Search Results
Found 1 results
510(k) Data Aggregation
(43 days)
Resuscitation International, LLC
To perform Cardiopulmonary Resuscitation (CPR) on adult patients and only adult patients in cases of clinical death as defined by a lack of spontaneous breathing and pulse.
The RMCC is an automated, portable chest compressor, which provides continuous chest compressions as an adjunct to performing manual CPR. It is powered by a battery powered Control Unit. The RMCC provides consistent CPR support for cardiac arrest patients under conditions which might otherwise hinder the effectiveness of manual techniques.
-
- Compressor Assembly: contains a motor that drives the compressor pad to depress a patient's chest.
-
- Control Unit: contains the battery, power ON/OFF switch, Function Control Panel and umbilical power cord connecting it to the compressor assembly. Note: The battery charger is a separate unit.
-
- Torso Restraint: is placed underneath and around the back of the patient to firmly secure the compressor assembly to the patient.
-
- Stabilizer: serves a dual purpose; providing a head support and also attaches to the torso restraint to provide stability during continuous operation of the RMCC especially during patient transport.
The provided text describes the ROSC-U Mini Chest Compressor (RMCC), a mechanical cardiopulmonary resuscitator, and makes claims of substantial equivalence to predicate devices, but it does not contain the detailed acceptance criteria for a device's performance, nor does it describe a study that explicitly proves the device meets specific performance acceptance criteria in the manner requested.
The document summarizes the device's technological characteristics and lists non-clinical performance tests. These tests are about fundamental safety and electromagnetic compatibility standards, not specific performance metrics like compression depth accuracy, rate stability, or physiological outcomes, which would typically be included in acceptance criteria for a device like this.
Therefore, I cannot fulfill your request for:
- A table of acceptance criteria and the reported device performance.
- Sample size and data provenance for a test set.
- Number and qualifications of experts for ground truth.
- Adjudication method for the test set.
- MRMC comparative effectiveness study results.
- Standalone performance results.
- Type of ground truth used.
- Sample size for the training set.
- How ground truth for the training set was established.
The document only mentions the following non-clinical performance tests:
Functional verification | PASS |
---|---|
IEC 60601-1 (2012) | PASS |
IEC 60601-1-2 (2007) | PASS |
It concludes that "After performing non-clinical performance studies, the data shows that the RMCC is substantially equivalent to the predicates as an external cardiac compressor." This suggests that the studies primarily aimed to show equivalence based on safety and general functional standards rather than defined performance acceptance criteria related to its CPR efficacy.
Ask a specific question about this device
Page 1 of 1