Search Results
Found 4 results
510(k) Data Aggregation
(183 days)
Respire Medical Holdings
Our devices are indicated to treat mild to moderate Obstructive Sleep Apnea (OSA) in patients, 18 years of age or older.
The Respire Pink AT (Hard, Hard/Soft, EF) is an oral appliance used in the treatment of mild to moderate obstructive sleep apnea. This device helps move a patient's jaw forward, thus opening their airways, and allowing them to breathe more easily throughout the night. The Respire Pink AT (Hard, Hard/Soft, EF) is made out of dental industry standard acrylic resins & metals, and is custom designed to comfortably fit each patients' unique oral anatomy.
The provided FDA 510(k) summary for the “Respire Pink AT (Hard, Hard/Soft, EF)” device focuses on demonstrating substantial equivalence to a predicate device, rather than providing detailed acceptance criteria and a study report for the device's performance against specific clinical or technical metrics for disease treatment. This summary primarily relies on bench testing and a comparison to its predicate device to show safety and effectiveness.
Therefore, many of the requested details, such as specific acceptance criteria for diagnostic performance, sample sizes for test/training sets, expert qualifications for ground truth, adjudication methods, or results from MRMC or standalone studies, are not available in the provided text. The device is an oral appliance for treating Obstructive Sleep Apnea (OSA), not a diagnostic tool where "acceptance criteria and reported device performance" as typically understood in AI/imaging devices would apply.
Here's a breakdown of the available information:
1. Table of Acceptance Criteria and Reported Device Performance:
The document does not explicitly state quantitative acceptance criteria for device performance in treating OSA (e.g., AHI reduction, oxygen saturation). Instead, it relies on demonstrating substantial equivalence to a predicate device and successful completion of non-clinical tests.
Topic | Acceptance Criteria (Implied) | Reported Device Performance |
---|---|---|
Mechanical Safety & Performance | Not explicitly stated as quantitative metric; assumed to meet industry standards for durability, functionality, etc. | "Performs as well as the predicate device" |
Biocompatibility | Not explicitly stated; assumed to meet ISO 10993 standards. | "Passed successfully" |
Shipping Validation | Not explicitly stated; assumed to ensure device integrity during transport. | "Passed successfully" |
Mandibular Advancement Range (Key change from predicate device) | Predicate: 6mm | |
Submitted Device: 7mm (allows for slightly more titration, consistent with other cleared devices) | Achieved 7mm maximum mandibular advancement range. | |
Clinical Efficacy (Indirect) | Assumed to be equivalent to the predicate device in treating mild to moderate OSA. | No direct clinical efficacy data provided for the submitted device; equivalence is based on similar design and mechanism of action to predicate. |
2. Sample size used for the test set and the data provenance:
- Sample size for test set: Not applicable/not provided. The "testing" mentioned is non-clinical bench testing.
- Data provenance: Not applicable. The testing is laboratory-based, not patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable/not provided. Ground truth in the context of diagnostic performance (e.g., expert consensus on images) is not relevant to this type of device and testing. The "ground truth" for the non-clinical tests would be the established engineering and safety standards.
4. Adjudication method for the test set:
- Not applicable/not provided. Adjudication typically applies to multi-reader studies for diagnostic devices.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This is not an AI-assisted diagnostic device, and no MRMC study was performed.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is a physical medical device, not an algorithm.
7. The type of ground truth used:
- For the non-clinical tests (mechanical, biocompatibility, shipping validation), the "ground truth" implicitly refers to established engineering standards, material safety specifications (e.g., ISO 10993), and regulatory requirements for medical devices. The device was assessed against these benchmarks.
8. The sample size for the training set:
- Not applicable. This is not an AI device that uses a training set of data.
9. How the ground truth for the training set was established:
- Not applicable.
Ask a specific question about this device
(80 days)
Respire Medical Holding
The Respire Pink Series intraoral appliances are intended to treat snoring and mild to moderate Obstructive Sleep Apnea (OSA) in adult patients 18 years of age or older.
Optionally, the DentiTrac® micro-recorder may be incorporated into a Respire Pink Series device. The micro-recorder is intended to measure patient compliance to oral appliance therapy in combination with the DentiTrac® system.
The Respire Pink Series with DentiTrac® refers to two devices: Respire Pink Series-Herbst and the Respire Pink Series-Herbst-EF, herein referred to collectively as Respire Pink Series. These devices function as mandibular advancement splints which hold the jaw in a forward position. The forward jaw placement moves the tongue and pharyngeal tissue into a position to help maintain an open airway, which allows the passage of more air per breath and helps in the treatment of snoring and mild to moderate Obstructive Sleep Apnea (OSA). Respire Pink Series appliances are customizable devices which are produced according to the individual patient anatomy and prescription.
The Respire Pink Series-Herbst consists of full hard acrylic upper and lower fitting surfaces, and the Respire Pink Series-Herbst-EF consists of hard acrylic fitting side plates and chrome (BEGO Wironit®) upper palatal and lower lingual plates. The Herbst hardware on the side of the device allows for forward and lateral jaw movement while restricting backward jaw movement, and allows the patient to open and close the mouth. The Herbst hardware contains an expansion screw for mandibular position adjustment, and is attached to the appliance using medical grade stainless steel fixing elements and screws.
This submission is to provide clinicians the option to incorporate a Braebon DentiTrac® micro-recorder into a Respire Pink Series device. The micro-recorder is encapsulated in Loctite epoxy and fully encased into the device acrylic, and is not patient contacting. The micro-recorder is utilized to track patient compliance to prescribed oral appliance therapy by collecting data for wear-time through oral temperature, movement tracking, and head position. The data can be uploaded to a cloud-based web application using the DeniTrac® system which allows clinicians to review the patient data. The inclusion of the microrecorder provides additional treatment information to clinicians without effecting the operating principles of the Respire Pink Series devices.
The provided document is a 510(k) summary for the Respire Pink Series with DentiTrac®, a medical device intended to treat snoring and mild to moderate Obstructive Sleep Apnea (OSA). The document describes the device, its intended use, comparison to predicate devices, and performance testing. However, it does not contain detailed acceptance criteria and a study proving the device meets those criteria in the typical format of a diagnostic performance study.
Instead, the document details a substantial equivalence determination, which is a different type of regulatory submission. In this context, "acceptance criteria" are related to demonstrating that the new device is as safe and effective as a legally marketed predicate device, rather than meeting specific quantifiable performance metrics against a defined ground truth in a clinical study.
Here's an analysis based on the information provided, highlighting what is and is not present:
Key Findings from the Document:
- Device Type: Intraoral appliance for snoring and OSA, with an optional micro-recorder for compliance tracking.
- Regulatory Pathway: 510(k) Premarket Notification, seeking substantial equivalence to predicate devices.
- Predicates:
- Key Difference of the New Device: The addition of the DentiTrac® micro-recorder to the Respire Pink Series (which previously existed without it).
- Performance Testing Focus: Non-clinical. The document explicitly states: "The addition of the DentiTrac® micro-recorder does not adversely affect the original design, therefore, not introducing any new issues of concern. A risk assessment shows that risks are mitigated to acceptable levels, and performance testing related to the DentiTrac® micro-recorder confirms that EMC and electrical safety, biocompatibility, and software elements were evaluated."
Addressing Your Specific Questions:
-
A table of acceptance criteria and the reported device performance:
No explicit table of quantitative acceptance criteria for clinical performance (e.g., sensitivity, specificity for OSA efficacy) is provided. The "performance" described is largely non-clinical, focusing on safety, electrical compatibility, biocompatibility, and software aspects, evaluated against standards or risk assessments. The goal of a 510(k) is to show substantial equivalence, not to re-prove clinical efficacy against new, explicit performance targets if the predicate device's efficacy is already established.
The "performance" aspect for the DentiTrac® micro-recorder relates to its ability to accurately measure compliance. The document states: "DentiTrac® performance testing is located in the Braebon DentiTrac® Master File MAF 2557. Process validation concerning the incorporation and functionality of the DentiTrac® micro-recorder is certified by Braebon." This implies performance testing for the micro-recorder itself occurred, but the details (acceptance criteria, results) are not in this document.
Acceptance Criterion (Implicit/General for 510(k)) Reported Device Performance (from document) Safety & Efficacy (Overall) Device is substantially equivalent to predicate devices (SomnoDent® with Micro-Recorder and Respire Pink Series-Herbst/Herbst-EF). The addition of the DentiTrac® micro-recorder does not adversely affect the original design or introduce new issues of concern. The new device utilizes the same materials, manufacturing procedures, principles of operation, and overall design as existing Respire Pink Series devices, with the DentiTrac® micro-recorder being identical to that used in the SomnoDent® predicate. Risk Mitigation A risk assessment showed that risks are mitigated to acceptable levels. EMC and Electrical Safety Evaluated (related to DentiTrac® micro-recorder). Specific data not in this document, but implied to meet standards. Biocompatibility Evaluated (related to DentiTrac® micro-recorder). Specific data not in this document, but implied to meet standards based on substantial equivalence claim and previous assessments of the Respire Pink Series. Software Elements Evaluated (related to DentiTrac® micro-recorder). Specific data not in this document. Micro-Recorder Functionality "Process validation concerning the incorporation and functionality of the DentiTrac® micro-recorder is certified by Braebon." Intended to measure patient compliance to oral appliance therapy by collecting data for wear-time through oral temperature, movement tracking, and head position. (Details of validation in Braebon DentiTrac® Master File MAF 2557, not in this document). Respire Medical performs 100% visual and functional inspections. Mechanical Integrity/Manufacturing Respire Medical performs 100% visual and functional inspections throughout the production process to ensure that the device meets manufacturing specifications for the Respire Pink Series device and specifications for the incorporation of the DentiTrac® micro-recorder. The device allows for forward and lateral jaw movement while restricting backward jaw movement, and allows the patient to open and close the mouth. The Herbst hardware contains an expansion screw for mandibular position adjustment. (These are design features, not "performance" in the sense of a clinical outcome metric). -
Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
Not applicable/Not provided in this document. This document focuses on substantial equivalence based on technical and non-clinical testing, particularly regarding the addition of the DentiTrac® micro-recorder. It does not describe a clinical performance study with a distinct "test set" of patients or data in the way a diagnostic AI device would. The performance testing for the DentiTrac® micro-recorder itself (which is embedded) is referenced to a Master File (MAF 2557), so details about its own test set would be there, but are not in this 510(k) summary.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
Not applicable. As there is no clinical "test set" in this document for evaluating the device's diagnostic or clinical efficacy, there's no mention of experts establishing a ground truth.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
Not applicable. No clinical "test set" or adjudication method is described.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This device is an intraoral appliance, not an AI-powered diagnostic tool that assists human readers. Therefore, an MRMC study is irrelevant to this submission.
-
If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
Not applicable. This is a physical medical device (intraoral appliance), not an algorithm. The DentiTrac® micro-recorder tracks compliance data, which is then uploaded to a web application for clinicians to review. It is a data collection tool, not a standalone diagnostic algorithm.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc):
Not applicable for clinical efficacy in this document. The "ground truth" for the non-clinical testing would be engineering specifications, material standards, electrical safety standards, and risk assessment thresholds. For instance, the "ground truth" for biocompatibility is successful adherence to ISO 10993 standards. For micro-recorder functionality, it would be the accuracy and reliability of its compliance tracking features (details in MAF 2557).
-
The sample size for the training set:
Not applicable. This is a physical medical device, not a machine learning algorithm that requires a "training set."
-
How the ground truth for the training set was established:
Not applicable. As above, no training set for an algorithm is involved.
Conclusion based on the provided text:
The document describes a 510(k) submission for a physical medical device (intraoral appliance) that has incorporated a compliance tracking micro-recorder (DentiTrac®). The "acceptance criteria" and "studies" are framed within the context of demonstrating substantial equivalence to existing predicate devices, focusing on non-clinical aspects like materials, manufacturing, design, principles of operation, electrical safety, biocompatibility, and software evaluation of the embedded micro-recorder. This is distinct from a clinical performance study with quantitative diagnostic metrics and explicit acceptance criteria against a defined ground truth, which is typically seen for new diagnostic or AI-driven medical devices. The document references a Master File (MAF 2557) for detailed performance testing of the DentiTrac® micro-recorder, suggesting that more specific performance data for that component exists elsewhere.
Ask a specific question about this device
(116 days)
Respire Medical Holding
The Respire Blue Scries - EF is indicated to treat mild to moderate OSA (Obstructive Sleep Apnea)
The Respire Blue Series – EF (Endurance Frameworks) – is available with a hard device fitting surface. The hard surface consists of Acrylic (side plates / trays) and chrome - Wironit material (upper / Palatal and lower / Lingual plates / trays). Refer to Figure 1 Representative Drawing. The device is retained with ball and clasps which are used to hold elastic bands which guide the jaw to the desired position. In addition there are expansion screws inside the Acrylic side plates / trays that allows for device expansion based on patient needs.
The upper and lower components are connected by an adjustable hinge, thus the patient can open and close the mouth while wearing the appliances
The provided text describes the Respire Blue Series - EF intraoral device, indicated for treating mild to moderate Obstructive Sleep Apnea (OSA). The 510(k) submission (K152292) is for a modification to a previously cleared device (K111207), specifically the material used for the top and bottom plates/trays.
Based on the provided text, the acceptance criteria and study information can be summarized as follows:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria Category | Specific Criteria/Test | Reported Device Performance |
---|---|---|
Biocompatibility | ISO 10993 Standards | "The results of testing determined that the material didn't cause skin irritation or allergenic sensitization." |
Cytotoxicity | MEM Elution testing | "The testing passed with a reactivity of 0 (zero) or none." |
Material Integrity (Strength) | Specimen elongation and load-bearing force to break | "The amount of energy shown in the results, still far exceeds the amount of energy required by biting to create such a failure which occurred at an average of 275.6 Ibf using five (5) sample devices. Thus, the proposed Wironit/Acrylic parts are sufficient to survive a load exposure exerted by a human oral cavity." |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: For the Material Integrity testing, five (5) sample devices were used.
- Data Provenance: The document does not specify the country of origin or whether the data was retrospective or prospective. It is assumed to be part of the manufacturer's internal testing as described in the 510(k) submission.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- The document does not mention the use of experts to establish ground truth for the non-clinical tests described. The tests appear to be laboratory-based material science tests.
4. Adjudication Method for the Test Set
- This information is not applicable as the described tests are laboratory-based material science tests, not clinical studies involving interpretation by multiple experts.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No MRMC comparative effectiveness study was done. This device is a physical intraoral device, not an AI-assisted diagnostic tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- No standalone algorithm performance study was done. This device is a physical intraoral device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- For biocompatibility and cytotoxicity, the "ground truth" is defined by the ISO 10993 standards and the specific reactivity thresholds of the MEM Elution test.
- For material integrity, the "ground truth" is established by the mechanical properties (elongation, load-bearing force) of the material compared to a benchmark (the predicate device and the force exerted by a human bite).
8. The sample size for the training set
- A "training set" is not applicable for the type of non-clinical tests described for this physical device. These are material characterization and performance validation tests, not machine learning model training.
9. How the ground truth for the training set was established
- This information is not applicable as there is no training set for a machine learning model.
Ask a specific question about this device
(174 days)
Respire Medical Holding
The Respire Pink Series - Herbst - EF is indicated to treat mild to moderate OSA.
The Respire Pink Series - Herbst - EF (Endurance Frameworks) - is available with a hard device fitting surface. The hard surface consists of Acrylic (side plates) and chrome - Wironit material ( upper / Palatal and lower / Lingual plates). Refer to Figure 1 Representative Drawing. The device is retained with ball and clasps which allows the device to be tightened if it becomes loose. The device is a mandibular advancement splint that holds the jaw in a forward position to help keep the tongue and supporting tissues in a position to help maintain an open airway, which helps in the treatment of snoring and mild to moderate obstructive sleep apnea. The Herbst hardware on the side of the device allows the patient to move forward and left and right, but not backwards. These movements give the patient some freedom to move which is important for their comfort and overall success of the device. The upper and lower components are connected by an adjustable hinge, thus patient can open and close while wearing the appliances.
{
"acceptance_criteria_and_study": {
"acceptance_criteria_table": [
{
"criterion": "Biocompatibility and cytotoxicity",
"reported_performance": "Demonstrated via biocompatibility and cytotoxicity testing for Wironit material."
},
{
"criterion": "Material integrity (specimen elongation and load bearing force to break)",
"reported_performance": "Points of failure were identified at the same location for both Wironit/Acrylic and Acrylic-only (predicate) devices. The amount of energy to cause failure was less for Wironit/Acrylic but still far exceeded the amount of energy required by an oral cavity to create such a failure (average of 275.6 lbf)."
}
],
"sample_size_test_set": "Five (5) sample devices for material integrity testing.",
"data_provenance": "Not specified (implied in-house testing by the submitter or a contracted lab).",
"number_of_experts_ground_truth": "Not applicable, as the evaluation involves material testing rather than expert-derived ground truth for clinical outcomes.",
"qualifications_experts": "Not applicable.",
"adjudication_method": "Not applicable.",
"mrmc_comparative_effectiveness_study": false,
"standalone_performance_study": true,
"type_of_ground_truth": "Laboratory testing data (biocompatibility, cytotoxicity, material integrity).",
"sample_size_training_set": "Not applicable, as this is a device modification evaluation with laboratory testing, not an AI/algorithm study requiring a training set.",
"ground_truth_training_set_establishment": "Not applicable."
}
}
Ask a specific question about this device
Page 1 of 1