Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    Powder Free Latex Patient Exam Glove, Smooth and Textured Natural Color (Off White) with Protein Labeling Claim (50 µg/dm² or less of water soluble protein)

    AI/ML Overview

    Acceptance Criteria and Device Performance for Powder Free Latex Patient Exam Glove

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicAcceptance Criteria (Standard)Reported Device Performance
    DimensionASTM Standard D3578Meets
    Physical PropertiesASTM Standard D3578Meets
    Freedom from Pinholes21 CFR 800.20; ASTM D5151Meets
    Powder ResidualASTM Standard D6124 (specifically, < 2 mg per glove)Meets (Results generated values below 2 mg of residual powder)
    BiocompatibilityPrimary Skin Irritation in rabbits (ISO 10993-10); Dermal Sensitization in the guinea pig (ISO 10993-10)Gloves are non-irritating; Gloves do not display any potential for sensitization
    Residual Protein LevelsASTM Standard D5712 (specifically, ≤ 50 µg/dm² of water soluble protein)Meets (Protein Labeling Claim: 50 µg/dm² or less of water soluble protein)

    2. Sample Size and Data Provenance

    The document does not explicitly state the specific sample sizes used for each test. However, the tests are conducted according to recognized ASTM and ISO standards, which typically specify appropriate sample sizes for such evaluations.

    • Data Provenance: Not explicitly stated, but assumed to be from laboratory testing conducted to meet the defined standards. The submission is for a Traditional 510(k), implying testing was performed to demonstrate substantial equivalence.

    3. Number of Experts and Qualifications for Ground Truth

    Not applicable. This device is a medical glove, and its performance is evaluated against engineering and biological standards, not through expert-driven image or data interpretation.

    4. Adjudication Method

    Not applicable. Performance is based on objective laboratory testing against established standards.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. This device is a medical glove, and its performance is evaluated against engineering and biological standards, not typically through MRMC studies.

    6. Standalone Performance (Algorithm Only)

    Not applicable. This is a physical device (medical glove), not an algorithm or software.

    7. Type of Ground Truth Used

    The ground truth for this device is based on established industry standards and regulatory requirements. This includes:

    • ASTM Standards: D3578 (dimensions, physical properties), D5151 (freedom from pinholes), D6124 (powder residual), D5712 (residual protein levels).
    • ISO Standard: 10993-10 (biocompatibility - irritation and sensitization).
    • CFR Regulations: 21 CFR 800.20 (freedom from pinholes).

    8. Sample Size for the Training Set

    Not applicable. As a physical medical device, there is no "training set" in the context of an AI or machine learning model. Performance is evaluated through laboratory testing of manufactured gloves.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. There is no training set for this device. The "ground truth" for evaluating the device's performance is defined by the requirements of the internationally recognized and regulatory standards listed above.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1