K Number
K162209Device Name
Sapphire NC Plus
Manufacturer
OrbusNeich Medical, Inc
Date Cleared
2016-10-06
(62 days)
Product Code
LOX
Regulation Number
870.5100Why did this record match?
Applicant Name (Manufacturer) :
OrbusNeich Medical, Inc
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Sapphire NC Plus Coronary Dilatation Catheter is indicated for:
- · balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion
- · balloon dilatation of a coronary artery occlusion for the treatment of acute myocardial infarction
- · in-stent restenosis
- · post-delivery expansion of balloon expandable coronary stents
Note: The subject device was tested on the bench with the OrbusNeich Blazer Cobalt-Chromium (CoCr) All stents should be deployed in accordance with the manufacturer's indications and instructions for use.
Device Description
The Sapphire NC Plus coronary dilatation catheter is a percutaneous transluminal coronary angioplasty (PTCA) balloon catheter with a working length of 140cm. The proximal shaft is a polymer coated stainless steel hypotube. Lubricious coatings are applied to the distal section. The non-compliant balloons, available in diameters from 2.0-4.0mm and lengths from 8-18mm, can be inflated by injecting dilute contrast media solution through the trailing hub of the catheter. Two radiopaque platinum marker bands are located within the balloon segment. The catheter is compatible with 5F or larger guiding catheters. The internal lumen of the catheter accepts a standard 0.014 inch PTCA guidewire. The proximal part of the guidewire enters the catheter tip and advances coaxially out the catheter proximal port, thereby allowing both coaxial guidance and rapid exchange of catheters with a single standard length guidewire. Two marked sections are located on the hypotube shaft to indicate catheter position relative to the tip of either a brachial or femoral guiding catheter. The design of this dilatation catheter does not incorporate a lumen for distal dye injections or distal pressure measurements. This Special 510(k) describes the modification to the Sapphire NC PTCA catheters including a change to the hydrophilic coating and also a change in the brand name to Sapphire NC Plus.
Ask a Question
K Number
K103808Device Name
SAPPHIRE NC
Manufacturer
ORBUSNEICH MEDICAL, INC.
Date Cleared
2011-09-01
(246 days)
Product Code
LOX
Regulation Number
870.5100Why did this record match?
Applicant Name (Manufacturer) :
ORBUSNEICH MEDICAL, INC
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Sapphire NC Coronary Dilatation Catheter is indicated for:
• balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion
• balloon dilatation of a coronary artery occlusion for the treatment of acute myocardial infarction
• in-stent restenosis
• post-delivery expansion of balloon expandable coronary stents
Note: The subject device was tested on the bench with the OrbusNeich Blazer Cobalt-Chromium (CoCr) Alloy Stent. All stents should be deployed in accordance with the manufacturer's indications and instructions for use.
Device Description
The Sapphire NC coronary dilatation catheter is a percutaneous transluminal coronary angioplasty (PTCA) balloon catheter with a working length of 140cm. The proximal shaft is a polymer coated stainless steel hypotube. Lubricious coatings are applied to the distal section. The non-compliant balloons, available in diameters from 2.0-4.0mm and lengths from 8-18mm, can be inflated by injecting dilute contrast media solution through the trailing hub of the catheter. Two radiopaque platinum marker bands are located within the balloon segment. The catheter is compatible with 5F or larger guiding catheters. The internal lumen of the catheter accepts a standard 0.014 inch PTCA guidewire. The proximal part of the guidewire enters the catheter tip and advances coaxially out the catheter proximal port, thereby allowing both coaxial guidance and rapid exchange of catheters with a single standard length guidewire. Two marked sections are located on the hypotube shaft to indicate catheter position relative to the tip of either a brachial or femoral guiding catheter. The design of this dilatation catheter does not incorporate a lumen for distal dye injections or distal pressure measurements.
Ask a Question
K Number
K103657Device Name
SAPPHIRE
Manufacturer
ORBUSNEICH MEDICAL, INC.
Date Cleared
2011-08-09
(237 days)
Product Code
LOX
Regulation Number
870.5100Why did this record match?
Applicant Name (Manufacturer) :
ORBUSNEICH MEDICAL, INC
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Sapphire Coronary Dilatation Catheter is indicated for:
- balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion
- balloon dilatation of a coronary artery occlusion for the treatment of acute myocardial infarction
Device Description
The Sapphire coronary dilatation catheter is a percutaneous transluminal coronary angioplasty (PTCA) balloon catheter with a working length of 140cm. The proximal shaft is a polymer coated stainless steel hypotube. Lubricious coatings are applied to the distal section. The semi-compliant balloons, available in diameters from 1.5-4.0mm and lengths from 10-30mm, can be inflated by injecting dilute contrast media solution through the trailing hub of the catheter. Two radiopaque platinum marker bands are located within the balloon segment with the exception of balloon diameters less than 2.0mm which incorporate a centrally positioned single marker band. The catheter is compatible with 5F or larger guiding catheters. The internal lumen of the catheter accepts a standard 0.014 inch PTCA guidewire. The proximal part of the guidewire enters the catheter tip and advances coaxially out the catheter proximal port, thereby allowing both coaxial guidance and rapid exchange of catheters with a single standard length guidewire. Two marked sections are located on the hypotube shaft to indicate catheter position relative to the tip of either a brachial or femoral guiding catheter. The design of this dilatation catheter does not incorporate a lumen for distal dye injections or distal pressure measurements.