Search Results
Found 2 results
510(k) Data Aggregation
(27 days)
SANTEC CORPORATION
ARGOS is a non-invasive, non-contact biometer based on swept-source optical coherence tomography (SS-OCT). The device is intended to acquire ocular measurements as well as perform calculations to determine the appropriate intraocular lens (IOL) power and type for implantation during intraocular lens placement. ARGOS measures the following 9 parameters: Axial Length, Corneal Thickness, Anterior Clamber Depth, Lens Thickness, K-values (Radii of flattest and steepest meridians), Astigmatism, White (corneal diameter) and Pupil Size. The Reference Image functionality is intended for use as a preoperative and postoperative image capture tool.
It is intended for use by ophthalmologists, physicians, and other eye-care professionals and may only be used under the supervision of a physician.
The ARGOS is substantially equivalent to the predicate device identified previously:
- the ARGOS (Santec Corporation) that was cleared by the FDA on October 2nd, 2015(K150754) as the primary predicate as it has the most similar intended use and characteristics
- the IOLMaster 700 (Carl Zeiss Meditec) that was cleared by the FDA on June 29, ● 2015 (K143275, K170171) as an additional (secondary) predicate to support a new feature of reference image capture on the version of ARGOS in this submission.
The predicate devices are Class 2 devices to premarket notification, as defined per regulation number 21 CFR 886.1850. In addition, the predicate devices have product codes of MXK(ARGOS), and HJO(IOLMaster700).
The version of ARGOS in this submission is a modified version of the Argos cleared under K150754 which is substantially equivalent with regard to intended use, operating principle, function, materials, and energy source. The differences from the predicate ARGOS (K150754) that are subject of this 510(k) submission are:
- An additional feature of reference image capture function ●
- Labeling change including change in the intended use adding the feature of reference . image capture.
- . This image can be transferred to image guided devices in order to support the execution of preoperative plan.
The changes described in this submission do not affect how the hardware is used to acquire measurements as a biometer, nor do these changes affect the principle of operation of the device.
The provided document describes the ARGOS device, a non-invasive, non-contact biometer based on swept-source optical coherence tomography (SS-OCT), and its 510(k) submission for clearance. The document focuses on demonstrating substantial equivalence to predicate devices, particularly regarding the ARGOS ver1.5 which includes a new "Reference Image functionality."
Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:
1. Acceptance Criteria and Reported Device Performance
The document does not explicitly present a "table of acceptance criteria and reported device performance" in terms of specific quantitative thresholds that the new ARGOS ver1.5 needs to meet for its added functionality. Instead, it argues for substantial equivalence primarily by:
- Comparing technological characteristics of the new ARGOS (ver1.5) with its primary predicate (ARGOS K150754) and a secondary predicate (IOLMaster 700 K143275, K170171).
- Stating that the new feature (reference image capture) is similar to a feature already present in a legally marketed predicate (IOLMaster 700).
- Referring to compliance with recognized consensus standards for performance and safety.
- Highlighting that the core biometry measurement capabilities did not change from the primary predicate. The in-vivo repeatability specifications for measured parameters (Axial Length, Corneal Thickness, Anterior Chamber Depth, Lens Thickness, Keratometry, Astigmatism, Pupil Size, White-to-White) are listed as identical to the primary predicate, implying these performance metrics were already accepted.
Implied Acceptance Criteria and Reported Performance (from comparison tables):
Parameter / Characteristic | Acceptance Criteria (Implied - by being equivalent to Predicate ARGOS) | Reported Device Performance (ARGOS ver1.5) |
---|---|---|
Biometry Measurements | Identical to Predicate ARGOS (K150754) | Identical to Predicate ARGOS (K150754) |
Axial Length Repeatability | SD 0.01mm (range 14-38mm) | SD 0.01mm (range 14-38mm) |
Corneal Thickness Repeatability | SD 10um (range 200-120um) | SD 10um (range 200-120um) |
Anterior Chamber Depth Repeatability | SD 0.01mm (range 0.7-8.0mm) | SD 0.01mm (range 0.7-8.0mm) |
Lens Thickness Repeatability | SD 0.02mm (range 0.5-10.0mm) | SD 0.02mm (range 0.5-10.0mm) |
Keratometry Repeatability | SD 0.02mm (range 5.5-10.0mm) | SD 0.02mm (range 5.5-10.0mm) |
Astigmatism Repeatability | SD 5deg (Cylinder>1D) (range 0-180deg) | SD 5deg (Cylinder>1D) (range 0-180deg) |
Pupil Size Repeatability | SD 0.09mm (range 2-13mm) | SD 0.09mm (range 2-13mm) |
White-to-White Repeatability | SD 0.06mm (range 7-15mm) | SD 0.06mm (range 7-15mm) |
Reference Image Functionality | Similar to IOLMaster 700 predicate | Implemented using white LED illumination |
Electrical Safety | Compliance with ANSI/AAMI ES 60601-1, IEC 60601-1-2 | Compliant |
Housing Protection | Compliance with IEC 60529 (IP20) | Compliant (IP20) |
Light Hazard | Compliance with ANSI.Z80-36 (Group 2), IEC 62471 (Exempt group) | Compliant (Group 2 for SS-OCT, Exempt for LED illumination) |
Laser Product Safety | Compliance with IEC 60825-1 (Laser Class 1) | Compliant (Laser Class 1) |
Ophthalmometer Accuracy | Compliance with ISO 10343 | Compliant |
Hardware Performance | Meet requirements for axial/lateral dist., SNR, depth attenuation | Met |
Software Evaluation | Compliance with FDA guidance, IEC 62366, ISO 14971, cybersecurity guidance | Verified and Validated |
Usability | Comparable to predicate Argos | Evaluated and found comparable |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify the sample size for any clinical test set in terms of patient data. It primarily relies on bench testing and software verification/validation. The comparison of in-vivo repeatability values references the existing performance of the predicate device, not new clinical data for the ARGOS ver1.5 to prove equivalent clinical performance.
- Sample Size for Test Set: Not explicitly stated for patient data. The "Performance Testing" section refers to "bench tests" and "software verification and validation," which typically involve internal testing and simulations rather than patient samples for the new features. The existing in-vivo repeatability data cited appears to be from the primary predicate, not new testing on the modified device regarding its core measurement functions.
- Data Provenance: The document does not provide details on the country of origin of data or whether it was retrospective or prospective. Given the focus on substantial equivalence through design and testing against standards rather than new clinical trials for the added feature, such details are not expected to be prominent.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
N/A. The document does not describe a clinical study involving human readers or expert-established ground truth for a test set, especially pertaining to the new "Reference Image functionality." The changes are assessed through engineering and software validation, and comparison to existing, already-cleared devices.
4. Adjudication Method for the Test Set
N/A. No clinical study with human readers and adjudication is described.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done
N/A. No MRMC study is described. The device is not an AI/CADe device that assists human readers in diagnosis. It's a measurement device with an added image capture utility.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
The device itself is a standalone measurement instrument. Its "performance" is its ability to accurately measure ocular parameters and capture images. The "Performance Testing" section (12.1 and 12.2) effectively describes this standalone performance evaluation, focusing on:
- Bench tests: ISO standards, hardware specifications (axial/lateral distance, SNR, depth attenuation).
- Software verification and validation: According to FDA guidance and ISO standards.
- Cybersecurity review.
These tests demonstrate the device's capability to function as intended without human intervention for the measurement process itself, or for the image capture function. The usability evaluation ensures the human-device interaction is acceptable.
7. The Type of Ground Truth Used
- For biometric measurements: The "ground truth" for proving performance (cited from the predicate) would typically be established based on highly accurate reference devices or physical models, although the document doesn't detail the predicate's original ground truth methodology. For the new device, the claim is that its measurement capabilities have not changed and are thus equivalent to the predicate.
- For reference image functionality: The ground truth for this new feature would be its ability to capture and transfer an image. This is validated by functional testing and comparison to the secondary predicate (IOLMaster 700), which already has this function. There isn't a "ground truth" in the clinical diagnostic sense for an image capture tool beyond its intended function of capturing an image.
- For safety and performance standards: The "ground truth" is compliance with the specifications and limits defined by the international and national consensus standards (e.g., ISO, IEC, ANSI/AAMI).
8. The Sample Size for the Training Set
N/A. This is not an AI/ML device that requires a training set in the typical sense of machine learning. It's a measurement instrument. The "Vision Planner software package" and "Argos UI software" are traditional software applications that undergo verification and validation, not model training.
9. How the Ground Truth for the Training Set Was Established
N/A. No training set is involved.
Ask a specific question about this device
(193 days)
SANTEC CORPORATION
ARGOS is a non-invasive, non-contact biometer based on swept-source optical coherence tomography (SS-OCT). The device is intended to acquire ocular measurements as well as perform calculations to determine the appropriate intraocular lens (IOL) power and type for implantation during intraccular lens placement. ARGOS measures the following 9 parameters: Axial Length, Corneal Thickness, Anterior Chamber Depth, Lens Thickness, K-values (Radii of flattest and steepest meridians), Astigmatism, White (corneal diameter) and Pupil Size. It is intended for use by ophthalmologists, physicians, and other eye-care professionals and may only be used under the supervision of a physician.
Argos is a swept-source based biometer that provides the biometry and keratometry of the eye prior to cataract surgery and aids in the selection of the appropriate IOL.
Argos processes 3 OCT images (B-scans) and 3 CMOS camera images per measurement in seconds - reducing procedure time and minimizing patient discomfort. From the OCT images the biometry parameters are evaluated by segmenting the cornea, iris, lens, and retina. The operator can edit the biometry parameters by manual adjustment: Axial length, Corneal thickness, Anterior chamber depth (or aqueous depth), Lens thickness, White-towhite (corneal diameter), Pupil size. The keratometry values: K-values (flattest and steepest meridians), Astigmatism (angle of flattest meridian), are evaluated from the CMOS camera image, in combination with the OCT information.
All distance/thickness parameters (Axial length, Corneal thickness, Anterior chamber depth, Lens thickness, White, Pupil size) are simultaneously measured from 2dimensional OCT images.
All the boundaries of ocular segments are detected and distances and thicknesses are calculated by taking into account the refractive indices of each medium.
Axial length is the distance from the corneal apex to the fovea or, more specifically, to the ILM (Internal Limiting Membrane). The calculation is performed as the sum of the thicknesses (after refraction correction) of the cornea, aqueous humor, lens and vitreous.
Corneal thickness is the distance between the anterior and the posterior apexes of the cornea divided by the cornea refractive index (1.375).
Aqueous depth is evaluated as the distance between the posterior surface of the cornea and the anterior surface of the lens divided by the refractive index (1.336).
Anterior chamber depth is the sum of corneal and aqueous humor distances.
Lens thickness is the distance between the anterior and the posterior surfaces of the lens divided by its refractive index (1.410).
Pupil size is the lateral distance between the two inner boundaries of the iris region.
White-to-white (Corneal diameter) is the lateral distance between the inner boundaries of the cornea-sclera interface.
K-values (Radii of flattest and steepest meridians) and Astigmatism are evaluated by the size and distance of the reflected images of the infrared LED ring projected onto the cornea.
Safety protocol is well integrated into the device in both hardware and software to ensure the safety of both the patient and the operator.
The control program on the computer runs on Windows. The operation panel on the display attached to the PC provides all the functionalities.
IOL calculation uses widely recognized formulas; Hoffer Q, Haigis, Holladay1, SRK/T, etc.
The provided document describes the ARGOS device, an optical biometer, and its performance evaluation. Here's a breakdown of the acceptance criteria and study details:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state formal "acceptance criteria" for the clinical studies in the format of a predefined threshold. Instead, it presents the device's performance in relation to a predicate device (Lenstar LS900) through various statistical measures, implying that performance comparable to the predicate or within acceptable clinical ranges constitutes "acceptance." The "repeatability" values can also be considered a form of performance criteria.
Parameter | Acceptance Criteria (Implied / Comparator Performance for Agreement) | Reported Device Performance (ARGOS) |
---|---|---|
Agreement Study (vs. Lenstar LS900) | ||
Axial Length (AL) | Similar mean, small mean difference (0.00 mm), small SD (0.05 mm), LoA of [-0.10, 0.09] mm, Deming slope close to 1 (1.03), high Spearman correlation (1.00) compared to predicate (Lenstar: Mean 23.43 (1.01), SD N/A, Spearman 1.00 for comparison, Imp. Repeatability SD: 0.035mm). | Mean: 23.44 (0.98), Mean Difference (Argos-Lenstar): 0.00 (0.05) mm, LoA: [-0.10, 0.09] mm, Deming Regression: Intercept -0.68 [-0.93,-0.43], Slope 1.03 [1.02,1.04], Spearman Correlation: 1.00. Repeatability (Precision Study): SD 0.01 mm, CV 0.02%. Reproducibility: SD 0.02 mm, CV 0.09%. |
Corneal Thickness (CCT) | Similar mean, small mean difference (0.00 mm), small SD (0.01 mm), LoA of [-0.02, 0.02] mm, Deming slope close to 1 (1.04), high Spearman correlation (0.97) compared to predicate (Lenstar: Mean 0.53 (0.04), SD N/A, Spearman 0.97 for comparison, Imp. Repeatability SD: 2.3um (0.0023mm)). | Mean: 0.53 (0.03), Mean Difference (Argos-Lenstar): 0.00 (0.01) mm, LoA: [-0.02, 0.02] mm, Deming Regression: Intercept -0.02 [-0.05,0.01], Slope 1.04 [0.99,1.10], Spearman Correlation: 0.97. Repeatability (Precision Study): SD 0.00 mm, CV 0.89%. Reproducibility: SD 0.01 mm, CV 1.19%. |
Aqueous Depth (AD) | Similar mean, small mean difference (0.05 mm), small SD (0.16 mm), LoA of [-0.27, 0.37] mm, Deming slope close to 1 (1.06), high Spearman correlation (0.89) compared to predicate (Lenstar: Mean 2.66 (0.41), SD N/A, Spearman 0.89 for comparison, Imp. Repeatability SD: 0.04mm). | Mean: 2.63 (0.39), Mean Difference (Argos-Lenstar): 0.05 (0.16) mm, LoA: [-0.27, 0.37] mm, Deming Regression: Intercept -0.10 [-0.29,0.09], Slope 1.06 [0.98,1.13], Spearman Correlation: 0.89. Repeatability (Precision Study): SD 0.01 mm, CV 0.37%. Reproducibility: SD 0.03 mm, CV 0.90%. |
Anterior Chamber Depth (ACD) | Similar mean, small mean difference (0.05 mm), small SD (0.17 mm), LoA of [-0.28, 0.37] mm, Deming slope close to 1 (1.07), high Spearman correlation (0.89) compared to predicate (Lenstar: Mean 3.19 (0.42), SD N/A, Spearman 0.89 for comparison, Imp. Repeatability SD: 0.04mm). | Mean: 3.16 (0.39), Mean Difference (Argos-Lenstar): 0.05 (0.17) mm, LoA: [-0.28, 0.37] mm, Deming Regression: Intercept -0.17 [-0.39,0.06], Slope 1.07 [1.00,1.14], Spearman Correlation: 0.89. Repeatability (Precision Study): SD 0.01 mm, CV 0.27%. Reproducibility: SD 0.03 mm, CV 0.77%. |
Lens Thickness (LT) | Similar mean, small mean difference (-0.18 mm), small SD (0.22 mm), LoA of [-0.61, 0.24] mm, Deming slope close to 1 (1.09), high Spearman correlation (0.87) compared to predicate (Lenstar: Mean 4.46 (0.45), SD N/A, Spearman 0.87 for comparison, Imp. Repeatability SD: 0.08mm). | Mean: 4.62 (0.41), Mean Difference (Argos-Lenstar): -0.18 (0.22) mm, LoA: [-0.61, 0.24] mm, Deming Regression: Intercept -0.58 [-1.10,-0.05], Slope 1.09 [0.97,1.20], Spearman Correlation: 0.87. Repeatability (Precision Study): SD 0.02 mm, CV 0.38%. Reproducibility: SD 0.06 mm, CV 1.45%. |
Pupil Size (PS) | Similar mean, small mean difference (-0.26 mm), small SD (0.46 mm), LoA of [-1.16, 0.64] mm, Deming slope close to 1 (0.81), high Spearman correlation (0.88) compared to predicate (Lenstar: Mean 4.10 (0.77), SD N/A, Spearman 0.88 for comparison, Imp. Repeatability SD: not specified). Note: Differences attributed to lack of ambient light control. | Mean: 4.38 (0.91), Mean Difference (Argos-Lenstar): -0.26 (0.46) mm, LoA: [-1.16, 0.64] mm, Deming Regression: Intercept 0.55 [-0.02,1.11], Slope 0.81 [0.68,0.95], Spearman Correlation: 0.88. Repeatability (Precision Study): SD 0.09 mm, CV 1.83%. Reproducibility: SD 1.01 mm, CV 20.31%. |
Corneal Diameter (CD) | Similar mean, small mean difference (-0.28 mm), small SD (0.69 mm), LoA of [-1.62, 1.06] mm, Deming slope close to 1 (0.57 - lower than 1 but still within CI), high Spearman correlation (0.48 - lower correlation acknowledged due to different measurement principle) compared to predicate (Lenstar: Mean 11.86 (0.49), SD N/A, Spearman 0.48 for comparison, Imp. Repeatability SD: 0.04mm). Note: Accuracy not well confirmed due to different measurement principle. | Mean: 12.11 (0.68), Mean Difference (Argos-Lenstar): -0.28 (0.69) mm, LoA: [-1.62, 1.06] mm, Deming Regression: Intercept 4.91 [1.79,8.04], Slope 0.57 [0.31,0.83], Spearman Correlation: 0.48. Repeatability (Precision Study): SD 0.06 mm, CV 0.46%. Reproducibility: SD 0.10 mm, CV 0.77%. |
Flattest Radius of Curvature (R1) | Similar mean, small mean difference (-0.01 mm), small SD (0.07 mm), LoA of [-0.15, 0.13] mm, Deming slope close to 1 (1.01), high Spearman correlation (0.97) compared to predicate (Lenstar: Mean 7.79 (0.30), SD N/A, Spearman 0.97 for comparison, Imp. Repeatability SD: 0.02mm). | Mean: 7.80 (0.29), Mean Difference (Argos-Lenstar): -0.01 (0.07) mm, LoA: [-0.15, 0.13] mm, Deming Regression: Intercept -0.09 [-0.45,0.26], Slope 1.01 [0.96,1.06], Spearman Correlation: 0.97. Repeatability (Precision Study): SD 0.01 mm, CV 0.14%. Reproducibility: SD 0.03 mm, CV 0.34%. |
Steepest Radius of Curvature (R2) | Similar mean, small mean difference (0.00 mm), small SD (0.07 mm), LoA of [-0.14, 0.14] mm, Deming slope close to 1 (0.97), high Spearman correlation (0.98) compared to predicate (Lenstar: Mean 7.61 (0.27), SD N/A, Spearman 0.98 for comparison, Imp. Repeatability SD: 0.03mm). | Mean: 7.61 (0.28), Mean Difference (Argos-Lenstar): 0.00 (0.07) mm, LoA: [-0.14, 0.14] mm, Deming Regression: Intercept 0.22 [-0.05,0.48], Slope 0.97 [0.94,1.01], Spearman Correlation: 0.98. Repeatability (Precision Study): SD 0.02 mm, CV 0.21%. Reproducibility: SD 0.04 mm, CV 0.48%. |
Average Radius of Curvature (Rav) | Similar mean, small mean difference (-0.01 mm), small SD (0.06 mm), LoA of [-0.12, 0.11] mm, Deming slope close to 1 (0.99), high Spearman correlation (0.97) compared to predicate (Lenstar: Mean 7.70 (0.28), SD N/A, Spearman 0.97 for comparison, Imp. Repeatability SD: N/A). | Mean: 7.70 (0.28), Mean Difference (Argos-Lenstar): -0.01 (0.06) mm, LoA: [-0.12, 0.11] mm, Deming Regression: Intercept 0.06 [-0.17,0.30], Slope 0.99 [0.96,1.02], Spearman Correlation: 0.97. Repeatability (Precision Study): SD 0.01 mm, CV 0.13%. Reproducibility: SD 0.03 mm, CV 0.35%. |
Astigmatism (Cyl. 1D) | Similar mean, small mean difference (4.36 degrees), small SD (9.71 degrees), LoA of [-14.67, 23.39] degrees, Deming slope close to 1 (1.05), high Spearman correlation (0.97) compared to predicate (Lenstar: Mean 81.44 (46.96), SD N/A, Spearman 0.97 for comparison, Imp. Repeatability SD: 11deg). | Mean: 78.40 (46.34), Mean Difference (Argos-Lenstar): 4.36 (9.71) Degrees, LoA: [-14.67, 23.39] Degrees, Deming Regression: Intercept 0.07 [-6.02,6.15], Slope 1.05 [0.96,1.15], Spearman Correlation: 0.97. Repeatability (Precision Study, All AST): SD 4.39 degrees, CV 5.59%. Reproducibility (All AST): SD 11.13 degrees, CV 14.17%. |
Enhanced Retinal Visualization (ERV) mode - Cataractous Eyes (vs. Lenstar for AL) | ||
Axial Length (AL) (ERV) | Statistically insignificant differences (p-value 1D): 36 eyes |
* **Data Provenance:** A single U.S. clinical site; collected prospectively for the agreement study.
-
Precision Study (Repeatability and Reproducibility - Table 17.2.2):
- 43 healthy eyes (for both repeatability and reproducibility).
- Data Provenance: Not explicitly stated, but generally part of the same prospective clinical study.
-
Enhanced Retinal Visualization (ERV) mode Validation Study (Table 17.2.3 and 17.2.4):
- Cataractous eyes: 45 eyes (23 right eyes) for comparison with Lenstar.
- Healthy eyes: 43 consecutive healthy eyes for repeatability of ERV mode.
- Data Provenance: Not explicitly stated, but generally part of the same prospective clinical study.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
The document does not mention the use of experts to establish ground truth for the test set beyond the clinical measurements themselves. The study primarily relies on the agreement between the ARGOS device and the predicate device (Lenstar LS900) as the reference for performance, and the precision of the ARGOS itself.
4. Adjudication Method for the Test Set
No adjudication method is described for the test set. The clinical studies compare measurements between devices and evaluate the precision of the ARGOS device.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size
No MRMC comparative effectiveness study involving human readers' improvement with or without AI assistance was done or described. This device is an optical biometer for objective measurements, not an image interpretation AI system for diagnosis where human reader studies are typical.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
Yes, the studies described are standalone performance evaluations of the ARGOS device. The ARGOS device is an automated measurement tool. The "Agreement Study" and "Precision Study" assess the device's inherent measurement capabilities. While an operator initiates measurements, the performance metrics (Axial Length, CCT, etc.) are generated solely by the device's algorithms and hardware.
7. The Type of Ground Truth Used
The "ground truth" for the clinical studies is implicitly established by:
- Predicate Device Measurements: The Lenstar LS900, an already legally marketed optical biometer, serves as the primary comparator. The assumption is that its measurements are an acceptable reference.
- Internal Consistency/Precision: The repeatability and reproducibility studies establish the device's consistency in its own measurements, which is a form of internal ground truth for precision.
There is no mention of "pathology" or external "outcomes data" being used as ground truth for parameter measurements in this context.
8. The Sample Size for the Training Set
The document does not specify a training set or its sample size. This is typical for medical devices that perform direct physical measurements rather than AI algorithms trained on large datasets for pattern recognition. The algorithms for biometry are based on physical principles (e.g., swept-source OCT) and established calculations, not machine learning model training in the conventional sense.
9. How the Ground Truth for the Training Set Was Established
As no training set is described for an AI model, there is no information on how its ground truth was established. The device relies on physical measurement principles and established formulas (e.g., Hoffer Q, Haigis, Holladay1, SRK/T) for IOL power calculation.
Ask a specific question about this device
Page 1 of 1