Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K970128
    Date Cleared
    1997-04-11

    (87 days)

    Product Code
    Regulation Number
    892.1200
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended use of the Dyna Camera 2 series collimator is to detect and image the distribution of high energy photons from an administered positron emitting radioactive agent in the human body. Same as predicate.

    Device Description

    Not Found

    AI/ML Overview

    This document describes the 510(k) summary for Park Medical Systems' 511 KeV collimator for their ISOCAM I and ISOCAM II nuclear imaging systems. The submission aims to demonstrate substantial equivalence to a predicate device, the Dyna camera 2 series high energy collimators, rather than proving the device meets specific acceptance criteria through a dedicated study with statistical endpoints.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in the traditional sense of a clinical trial or performance study with defined thresholds for sensitivity, specificity, etc. Instead, it presents a comparison table of technological characteristics and performance metrics against a predicate device to demonstrate substantial equivalence.

    Feature/Spec.Predicate (Dyna Camera 2 Series High Energy Collimator)Park 511 KeV Collimator (ISOCAM I/II)Acceptance Criteria (Implied by Substantial Equivalence to Predicate)
    Intended UseDetect and image the distribution of high energy photons from an administered positron emitting radioactive agent in the human body.Same as predicate.Be suitable for the same intended use.
    Physical Description:
    1) NameHigh Energy CollimatorUltra High Energy CollimatorBe a collimator for high energy imaging.
    2) No. of holes1,100 (round)7,500 (hex)Number and shape of holes is a design choice, not a direct performance criterion for equivalence.
    3) Energy Range44 - 525 KeV50 - 562 KeVOperate within a similar high-energy range.
    4) Thickness2.5 inches = 63.5 mm76 mmHave a physical thickness suitable for high-energy attenuation.
    5) Hole Size0.20 inches = 5.08 mm4 mmHole size is a design choice affecting resolution and sensitivity trade-offs.
    6) Septa thickness0.156 inches = 3.94 mm2 mmSepta thickness is a design choice affecting resolution and septal penetration.
    7) Field Size11.8 inches = 300 mm (diameter)419 mm × 566 mm (rectangular)Have a field size suitable for nuclear imaging.
    PerformanceResolutionResolutionResolution performance comparable to the predicate.
    FWHM 99mTc (surface)5.59 mmN/A
    Ask a Question

    Ask a specific question about this device

    K Number
    K964834
    Date Cleared
    1997-02-28

    (88 days)

    Product Code
    Regulation Number
    892.1200
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The “ISOCAM II” manufactured by Park Medical Systems Inc. is a Single Photon Emission Computed Tomography (SPECT) gamma camera which is intended to image the distribution of radionuclides in the body by means of a single photon radiation detector. This system includes signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories. Same as ISOCAM II.

    Device Description

    The ISOCAM II (Dual Head Gamma Camera) is a Single Photon Emission Computed Tomography (SPECT) gamma camera. This 510(K) is being submitted because of a modification to our ISOCAM II, more specifically, an additional collimator for the ISOCAM II, that allows better images. This modification changes the effectiveness of the ISOCAM II, therefore, we are submitting another premarket notification. The collimator is called MCAT™, which stands for Modular Coded Apperture Technology.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information based on the provided text for the ISOCAM II with MCAT™:

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance MetricAcceptance Criteria (ISOCAM II with MCAT™)Reported Device Performance (ISOCAM II with MCAT™)Baseline Performance (ISOCAM II)
    System Sensitivity≥ 1000 cpm/μCi/min1000 cpm/μCi/min (implied by "better sensitive measurements" and meeting the "≥" criteria)190 cpm/μCi/min (with LEHR collimator)
    Image Resolution (0-20 cm)5-8 mm FWHM5-8 mm FWHM3.7-12.3 mm FWHM

    Note: The document explicitly states "better sensitivity and resolution measurements change the effectiveness of the ISOCAM II system." It then lists the specific values for the MCAT™ version, implying these are the performance goals and achievements.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided text does not contain information about a specific test set in terms of patient data or clinical trials for performance evaluation. The "study" described appears to be a technical comparison and verification of the device's physical and technical characteristics against a predicate device (the ISOCAM II without MCAT™).

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This information is not provided in the document. Given the nature of the evaluation (technical performance metrics for a gamma camera), it's unlikely that traditional "ground truth" established by clinical experts in the context of diagnostic accuracy was involved in the same way it would be for AI algorithms. The performance metrics (sensitivity and resolution) are typically measured using physical phantoms and established scientific methodologies.

    4. Adjudication Method

    This information is not applicable to the type of evaluation described. There's no mention of a human-in-the-loop diagnostic task requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    A MRMC comparative effectiveness study was not conducted, or at least not described in this submission. The focus is on the technical specifications of the imaging equipment itself, not on comparing diagnostic outcomes with human readers.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    This is not applicable as the device is a gamma camera, not an AI algorithm. The performance metrics listed are inherent to the machine's imaging capabilities.

    7. Type of Ground Truth Used

    The ground truth used for evaluating the performance metrics (sensitivity and resolution) would be based on physical measurements using standardized phantoms and radioactive sources. This is a common practice in imaging device evaluation to objectively quantify their technical capabilities. There are no mentions of expert consensus, pathology, or outcomes data being used for this specific evaluation.

    8. Sample Size for the Training Set

    This information is not applicable as the device is a gamma camera and not an AI algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable for the same reason as above.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1