Search Results
Found 2 results
510(k) Data Aggregation
(569 days)
K3 is intended to supply power to and serve as a base for dental devices and accessories. This product includes a dental chair. The dental treatment unit is intended for use in the dental clinic environment and is used by trained dentists and/or dental assistants.
K3, Dental Unit and Chair, is designed for dental treatment of children and adults. It is a dental treatment unit in accordance with ISO 7494-1, ISO 7494-2 and ISO 6875. The 3way syringe is a dental instrument in accordance with EN 1639. It aids the dental application in the mouth of the patient by supplying air, water or spray. This product is designed for use in dentistry only and may only be used by trained medical personnel K3 is similar to other commercially available products based on the intended use, the technology used, the claims, the electrical power and performance characteristics. It is substantially equivalent in design, function and intended use to the predicate devices.
This FDA 510(k) summary describes the K3 dental unit, which is a dental operative unit and accessories, including a dental chair. The submission aims to establish substantial equivalence to a predicate device, the Sirona C8 (K983242).
Here's the breakdown of the acceptance criteria and study information:
Acceptance Criteria and Reported Device Performance
The document does not explicitly state acceptance criteria in terms of specific performance metrics (e.g., sensitivity, specificity, accuracy) that the device must meet for a particular clinical task. Instead, it focuses on demonstrating substantial equivalence to a predicate device by comparing technical specifications, safety standards, and intended use. The "reported device performance" is largely framed within compliance with various international standards and the comparison table.
Table of Acceptance Criteria and Reported Device Performance (as inferred from the document):
| Acceptance Criteria (Inferred) | Reported Device Performance |
|---|---|
| 1. Intended Use Equivalence | K3 is intended to supply power to and serve as a base for dental devices and accessories, including a dental chair, for use in a dental clinic by trained dentists and/or dental assistants. This is compared directly to the predicate device's intended use. |
| 2. Technical Characteristics Equivalence | Power and Utility Supply: AC 100-120/220-240V, 50/60Hz, compressed air and water (Predicate: 110V/220V AC, compressed air and water). Protection Class: Class 1 equipment (Same as predicate). Applied parts: Type B (Same as predicate). Air Pressure: 500kPa(min)/750kPa(max) (Predicate: 550 kPa/750 kPa). Water Pressure: 250kPa(min)/600 kPa(max) (Same as predicate). Water System: City water supply (Predicate: User may select self-contained or city water supply). Patient load: Max. 135kg (Same as predicate). Chair Height: Max. 790±10mm, Min. 450±10mm (Predicate: Max.755mm, Min. 395mm). Back Rest: 0° to 66° (Predicate: 0° to 60°). Head Rest: -10° to 45° (Same as predicate). |
| 3. Electrical Safety Compliance | Complies with IEC 60601-1 (Same as predicate). |
| 4. EMC Compliance | Complies with IEC 60601-1-2, EN55011, EN 61000-3-2, EN 61000-3-3 (Same as predicate on IEC 60601-1-2). |
| 5. Biocompatibility (for contacting components) | Components (upholstery, waterlines, airlines) tested for cytotoxicity (ISO 10993-5) and found acceptable. Considered external communicating device for <24 hours. |
| 6. Cleaning and Sterilization Validation (for relevant parts) | 3-Way syringe components tested to validate steam sterilization process and acceptable sterility (ISO 17665-1:2006, ISO 17665-2). Waterlines cleaning/disinfection validated (ISO 16954:2015, ISO 19458:2006, ISO/TS 11080:2009 for microbiological simulation and physical/chemical tests). |
| 7. Software Verification and Validation | Conducted in accordance with FDA guidance for "moderate" level of concern software (ISO 14971, IEC 62304, IEC 60601-1). |
| 8. Usability Engineering | Conducted in accordance with EN/IEC 60601-1-6 and EN/IEC 62366. |
| 9. Mechanical Safety and Performance | Complies with IEC 60601-1, ISO 7494-1, ISO 7494-2, ISO 6875, IEC 80601-2-60. Hydraulic motor durability test to ISO 13485. |
Study Details:
This document is a 510(k) summary for a Class I medical device (Dental Operative Unit and Accessories). For such devices, clinical studies and elaborate performance evaluations focusing on diagnostic accuracy (like those for AI/ML devices) are typically not required. The primary mechanism for clearance is demonstrating substantial equivalence to a legally marketed predicate device through non-clinical performance testing and comparison of technical specifications and intended use.
Therefore, many of the requested details about "acceptance criteria and the study that proves the device meets the acceptance criteria" in the context of an AI/ML device (e.g., sample sizes for test sets, ground truth establishment, MRMC studies) are not applicable to this submission.
Here are the answers based on the provided text, indicating "Not Applicable" where appropriate for this type of device submission:
-
A table of acceptance criteria and the reported device performance: See the table above based on inferred criteria and the comparison tables in the document.
-
Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
- Sample Size for Test Set: Not applicable. The document describes non-clinical engineering and safety compliance testing (e.g., electrical safety, EMC, biocompatibility, sterilization, cleaning validation) rather than testing on a "test set" of patient data as would be used for an AI/ML diagnostic device. These tests are performed on either components or the assembled device.
- Data Provenance: Not applicable. The tests are laboratory-based and standard compliance tests. The manufacturer is Osstem Implant Co. Ltd. from Seoul, Republic of Korea.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
- Number of Experts: Not applicable. Ground truth as typically defined for diagnostic performance is not relevant here. Compliance with standards and technical specifications is verified by testing facilities and potentially by internal engineering teams.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Adjudication method: Not applicable.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- MRMC Study: No. This is a physical dental unit, not an AI/ML diagnostic or assistive device that would involve human readers.
-
If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Standalone Performance: No. This is not an algorithm; it's a dental chair and associated components.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Type of Ground Truth: Not applicable. The "ground truth" here is compliance with established international standards (e.g., IEC, ISO, EN) for safety, electrical characteristics, biocompatibility, cleaning, and sterilization.
-
The sample size for the training set:
- Sample Size for Training Set: Not applicable. This device does not involve a training set as it is not an AI/ML algorithm.
-
How the ground truth for the training set was established:
- Ground Truth for Training Set: Not applicable.
Ask a specific question about this device
(289 days)
The US SA Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and final or temporary abutment support for fixed bridgework. It is intended for delayed loading. The abutment is intended for use with a dental implant fixture to provide support for prosthetic restorations such as bridges, or overdenture.
The US SA Implant System is a dental implant made of titanium metal intended to be surgically placed in the bone of the upper or lower jaw arches. Fixture is made of pure titanium metal and supplied sterile. The surface is SA, Sandblasting and Acid etching, treated. The US Multi Angled Abutment is device made of titanium alloy intended for use as an aid in prosthetic restoration.
The provided text describes a 510(k) premarket notification for the "US SA Implant System," a dental implant device. The document focuses on demonstrating substantial equivalence to previously marketed predicate devices, rather than establishing acceptance criteria or reporting performance from a clinical study for a new device on its own merits against specific criteria.
Therefore, many of the requested details about acceptance criteria, specific device performance numbers, sample sizes for test/training sets, expert qualifications, adjudication methods, MRMC studies, or standalone algorithm performance are not directly applicable or available in this type of submission document.
However, I can extract information related to the non-clinical testing performed to support the substantial equivalence claim.
Here's a breakdown of the information that can be gleaned from the document based on your request:
1. Table of Acceptance Criteria and Reported Device Performance:
This information is not provided in the document in the format of specific numerical acceptance criteria and corresponding device performance metrics. The document instead focuses on demonstrating substantial equivalence through comparison of materials, design, indications for use, and non-clinical testing results against predicate devices. The fatigue testing was conducted according to a guidance document, implying the device met the requirements outlined in that guidance, but specific numerical thresholds and results are not detailed.
2. Sample size used for the test set and the data provenance:
- Test set sample size: Not specified. The document states "fatigue testing was conducted." The number of devices or iterations in this testing is not mentioned.
- Data provenance: The testing was conducted by the manufacturer, OSSTEM IMPLANT Co., Ltd., likely in a laboratory setting. The country of origin for this specific testing is implied to be Republic of Korea, where the company is based. The nature of the testing is non-clinical (laboratory).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable. This relates to clinical studies and human interpretation of data. The submission focuses on non-clinical engineering testing (fatigue, sterilization).
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable. This relates to clinical studies and human interpretation of data.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, an MRMC study was not done. This type of study is for comparing human performance, often in diagnostic imaging, with and without AI. This submission is for a dental implant, which is a physical device, and does not involve AI.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Not applicable. This device is a physical dental implant, not an algorithm or AI system.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- For the fatigue testing, the "ground truth" would be the engineering standards and requirements for mechanical strength and durability of dental implants, aligned with ISO standards and FDA guidance. The determination of "success" or "failure" would be based on whether the device passed these mechanical tests without fracture or significant deformation.
- For sterilization validation, the "ground truth" is established by adherence to recognized international standards (ISO 11137 series for gamma sterilization and ISO 17665 series for steam sterilization), demonstrating acceptable sterility assurance levels.
8. The sample size for the training set:
- Not applicable. There is no "training set" as this is a physical device, not an AI model or a system that requires machine learning training.
9. How the ground truth for the training set was established:
- Not applicable. As above, there is no training set for this type of device submission.
Summary of Non-Clinical Testing Performed (from the document):
- Fatigue testing: Conducted according to the "Guidance for industry and FDA staff Class II Special Controls Guidance Document Root-form Endosseous Dental Implants and Endosseous Dental Abutment." This testing confirms the mechanical strength and durability of the implant and abutment components. The document indicates that a fatigue test was specifically performed for the US Multi Angled Abutment due to differences in connection structure and shape compared to a predicate, to confirm its strength.
- Gamma Sterilization Validation: Performed with the Fixture System according to ISO 11137-1, ISO 11137-2, and ISO 11137-3.
- Steam Sterilization Validation: Provided according to ISO 17665-1 and ISO 17665-2.
- Biocompatibility and Surface Coating Characterization: The document states that the subject devices are made of the same materials, manufacturing process, chemical composition, and body contact as predicate devices (TS Fixture System, K121995 and Multi Angled Abutment, K123755), thus demonstrating substantial equivalence in these areas.
In conclusion, this 510(k) submission primarily relies on demonstrating substantial equivalence to predicate devices through comparisons of design, materials, indications for use, and adherence to recognized non-clinical testing standards (fatigue, sterilization) rather than presenting a clinical study with detailed acceptance criteria and performance statistics for a completely novel device.
Ask a specific question about this device
Page 1 of 1