Search Results
Found 1 results
510(k) Data Aggregation
K Number
K192747Device Name
Yxoss CBR®
Manufacturer
Date Cleared
2020-06-17
(261 days)
Product Code
Regulation Number
872.4760Why did this record match?
Applicant Name (Manufacturer) :
ReOss GmbH
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
Yxoss CBR® is used for reconstruction of alveolar bone deficits prior to placing a dental implant, shaping alveolar bone and as a support to the augmented bone volume in the regeneration of bone defects that may include:
- · extraction sites
- · horizontal and/or vertical augmentation of the alveolar ridge
- · reconstruction of bone defects in the maxillofacial area
Device Description
Yxoss CBR is a titanium scaffold, which will be surgically inserted over bone defects as a volume support for contouring newly formed bone over a bone defect. The titanium scaffold is held in place to the existing bone with standard titanium screw(s). The following screws are compatible with the Yxoss CBR device system (informational purposes only, not devices subject to this sub-mission, K192747): Synthes 1.3mm self-drilling screws, K983485, and Salvin Dental 1.5mm Tent-ing Screws, K161857. Yxoss CBR thus stabilize the bone graft in the defect area and specifies the shape of the bone to be augmented.
The Yxoss CBR may be used in a one-stage approach with simultaneous placement of a dental implant. The Standard Yxoss scaffold (without integrated implant positioning) is used to cover over the simultaneously placed bone level dental implant with cover screw. The Standard Yxoss scaffold is explanted after healing occurs, leaving in place the dental implant.
In a two-stage approach, the dental implant will be placed after revascularization of the augmented bone. In a two-stage approach, there are two options available:
- . Standard Yxoss scaffold (without integrated implant positioning)
- . Backward Yxoss scaffold (with integrated implant positioning)
For the two-stage approach, the Standard Yxoss scaffold will be removed before placing the dental implant. For the Backward Yxoss scaffold using the two-stage approach, the pre-holes are used as guide for drilling the seat before inserting the dental implant chosen by the surgeon. Then, the Backward Yxoss scaffold will be explanted.
The device is manufactured by SLS (Selective Laser Sintering) additive manufacturing.