Search Filters

Search Results

Found 9 results

510(k) Data Aggregation

    K Number
    K243988
    Device Name
    RootMend MRR
    Date Cleared
    2025-03-20

    (84 days)

    Product Code
    Regulation Number
    N/A
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Riverpoint Medical LLC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The RootMend MRR devices are intended to be used for suture (soft tissue) fixation to bone in the knee for meniscal root repair.

    Device Description

    The RootMend Meniscal Root Repair (MRR) devices are composed of uncoated UHMWPE (ultra-high molecular weight polyethylene) with UHMWPE tracer material. RootMend MRR devices are available with a fixation button composed of a titanium (Ti-6Al-4V) button, or with an attached UHMWPE fixation button. RootMend MRR devices will be provided sterile using ethylene oxide, for single use only, and are not to be resterilized. The UHMWPE is available undyed (white), black, or blue, with trace filaments of black or blue. RootMend MRR devices are intended to be used for suture (soft tissue) fixation to bone in the knee for meniscal root repair.

    AI/ML Overview

    The provided text from the FDA 510(k) summary for the RootMend MRR device states that "Non-clinical mechanical testing was performed to verify the fixation strength of the RootMend MRR devices and is compared to the predicate device. Testing conducted includes cyclic and ultimate tensile strength (UTS) testing, and assessment of displacement for the subject button and predicate anchor constructs. Usability engineering testing with simulated use in cadaver lab was performed on the subject device per EN62366-1. In all instances of the testing referenced above, the acceptance criteria were met, and the proposed RootMend MRR devices performed as intended."

    However, the provided document does not contain the specific acceptance criteria or the detailed results (e.g., reported device performance values) from the mechanical testing and usability testing. It also lacks information regarding sample size, ground truth establishment, expert qualifications, or MRMC studies that would be relevant for an AI/software device evaluation.

    Therefore, I cannot fulfill the request the following information based on the provided text:

    • A table of acceptance criteria and the reported device performance: Not provided.
    • Sample sized used for the test set and the data provenance: Not provided.
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable to this type of device and testing described.
    • Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
    • If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this is a hardware device, not an AI/software device assisting human readers.
    • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    • The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For mechanical testing, the "ground truth" is the specified performance thresholds (acceptance criteria) and the measured physical properties. For usability testing, the 'ground truth' is successful completion of tasks by users in a cadaver lab. Specifics are not detailed beyond "performed as intended."
    • The sample size for the training set: Not applicable, as this is not an AI/machine learning device.
    • How the ground truth for the training set was established: Not applicable.

    Based on the available information, here is what can be inferred or directly stated:

    1. Acceptance Criteria and Device Performance:

    The document states: "In all instances of the testing referenced above, the acceptance criteria were met, and the proposed RootMend MRR devices performed as intended."

    • Acceptance Criteria: Not explicitly stated in quantitative terms. They likely relate to minimum cyclic and ultimate tensile strength, and maximum displacement relative to the predicate device.
    • Reported Device Performance: Not explicitly stated in quantitative terms. The performance met the unstated acceptance criteria.

    2. Sample Size and Data Provenance:

    • Sample Size: Not specified for any of the tests (cyclic, UTS, displacement, usability).
    • Data Provenance: The tests were "non-clinical mechanical testing" and "usability engineering testing with simulated use in cadaver lab." This implies laboratory/bench testing rather than human clinical data. The location of the testing or data origin (country) is not specified.

    3. Experts for Ground Truth / Adjudication / MRMC studies / Standalone Performance:

    • These concepts (experts, adjudication, MRMC, standalone performance) are typically relevant for AI/software-based medical devices that interpret medical images or data. The RootMend MRR is a physical orthopedic implant (metallic bone fixation fastener). Therefore, these criteria are not applicable to the device and studies described.

    4. Type of Ground Truth:

    • For mechanical testing: The ground truth is established by engineering specifications and comparative performance to a predicate device. This would involve objective measurements of strength, fatigue, and displacement against predefined thresholds.
    • For usability testing: The ground truth is successful completion of surgical tasks and safe operation in a simulated environment (cadaver lab) as assessed by the test protocol and observations.

    5. Training Set Information:

    • Sample Size for Training Set: Not applicable. This is not an AI/machine learning device that uses a "training set."
    • How Ground Truth for Training Set was Established: Not applicable.

    In summary, the provided FDA 510(k) summary provides a high-level statement that the device met acceptance criteria through specified non-clinical mechanical and usability testing, but it does not disclose the quantitative details of those criteria, the measured performance, or the sample sizes used. The remaining questions are not relevant given the nature of the device as a physical implant rather than an AI/software product.

    Ask a Question

    Ask a specific question about this device

    K Number
    K242494
    Date Cleared
    2024-11-26

    (96 days)

    Product Code
    Regulation Number
    888.3040
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Riverpoint Medical LLC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OsseoFit Interfacing Anchor and OsseoFit Bone Tunnel Construct are intended for use in soft tissue to bone fixation for the following indications: Shoulder Rotator Cuff Repair

    Device Description

    The Riverpoint Medical Interfacing Anchor is comprised of a pre-deployed all-suture anchor with attached repair sutures and 302 stainless steel needles. The Interfacing Anchor is inserted into either the humerus or into passing slots available in devices intended for total shoulder arthroplasty. The Riverpoint Medical Bone Tunnel Construct is comprised of a double armed Ultra-High Molecular Weight Polyethylene (UHMWPE) suture with a preassembled polypropylene passing loop and 302 stainless steel needles. Suture supplied meet United States Pharmacopeia (USP) requirements for nonabsorbable suture except for diameter. Suture dyes are FDA authorized. The device is sterilized by ethylene oxide gas and is provided sterile for single use. Interfacing Anchor and Bone Tunnel Construct are available in common sizes and lengths with preattached 302 stainless steel needles and will be sold sterile for single use. The device is intended for use in a hospital/clinic/surgical setting.

    AI/ML Overview

    The provided text is a 510(k) summary for the OsseoFit Interfacing Anchor and OsseoFit Bone Tunnel Construct. This document details the regulatory review process for a medical device and its substantial equivalence to a predicate device.

    It does NOT describe an AI/ML-driven medical device, nor does it contain information about diagnostic performance metrics (e.g., sensitivity, specificity, AUC), test set characteristics, expert adjudication processes, or MRMC studies.

    The "Performance Data" section of the document refers to non-clinical performance testing for a surgical device (an anchor and tunnel construct), not a diagnostic AI. The tests mentioned are:

    • Usability engineering validation with simulated use in cadaveric models (EN62366:2015)
    • Endotoxin/pyrogenicity testing (ANSI/AAMI ST72:2019, USP , USP , USP )
    • Mechanical testing to verify fixation strength (cyclic and pullout testing)

    Therefore, I cannot provide the requested information regarding acceptance criteria and study details for an AI/ML device's performance, as the provided text pertains to a physical surgical device.

    To directly answer your prompt based on the provided text, the following cannot be extracted:

    • A table of acceptance criteria and reported device performance related to diagnostic accuracy (as it's not a diagnostic device). The acceptance criteria mentioned are for mechanical and usability performance of a surgical implant.
    • Sample sizes used for a "test set" in the context of image analysis or diagnostic performance. The "test set" here refers to cadaveric models and mechanical testing samples.
    • Data provenance (country of origin, retrospective/prospective) for a diagnostic dataset.
    • Number of experts used to establish ground truth or their qualifications.
    • Adjudication method for a test set.
    • MRMC comparative effectiveness study, including effect size.
    • Standalone (algorithm only) performance.
    • Type of ground truth (expert consensus, pathology, outcomes data).
    • Sample size for a training set.
    • How ground truth for the training set was established.

    However, based on the provided text, here's what can be inferred or directly stated about the "acceptance criteria" and "study" for this specific device (a surgical implant):

    Acceptance Criteria and Device Performance (for a surgical implant):

    Acceptance Criterion (Type)Reported Device Performance
    Material/Suture PropertiesSuture supplied meets USP requirements (except for diameter).
    UHMWPE sutures tested per USP for needle attachment & tensile strength.
    Biocompatibility/SterilityEndotoxin/pyrogenicity testing per various standards; results below recommended limits.
    Mechanical Performance (Fixation Strength)Performed comparably to the predicate device in cyclic and pullout testing.
    Usability/Clinical PerformanceUsability engineering validation with simulated use in cadaveric models performed as per EN62366:2015. Results support use and IFU.
    General Performance (Overall Summary)"In all instances of the testing referenced above, the acceptance criteria were met, and the proposed OsseoFit Interfacing Anchor and OsseoFit Bone Tunnel Construct performed as intended."
    "Validations performed demonstrated that the OsseoFit Interfacing Anchor and OsseoFit Bone Tunnel Construct met all requirements for its intended use."

    Study Details (for this surgical implant):

    1. Sample sizes used for the test set and data provenance:

      • The "test set" refers to samples for mechanical testing and cadaveric models for usability. Specific numerical sample sizes are not provided in this summary.
      • Data provenance is not described (e.g., where the cadavers came from, or if the mechanical tests were done by a specific lab).
      • The studies were non-clinical ("bench testing" and "simulated use").
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

      • Not applicable in the context of diagnostic "ground truth." The "ground truth" for a surgical implant's mechanical performance is defined by engineering standards and comparative testing to predicates. Usability testing would involve typical users (surgeons, OR staff), but the number or qualifications are not specified.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not applicable as this is not a diagnostic study requiring human expert consensus/adjudication of results. Performance is measured against physical test standards and engineering specifications.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, a MRMC comparative effectiveness study was not done. This is a physical surgical device, not an AI/ML diagnostic tool.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable as this is a physical surgical device, not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For mechanical testing: Engineering specifications and comparative performance to a legally marketed predicate device.
      • For usability testing: Performance in simulated use conditions in cadaveric models against predefined usability objectives (derived from EN62366:2015).
    7. The sample size for the training set:

      • Not applicable. This is not an AI/ML device with a "training set."
    8. How the ground truth for the training set was established:

      • Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K232411
    Date Cleared
    2023-10-06

    (57 days)

    Product Code
    Regulation Number
    888.3040
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Riverpoint Medical, LLC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    JuggerKnot Soft Anchor OC are intended for use in soft tissue to bone fixation for the following indications: Knee MPFL Knee Patellar tendon repair Knee MCL Knee Quadriceps tendon repair Hip Acetabular labral repair Hip Proximal hamstring repair Hip Hip Labral reconstruction Foot and Ankle Achilles tendon repair Foot and Ankle Medial/lateral repair and reconstruction Foot and Ankle Plantar plate repair Foot and Ankle Mid- and forefoot repair Foot and Ankle Metatarsal ligament/tendon repair or reconstruction Shoulder Rotator Cuff Shoulder Shoulder Instability Shoulder Biceps Tenodesis Elbow Lateral epicondylitis repair Elbow Biceps tendon reattachment

    Device Description

    The JuggerKnot Soft Anchor OC is identical to the previously cleared JuggerKnot Soft Anchor OC except for the presence of a bioceramics embedded in the anchor portion of the device. The JuggerKnot Soft Anchor OC is comprised of a suture sleeve structure and working suture. Nonabsorbable braided ultra-high molecular weight polyethylene (UHMWPE) sutures are spliced through a suture anchor sleeve comprised of non-absorbable braided polyester. Up to two nonabsorbable round or flat braided UHMWPE working sutures can be added inside the suture anchor sleeve. The UHMWPE sutures are available undyed (white) and black. The hydroxyapatite and bioceramic particles are embedded in the surface of the suture anchor material. Sutures supplied meet United States Pharmacopeia (USP) requirements for non-absorbable suture except for diameter. Suture dyes are FDA approved. The inserter is comprised of a metallic shaft with over molded handle. The device is sterilized by ethylene oxide gas and is provided sterile for single use. The JuggerKnot Soft Anchor OC are available in common sizes and lengths and will be sold sterile for single use. The device is intended for use in a hospital/clinic/surgical setting.

    AI/ML Overview

    The provided text describes a medical device, the JuggerKnot Soft Anchor OC, and its clearance under a 510(k) premarket notification. This type of submission relies on demonstrating substantial equivalence to a legally marketed predicate device, rather than conducting a full clinical trial to prove efficacy and safety from scratch.

    Therefore, the specific information requested about acceptance criteria and a study proving the device meets those criteria (especially in terms of diagnostic performance, human reader improvement with AI assistance, standalone algorithm performance, and ground truth establishment for training and test sets) is not applicable in the context of this 510(k) submission.

    The document focuses on non-clinical performance testing and substantial equivalence to a predicate device, as opposed to a diagnostic device for which such performance metrics would be relevant.

    Here's a breakdown based on the information provided and what is implied by a 510(k) submission for this type of device:

    1. A table of acceptance criteria and the reported device performance:

      • Acceptance Criteria: For this device (a bone fixation fastener), acceptance criteria are primarily related to mechanical performance, biocompatibility, sterilization, and material properties, demonstrating it is as safe and effective as the predicate.
      • Reported Device Performance:
        • Mechanical Testing: "performed comparably to the predicate device in insertion, cyclic and pullout testing."
        • Sterilization: "sterilization adoption validation"
        • Biocompatibility: "biocompatibility testing per ISO10993- 1:2018"
        • Stability: "stability testing on the product packaging per ISO 11607-1:2006"
        • Usability Engineering: "usability engineering validation with simulated use in a cadaveric models performed per EN62366: 2015"
        • Endotoxin/Pyrogenicity: "Endotoxin/pyrogenicity testing was performed per ANSI/AAMI ST72:2019, USP , USP and USP ."
    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

      • Test Set (for mechanical and other non-clinical tests): Not explicitly stated numbers for each test, but standard engineering and biological testing samples would have been used.
      • Data Provenance: The tests were likely conducted in laboratories or facilities chosen by the manufacturer, Riverpoint Medical, LLC, which is based in Portland, Oregon, USA. The studies are prospective in the sense that they were conducted for the purpose of this submission.
      • Animal Study: "An animal study was performed to evaluate the biological safety and in vivo performance associated with the JuggerKnot Soft Anchor OC." No specific animal numbers are provided.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

      • This is not applicable as the device is a medical implant (bone fixation fastener), not a diagnostic device requiring expert interpretation of images or other medical data for ground truth. Ground truth for this device relates to physical and biological properties.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • This is not applicable for the reasons stated above. Adjudication methods are typically for diagnostic interpretations.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • This is not applicable. The device is an implantable fastener, not an AI-assisted diagnostic tool.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • This is not applicable. The device is a physical implant, not an algorithm.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For the non-clinical tests, the "ground truth" is based on established engineering standards, material science properties, and biological safety parameters. For the animal study, the "ground truth" would be the observed biological response to the implant in vivo.
    8. The sample size for the training set:

      • This is not applicable. This is a physical device submission; there is no "training set" in the context of machine learning or AI.
    9. How the ground truth for the training set was established:

      • This is not applicable for the reason stated above.
    Ask a Question

    Ask a specific question about this device

    K Number
    K231078
    Date Cleared
    2023-10-05

    (171 days)

    Product Code
    Regulation Number
    888.3040
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Riverpoint Medical, LLC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Iconix® HA+™ anchors are intended to be used for soft-tissue to bone fixation in the foot, ankle, knee, hip, hand, wrist, elbow and shoulder. Specific indications are listed below.

    Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction

    Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Repair

    Hand Wrist: Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP, and MCP joints for all digits, Digital Tendon Repar

    Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus Reconstruction, Digital Tendon Transfers, Mid-foot Reconstruction

    Knee: Medial Collateral Ligament Repair, Lateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair. Iliotibial Band Tenodesis. Medial Patellofemoral Ligament (MPFL) Repair/ Reconstruction, Quadriceps Tendon Repair

    Hip: Capsular Repair, Acetabular Labral Repair, Gluteal Tendon Repair, Proximal Hamstring Repair

    Device Description

    The Iconix® HA*TM Anchor is comprised of a suture sleeve structure and working suture. Nonabsorbable braided ultra-high molecular weight polyethylene (UHMWPE) sutures are spliced through a suture anchor sleeve comprised of non-absorbable braided polyester and bioceramics. Up to two non- absorbable round or flat braided UHMWPE working sutures can be added inside the suture anchor sleeve. The UHMWPE sutures are available undyed (white) and black.

    Sutures supplied meet United States Pharmacopeia (USP) requirements for non-absorbable suture except for diameter. Suture dyes are FDA approved. The inserter is comprised of a metallic shaft with over molded handle. The device is sterilized by ethylene oxide gas and is provided sterile for single use. Iconix® HA*M Anchors are available in common sizes and lengths and will be sold sterile for single use with no components or accessories. The device is intended for use in a hospital/clinic/surgical setting.

    AI/ML Overview

    This document describes the validation study for the Iconix® HA+TM Anchor, a medical device used for soft-tissue to bone fixation. This is a 510(k) Pre-market Notification, which focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving novel effectiveness. Therefore, the "acceptance criteria" and "device performance" are primarily demonstrated through comparison to the predicate and established performance benchmarks for such devices, rather than a clinical accuracy study for AI/software.

    Here's the breakdown of the information requested, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a mechanical device, not an AI/software device, the "acceptance criteria" and "reported device performance" are primarily based on meeting established standards and demonstrating comparable mechanical properties to a predicate device.

    Acceptance Criteria CategorySpecific Test/StandardReported Device Performance
    Material PerformanceUHMWPE sutures meet USP requirements for non-absorbable suture (except for diameter)UHMWPE sutures tested per USP performance requirements for tensile strength. (Implicitly, the device met these, as it concluded substantial equivalence.)
    BiocompatibilityISO 10993-1:2018 - Biological Evaluation of Medical DevicesBiocompatibility testing performed per ISO10993-1:2018. An animal study was also performed. (Implicitly, the device met these, demonstrating biological safety.)
    SterilizationSterilization adoption validation (likely internal protocol based on recognized standards)Sterilization adoption validation performed. Device is sterilized by ethylene oxide gas and provided sterile. (Implicitly, the device met these.)
    PackagingISO 11607-1:2006 - Packaging for terminally sterilized medical devicesStability testing on product packaging per ISO 11607-1:2006. (Implicitly, the device met these.)
    UsabilityEN 62366: 2015 - Medical devices - Application of usability engineeringUsability engineering validation with simulated use in cadaveric models performed per EN62366:2015. (Implicitly, the device met these.)
    Endotoxin/PyrogenicityANSI/AAMI ST72:2019, USP , USP , USPEndotoxin/pyrogenicity testing performed per specified standards. (Implicitly, the device met these.)
    Mechanical FixationComparison to predicate device (K173074) for insertion, cyclic, and pullout testing"Results of performance testing for the Iconix® HA+TM Anchor device concluded that the device performed comparably to the predicate device and to other currently marketed All-Suture anchor devices in insertion, cyclic and pullout testing and the validations performed demonstrated that the Iconix® HA+TM Anchor met all requirements for its intended use."

    2. Sample Size Used for the Test Set and Data Provenance

    The provided document describes non-clinical performance testing and an animal study. It does not describe a clinical study with a human-derived test set in the way an AI/software device would.

    • Non-clinical Mechanical Testing: The sample sizes for insertion, cyclic, and pullout testing are not specified. The studies were performed to compare against the predicate device.
    • Animal Study: The sample size for the animal study is not specified. The purpose was to evaluate biological safety and in vivo performance.
    • Data Provenance: The document implies these were prospective laboratory and animal studies conducted by the manufacturer, Riverpoint Medical, LLC. There's no mention of country of origin for the data or whether it was retrospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This question is not applicable in the context of this device and study type.

    • For non-clinical mechanical testing, the "ground truth" is typically defined by engineering specifications and objective measurements (e.g., force, displacement), not expert interpretation.
    • For the animal study, "ground truth" would be established through histological analysis, observation of biological responses, and potentially pathological evaluation, conducted by veterinary pathologists or researchers, but this is not explicitly detailed as "experts establishing ground truth" in the sense of a diagnostic interpretation task.

    4. Adjudication Method for the Test Set

    This is not applicable. There was no human "test set" requiring adjudication in the context of an AI/software performance study. Mechanical and biological test results are typically objectively measured against defined criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If So, What was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    This is not applicable. This device is a physical bone anchor, not an AI/software diagnostic tool. There were no human "readers" involved in interpreting findings from the device itself.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    This is not applicable. This device is a physical bone anchor, not an algorithm or software. The mechanical and biological tests assess the device's inherent function, which is a "standalone" evaluation of its physical properties.

    7. The Type of Ground Truth Used

    • Mechanical Testing: Ground truth was based on objective engineering measurements (e.g., load-to-failure, displacement) and comparison to the predicate device's established performance and industry standards for bone anchors.
    • Biocompatibility/Animal Study: Ground truth was based on biological responses and safety profiles observed in the animal model, likely through histological analysis, gross observations, and clinical pathology, evaluated against established safety benchmarks for implantable materials.
    • Suture Testing: Ground truth was based on United States Pharmacopeia (USP) requirements for tensile strength.

    8. The Sample Size for the Training Set

    This is not applicable. This is a physical device, not an AI/software model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable, as there was no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231278
    Date Cleared
    2023-08-01

    (90 days)

    Product Code
    Regulation Number
    888.3040
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Riverpoint Medical, LLC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Iconix® Knotless Anchors are intended to be used for soft-tissue to bone fixation in the foot, ankle, knee, hip, hand, wrist, elbow and shoulder. Specific indications are listed below.

    Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction

    Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Repair

    Hand/Wrist: Scapholunate Ligament Reconstruction, Carpal Ligament Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP, and MCP joints for all digits, Digital Tendon Repair

    Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus Reconstruction, Digital Tendon Transfers, Mid-foot Reconstruction

    Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis, Medial Patellofemoral Ligament (MPFL) Repair/ Reconstruction, Quadriceps Tendon Repair

    Hip: Capsular Repair, Acetabular Labral Repair, Gluteal Tendon Repair, Proximal Hamstring Repair

    Device Description

    The Iconix® Knotless Anchors are soft-tissue fixation devices, provided preloaded on a disposable inserter. The device is composed of a braided polyester anchor body that contains one working suture, also referred to as the repair strand, and a shuttle strand that is used to shuttle the repair strand around tissue and through the braided anchor body.

    Sutures supplied meet United States Pharmacopeia (USP) requirements for non-absorbable suture except for diameter. Suture dyes are FDA approved. The inserter is comprised of metallic shaft with over molded handle. The device is sterilized by ethylene oxide gas and is provided sterile for single use. The anchor size will be available in 1.4mm with working sutures in standard USP size 2. Iconix® Knotless Anchor will be sold sterile for single use with no components or accessories.

    AI/ML Overview

    The provided document is a 510(k) summary for the Iconix® Knotless Anchor, a medical device used for soft-tissue to bone fixation. It describes the device, its intended use, and its equivalence to a predicate device (Stryker ICONIX Anchors).

    However, this document does not contain any information about a study involving AI/algorithm performance or human-in-the-loop studies. It focuses solely on the mechanical and biological safety of the physical medical device (the knotless anchor itself) in comparison to a predicate device.

    Therefore, I cannot fulfill your request to describe the acceptance criteria and the study that proves the device meets the acceptance criteria in the context of AI/algorithm performance. The document only discusses non-clinical mechanical testing (insertion, cyclic, and pullout testing) to demonstrate performance comparability of the physical anchor to a predicate device, and standard safety validations (sterilization, biocompatibility, stability, usability engineering).

    In summary, there is no information in the provided text that relates to:

    • Acceptance criteria for an AI/algorithm (e.g., sensitivity, specificity, AUC)
    • Sample sizes for algorithm testing (test sets, training sets)
    • Data provenance for AI/algorithm development or testing
    • Number/qualifications of experts for AI ground truth
    • Adjudication methods for AI ground truth
    • MRMC comparative effectiveness studies
    • Standalone algorithm performance
    • Type of ground truth for AI (pathology, outcomes data, etc.)
    • Training set size or ground truth establishment for an AI algorithm

    The document focuses on the equivalency of a physical medical device (a bone anchor) and its mechanical properties.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220765
    Date Cleared
    2023-07-26

    (497 days)

    Product Code
    Regulation Number
    888.3010
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Riverpoint Medical, LLC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HS Fiber Cerclage sutures are indicated for use in general soft tissue approximation and/or ligation. These sutures may be used in cardiovascular surgeries, and orthopedic surgeries using allograft tissue. When used as a bone fixation cerclage, the sutures are intended for:

    • · Trauma surgery indications including olecranon, ankle, patella, and some shoulder rewiring.
    • · Repair of long bone fracture due to trauma or reconstruction.
    Device Description

    The Riverpoint Medical HS Fiber® Cerclage sutures are non-absorbable, sterile, surgical sutures composed of multiple single strands of ultra-high molecular weight polyethylene (UHMWPE) braided together to form the implant. HS Fiber Cerclage sutures are available in common sizes and lengths with or without pre-attached needles. Suture supplied meet United States Pharmacopeia (USP) requirements for non-absorbable suture except for diameter. The device is sterilized by ethylene oxide gas, and is provided sterile for single use. The device is intended for use in a hospital/clinic/surgical setting.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Riverpoint Medical HS Fiber® Cerclage, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Test/RequirementReported Device Performance
    Material PropertiesSuture MaterialMultiple single strands of ultra-high molecular weight polyethylene (UHMWPE) braided together.
    SterilizationSterilized by ethylene oxide gas. Meets EN ISO14937:2009.
    BiocompatibilityMeets ISO 10993-1:2018.
    Device PerformanceNeedle AttachmentMeets USP performance requirements.
    Tensile StrengthMeets USP performance requirements. Performed comparably to the predicate device.
    Knot StrengthTested. No specific numerical performance provided, but "met all requirements for its intended use."
    Fatigue StrengthPerformed comparably to the predicate device.
    CreepPerformed comparably to the predicate device.
    Wear DebrisPerformed comparably to the predicate device. Particle analysis performed.
    Packaging & StabilityProduct & Packaging StabilityMeets ISO 11607-1:2019.
    Human Factors/UsabilityUsability ValidationMet all acceptance criteria per EN62366-1: 2015.
    Regulatory ComplianceUSP RequirementsMeets USP requirements for non-absorbable suture, except for diameter.
    FDA Guidance ComplianceFollowed "Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA."

    2. Sample Size Used for the Test Set and Data Provenance

    The provided text does not explicitly state the sample sizes used for each specific test set. It broadly mentions tests performed, but not the number of units or samples involved in, for example, the tensile strength or fatigue strength testing.

    Regarding data provenance, the study appears to be non-clinical, laboratory-based testing conducted by the manufacturer, Riverpoint Medical, LLC. No information is given about the country of origin of the data beyond the manufacturer's location in Portland, Oregon, USA. The studies are retrospective in the sense that they are conducted on manufactured devices as part of the submission process, not on a prospective cohort of patients.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable as the studies described are non-clinical, mechanical, and material characteristic tests, not studies requiring expert interpretation of medical images or patient data to establish a ground truth.

    4. Adjudication Method for the Test Set

    This information is not applicable for the same reasons as point 3.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC comparative effectiveness study was not done. The device is a physical medical device (suture) and not an AI/software-based medical device that would involve human readers or AI assistance.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    A standalone performance study was not done. Again, this is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the non-clinical performance studies were:

    • Established standards and specifications: United States Pharmacopeia (USP) requirements for sutures (primarily for needle attachment and tensile strength), and various ISO and EN standards for sterilization, biocompatibility, packaging, and usability.
    • Predicate device performance: Comparative testing was conducted against a predicate device (Arthrex FiberTape Cerclage - K170206) to demonstrate comparable performance in areas like tensile strength, fatigue strength, creep, and wear debris.

    8. The Sample Size for the Training Set

    This information is not applicable as there is no "training set" for this type of physical medical device. The device is manufactured based on design specifications and then tested against these specifications and regulatory standards.

    9. How the Ground Truth for the Training Set was Established

    This information is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231128
    Date Cleared
    2023-05-17

    (27 days)

    Product Code
    Regulation Number
    888.3040
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Riverpoint Medical LLC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    JuggerKnot® Soft Anchors are intended for soft tissue to bone fixation for the following indications: Knee MPFL Knee Patellar tendon repair Knee MCL Knee Quadriceps tendon repair Foot and Ankle Achilles tendon repair Foot and Ankle Medial/lateral repair and reconstruction Foot and Ankle Plantar plate repair Foot and Ankle Mid- and forefoot repair Foot and Ankle Metatarsal ligament/tendon repair or reconstruction Shoulder Rotator Cuff Shoulder Shoulder Instability Shoulder Biceps Tenodesis Elbow Lateral epicondylitis repair Elbow Biceps tendon reattachment

    Device Description

    The JuggerKnot® Soft Anchor is comprised of a suture sleeve structure and working suture. Non-absorbable braided ultra-high molecular weight polyethylene (UHMWPE) sutures are spliced through a non-absorbable braided polyester suture anchor sleeve. Up to three nonabsorbable round and flat braided UHMWPE working sutures can be added inside the suture anchor sleeve. The UHMWPE sutures are available undyed (white), blue, black or with or without a stainless steel needle attached. Available Suture sizes are standard according to USP requirements (dependent on suture type). Suture supplied meet United States Pharmacopeia (USP) requirements for non-absorbable suture except for diameter. Suture dyes are FDA approved. The inserter is comprised of a metallic shaft with a self-punching tip and an overmolded handle. The device is sterilized by ethylene oxide gas, and is provided sterile for single use. JuggerKnot® Soft Anchors are available in common sizes and lengths with or without pre-attached 302 stainless steel needles and will be sold sterile for single use with no components or accessories. The device is intended for use in a hospital/clinic/surgical setting.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device (JuggerKnot® Soft Anchor) that has received FDA clearance. It focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed study proving the device meets specific acceptance criteria in the manner of an AI/ML algorithm or diagnostic test.

    Therefore, the information required to populate the fields related to AI/ML device performance (like "acceptance criteria table," "sample size for the test set," "data provenance," "number of experts for ground truth," "adjudication method," "MRMC study," "standalone performance," "type of ground truth," "training set sample size," and "ground truth for training set") is not present in the provided document.

    The document describes non-clinical mechanical testing and usability engineering validation, but these are not the types of studies that would generate the performance metrics typically requested for an AI/ML or diagnostic device's acceptance criteria table.

    Here's a breakdown of what is available and what is not:

    Information Present in the Document (relevant to general device clearance, but not AI/ML specific):

    • Device Name: JuggerKnot® Soft Anchor
    • Intended Use/Indications for Use: Detailed list of soft tissue to bone fixation indications (e.g., Knee MPFL, Shoulder Rotator Cuff).
    • Predicate Device: K203740 – Riverpoint JuggerKnot® Soft Anchor.
    • Performance Data (General):
      • Sutures meet USP requirements (except diameter).
      • UMHWPE sutures tested for needle attachment and tensile strength per USP.
      • Followed FDA Guidance documents for "Bone Anchors" and "Surgical Sutures."
      • Biocompatibility, sterilization method, packaging, material stability are the same as the predicate.
      • Non-clinical performance testing:
        • Usability engineering validation with simulated use in cadaveric models per EN62366:2015.
        • Endotoxin/pyrogenicity testing per ANSI/AAMI ST72:2019, USP , USP , USP .
        • Mechanical testing (insertion, cyclic, and pullout testing) comparing to predicate and other marketed soft anchor devices.
    • Conclusion: Device performed comparably to the predicate and met all requirements for its intended use, demonstrating substantial equivalence.

    Information NOT Present in the Document (required for the requested AI/ML acceptance criteria study description):

    1. A table of acceptance criteria and the reported device performance: This document does not present performance in terms of sensitivity, specificity, accuracy, or similar metrics typically seen for AI/ML or diagnostic devices. Instead, it discusses mechanical performance relative to a predicate.
    2. Sample sized used for the test set and the data provenance: No "test set" in the context of an AI/ML algorithm is mentioned. The mechanical testing implicitly used a "test set" of anchors, but the size and provenance are not detailed, nor is it the type of test set requested (e.g., patient data).
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, as there's no diagnostic ground truth being established.
    4. Adjudication method: Not applicable.
    5. Multi Reader Multi Case (MRMC) comparative effectiveness study: Not applicable. This is not an AI-assisted diagnostic device.
    6. Standalone performance: While the device has "standalone performance" as a mechanical anchor, this isn't relevant to "algorithm only without human-in-the-loop performance."
    7. The type of ground truth used: Not applicable in the AI/ML context. Ground truth for a mechanical device is its physical performance characteristics.
    8. The sample size for the training set: Not applicable, as this is not an AI/ML device that requires a training set.
    9. How the ground truth for the training set was established: Not applicable.

    In summary, the provided document describes the clearance of a mechanical medical device (a surgical anchor) based on its substantial equivalence to an existing predicate. It does not pertain to an AI/ML algorithm or a diagnostic device, and therefore, the requested information regarding AI/ML acceptance criteria and study details cannot be extracted from this text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230212
    Device Name
    OrthoButton AL
    Date Cleared
    2023-02-23

    (28 days)

    Product Code
    Regulation Number
    888.3040
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Riverpoint Medical, LLC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Riverpoint Medical OrthoButton® AL is intended for use in the fixation of bone and soft tissue in orthopedic procedures requiring ligament or tendon reconstruction.

    Device Description

    The Riverpoint Medical OrthoButton® AL is comprised of a braided ultra-high molecular weight polyethylene ("UHMWPE") adjustable loop combine with a titanium (Ti-6A1-4V ELI per ASTM F136) plate. The device is provided sterile for single use. The device is intended for use in a hospital/clinic/surgical setting.

    AI/ML Overview

    This document does not describe an AI/ML powered device, nor does it describe software that provides automated analysis or diagnosis. The device, OrthoButton® AL, is a physical medical device (Fastener, Fixation, Non-Degradable, Soft Tissue) used in orthopedic procedures. Therefore, the questions related to acceptance criteria, AI performance, ground truth, and study design for an AI/ML system are not applicable to this submission.

    The document discusses performance data related to the physical device's characteristics (sterilization, biocompatibility, stability, usability) and its substantial equivalence to a predicate device.

    To answer your request, if this were an AI/ML device, the relevant information would typically be found in a section detailing "Software Verification and Validation" or "Clinical Performance Studies" specifically for the algorithmic aspects. This document focuses on the physical properties and intended use of a mechanical medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K202242
    Device Name
    HS Fiber
    Date Cleared
    2020-09-10

    (31 days)

    Product Code
    Regulation Number
    878.5000
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Riverpoint Medical, LLC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HS Fiber sutures are indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular surgery, and the use of allograft tissues for orthopedic procedures.

    Device Description

    The Riverpoint Medical HS Fiber® sutures are non-absorbable, sterile, surgical sutures composed of multiple single strands of ultra-high molecular weight polyethylene (UHMWPE) braided together to form the implant. HS Fiber sutures are available in common sizes and lengths with or without pre-attached needles.

    AI/ML Overview

    This submission is for a medical device (HS Fiber surgical suture) and not an AI/ML powered device, therefore most of the requested information cannot be extracted from the text. The document describes the device, its intended use, and its substantial equivalence to a previously cleared predicate device.

    However, based on the provided text, here is the information related to the acceptance criteria and performance data for the device:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Standard)Reported Device Performance
    United States Pharmacopeia (USP) requirementsMeets USP performance requirements
    USP performance requirements for needle attachmentTested per USP performance requirements for needle attachment
    USP performance requirements for tensile strengthTested per USP performance requirements for tensile strength
    Endotoxin quantities below recommended limits outlined in FDA Guidance "Pyrogens and Endotoxins Testing: Questions and Answers"LAL endotoxin quantification assessments demonstrate endotoxin quantities below recommended limits
    Biological evaluation per ISO 10993-1:2018 - Biological Evaluation of Medical DevicesMaterials used were evaluated per ISO 10993-1:2018

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify the exact sample sizes used for the testing. It generally states that the device "meets requirements" and "is tested," but doesn't provide specific numbers of units tested. The provenance of the data is not mentioned (e.g., country of origin, retrospective/prospective).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable as the submission is for a physical medical device (suture) and not an AI/ML powered device that would require expert-established ground truth for a test set.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable as the submission is for a physical medical device (suture). Testing involved adherence to established physical and biological standards, not image or data interpretation requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable as the submission is for a physical medical device (suture) and not an AI/ML powered device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable as the submission is for a physical medical device (suture).

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the device's performance is based on established, objective, and quantitative performance standards set by:

    • United States Pharmacopeia (USP) standards for physical properties like tensile strength and needle attachment.
    • ISO 10993-1:2018 (Biological Evaluation of Medical Devices) for biocompatibility.
    • FDA Guidance "Pyrogens and Endotoxins Testing: Questions and Answers" for endotoxin levels.

    8. The sample size for the training set

    This is not applicable as the submission is for a physical medical device (suture) and not an AI/ML powered device that requires a training set. The device is manufactured according to specifications and tested for compliance.

    9. How the ground truth for the training set was established

    This is not applicable as the submission is for a physical medical device (suture) and not an AI/ML powered device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1