510(k) Data Aggregation
(399 days)
The DELL U3014 with QUBYX PerfectLum is intended to be used for displaying and viewing of digital images, for review and analysis by trained medical practitioners. The DELL U3014 must only be used in conjunction with QUBYX PerfectLum. The device must not be used in primary image diagnosis in mammography. The device can not be used for a life-support system and does not contact with the patient.
The DELL U3014 with QUBYX PerfectLum is a 30" color display for medical viewing. It provides 2560x1600 resolution with an adjustable Look Up Table and a 10 bit panel. It is combined with QUBYX PerfectLum and PerfectLum remote management, a userfriendly DICOM calibration and AAPM TG18 verification software suite. The software allows setting the display function to DICOM, displaying test pattern and performing acceptance and constancy tests.
The acceptance criteria and study proving the device meets those criteria are detailed below. It's important to note that this document describes a medical display system, not an AI algorithm for diagnostic interpretation. Therefore, some standard questions related to AI studies (like MRMC, training set details, or deep learning specific ground truth) are not applicable in this context.
1. Table of Acceptance Criteria and Reported Device Performance
| Criterion | Acceptance Standard | Reported Device Performance |
|---|---|---|
| DICOM Conformance | DICOM Part 14 GSDF standard | Successfully passed DICOM conformance test; compliant with DICOM Part 14 GSDF standard. |
| AAPM TG18 Conformance | AAPM TG18 standard acceptance test requirements | Successfully passed AAPM TG18 acceptance test; compliant with AAPM TG18 standard. |
| Intended Use | Displaying and viewing medical images for review and analysis by trained medical practitioners. Not for primary image diagnosis in mammography. Not a life-support system. No patient contact. | Meets all stated indications for use. |
| Technical Characteristics (key) | Similar to predicate devices (DELL U3011, NEC MD301C4) | Substantially equivalent to predicate devices, with minor differences (e.g., backlight type, slight luminance variation). |
| Software Bundle | Must be used in conjunction with QUBYX PerfectLum | Bundled with QUBYX PerfectLum for calibration and verification. |
2. Sample Size Used for the Test Set and Data Provenance
The "test set" in this context refers to the physical device undergoing conformance testing, not a dataset of medical images.
- Sample Size for Test Set: One unit of the DELL U3014 with QUBYX PerfectLum bundle.
- Data Provenance: The device performance data was generated through tests conducted by QUBYX and verified by the University of Arizona. This is a prospective evaluation of a specific hardware and software combination.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
This question is not directly applicable. The "ground truth" for a medical display system's performance is objective compliance with technical standards (DICOM Part 14 GSDF and AAPM TG18). These standards define measurable characteristics and visual patterns for evaluation.
- The tests were performed using instrumentation (X-Rite i1 Display Pro measurement device) and software (QUBYX PerfectLum) according to the specified standards.
- One "user" was involved in the visual steps of the AAPM TG18 test. While not explicitly stated to be an "expert" in the same way a radiologist diagnoses an image, this user's role was to assess test patterns as per the AAPM standard, implying knowledge of what to look for based on documented criteria.
- The overall compliance was verified by the University of Arizona, implying expert oversight in the validation process, though specific qualifications of individuals involved are not provided in this summary.
4. Adjudication Method for the Test Set
Not directly applicable in the typical sense of agreement on diagnostic findings. The adjudication of the device's technical performance was based on:
- Objective Measurement: The X-Rite i1 Display Pro measurement device performed objective measurements of display characteristics.
- Software Analysis: The QUBYX PerfectLum software analyzed these measurements against target values defined by DICOM and AAPM standards.
- User Input (for visual tests): For visual steps of the AAPM TG18 test, a user's answers were recorded and analyzed by the software against the standard's criteria.
- Verification: The results of these tests were verified by the University of Arizona. There is no mention of a multi-reader adjudication method for discrepancies, as the tests themselves are designed to be objective or based on clearly defined visual criteria.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, an MRMC comparative effectiveness study was not performed. This type of study is relevant for evaluating the impact of an AI algorithm on human reader performance for diagnostic tasks using medical images. This submission is for a medical display system, where regulatory clearance is based on its technical compliance with established standards for image presentation, not its impact on diagnostic accuracy through reader studies.
6. If a Standalone (i.e. Algorithm Only Without Human-in-the-Loop Performance) Was Done
This question is not applicable. The device is a display system, not a diagnostic algorithm. Its "standalone performance" refers to its ability to meet display standards, which was evaluated through technical tests (DICOM and AAPM compliance) as described.
7. The Type of Ground Truth Used
The ground truth used for evaluating the display system's performance was:
- Established Industry Standards: DICOM Part 14 Grayscale Standard Display Function (GSDF) and AAPM TG18 display quality control guidelines.
- These standards define objective requirements for luminance response, uniformity, resolution, and other visual characteristics essential for medical image display.
8. The Sample Size for the Training Set
This question is not applicable. The device is a medical display system, not an AI algorithm that requires a training set of data.
9. How the Ground Truth for the Training Set Was Established
This question is not applicable for the reasons stated above.
Ask a specific question about this device
(258 days)
The DELL UltraSharp U3011 with QUBYX PerfectLum is intended to be used in displaying and viewing of digital images, for review and analysis by trained medical practitioners.
The DELL UltraSharp U3011 must only be used in conjunction with QUBYX PerfectLum 3.0
These devices must not be used in primary image diagnosis in mammography.
The device can not be used for a life-support system.
The DELL UltraSharp U3011 with QUBYX PerfectLum is a 30" color display for medical viewing. It provides 4 mega pixel (2560x1600) resolution with a adjustable Look Up Table and a 10 bit Panel.
It is combined with QUBYX PerfectLum 3.0 and PerfectLum remote QA, a user-friendly DICOM calibration and AAPM TG18 verification software suite. The software allows to set the display function to DICOM, display testpattern and perform acceptance and constancy tests.
The provided text is a 510(k) summary for the DELL UltraSharp U3011 with QUBYX PerfectLum bundle. This document focuses on demonstrating substantial equivalence to a predicate device rather than presenting a performance study with specific acceptance criteria and detailed results.
Therefore, the requested information regarding acceptance criteria, study details, sample sizes, ground truth establishment, expert qualifications, adjudication methods, and MRMC effectiveness studies cannot be extracted directly from this document. The document describes a comparison, not an independent performance study against set criteria as would typically be seen for a new diagnostic algorithm or device requiring such validation.
The primary method used to "prove" the device meets acceptance criteria in this context is through substantial equivalence to an existing legally marketed device (the predicate device). The acceptance criteria, in this case, are implicitly that the new device performs at least as well as, or is sufficiently similar to, the predicate device in its intended use and technical specifications.
Here's a breakdown of what can be extracted and explanations for what cannot:
1. Table of acceptance criteria and the reported device performance:
As explained above, explicit acceptance criteria in the sense of performance metrics (sensitivity, specificity, accuracy, etc.) are not provided. The "performance" is a comparison to a predicate device.
| Characteristic | DELL U3011 with PerfectLum version 3 (New Device) | Predicate device NEC LCD3090WQXI (K083916) | Implicit Acceptance/Comparison |
|---|---|---|---|
| Panel Type | IPS | IPS | Equivalent |
| Panel size | 30" viewable | 29.8" viewable | Similar |
| Native Resolution | 2560 x 1600 | 2560 x 1600 | Equivalent |
| Pixel Pitch | 0.25 mm | 0.25 mm | Equivalent |
| Brightness (typical) | 370 cd/m2 | 350 cd/m2 | New device is slightly better |
| Contrast Ratio (typical) | 1000:1 | 1000:1 | Equivalent |
| Viewing Angle (typical) | 178° Vert., 178° Hor. | 178° Vert., 178° Hor. | Equivalent |
| Displayable Colors | 1.07 billion colors | 16.7 million | New device is significantly better |
| DICOM calibration software and AAPM verification software | bundled | optional | New device offers as standard, old as option |
| Panel bit depth | 10 bit | 8 bit | New device is better |
| Color Gamut | 117% | 102% | New device is better |
The "acceptance criteria" here are that the new device should not be worse than the predicate in critical areas and can be better. The conclusion states: "The new and predicate device are substantially equivalent in the areas of technical characteristics, general function, application and indented use."
2. Sample size used for the test set and the data provenance:
- Not applicable/Not provided. This document describes a technical comparison of display specifications and software features, not a diagnostic performance study using a test set of medical images.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable/Not provided. No clinical ground truth establishment is described.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable/Not provided.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable/Not provided. This is a medical display, not an AI-assisted diagnostic tool requiring MRMC studies.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Not applicable/Not provided. This is a medical display, not a standalone diagnostic algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- Not applicable/Not provided. The "ground truth" for this submission would be the published specifications and performance of the predicate device and the measured specifications of the new device.
8. The sample size for the training set:
- Not applicable/Not provided. This is a hardware/software bundle, not a machine learning model that requires a training set.
9. How the ground truth for the training set was established:
- Not applicable/Not provided.