Found 2 results

510(k) Data Aggregation

    K Number
    K132066
    Device Name
    INFLAMMA DRY
    Manufacturer
    RAPID PATHOGEN SCREENING, INC.
    Date Cleared
    2013-11-21

    (141 days)

    Product Code
    PFQ, JJX
    Regulation Number
    862.1540
    Why did this record match?
    Applicant Name (Manufacturer) :

    RAPID PATHOGEN SCREENING, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
    Intended Use
    InflammaDry is a rapid, immunoassay test for the visual, qualitative in vitro detection of elevated levels of the MMP-9 protein in human tears from patients suspected of having dry eye to aid in the diagnosis of dry eye in conjunction with other methods of clinical evaluation. This test is intended for prescription use at point-of-care sites. InflammaDry External Controls are QC materials used for verifying the performance of the InflammaDry test reagents and assay. These controls can also be used to assist in operator training and troubleshoot invalid results.
    Device Description
    InflammaDry™ consists of three (3) parts: a sterile Sample Collector, an immunoassay test strip in a plastic Test Cassette housing, and Buffer in a vial. The Sample Collector is used to take a sample of human tears. The separately packaged and sterile Sample Collector has a contoured end with a Dacron fleece to collect the tear sample from the inside of the lower eyelid. The plastic housing of the Test Cassette body protects the strip from unintended physical influence. Additionally, the housing guarantees correct sample transfer onto the lateral flow assay strip. The Buffer vial contains a buffered salt solution containing proteins, detergents and preservatives. The Buffer functions as the solution that initiates the test, carries antigen through a microfiltration process to remove unwanted cellular debris, and transports the immune complex and the control conjugate to the Test and Control Lines on the test strip membrane. InflammaDry External Controls are to be used with the InflammaDry test only and are intended to verify that the test reagents are working and that the test is performed correctly. Both Negative and Positive external controls for InflammaDrv™ are supplied as lyophilized powder in small glass vials with screw caps. 200 ul of recombinant MMP-9 in Stabilizing buffer solution is quickly frozen and lyophilized under vacuum. A soft and pliable plastic dropper bottle filled with a Dl water diluent is provided with each set of external controls. The InflammaDry external controls are sold as a separate catalog item. The InflammaDry™ test is based on the principle of lateral flow immunoassays using Direct Sampling Micro-Filtration technology. Matrix metalloproteinase-9 (MMP-9) present in the tear fluid is captured between two (2) highly specific anti-MMP-9 antibodies: a monoclonal mouse anti-MMP-9 antibody and a polyclonal goat antihuman antibody. This antigen-antibody complex is captured at an immobilized Test Line. The formation of a blue color line at the control zone line with a red color line at the test zone line is considered as a positive result, a blue color line at the control zone only is considered as a negative result, if a blue color line in the control zone does not appear the test is considered invalid. The test is a disposable, rapid test requiring 10-15 minutes for a result.
    Ask a Question
    K Number
    K110722
    Device Name
    RPS ADENO DETECTOR PLUS
    Manufacturer
    RAPID PATHOGEN SCREENING, INC.
    Date Cleared
    2011-05-17

    (63 days)

    Product Code
    GOD, REQ
    Regulation Number
    866.3020
    Why did this record match?
    Applicant Name (Manufacturer) :

    RAPID PATHOGEN SCREENING, INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
    Intended Use
    The RPS Adeno Detector Plus is a rapid immunoassay test for the visual, qualitative in vitro detection of Adenoviral antigens (hexon protein) directly from human eye fluid. The test is intended for professional use as an aid in the rapid differential diagnosis of acute conjunctivitis. Negative results do not preclude Adenovirus infection nor are thev intended to rule out other microbial-caused infections of the conjunctiva. and should not be used as the sole basis for treatment or other management decisions.
    Device Description
    The RPS Adeno Detector Plus™ consists of three (3) parts: a Sample Collector, an immunoassay test strip in a plastic Test Cassette housing, and a Buffer. The Sample Collector is used to take a sample of ocular fluid. The separately packaged and sterile Sample Collector has a contoured end with a Dacron fleece to collect the samples. The plastic housing of the Test Cassette body protects the strip from unintended physical influence. Additionally the housing guarantees correct sample transfer onto the lateral flow assay strip. The Buffer is a buffered salt solution containing proteins, detergents and preservatives. The Buffer functions as the solution that initiates the test, extracts the Adenoviral proteins, filters unwanted cellular debris, and transports the immune complex and the control conjugate to the Test and Control Lines on the test strip membrane. Mechanism of action - RPS Adeno Detector Plus™ is based on the principle of lateral flow immunoassays using Direct Sampling Micro-filtration technology. Viral particles or virus antigens are captured by an antigen specific antibody. A single monoclonal antibody highly specific to the Adenoviral hexon protein is labeled with colloidal gold and also is immobilized as the Test Line.