Search Results
Found 3 results
510(k) Data Aggregation
(239 days)
Orthofix US LLC
The OFIX MIS App software assists healthcare professionals in viewing, storing, and measuring images as well as planning orthopedic surgeries. The app allows service providers to plan surgical procedures by making measurements for the placement of surgical implants. Clinical judgement and experience are required to properly use the software.
The OFIX MIS App will provide a novel mobile phone solution to supplement the use of calipers when selecting rods used in immobilization of pedicle screw type spinal systems. The App will rely on capturing 2D images used in spinal implant procedures through use of mobile phone camera. The App measures the difference between known Orthofix pedicle screw diameters and provides feedback useful in determining appropriate rod length with corresponding Orthofix product number.
The OFIX MIS App is a software designed to assist healthcare professionals in viewing, storing, measuring images, and planning orthopedic surgeries, specifically by measuring anatomical components for the placement of surgical implants. The device helps determine appropriate rod length with corresponding Orthofix product numbers by capturing 2D images used in spinal implant procedures via a mobile phone camera.
Here's an analysis of the acceptance criteria and the study proving the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance
The provided text does not explicitly detail specific quantitative acceptance criteria (e.g., minimum accuracy percentages, maximum error margins) and their corresponding reported device performance values. Instead, it broadly states that "The testing demonstrated that the App meets the required specifications." and "The OFIX MIS App demonstrated accuracy and usability with bench top and simulated use testing."
However, based on the device's function, the implied acceptance criteria would revolve around the accuracy of measurements and the usability of the app for surgical planning. The provided information indicates compliance with these general requirements.
Acceptance Criteria (Implied) | Reported Device Performance (Summary from text) |
---|---|
Accuracy of measurements | "demonstrated accuracy" |
Usability for surgical planning | "demonstrated usability" |
Meeting required specifications | "meets the required specifications" |
Safety and effectiveness | "testing data support the safety and effectiveness" |
Performance as intended | "demonstrate performance as intended in the specified use conditions" |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify the sample size used for the test set. It also does not explicitly state the data provenance (e.g., country of origin, retrospective or prospective nature of the images used for testing). The testing is described as "bench top and simulated use testing," which implies either synthetic data, controlled clinical images, or a mix, but no details are given.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications
The document does not provide information regarding the number of experts used to establish ground truth for the test set or their specific qualifications (e.g., "radiologist with 10 years of experience").
4. Adjudication Method for the Test Set
The document does not describe the adjudication method used for the test set (e.g., 2+1, 3+1, none).
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size
The document does not indicate that an MRMC comparative effectiveness study was done to assess how much human readers improve with AI vs. without AI assistance. The study focuses on the standalone performance and usability of the OFIX MIS App.
6. If a Standalone Study (Algorithm Only Without Human-in-the-Loop Performance) Was Done
Yes, a standalone study was implicitly done. The "Software Verification and Validation Testing" and "Usability Testing" sections describe the evaluation of the App itself to ensure it meets specifications and demonstrates accuracy and usability. The statement "The OFIX MIS App demonstrated accuracy and usability with bench top and simulated use testing" suggests testing of the algorithm's performance and the app's functionality independent of clinical human performance metrics.
7. The Type of Ground Truth Used
The document does not explicitly state the type of ground truth used (e.g., expert consensus, pathology, outcomes data). Given the nature of the device (measurement and planning), it is highly probable that a "gold standard" of measurement, likely derived from manual measurements by qualified experts or using high-precision instruments on the test images, would have been used as the ground truth.
8. The Sample Size for the Training Set
The document does not specify the sample size for the training set. It mentions the algorithm is "not machine-learning based," which implies that there might not have been a traditional "training set" in the machine learning sense. Instead, the algorithm relies on "information specific to Orthofix implant systems."
9. How the Ground Truth for the Training Set Was Established
Since the algorithm is explicitly stated as "not machine-learning based" and uses "information specific to Orthofix implant systems," there wouldn't be a typical training set with corresponding ground truth established in the way machine learning models are trained. Instead, the "ground truth" or foundational data for the algorithm would be the known, precise specifications and dimensions of Orthofix's pedicle screws and rods, likely derived from their engineering and manufacturing data. The algorithm would then use these pre-defined parameters to process the captured images and recommend appropriate rod lengths.
Ask a specific question about this device
(253 days)
Orthofix US LLC
The 3° Anterior Cervical Plating System is intended for anterior fixation to the cervical spine from C2 to C7. The specific clinical indications include:
- Degenerative disc disease (defined as back pain of discogenic origin with degenerative disc confirmed by patient history and radiographic studies).
- Spondylolisthesis
- Fracture
- Spinal stenosis
- Deformities (i.e., scoliosis, kyphosis, and/or lordosis)
- Tumor
- Pseudoarthrosis
- Revision of previous surgery
The Reliant Anterior Cervical Plating System is intended for anterior fixation to the cervical spine from C2 to C7. The specific clinical indications include:
- Degenerative disc disease (defined as back pain of discogenic origin with degenerative disc confirmed by patient history and radiographic studies).
- Spondylolisthesis
- Fracture
- Spinal stenosis
- Deformities (i.e., scoliosis, kyphosis, and/or lordosis)
- Tumor
- Pseudoarthrosis
- Revision of previous surgery
The Unity Lumbosacral Fixation Plate is indicated for use as an anteriorly placed supplemental fixation device for the lumbosacral (L5-S1) level below the bifurcation of the vascular structures. The Unity LX Lumbar Fixation Plate is indicated for use as an anteriorly or anterolaterally placed supplemental fixation device for the lumbar region of the spine above the bifurcation of the vascular structures. When properly used, the system will help provide temporary stabilization until a solid spinal fusion develops. Specific indications include:
- Degenerative disc disease (defined as back pain of discogenic origin with degenerative disc confirmed by patient history and radiographic studies).
- Pseudoarthrosis.
- Spondylolysis.
- Spondylolisthesis.
- Fracture.
- Neoplastic disease.
- Unsuccessful previous fusion surgery.
- Lordotic deformities of the spine.
- Idiopathic thoracolumbar or lumbar scoliosis.
- Deformities (i.e., scoliosis, kyphosis, and/or lordosis) associated with deficient posterior elements such as that resulting from laminectomy, spina bifida, or myelomenigocele.
- Neuromuscular deformity (i.e., scoliosis, lordosis, and/or kyphosis) associated with pelvic obliquity.
The Hallmark Anterior Plate System is a temporary implant intended for anterior fixation to the cervical spine from C2 to C7. The specific clinical indications include:
- Degenerative disc disease (defined as back pain of discogenic origin with degenerative disc confirmed by patient history and radiographic studies).
- Spondylolisthesis
- Fracture
- Spinal stenosis
- Deformities (i.e., scoliosis, kyphosis, and/or lordosis)
- Tumor
- Pseudoarthrosis
- Revision of previous surgery
The NewBridge Laminoplasty Fixation System is indicated for use in the lower cervical and upper thoracic spine (C3-T3) after a laminoplasty has been performed. The system holds or buttresses the allograft in place in order to prevent expulsion of the allograft or impingement of the spinal cord.
The CETRA Anterior Cervical Plate System is a temporary implant intended for anterior fixation to the cervical spine from C2 to C7. The specific clinical indications include:
- Degenerative disc disease (defined as back pain of discogenic origin with degenerative disc confirmed by patient history and radiographic studies).
- Spondylolisthesis
- Fracture
- Spinal stenosis
- Deformities (i.e., scoliosis, kyphosis, and/or lordosis)
- Tumor
- Pseudoarthrosis
- Revision of previous surgery
3° Anterior Cervical Plating System - The 3° Anterior Cervical Plating System is a temporary titanium alloy (Ti6Al-4V ELI, per ASTM F136) system comprised of a variety of non-sterile, single use components that allow the surgeon to build an anterior cervical implant construct. The system's design is intended to stabilize the cervical spinal operative site during the fusion process of a bone graft in the disc space. The system is attached to the anterior aspect of the vertebral body by means of screws to the cervical spine. The system consists of an assortment of screws, plates and associated instrumentation which assists in the surgical implantation of the devices. The system is provided non-sterile and requires sterilization prior to use.
Reliant Anterior Cervical Plating System - The Reliant Anterior Cervical Plating System is temporary, titanium alloy system comprised of a variety of single-use components that allow the surgeon to build an anterior cervical implant construct. The system is attached to the anterior aspect of the vertebral body by means of screws to the cervical spine. The system consists of an assortment of screws, plates and associated instrumentation which assist in the surgical implantation of the devices. The system is provided non-sterile and requires sterilization prior to use.
Unity Lumbosacral Fixation System - The Unity Lumbosacral Fixation System consists of the Unity 51 Lumbosacral Fixation Plate for lumbosacral fixation and the Unity LX Lumbar Fixation Plate for anterior or anterolateral fixation above L5-S1.
The Unity 51 Lumbosacral Fixation Plate is designed specifically for supplemental fixation of anterior lumbar fusions at the L5-S1 level. The plate is contoured to the unique anatomy of the L5-S1 segment. The Unity 51 Plate is available in six heights: 17mm to 25mm in 2mm increments and one additional plate at 28mm.
The Unity LX Lumbar Fixation Plate is designed for lumbar levels above L5-S1 in 2 configurations - anterior and anterolateral. In the anterolateral configuration, the plate allows surgeons lateral-to-medial placement of the self-neuro foramen. The Unity LX Plate is available in six heights: 19mm to 25mm in 2mm increments and two additional plates at 28mm and 31mm.
Hallmark Anterior Plate System – The Hallmark Anterior Plate System is comprised of a variety of non-sterile, single use, titanium alloy (6AL-4V-ELI, per ASTM F136) components that allow a surgeon to build an anterior cervical implant construct. The system is attached to the anterior aspect of the vertebral body by means of screws to the cervical spine. The system consists of an assortment of screws, plates and associated instrumentation, which assists in the surgical implantation of the devices.
NewBridge Laminoplasty Fixation System – The NewBridge Laminoplasty Fixation System is a device comprised of non-sterile, single use, titanium alloy components. The specially shaped plates, made of commercially pure (CP) titanium conforming to ASTM F67, are designed to fit the anatomy of a dorsally elevated lamina. The plates have screw holes on both ends, which allow for attachment to the vertebral body, and a screw hole in the center for attachment to the allograft.
The screws, made of titanium alloy (Ti-6AI-4V ELI, per ASTM F136) are available in a variety of lengths and diameters in order to meet individual anatomical requirements.
The NewBridge Laminoplasty Fixation System must always be used with an allograft.
CETRA Anterior Cervical Plate System – The CETRA Anterior Cervical Plate System is comprised of an assortment of non-sterile, single use, titanium alloy (Ti-6AI-4V ELI per ASTM F136) plates and screws that allow a surgeon to build a temporary anterior cervical implant construct. The plate is attached to the anterior aspect of the vertebral body, by means of screws to the cervical spine. The system includes the necessary instrumentation to assist in the surgical implantation of the devices.
The provided document describes the addition of MR Conditional labeling to several spinal fixation systems and details the non-clinical tests performed to support this claim. It does not contain information about a study proving the device meets clinical acceptance criteria or comparative effectiveness with human readers. The information below is based only on the provided text, focusing on the MR Conditional aspect.
Here's the breakdown of the acceptance criteria and the study that proves the device meets the acceptance criteria for MR Conditional labeling:
1. Table of Acceptance Criteria and Reported Device Performance (for MR Conditional Labeling)
Acceptance Criteria (Standard Test Methods) | Purpose (Performance Measure) | Reported Device Performance (Implied by clearance) |
---|---|---|
ASTM F2052-15 | Measurement of magnetically induced displacement force | Met requirements for MR Conditional labeling |
ASTM F2213-17 | Measurement of magnetically induced torque | Met requirements for MR Conditional labeling |
ASTM F2119-07 | Evaluation of MR image artifacts | Met requirements for MR Conditional labeling |
ASTM F2182-19E2 | Measurement of radio frequency induced heating | Met requirements for MR Conditional labeling |
ASTM F2503 | Marking medical devices for safety in MR environment | Applied appropriate MR safety markings |
Note: The document states that the devices "met requirements necessary for MRI Conditional labeling", implying that the performance in each test fell within the acceptable limits defined by the respective ASTM standards for MR Conditional compatibility. Specific quantitative values for displacement force, torque, artifact size, or temperature rise are not provided in this summary.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Not explicitly stated for each test. The testing was conducted on the "subject devices," referring to the various components of the 3° Anterior Cervical Plating System, Reliant Anterior Cervical Plating System, Unity Lumbosacral Fixation System, Hallmark Anterior Plate System, NewBridge Laminoplasty Fixation System, and CETRA Anterior Cervical Plate System. Typically, MR safety testing involves multiple samples of the actual devices or representative worst-case configurations.
- Data Provenance: The tests conducted are non-clinical, controlled laboratory experiments. No human patient data (retrospective or prospective) or country of origin for such data is relevant for these specific MR safety tests.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts
- Not applicable. The ground truth for MR Conditional testing is established by recognized international standards (ASTM). The tests are performed by qualified testing laboratories or personnel following these standardized protocols, not by clinical experts establishing a ground truth based on patient cases.
4. Adjudication Method for the Test Set
- Not applicable. Adjudication methods like 2+1 or 3+1 are typically used for clinical study endpoints, especially when expert consensus is needed to establish ground truth or resolve discrepancies in interpretations of medical images or patient outcomes. For non-clinical, standardized engineering tests, the results are objectively measured against the criteria defined in the ASTM standards.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
- No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This submission is for the addition of MR Conditional labeling, which relies on non-clinical engineering tests to ensure device safety in the MR environment. It does not involve AI or human-in-the-loop performance evaluation.
6. If a Standalone Study (i.e., algorithm only without human-in-the-loop performance) Was Done
- No, a standalone study in the context of an algorithm or AI performance was not done. The tests performed are engineering-based evaluations of physical device properties in an MR environment.
7. The Type of Ground Truth Used
- The ground truth used for this determination is explicit, objective criteria defined by internationally recognized consensus standards for MR safety: ASTM F2052-15, ASTM F2213-17, ASTM F2119-07, ASTM F2182-19E2, and ASTM F2503. These standards specify the methods and acceptable limits for magnetically induced displacement, torque, image artifacts, and radiofrequency induced heating.
8. The Sample Size for the Training Set
- Not applicable. This is not an AI or machine learning device. Therefore, there is no "training set" in the conventional sense.
9. How the Ground Truth for the Training Set Was Established
- Not applicable. Since there is no training set, the establishment of its ground truth is not relevant.
Ask a specific question about this device
(19 days)
Orthofix US LLC
FORZA XP Expandable Spacer System is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels in the lumbar spine (L2-S1). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies. DDD patients may also have up to Grade 1 spondylolisthesis at the involved levels. These patients may have had a previous non-fusion surgery at the involved level(s).
FORZA XP Expandable Spacer System is intended for use with autograft comprised of cancellous and/or corticocancellous bone graft and supplemental fixation.
Patients must have undergone a regimen of at least six months of non-operative treatment prior to being treated with FORZA XP Expandable Spacer System.
The FORZA XP Expandable Spacer System is comprised of an assortment of single use spacers made from titanium alloy and PEEK Polymer with height expansion capability. The implants feature a bulleted nose for ease of insertion and anti-migration ripples on both the inferior and superior surfaces to provide increased stability and help prevent anterior/posterior movement of the device.
The implants are offered in non-lordotic, lordotic, and hyperlordotic configurations to help restore the natural curvature of the spine. The implants can be used in single placement or pairs with typical approaches being transforaminal lumbar interbody fusion (TLIF) and posterior lumbar interbody fusion (PLIF).
The FORZA XP Expandable Spacer System is not intended to be used as a stand-alone device and must be used with a supplemental fixation system.
The provided text describes a 510(k) premarket notification for the FORZA XP Expandable Spacer System. However, it does not contain any information about acceptance criteria or a study proving the device meets performance criteria, especially regarding AI/algorithm performance.
The document explicitly states:
- "No new mechanical testing was performed for the FORZA XP Expandable Spacer System because there were no design changes to the device."
- "The proposed sterile package offering does not change the design, materials, function, performance for the FORZA XP implants most recently cleared thru 510(k) K211704."
- "FORZA XP Expandable Spacer System has the same intended use, indications for use, technological characteristics, materials, the same principles of operation and same design as the predicate device FORZA XP Expandable Spacer System (K211704)."
This 510(k) submission is a "Product Line Extension" for a physical medical device (an expandable spinal spacer), primarily addressing a sterile packaging configuration change from a previously cleared device (K211704). It does not involve any AI/algorithm component.
Therefore, I cannot fulfill your request for information about acceptance criteria or a study related to AI/algorithm performance, multi-reader multi-case studies, standalone algorithm performance, or ground truth establishment for AI, as this information is not present in the provided document.
If you have a document describing an AI/algorithm-based device, please provide that document, and I would be happy to describe the acceptance criteria and study as requested.
Ask a specific question about this device
Page 1 of 1