic properties of material | Complies |
USP37-NF32 Pyrogen test | Complies |
Biocompatibility requirements of ISO 10993 series Standards | Conforms to the requirements (for both proposed and predicate devices) |
2. Sample Size Used for the Test Set and Data Provenance:
The document states "Non clinical tests were conducted to verify that the proposed device met all design specifications..." and lists various standards for these tests. However, it does not specify the sample sizes used for these non-clinical tests (e.g., number of syringes tested for seal strength or internal pressurization). The data provenance is generally from laboratory testing performed by the manufacturer to demonstrate compliance with the listed standards. No country of origin is explicitly stated for the testing itself, but the manufacturer is based in China. The data is prospective in the sense that the manufacturer conducted these tests to support the 510(k) submission.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
This type of information is not applicable to this 510(k) submission. The document focuses on the engineering and material safety aspects of a medical device (angiographic syringes), not on diagnostic accuracy where expert ground truth would be established. The "ground truth" here is compliance with established engineering and biological safety standards, which are verified through objective testing methods rather than expert consensus on images or case outcomes.
4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set:
Not applicable. Adjudication methods like "2+1" typically refer to the process of resolving discrepancies in expert interpretations (e.g., in reading medical images). Since no expert interpretations are involved in establishing the "ground truth" for these engineering and biological tests, no such adjudication method was used.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:
No, an MRMC comparative effectiveness study was not done. The document explicitly states: "No clinical study is included in this submission." This type of study is relevant for assessing the clinical utility or diagnostic effectiveness of devices, often involving human readers interpreting medical data. This syringe device is a non-active, sterile, single-use product, and its evaluation focuses on its physical and biological integrity, not its interaction with clinical interpretation.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done:
Not applicable. This device is a physical medical product (angiographic syringe), not an algorithm or software. Therefore, the concept of "standalone performance" for an algorithm is not relevant here.
7. The Type of Ground Truth Used:
The "ground truth" for this device's evaluation is primarily based on compliance with recognized national and international consensus standards for medical device physical performance, sterility, and biocompatibility. This involves objective measurements against predefined thresholds and specifications outlined in standards like ISO 7886, ISO 10993, ASTM F88, etc.
8. The Sample Size for the Training Set:
Not applicable. The concept of a "training set" is relevant for machine learning algorithms. This submission is for a physical medical device, not an AI/ML product. The manufacturer did not develop this device using machine learning.
9. How the Ground Truth for the Training Set was Established:
The concept of a training set and its ground truth is not applicable for this device submission.
Ask a specific question about this device
Page 1 of 1