Search Filters

Search Results

Found 10 results

510(k) Data Aggregation

    Why did this record match?
    Applicant Name (Manufacturer) :

    pro med instruments GmbH

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K240319
    Date Cleared
    2024-04-24

    (82 days)

    Product Code
    Regulation Number
    882.4460
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Pro Med Instruments GmbH

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DORO LUCENT® Skull clamp is part of a mechanical support system, which is used during head or neck surgery. The product can be used in an intra-operative setting for CT and X-ray imaging.

    The DORO LUCENT® Skull clamp Pediatric Set is part of a mechanical support system, which is used during head or neck surgery. The product can be used in an intra-operative setting for CT and X-ray imaging. The target patient population includes children over 5 years of age.

    Device Description

    The DORO LUCENT® Skull Clamp ensures an adequate positioning of a patient's head and provides rigid fixation for neurosurgery. The device is suitable for intra-operative X-ray and CT based imaging procedures.

    The DORO LUCENT® Skull Clamp consists of a fixed arm, an adjustable arm and modular pin holders. Thus, using the DORO LUCENT® Skull Clamp Pediatric Set, the skull clamp can be configured as three-pin or four-pin pediatric set up.

    AI/ML Overview

    The provided text describes the 510(k) clearance for the DORO LUCENT® Skull Clamp and DORO LUCENT® Skull Clamp Pediatric Set. However, it does not contain information about a study based on AI/ML or any performance testing related to software or AI. The tests mentioned are for mechanical performance, usability, and reprocessing of the physical device.

    Therefore, for aspects related to AI/ML device performance, such as sample size for test/training sets, data provenance, expert ground truth, adjudication methods, MRMC studies, standalone performance, and ground truth establishment for training, the information is not available in the provided document.

    Here's a breakdown of what is available based on the mechanical device performance testing:

    Device: DORO LUCENT® Skull Clamp (1101.001) and DORO LUCENT® Skull Clamp Pediatric Set (1101.040)

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance Criteria / Benchmark (Implicit)Reported Device Performance
    Maximum Load Test (System Test)The skull clamp must be capable of resisting the forces imposed by the patient and surgeon when in use and in locked position without movement in any connection, joint and/or plastic deformation.Pass. The skull clamp is capable of withstanding the given force without movement in any connection, joint and/or plastic deformation.
    UsabilityThe usability of the skull clamp must be given.Pass. The usability of the skull clamp is given.
    Static load (Latching teeth mechanism)The skull clamp must be able to withstand a maximum static load for a given duration without signs of permanent deformation, fracture, or breakage.Pass. The skull clamp withstands a static load for the given duration without signs of permanent deformation, fracture, or breakage.
    CreepThe skull clamp must maintain the applied maximum force without a force deviation from the initially applied load by a defined value.Pass. The skull clamp can maintain the applied maximum force without a force deviation from the initially applied load by a defined value.
    Pin force accuracyThe skull clamp force delivery component must be verified at each force level throughout its range to deliver the stated force within the actual setting.Pass. The skull clamp force delivery component is verified at each force level throughout its range to deliver the stated force within the actual setting.
    Torque load resistanceThe skull clamp must withstand a defined torque for a given amount of time without deformation or any structural failures.Pass. The skull clamp withstands a defined torque for a given amount of time without deformation or any structural failures.
    CT EvaluationThe skull clamp must produce fewer artifacts than the set benchmark for adequate imaging results.Pass. The skull clamp produces fewer artifacts than the set benchmark as defined in the test report.
    ReprocessingReprocessing must be validated according to specified standards (ANSI/AAMI ST98:2022, DIN EN ISO 17664-1:2021, DIN EN ISO 17664-2:2021, AAMI TIR 12: 2020) including accumulation of residue, cleaning effectiveness, and disinfection effectiveness.Pass. The test data shows the reprocessing of DORO LUCENT® Skull Clamp is validated and a test report in support of the statement is provided.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: Not explicitly stated for each test. For mechanical/physical performance tests, the "sample size" typically refers to the number of units tested. The document indicates "Tests were performed and the results are shown in the table below," implying a sufficient number of tests were conducted to confirm performance.
    • Data Provenance: Not specified in terms of country of origin or retrospective/prospective for these physical device tests. These are likely internal laboratory tests conducted by the manufacturer.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    • Not applicable. This device is a mechanical skull clamp, not an AI/ML diagnostic device requiring expert interpretation for ground truth. The "ground truth" for these tests is the physical measurement and observation of the device's performance against defined engineering and safety standards.

    4. Adjudication Method for the Test Set:

    • Not applicable. See point 3.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No. This type of study is relevant for AI/ML diagnostic tools being compared to human performance, which is not the case for this mechanical device.

    6. Standalone (Algorithm Only) Performance:

    • Not applicable. There is no algorithm or software for standalone performance in this mechanical device.

    7. Type of Ground Truth Used:

    • Engineering and Safety Standards/Specifications: The ground truth for these tests is established by the device's design specifications and compliance with relevant industry standards (e.g., for mechanical strength, CT compatibility, and reprocessing).

    8. Sample Size for the Training Set:

    • Not applicable. This is a mechanical device, not an AI/ML product developed through a training process with a data set.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable. See point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K203505
    Date Cleared
    2021-03-29

    (119 days)

    Product Code
    Regulation Number
    882.4460
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    pro med instruments GmbH

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DORO QR3 Headrest System is placed on the patient's skull to hold their head and neck securely in a particular position when rigid fixation is desired. The clamp is indicated for use in open and percutaneous craniotomies as well as spinal surgery when rigid skeletal fixation is necessary.

    Device Description

    DORO® QR3 Headrest System consists of 3 components: 1. Skull clamp, 2. Swivel Adapter, 3. Adjustable Base Unit. DORO® QR3 Headrest System (Aluminum) is a cranial stabilization device, designed to provide rigid skeletal fixation. The DORO Headrest System is placed on the patient's skull to hold their head and neck securely in a particular position when rigid fixation is desired. The clamp is indicated for use in open and percutaneous craniotomies as well as spinal surgery when rigid skeletal fixation is necessary. The Swivel Adaptor connects the Base Unit (by means of the Transitional Member) with the Skull Clamp. The Swivel Adaptor provides 360 degrees rotation. This allows a fully flexible adjustment of the DORO® Headrest System to the patient's position. The Base Unit is designed for patient positioning in prone or supine, lateral and sitting positions.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device called the DORO QR3 Headrest System (Aluminum). This document primarily focuses on establishing substantial equivalence to a predicate device and outlines performance testing for safety and effectiveness, rather than detailing clinical study results for an AI/algorithm-based device.

    Therefore, many of the requested categories related to AI/algorithm performance (e.g., sample size for test set, number of experts, adjudication methods, MRMC studies, standalone performance, training set details) are not applicable to this submission.

    Here's the information that can be extracted relevant to acceptance criteria and device performance based on the provided text:

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implied by the "Pass" results of the performance tests, indicating that the device met the established benchmarks for each test. The study performed is a series of engineering and usability tests.

    Acceptance Criteria (Test)Reported Device Performance (Result)
    System Test (in accordance with ASTM F3395 / F3395M-19)Pass
    Torque Load Resistance Test (in accordance with ASTM F3395 / F3395M-19)Pass
    Force Delivery Verification (in accordance with ASTM F3395 / F3395M-19)Pass
    Creep Test (in accordance with ASTM F3395 / F3395M-19)Pass
    Static Load Test (in accordance with ASTM F3395 / F3395M-19)Pass
    Fourfold Load - Skull Clamp (in accordance with DIN EN 60601-2-46)Pass
    Fourfold Load - Interface Skull Clamp to Swivel Adaptor (in accordance with DIN EN 60601-2-46)Pass
    Usability Test (in accordance with IEC 62366 as well as FDA-2011-D-0469)Pass
    Automated cleaning including disinfection (in accordance with EN ISO 15883-1:2014 – 10, RKI Guideline: 2012, AAMI TIR 30: 2011, Guideline DGKH, DGSV, AKI: 2014)Pass

    Additional Information:

    1. Sample size used for the test set and the data provenance: Not applicable. These were engineering performance tests of the physical device, not a test set of medical data.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for engineering tests is typically based on established physical standards and measurements, not expert human interpretation.
    3. Adjudication method for the test set: Not applicable.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This document is for a physical medical device, not an AI/algorithm.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    6. The type of ground truth used:
      • For engineering tests (Load, Torque, Creep, System Tests): The ground truth is the performance specified by the referenced ASTM and DIN EN standards. The "Pass" results indicate the device met these engineering specifications.
      • For Usability Test: The ground truth would be adherence to the usability requirements outlined in IEC 62366 and FDA-2011-D-0469.
      • For Cleaning and Disinfection Test: The ground truth is compliance with the referenced ISO, RKI, AAMI, and DGKH/DGSV/AKI guidelines for reprocessing medical devices.
    7. The sample size for the training set: Not applicable. This is for a physical device, not an AI/algorithm.
    8. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K193438
    Date Cleared
    2020-06-02

    (174 days)

    Product Code
    Regulation Number
    882.4460
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Pro Med Instruments GmbH

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Indicated for use in combination with a skull clamp in open and percutaneous craniotomies as well as spinal surgery when rigid fixation is necessary.

    Device Description

    DORO® Sterile Disposable Skull Pins are a disposable device. These are delivered sterile (gamma-sterilization), ready for use, and are intended to be disposed after one use. DORO® Sterile Disposable Skull Pins are available in 2 different types (Stainless steel & Titanium) and 2 different sizes (Adult & Pediatric). DORO® Sterile Disposable Skull Pins are used together with the DORO ® Headrest Systems, intended as a neck and head support to stabilize the patient's head during neurosurgical procedures. Two DORO® Sterile Disposable Skull Pins are inserted into the rocker arm side of the skull clamp and one single skull pin is inserted in the opposite side. DORO® Sterile Disposable Skull Pins (Titanium) are used when Intra-Operative X-Ray-, CTor MR Imaging of the patient is used.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for the DORO® Sterile Disposable Skull Pins. This document asserts the substantial equivalence of the new device to legally marketed predicate devices, rather than proving the device meets specific performance acceptance criteria through the kind of study described in the prompt.

    Therefore, I cannot provide a table of acceptance criteria and reported device performance directly from this document, nor can I describe an MRMC comparative effectiveness study or details about the training set.

    The document focuses on demonstrating substantial equivalence for a physical medical device (skull pins) by comparing its technological characteristics and performance testing results to those of predicate devices. It does not involve AI/ML components or studies of human reader performance.

    However, I can extract information related to the performance data provided for the physical device:

    1. Table of Acceptance Criteria and Reported Device Performance (as listed in the document for the physical device):

    TestAcceptance Criteria (Implied by action 'Confirms ability to withstand' or 'Confirms acceptance', and 'Pass' result)Reported Device Performance
    DORO® Sterile Disposable Skull Pins (Stainless Steel & Titanium)
    Axial Load, and Creep TestAbility to withstand typical loads with additional safety factor without plastic deformation and with an accepted distension.Pass: The Pins are able to withstand typical static loads w/ additional safety factor w/o plastic deformation and an accepted distension.
    Shear TestAbility of the skull pins and its interface with the cranial bone to withstand typical shear force with additional safety factor without spatial locomotion.Pass: The Pins are able to withstand typical shear force w/ additional safety factor w/o spatial locomotion.
    Transport ValidationCapability of the transport packaging as well as the product packaging to protect the skull pins against hazards, which may occur during handling, storage, and transport by air or ground.Pass: The utilized packing system for the skull pins is capable to protect those against visible damage.
    Packing Integrity TestCapability of the sterile barrier system of the blister pack as well as the blister pack itself utilized for the skull pins whether it is effective, efficient and safe to use.Pass: The utilized sterile barrier system as well as the blister pack was proven to perform efficiently, safely, and effectively.
    System Handling TestSafe and effective utilization of the skull pins with skull clamps without negative impact on utilized single-use gloves.Pass: Skull Pins performed safely and effectively.
    Input MaterialsAcceptance of utilized materials of the skull pins for medical use prior biocompatibility test.Pass: All utilized materials are accepted for medical use.
    Biocompatibility TestBiocompatibility of the skull pins by cytotoxicity test and chemical analysis of utilized materials (no cell growth inhibiting character, no chemical substance above harmful threshold).Pass: The skull pins obtain no cell growth inhibiting character. No chemical substance above harmful threshold was detected.
    Color CodingColoring of the plastic pin collets that enables the user to differentiate between MR conditional and MR unsafe skull pins as well as between skull pins for the adult population and pediatric population.Pass: Differentiation between designated patient populations as well as different applications is possible due to different colors utilized for the plastic pin collets.
    Sterility TestSterility of the skull pins as well as the maintained sterility of the skull pins at the end of the product's shelf life.Pass: VDmax validation as well as dye penetration test have proven the sterility of the product at delivery status as well as at the end of shelf life.
    DORO® Sterile Disposable Skull Pins (Titanium)
    MR-CompatibilityMR-Compatibility of the skull pins dedicated for the utilization in conjunction with MRT scanner utilizing the worst-case product.Pass: The skull pins are MR conditional.

    2. Sample size used for the test set and the data provenance:

    The document describes performance testing of physical properties of the skull pins (e.g., axial load, shear, sterility, biocompatibility) and packaging, not a study involving a "test set" of medical images or patient data in the typical sense of AI/ML evaluation. The specific sample sizes for these engineering and material tests are not detailed in this 510(k) summary, beyond indicating that "worst-case product" was utilized for some tests. The data provenance is from the manufacturer's internal testing (Pro Med Instruments GmbH, Germany). The tests are not clinical studies, so terms like "retrospective or prospective" do not apply.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This is not applicable to the type of device and testing described. Ground truth for a physical device's performance tests would typically involve established engineering standards, material specifications, and validated laboratory procedures, not expert consensus on medical images.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not applicable. This is not a study involving human interpretation or adjudication of medical data.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This document is for a physical medical device (skull pins), not an AI/ML diagnostic or assistive device. Thus, there is no mention of MRMC studies or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This device is not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    For the physical device, the "ground truth" for the performance tests would be defined by engineering specifications, material standards (e.g., ISO, ASTM), and validated test methodologies. For example, for the "Axial Load and Creep Test," the ground truth is the specified load and deformation limits that the device must withstand. For "Biocompatibility," the ground truth is the absence of cytotoxicity or harmful chemical substances as per established biocompatibility standards.

    8. The sample size for the training set:

    Not applicable. This device is not an AI/ML model.

    9. How the ground truth for the training set was established:

    Not applicable. This device is not an AI/ML model.

    In summary, the provided document is a regulatory submission for a physical medical device focusing on substantial equivalence based on material properties, design, and mechanical performance testing, rather than an AI/ML device requiring clinical validation or human reader studies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191740
    Date Cleared
    2020-02-21

    (235 days)

    Product Code
    Regulation Number
    882.4460
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Pro Med Instruments GmbH

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DORO LUCENT radiolucent/MRI Compatible Headrest System with Skull Pins are components of a mechanical support system, which is used in cranial and spine surgery when rigid skeletal fixation is required for cranial stabilization and when intra-operative imaging is used.

    The DORO LUCENT radiolucent/MRI Compatible Headrest System with noninvasive head positioning or noninvasive cranial stabilization device are components of a mechanical support system, which is used in cranial and spine surgery when noninvasive head positioning or noninvasive cranial stabilization is required and when intra-operative imaging is used.

    The DORO LUCENT Headrest System provides an interface for accessories like retractor systems, navigation adaptors or other items.

    Device Description

    The DORO LUCENT® iXI and iMRI Headrest System ensures an adequate positioning of a patient's head for neurosurgery. Additional intra-operative imaging can be performed.

    The DORO LUCENT® iXI and iMRI Headrest System consists of the following: Skull Clamp, Skull Pins, Parallelogram Adaptor and Headplate. The Parallelogram Adaptor is used to connect the Skull Clamp (including Skull Pins) or the Headplate to the OR-Table/ Transfer Board.

    Additional accessories like the Adjusting wrench and Transfer Collision Indicator supports the performance of the Headrest System.

    These special Headrest Systems are developed for selected OR-Table/ MRI scanner combinations. They are separated in following sets:

    • . 4003.200 DORO LUCENT® Headrest System TRUMPF for SIEMENS Aera/Skyra MRI Systems
    • 4003.300 DORO LUCENT® iMRI Headrest System MAQUET for SIEMENS/PHILIPS/GE ●
    • . 4003.500 DORO LUCENT® iXI Headrest System

    By the use of the Navigation Adaptors the performance of the Headrest System is supplemented. The intra-operative navigation is feasible because of the provided interface to the navigation device.

    AI/ML Overview

    The provided FDA 510(k) summary describes a medical device, the DORO LUCENT® iXI and iMRI Headrest System. This document focuses on the substantial equivalence of the new device to a predicate device rather than presenting a clinical study to evaluate diagnostic performance with a human-in-the-loop or standalone algorithm. Therefore, many of the requested criteria related to AI/algorithm performance and clinical study details are not applicable or not present in this submission.

    Here's the information that can be extracted from the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally demonstrated through "Pass" results in various performance tests.

    Acceptance Criteria (Test)Reported Device Performance
    DORO LUCENT® iXI and iMRI Headrest System
    Static load: Ability to sustain a certain load with an additional safety factor.Pass
    MR-Compatibility: Does not impair the function of the MRI system.Pass
    CT/Angio-Compatibility: No new or additional shadowing on the image.Pass
    X-Ray-Compatibility: No new or additional shadowing on the image.Pass
    DORO LUCENT® Skull Clamp
    Static load (Latching teeth mechanism): Ability to sustain a certain load with an additional safety factor without damage/malfunction.Pass
    Torque (Rocker Arm): Ability to resist applied torque without damaging, opening, or malfunction of the Open-Lock mechanism.Pass
    DORO LUCENT® Skull Pins
    Mechanical stability (scratch test): Withstands an applied radial force when an axial force is applied.Pass
    MR-Compatibility: MR conditional.Pass
    Sterility: Packaged sterile and stays sterile for shelf life.Pass
    Biocompatibility: Biocompatible.Pass
    DORO LUCENT® Headplate
    Static load: Ability to sustain a certain load with an additional safety factor without mechanical failure.Pass
    Biocompatibility: Biocompatible.Pass

    2. Sample size used for the test set and the data provenance

    This is a physical medical device (headrest system), not an AI/software device. The "test set" here refers to the physical units of the device and its components undergoing engineering and materials testing. The document does not specify the exact number of units or components tested for each criterion. The provenance of the data is from pro med instruments GmbH, the manufacturer, and the tests were likely conducted internally or by accredited third-party labs. The study is a prospective testing of manufactured devices against pre-defined engineering and safety standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. Ground truth for a physical device's performance often relates to instrumented measurements (e.g., force, torque, imaging artifacts) and material properties, not expert consensus as would be used for diagnostic imaging algorithms.

    4. Adjudication method for the test set

    Not applicable. Adjudication methods like "2+1" or "3+1" are used in clinical studies for diagnostic accuracy, where expert opinions need to be reconciled. For device performance testing, results are quantitative or pass/fail based on pre-defined specifications.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device.

    7. The type of ground truth used

    The ground truth for this device's performance is based on engineering specifications, material science, sterility standards, and regulatory requirements (e.g., MRI compatibility standards). For example, "MR-Compatibility" is verified against established standards for how metallic or non-metallic objects interact with MRI fields. "Static load" is verified against a specified weight with a safety factor.

    8. The sample size for the training set

    Not applicable. This is a physical medical device and does not involve AI/machine learning training sets.

    9. How the ground truth for the training set was established

    Not applicable. As there is no training set for an AI algorithm, no ground truth needs to be established for it.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191979
    Date Cleared
    2019-10-22

    (90 days)

    Product Code
    Regulation Number
    882.4460
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Pro Med Instruments GmbH

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DORO® QR3 XTom Headholder System is a mechanical support system, which is used in cranial and spine surgery when rigid skeletal fixation is required for cranial stabilization and when intra-operative imaging with a CT-Scanner is used.

    Device Description

    The DORO® QR3 XTom Headholder System ensures an adequate positioning of a patient's head for neurosurgery. Due to the utilized material the device can be used for intraoperative CT imaging procedures. The DORO® QR3 XTom Headholder System consists of the following: Skull Clamp, Skull Pins, Base Unit, Torque Screw Driver and U-Belt. The Base Unit is used to connect the Skull Clamp (including Skull Pins) to the OR-Table. Additional components like the Torque Screw Driver and U-Belt supports the performance of the Headholder System.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the DORO® QR3 XTom Headholder System.

    It's important to note that the provided text is an FDA 510(k) summary for a mechanical support system (Neurosurgical Head Holder/Skull Clamp) and not an AI/ML medical device. Therefore, the questions related to AI/ML specific aspects like training data, expert consensus for ground truth, MRMC studies, and various performance metrics like sensitivity, specificity, AUC are not applicable to this type of device. The acceptance criteria and performance are based on mechanical and usability testing.


    Acceptance Criteria and Device Performance

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance Criteria (Implied by "Pass" Result)Reported Device Performance
    DORO® QR3 XTom Headholder System (System Test)Ability of the system to sustain a certain load without mechanical failure.Pass: The System supports the static load without mechanical failure.
    DORO® QR3 XTom Headholder System (Usability)Usability of the System is given.Pass: The usability of the System is given.
    DORO® QR3 XTom Skull Clamp (Static Load - Latching teeth mechanism)The interface must withstand the static load over the defined duration without damage or malfunction.Pass: The interface must withstand the static load over the defined duration without damage or malfunction.
    DORO® QR3 XTom Skull Clamp (Torque - Rocker Arm)The Rocker Arm must withstand the torque without damaging, opening, or malfunction of the Open-Lock mechanism.Pass: The Rocker Arm must withstand the torque without damaging, opening or malfunction of the Open-Lock mechanism.
    DORO® QR3 XTom Skull Clamp (Creep Test)The skull clamp must maintain the applied maximum force for a defined time without a force deviation from the initially applied load by a defined value.Pass: The skull clamp must maintain the applied maximum force for a defined time without a force deviation from the initially applied load by a defined value.
    DORO® QR3 XTom Skull Clamp (Force Delivery Accuracy Verification)The skull clamp force delivery component must be verified at each major graduation throughout its range for a defined time to deliver the stated force within a defined range of the actual setting.Pass: The skull clamp force delivery component must be verified at each major graduation throughout its range for a defined time to deliver the stated force within a defined range of the actual setting.
    DORO® XTom U-Belt (Dynamic Load)The Belt must withstand the dynamic load without breakage or opening of the Belt.Pass: The Belt must withstand the dynamic load without breakage or opening of the Belt.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify the exact number of units tested. It states "The devices have been tested as a system and single device," implying that at least one of each (system, skull clamp, U-Belt) was subjected to testing. For mechanical tests, the sample size might be small, typically 1 to 3 units, as per engineering testing standards, unless statistical significance for variability is a concern.
    • Data Provenance: The manufacturer is "pro med instruments GmbH" located in "Freiburg im Breisgau, Germany." This suggests the testing and data likely originated from Germany. The document indicates these were direct performance tests on the physical device, not retrospective or prospective patient data studies.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    • Not Applicable (N/A). As this is a mechanical device, ground truth for its performance is established through objective physical measurements and engineering standards, not through expert human interpretation or consensus like in AI/ML image analysis.

    4. Adjudication Method for the Test Set

    • N/A. Adjudication methods (e.g., 2+1, 3+1) are typically used in studies involving human interpretation (like radiology reads) to resolve disagreements or establish consensus. For mechanical performance tests, the results are quantitative and objective, not subject to human interpretation disagreement in the same way.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No MRMC study was done. This type of study is relevant for diagnostic imaging AI/ML devices to assess the impact on human reader performance. This device is a mechanical support system, not an imaging interpretation aid.
    • Effect Size: N/A, as no MRMC study was conducted.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance)

    • N/A. This is a mechanical device, not an algorithm or software. Its performance is inherent to its physical design, materials, and manufacturing, not dictated by an algorithm.

    7. The Type of Ground Truth Used

    • Objective Mechanical Measurements and Engineering Standards: The "ground truth" for this device's performance is derived from its ability to meet predefined physical requirements (e.g., load bearing, torque resistance, force accuracy, maintaining clamping force over time) as measured by calibrated testing equipment and adherence to established engineering principles and safety margins. The "Pass" results confirm it met these criteria.

    8. The Sample Size for the Training Set

    • N/A. This device does not involve machine learning; therefore, there is no "training set."

    9. How the Ground Truth for the Training Set was Established

    • N/A. Since there is no training set, this question is not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K063494
    Date Cleared
    2007-05-21

    (182 days)

    Product Code
    Regulation Number
    882.4460
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    PRO-MED INSTRUMENTS GMBH

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Radiolucent / MRI Compatible Skull Clamp Headrest System with Skull Pins are components of a mechanical support system which is used in head and neck surgery when rigid skeletal fixation is required for cranial stabilization and when intra-operative CT or MR Imaging is used.

    The DORO® Radiolucent / MRI Compatible Horseshoe Headrest System are components of a mechanical support system which is used in head and neck surgery when non-invasive head support is required and when intra-operative CT or MR Imaging is used.

    The DORO® Radiolucent / MRI Compatible Halo System is a system of radiolucent halo rings of varying size and styles that may be used in neurosurgical applications as an arm rest during head and neck surgery when intra-operative CT or MR Imaging is required.

    Device Description

    The DORO® Radiolucent / MRI Compatible Cranial Stabilization and Halo Systems are for use in open and percutaneous craniotomies and spinal surgeries when rigid skeletal fixation is necessary (Skull Clamp with Skull Pins) or when secured noninvasive cranial stabilization (Horseshoe Headrest System) is required and when Intra-Operative CT or MR Imaging is used.

    Main components:

    • DORO® Radiolucent / MRI Compatible Skull Clamp Headrest System with Skull Pins
    • DORO® Radiolucent / MRI Compatible Horseshoe Headrest Systems
    • DORO® Radiolucent / MRI Compatible Halo System

    The DORO® Radiolucent / MRI Compatible Skull Clamp Headrest System uses a three-point fixation of the head during surgery in the prone, supine, lateral, and sitting positions via the radiolucent Skull Clamp with Skull Pins. The DORO® Radiolucent / MRI Compatible Horseshoe Headrest System is an alternative noninvasive gel-pad cranial stabilization system for prone and supine positioning. The Skull Clamp or Horseshoe Headrests can be connected to Swivel Adaptor which in turn can be connected to the appropriate Transitional Members (of varying lengths). These are all MR-Safe.

    The Radiolucent Halo System is a series of radiolucent rings (full, half and quarter rings) with fixation rods to attach to the radiolucent skull clamp described above. It provides a hand rest for the Surgeon and a mounting place for a variety of surgeon tools, such as brain Retractor and a holding tray. The Halo Rings and the Rod Fixation Assembly are all MR-Safe.

    The DORO® Radiolucent / MRI Compatible Cranial Stabilization and Halo Systems are sold non-sterile. The Swivel Adaptor, Skull Clamp and Horseshoe Headrests are intended to be cleaned by the user between uses. For rigid skeletal fixation, the Skull Clamp uses the disposable single-use ceramic (Yttried Zirconium) or Titanium Skull Pins, which are both x-ray and MRI-compatible.

    The components of the DORO® Headrest Systems are made of the following materials:

    • The Skull Clamp is made of NOVOTEX laminated fabric with phenolic resin (GRP) colored with BASANTOL black X82 liquid and POM (Delrin), PEEK and Polyurethan. MR-Safe.
    • The Swivel Adaptor and the Horseshoe Headrest in adult and pediatric versions, the Transitional Members and the Halo Systems are made of Novotex, PEEK and POM. MR-Safe.
    • DORO® Radiolucent Disposable Single-Use Skull Pins of Yttried Zirconium or Titanium are X-ray and MRI compatible. MR-Safe.

    The DORO® Radiolucent / MRI Compatible Cranial Stabilization and Halo Systems can be used with accessories from the predicate DORO® Headrest System (K001808) with Gel Pads for adults, Headrest supports with one or two Gel Pads, single Pin Holders, dual Pin Holders for adults, dual Pin Holders for children, a Multi-Purpose Skull Clamp with six Pin fixation or three Gel Pads.

    AI/ML Overview

    The provided text is a 510(k) summary for a medical device (DORO® Radiolucent / MRI Compatible Cranial Stabilization and Halo Systems). This document is a regulatory submission for premarket notification to the FDA, demonstrating that a new device is substantially equivalent to a legally marketed predicate device.

    It does not contain the acceptance criteria and study information typically found in a clinical performance study report for a diagnostic or AI-driven device. The document focuses on showing substantial equivalence based on design, construction, intended use, and performance characteristics compared to existing devices, particularly concerning its MRI compatibility.

    Therefore, many of the requested categories (such as reported device performance, sample size for test set, data provenance, number of experts for ground truth, adjudication method, MRMC study, standalone performance, type of ground truth, training set size, and ground truth for training set) are not applicable to this type of regulatory submission for this device.

    The "performance" aspect here refers more to the functional performance and safety of a mechanical device, rather than the accuracy of a diagnostic algorithm.

    Here's a breakdown of what can be extracted and what is not available from the provided text:


    Acceptance Criteria and Study for DORO® Radiolucent / MRI Compatible Cranial Stabilization and Halo Systems

    This 510(k) summary (K063494) for the DORO® Radiolucent / MRI Compatible Cranial Stabilization and Halo Systems focuses on establishing substantial equivalence to predicate devices, rather than presenting a clinical study with acceptance criteria for diagnostic performance. The "performance" discussed primarily relates to the device's mechanical integrity, radiolucency, and MRI compatibility.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria (from document)Reported Device Performance (from document)
    Material CompositionUse of specific materials for radiolucency and MRI compatibility.Skull Clamp: NOVOTEX (laminated fabric with phenolic resin, colored with BASANTOL black X82 liquid), POM (Delrin), PEEK, and Polyurethane. MR-Safe.
    Swivel Adaptor, Horseshoe Headrest, Transitional Members, Halo Systems: Novotex, PEEK, and POM. MR-Safe.
    Skull Pins: Yttried Zirconium or Titanium. X-ray and MRI compatible. MR-Safe.
    Intended Use EquivalenceSimilar use cases as predicate devices for cranial stabilization and support during neurosurgery.Used for rigid skeletal fixation (Skull Clamp with Skull Pins) or non-invasive cranial stabilization (Horseshoe Headrest System) and as an armrest/tool mount (Halo System) in head and neck surgery, specifically when intra-operative CT or MR Imaging is used. This is an additional application compared to some predicates, enabled by the device's MRI compatibility.
    Biocompatibility (for contacting parts)No biocompatibility issues should be raised for patient-contacting components.Radiolucent components (non-sterile): Do not contact the patient. Materials: NOVOTEX colored with BASF Basantal, POM, and Polyurethane.
    DORO® Radiolucent Disposable Single-Use Skull Pins: Contact the patient, but "there is no biocompatibility issues raised." (This implies a review and satisfaction of biocompatibility, often through adherence to ISO 10993 or similar standards, though specific data is not provided in this summary).
    MRI CompatibilityDevice components must be MR-Safe.All specified components (Skull Clamp, Swivel Adaptor, Horseshoe Headrest, Transitional Members, Halo Systems, Skull Pins of Yttried Zirconium or Titanium) are explicitly stated as "MR-Safe."
    Cleaning/SterilizationInformation on cleaning/sterilization status.Sold non-sterile. Swivel Adaptor, Skull Clamp, and Horseshoe Headrests are "intended to be cleaned by the user between uses." Skull Pins are "disposable single-use."
    Compatibility with existing systemsAbility to integrate with predicate accessories.Can be used with accessories from the predicate DORO® Headrest System (K001808), including Gel Pads, Headrest supports, Pin Holders, and a Multi-Purpose Skull Clamp.
    General Equivalence"Similar in design, construction, intended use and performance characteristics to the predicate devices." (K032331, K001808, K051501xc for specific materials).The summary states: "This device is almost similar in design, construction, intended use and performance characteristics to the predicate devices."

    2. Sample size used for the test set and the data provenance

    • Not Applicable. This is a 510(k) for a mechanical device, not a diagnostic algorithm that requires a "test set" in the context of diagnostic accuracy. Data provenance would refer to where clinical data for performance assessment came from, which is not relevant here.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable. No ground truth for a diagnostic test set is established. This device's performance is assessed through material properties, mechanical testing, and comparison of design features, not expert interpretation of outputs.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. No test set or adjudication process as defined for diagnostic performance is mentioned.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This is a mechanical surgical device, not an AI-powered diagnostic tool. MRMC studies are not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This device does not involve an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not Applicable. The "ground truth" for this device's regulatory review would be adherence to material specifications, engineering standards, biocompatibility testing results (implied), and demonstration of functional equivalence (e.g., rigid fixation, MRI safety) to predicate devices rather than a medical "truth" like disease presence.

    8. The sample size for the training set

    • Not Applicable. No training set for an algorithm exists for a mechanical device.

    9. How the ground truth for the training set was established

    • Not Applicable. No training set for an algorithm exists for a mechanical device.

    In summary: The provided document is a 510(k) summary for a mechanical neurosurgical device. It addresses regulatory requirements by demonstrating substantial equivalence to previously cleared devices, primarily focusing on materials, construction, intended use, and functional properties like MRI compatibility. The questions regarding acceptance criteria, sample sizes, expert ground truth, and AI/algorithm performance are not applicable to the information contained within this type of submission for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K070997
    Date Cleared
    2007-05-16

    (37 days)

    Product Code
    Regulation Number
    878.4400
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    PRO-MED INSTRUMENTS GMBH

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DORO® Non-Stick Bipolar Forceps are designed to grasp, manipulate and coagulate selected tissue and are intended for use in general surgical procedures. They are connected through a suitable bipolar cable with the bipolar output of an electrosurgical generator. Bipolar forceps must only be used with bipolar coagulation is achieved using electrosurgical energy generated by the electrosurgical generator unit and activated by a footswitch.

    The DORO® NON-STICK Bipolar Forceps have not been shown to be effective for tubal sterliization or tubal coagulation for sterilization procedures and should not be used for these procedures.

    The DORO® Bipolar Reusable Cables are designed to conduct electrical power from the output of a high frequency electrosurgical generator to the instrument.

    Device Description

    The DORO® Non-Stick Bipolar Forceps are for use in general surgical procedures. These devices are connected through a suitable bipolar cable with the bipolar output of an electrosurgical generator. The forceps are designed to grasp and manipulate selected tissues. Coagulation is achieved using electrosurgical energy generated by the electrosurgical generator unit and activated by a footswitch.

    During the coagulation of tissue, the coagulated tissue may stick to the forceps. This undesirable effect can be eliminated by the non-stick properties of the unique solid silver alloy tip forceps of the DORO line. The solid silver allov tip provides excellent thermal properties and eliminates the difficult and time-consuming of the forceps during a procedure. The non-stick properties are permanent since it is not a coating and the non-stick properties are retained through repeated and frequent sterilizations.

    Reusable Standard two-pin round connector cables are designed to work with all foot-switching bipolar electrosurgical generators.

    The DORO® Bipolar Reusable Cables are non-sterile, reusable bipolar cables of 3 and 5 meters. They are designed to transport electrical power from a high-frequency electrosurgical generator by a foot-activated switch. The cable fittings are compatible for use with: Erbe, Martin, Valleylab, Aesculap and other ES Generator Units with twinpin connectors (2-banana plug). The cables are comprised of silicone (outer insulation), PTFE or FEP (core insulation), and Cu-wire poreless silver plated (wire).

    The cable consists of a twin-pin connector (2-banana plug) that is comprised of silver plated bronze and insulated with silicone. The connector cable at the forceps end is comprised of bronze or brass - silver or nickel plated, Stainless Steel 1.4310, PP (Polypropylene or TPE).

    AI/ML Overview

    Here's an analysis of the provided text regarding the DORO® Non-Stick Bipolar Forceps and DORO® Bipolar Reusable Cables, focusing on acceptance criteria and study information:

    Analysis:

    The provided 510(k) Notification (K070997) does not contain a study that proves the device meets specific performance acceptance criteria in the manner typically seen for AI/ML or diagnostic devices.

    Instead, this submission seeks substantial equivalence for the DORO® devices based on their similarity to legally marketed predicate devices and compliance with recognized performance standards for electrosurgical equipment. The "performance" discussed primarily relates to safety and functional standards, not to a measurement of diagnostic accuracy or clinical effectiveness against a set of predefined acceptance metrics for a novel technology.

    Therefore, many of the requested categories (e.g., sample size for test/training sets, number of experts for ground truth, MRMC study, standalone performance) are not applicable to this type of device and submission.


    Detailed Breakdown based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

      Acceptance CriteriaReported Device Performance/Statement
      General Safety and Effectiveness (Implicit)Substantial equivalence to predicate devices (K51429, K992931 for forceps; K061835, K051429 for cables).
      "Non-Stick" Property (for Forceps)"This undesirable effect [tissue sticking] can be eliminated by the non-stick properties of the unique solid silver alloy tip forceps of the DORO line. The solid silver alloy tip provides excellent thermal properties and eliminates the difficult and time-consuming of the forceps during a procedure. The non-stick properties are permanent since it is not a coating and the non-stick properties are retained through repeated and frequent sterilizations." (This is a descriptive claim, not a quantitative performance metric with a defined acceptance threshold within this document).
      Compatibility (for Cables)"The cable fittings are compatible for use with: Erbe, Martin, Valleylab, Aesculap and other ES Generator Units with twin-pin connectors (2-banana plug)."
      Sterilization (by user)"The user may sterilize these devices by using a validated and applicable sterilization process." "Only a validated steam-sterilization process according DIN EN 554 / ISO 11134 that uses a sterilization cycle of 137°C / 280°F, 3 bar, for min. 15 minutes should be used." (Implicit acceptance is that the device withstands this process and remains functional and sterile.)
      Compliance with Recognized StandardsDIN EN 60601-1: Medical electrical equipment - Part 1: General requirements for safety (IEC 60601-1:1988 + A1:1991 + A2:1995); German version EN 60601-1:1990 + A2:1995; Version: 01-Mar-1996;

    DIN EN 60601-2-2: Medical electrical equipment - Part 2-2: Particular requirements for the safety of high frequency surgical equipment (IEC 60601-2-2:1998); German version EN 60601-2-2:2000; Version: 01-Aug-2001;
    ANSI/AAMI HF18-2001: Electrosurgical Devices: Version: 01-May-2001.
    (The device's performance is implicit in meeting these standards, but specific criteria and test results are not detailed in this summary.) |
    | Functionality (Forceps) | "designed to grasp and manipulate selected tissues. Coagulation is achieved using electrosurgical energy generated by the electrosurgical generator unit and activated by a footswitch." |
    | Functionality (Cables) | "designed to transport electrical power from a high-frequency electrosurgical generator by a foot-activated switch." |
    | Limitations (Forceps) | "The DORO® NON-STICK Bipolar Forceps have not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures and should not be used for these procedures." (This is a contraindication/limitation, not a performance metric itself.) |

    1. Sample size used for the test set and the data provenance:

      • Not applicable / Not provided. The submission focuses on substantial equivalence to existing devices and compliance with recognized standards rather than a clinical or performance study with a test set of data. The "tests" would have been conducted against the standards themselves (e.g., electrical safety, mechanical durability, sterilization validation), but the details of those internal test sets are not in this summary.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable / Not provided. Ground truth establishment by experts is not relevant for this type of device submission where safety and functional compliance to standards are primarily demonstrated.
    3. Adjudication method for the test set:

      • Not applicable / Not provided.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This is not an AI/ML device, nor a diagnostic device that involves human readers interpreting results.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This is not an AI/ML device.
    6. The type of ground truth used:

      • Not applicable. For this type of device, the "ground truth" would be the successful demonstration of compliance with the specified engineering and safety standards (e.g., electrical resistance within limits, mechanical strength, material biocompatibility, successful sterilization cycles), rather than clinical outcomes or expert consensus.
    7. The sample size for the training set:

      • Not applicable. There is no "training set" in the context of this traditional medical device submission.
    8. How the ground truth for the training set was established:

      • Not applicable.

    Summary of the Study (or lack thereof, in the context of typical "device performance studies"):

    The 510(k) K070997 for the DORO® Non-Stick Bipolar Forceps and DORO® Bipolar Reusable Cables relies on a Substantial Equivalence pathway. This means that no specific clinical or performance study (in the sense of a trial with a defined 'test set' against a 'ground truth' for an AI/ML or diagnostic device) is presented within this summary.

    Instead, the "study" is an assertion of equivalence and adherence to established regulatory and engineering standards. The key elements for proving acceptance were:

    • Predicate Device Comparison: The devices are compared to previously cleared, legally marketed bipolar forceps and reusable cables (e.g., Günter Bissinger Medizintechnik GmbH (K51429), e.g. Link Technology, Inc. (K992931) for forceps; Stryker Silverglide Bipolar Forceps Reusable Cables (K061835) and CLARIS bipolar forceps (K051429) for cables). The claim is that the DORO devices have the same intended use, similar technological characteristics (e.g., material for non-stick properties, cable construction, connectivity), and raise no new questions of safety or effectiveness.
    • Compliance with Recognized Standards: The devices are stated to comply with several international and national standards for medical electrical equipment and electrosurgical devices (DIN EN 60601-1, DIN EN 60601-2-2, ANSI/AAMI HF18-2001). This compliance implicitly means they have undergone testing to meet the safety and performance requirements outlined in these standards.
    • Sterilization Validation (by user): The document specifies a validated steam-sterilization process, implying that the device was tested to withstand and allow for effective sterilization by the end-user.

    In essence, the "proof" is the argument for substantial equivalence and adherence to existing safety and performance standards for electrosurgical accessories, rather than a novel study demonstrating a new type of performance metric.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Applicant Name (Manufacturer) :

    PRO-MED INSTRUMENTS GMBH

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DORO Radiolucent Headrest System is used in open and percutaneous craniotomies and spinal surgeries for rigid cranial fixation and when intraoperative CT imaging is used.

    The DORO® J-ARM Retractor Accessory is used to fix tissue retractor during neurosurgical procedures.

    Device Description

    The DORO® Radiolucent Headrest System is a device system used as a support during head and neck surgery. It is composed of several components which are fabricated of materials that allow it to be radiolucent on x-ray. These components are:

      1. the adjustable Base Unit,
      1. the Skull Clamp, and
      1. the Swivel Adaptor,
      1. Horseshoe Headrest Adult and Children

    It is suitable for use in adults and children when the appropriate components are selected.

    The DORO System uses a three-point fixation of the head during surgery in the prone, supine, lateral, and sitting positions. The double clamping of the Adjustable Base unit provides simultaneous fixation of the lateral and vertical positions. The Crossbar is used for connection to the side rails of standard operating tables for fixation in a sitting position. The Swivel Adaptor is used in the base unit to provide 360 degree rotation.

    The DORO Headrest System, except for the disposable Skull Pins, is sold non-sterile. The Base Unit, Swivel Adaptor and Skull Clamp are intended to be cleaned by the user between uses. It is intended to be used with Disposable, Skull Pins which were cleared with the DORO Headrest System (K001808).

    The components of the DORO Headrest System are made of the following materials:

    Skull Clamp is made of NOVOTEX laminated fabric with phenolic resin (GRP) colored with BASANTOL black X82 liqu1d and POM (Delrin), PEEK and Polyurethan.

    The Swivel Adaptor and the Horseshoe Headrest in adult and pediatric versions, made of Novotex, PEEK and POM

    The Base is made of Novotex, PEEK, POM and aluminum.

    The J-Arm Retractor System is made of aluminum alloy and is not radiolucent.

    The DORO Radiolucent Headrest System can also be used with accessories from the predicate DORO Headrest System (K001808) with Elastic Pads for adults, Headrest supports with one or two Elastic Pads, single Pin Holders, dual Pin Holders for adults, dual Pin Holders for children, a Multi-Purpose Skull Clamp with six Pin fixation or three Elastic Pads).

    The DORO Radiolucent Headrest System can also be used with the J-Arm Retractor and Halo Retractor Assembly (K001808) which allows for use as a support and retraction device during neurosurgical procedures where retraction of the brain tissue and handrest are needed.

    AI/ML Overview

    The provided 510(k) summary for the DORO® Radiolucent Headrest System does not contain information about specific acceptance criteria, device performance metrics, or any detailed studies demonstrating the device meets such criteria.

    The document primarily focuses on demonstrating substantial equivalence to predicate devices based on design, construction, intended use, and performance characteristics, as required for a 510(k) submission. It briefly mentions materials and intended use but does not present quantitative performance data or a study design.

    Therefore, the following information cannot be extracted from the provided text:

    1. A table of acceptance criteria and the reported device performance: No specific acceptance criteria or performance metrics are listed. The submission asserts "similar performance characteristics" to predicate devices without providing explicit data.
    2. Sample size used for the test set and the data provenance: No test set is described, and thus no sample size or data provenance is available.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: No test set or ground truth establishment process is described.
    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable as no test set is described.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. The device is a physical medical instrument (surgical head holder), not an AI-powered diagnostic or assistive tool for human readers.
    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an algorithmic device.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable as no ground truth is discussed for a study.
    8. The sample size for the training set: Not applicable as there is no training set for a study of this nature.
    9. How the ground truth for the training set was established: Not applicable.

    The document states that "Performance standards for sheath introducers have not been established by the FDA under Section 514 of the Food, Drug and Cosmetic Act." This implies that the submission relies on substantial equivalence rather than meeting specific pre-defined performance standards or conducting extensive performance studies beyond what's needed for safety and equivalence claims.

    Ask a Question

    Ask a specific question about this device

    K Number
    K001808
    Date Cleared
    2000-08-04

    (50 days)

    Product Code
    Regulation Number
    882.4460
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    PRO-MED INSTRUMENTS GMBH

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1