    K Number
    K172677
    Device Name
    Advantage Plus Pass-Thru
    Manufacturer
    Medivators Inc.
    Date Cleared
    2017-12-19

    (105 days)

    Product Code
    FEB
    Regulation Number
    876.1500
    Why did this record match?
    Applicant Name (Manufacturer) :

    Medivators Inc.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
    Intended Use
    Advantage Plus Pass-Thru Endoscope Reprocessing System tests, washes, disinfect and rinses endoscopes, such as fiberoptic and video endoscopes between patient uses. The Advantage Plus Pass-Thru system is indicated to provide high level disinfection of heat sensitive semi-critical endoscopes and related accessories. Manual cleaning of endoscopes is required prior to disinfection in the Advantage Plus Pass-Thru System. The Advantage Plus Pass-Thru Endoscope Reprocessing System uses Rapicide PA High Level Disinfectant to provide high level disinfection of endoscopes when used according to the directions for use. The system uses Intercept Plus Detergent in its washing cycle at a concentration of 0.5%. Rapicide PA Test Strips are used after the disinfection cycle to ensure that the used disinfectant is above the minimum recommended concentration (MRC) of 850 ppm peracetic acid; this ensures that the disinfectant was above MRC during the entire disinfection cycle. Rapicide PA should be used under the following contact conditions -Contact Time: 5 minutes Temperature: 30°C MRC: 850ppm
    Device Description
    The Advantage Plus Pass-Thru is an electro-mechanical Automatic Endoscope Reprocessor (AER) intended to high-level disinfect pre-cleaned endoscopes and their related accessories by exposing them to a peracetic acid based disinfectant – Rapicide PA disinfectant solution (cleared under 510k – K082988). The subject device is closely related to its predicate the Advantage Plus Endoscope Reprocessing System cleared in 510k – K082988 and K102996 but differs in that it is designed in a "pass-thru" configuration. The subject device is intended to be built into a wall which separates 'clean' and 'dirty' sides. It provides the ability to load an endoscope from the dirty side and unload the disinfected endoscope from the clean side, thus preventing potential mix-up between high level disinfected and contaminated endoscopes. Contaminated endoscopes are placed in the subject device from the 'dirty' side and appropriate connections to the endoscope channels are made. The system is designed to test for proper channel connectivity, channel blockages and leaks in the outer endoscope sheath. After the endoscope passes these tests, the system proceeds to the washing cycle with Intercept Plus detergent, and the highlevel disinfection cycle. In the high-level disinfection cycle, Rapicide PA HLD is diluted by the incoming water to a minimum concentration of 850 ppm of peracetic acid. The system measures and maintains the temperature of this disinfectant solution to ensure that it is above the minimum recommended temperature of 30℃ for Rapicide PA for effective High Level Disinfection to occur. Following the 5 minute disinfectant contact time, the user takes a sample to test for MRC (850 ppm of peracetic acid). Once the MRC test passes, the disinfectant is emptied from basin and drained out of the AER. The endoscope is then rinsed, dried and removed from the 'clean side' of the AER.
    Ask a Question
    K Number
    K162128
    Device Name
    Scope Buddy Plus Endoscope Flushing Aid
    Manufacturer
    MEDIVATORS INC.
    Date Cleared
    2016-12-21

    (142 days)

    Product Code
    FEB
    Regulation Number
    876.1500
    Why did this record match?
    Applicant Name (Manufacturer) :

    MEDIVATORS INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
    Intended Use
    Scope Buddy Plus Endoscope Flushing Aid is an electro-mechanical to pump fluids through channels of flexible, immersible endoscopes during the endoscope manual cleaning process.
    Device Description
    The subject device is intended to provide fluid delivery to endoscope channels in the same manner as a manual syringe would be used during the manual cleaning phase of endoscope reprocessing as defined by the endoscope manufacturer's instructions. The subject device utilizes an external peristaltic pump and 24-hour multi-use connection tubing (tubing used for 24 hours with multiple endoscopes and then discarded) to deliver specific volumes of detergent solution and rinse water to the endoscope channels during the manual cleaning phase. These fluid volumes are defined by the endoscope manufacturer. Scope Buddy Plus is intended to be used only during the manual cleaning phase of endoscope reprocessing, in conjunction with the instructions and labeling provided by the endoscope manufacturer, to deliver an endoscope manufacturer specified volume of fluid to the endoscope channels. The subject device is not an endoscope washer-disinfector and does not bear labeling claims for direct cleaning efficacy or high-level disinfection efficacy for endoscope reprocessing.
    Ask a Question
    K Number
    K152394
    Device Name
    Rapicide PA High-Level Disinfectant Test Strips
    Manufacturer
    MEDIVATORS INC.
    Date Cleared
    2016-03-11

    (200 days)

    Product Code
    JOJ
    Regulation Number
    880.2800
    Why did this record match?
    Applicant Name (Manufacturer) :

    MEDIVATORS INC.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
    Intended Use
    The Rapicide PA High-Level Disinfectant Test Strips are used as a chemical indicator after the disinfection cycle to ensure that the Rapicide PA High-Level Disinfectant Solution is above minimum recommended concentration (850ppm peracetic acid); this ensures the disinfectant was above MRC during the entire disinfection cycle.
    Device Description
    Rapicide PA High-Level Disinfectant Test Strips have the ability to measure the disinfectant use solution concentration above 850ppm PAA. This is the minimum recommended concentration (MRC) of PAA for Rapicide PA high level disinfectant. If the solution is at or below MRC, the test strip pad will indicate a failure by turning dark grey, violet grey, light grey, or white (no change in color). A passing result will be indicated by the test strip pad turning solid black color.