Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K111413
    Date Cleared
    2012-01-30

    (255 days)

    Product Code
    Regulation Number
    878.4460
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    MOLNLYCKE HEALTHCARE, US LLC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Biogel® PI Indicator® Underglove is a disposable device made of polyisoprene, blue in color, that is intended to be worn on the hands, usually in a surgical setting, to provide a barrier against potentially infectious material and other contaminants.

    Device Description

    The proposed device, the Biogel PI Indicator Underglove is manufactured of polyisoprene colored with blue pigmentation. The Biogel PI Indicator Underglove is manufactured of the exact same material and coated with the Biogel® Coating which is used on the currently cleared device that has been legally marketed by Molnlycke Health Care for many years with the addition of a surfactant.

    AI/ML Overview

    This describes the Biogel® PI Indicator® Underglove, a surgical glove. As a Class I device, it is subject to general controls and does not require extensive clinical studies or complex acceptance criteria typically seen with higher-risk medical devices or AI-powered diagnostics.

    Here's an analysis based on the provided document:

    Acceptance Criteria and Reported Device Performance

    CharacteristicAcceptance Criteria (Standard)Reported Device Performance
    DimensionsMeets ASTM D3577Meets ASTM D3577
    Physical PropertiesMeets ASTM D3577Meets ASTM D3577
    Freedom from HolesMeets ASTM D3577Meets ASTM D3577
    BiocompatibilityMeets ISO 10993-1Meets ISO 10993-1
    LAL Test ResultsMeets ASTM D7102Meets ASTM D7102

    Note: The document explicitly states, "No clinical data was required." This indicates that the device's performance was evaluated against established standards for surgical gloves, rather than through
    comparative clinical trials with human subjects.

    1. Sample size used for the test set and the data provenance:

    • Sample Size: The document does not specify the exact sample size used for each test (Dimensions, Physical Properties, Freedom from Holes, Biocompatibility, LAL Test). It only states that the device "Meets" the respective ASTM and ISO standards. These standards themselves specify sampling plans and test methodologies.
    • Data Provenance: Not specified. The testing was likely conducted in a lab setting by the manufacturer or a contracted lab to confirm compliance with the standards.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This is not applicable as the evaluation is based on objective physical and chemical testing against established engineering standards (ASTM, ISO), not on expert subjective assessment or medical diagnosis.

    3. Adjudication method for the test set:

    This is not applicable. The assessment is based on objective measurements against pass/fail criteria defined by the ASTM and ISO standards. There is no subjective interpretation requiring adjudication.

    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable. This device is a surgical glove, not an AI-powered diagnostic or decision-support system. Therefore, no MRMC comparative effectiveness study was performed or is relevant.

    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    This is not applicable. This device is a physical product (surgical glove), not a software algorithm.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" used for this device's acceptance is adherence to established industry standards and regulatory requirements for surgical gloves (ASTM D3577, ISO 10993-1, ASTM D7102) via objective physical and chemical testing. These standards define the acceptable performance characteristics for devices of this type.

    7. The sample size for the training set:

    This is not applicable. This is a physical medical device, not a machine learning model that requires a training set.

    8. How the ground truth for the training set was established:

    This is not applicable for the same reason as above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K112089
    Date Cleared
    2011-08-24

    (34 days)

    Product Code
    Regulation Number
    878.4460
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    MOLNLYCKE HEALTHCARE, US LLC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Biogel® OrthoPro® Brown Surgical Glove is a disposable device made of natural rubber latex that is intended to be worn on the hands, usually in a surgical setting, to provide a barrier against potentially infectious material and other contaminants.

    Device Description

    The proposed device, the Biogel® OrthoPro® Brown Surgical Glove is manufactured of natural rubber latex colored with brown nigmentation. The Biogel® OrthoPro® Brown Surgical Glove is manufactured of the exact same material and coated with the Biogel® Coating which is used on the currently cleared device(s) that have been legally marketed by Mölnlycke Health Care for many years with the addition of the brown colorant. The glove former design used in the manufacture of this glove allows the Biogel® OrthoPro® Brown Surgical Glove to provide the user with this additional feature: Slightly Curved Former with Independent/Displaced thumb.

    AI/ML Overview

    The document describes the Biogel® OrthoPro® Brown Surgical Glove.

    1. Table of Acceptance Criteria and Reported Device Performance:

    TestAcceptance Criteria (requirements from ASTM D3577, 21 CFR 800.20, and ASTM D5712)Reported Device Performance (Result)
    Primary Skin IrritationNot explicitly stated (implied non-irritating)Gloves are non-irritating.
    ISO Closed Patch SensitizationNot explicitly stated (implied no sensitization potential)Gloves do not display any potential for sensitization.
    DimensionsRequirements of ASTM D3577Gloves meet requirements of ASTM D3577.
    Physical PropertiesRequirements for rubber surgical gloves per ASTM D3577Gloves meet requirements for rubber surgical gloves per ASTM D3577.
    Freedom from HolesRequirements of 21 CFR 800.20 and ASTM D3577Gloves exceed the requirements of 21 CFR 800.20 and ASTM D3577.
    Protein ExtractablesRequirements of ASTM D5712Gloves meet requirements of ASTM D5712.
    Powder ResidualPowder level requirements for "Powder-Free" designation per ASTM D3577 (using ASTM standard D6142, presumably
    Ask a Question

    Ask a specific question about this device

    K Number
    K071700
    Date Cleared
    2007-09-11

    (82 days)

    Product Code
    Regulation Number
    878.4460
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    MOLNLYCKE HEALTHCARE, US LLC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Biogel® Sensor™ Surgeon's Glove is a sterile disposable device made of natural rubber latex that is intended to be worn on the hands, usually in a surgical setting, to provide a barrier against potentially infectious material and other contaminants.

    Device Description

    The Biogel® Sensor™ Surgeon's Glove is a sterile powder free, polymer coated latex surgical glove. The glove contains 50 micrograms or less of total water extractable protein per gram.

    AI/ML Overview

    The provided document is a 510(k) summary for the Biogel® Sensor™ Surgeon's Glove. This device is a Class I surgical glove, which is a low-risk device. As such, the "study" demonstrating its performance is primarily based on meeting established industry standards for surgical gloves, rather than complex clinical trials or AI-specific validation studies.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicAcceptance Criteria (Standard)Reported Device Performance
    DimensionsMeets ASTM D3577Meets ASTM D3577
    Physical PropertiesMeets ASTM D3577Meets ASTM D3577
    Freedom from HolesMeets ASTM D3577Meets ASTM D3577
    BiocompatibilityMeets ISO 10993-1Meets ISO 10993-1
    LAL Test - Endotoxin Concentration≤ 0.50 EU/ml≤ 0.50 EU/ml
    Total Water Extractable Protein≤ 50 micrograms/gram(Implicitly meets, as it's part of the device name and claim)

    Note: The document states "The Biogel® Sensor™ Surgeon's Glove meets the technological characteristics of ASTM D3577 and is substantially equivalent to the predicate devices cited in this 510(k) summary." For protein content, the device name explicitly includes "Protein Content Labeling Claim (50 micrograms or less)," indicating it meets this claim.

    2. Sample size used for the test set and the data provenance

    The document does not specify general "test set" sample sizes or data provenance (e.g., country of origin, retrospective/prospective). For surgical gloves, compliance with ASTM and ISO standards typically involves standardized testing of batches of gloves, but the specific sample sizes for these tests are not provided in this summary.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This is not applicable. For surgical gloves, the "ground truth" is adherence to established engineering and material science standards (ASTM D3577, ISO 10993-1) and specific chemical limits (LAL test, protein content) rather than expert interpretation of a clinical dataset. These standards define objective pass/fail criteria.

    4. Adjudication method for the test set

    Not applicable. As noted above, the evaluation is based on objective measurements against established standards, not on expert adjudication of subjective assessments.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a medical device (surgical glove), not an AI algorithm or an imaging/diagnostic tool requiring human reader studies.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device (surgical glove), not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this device is based on objective measurements against established national and international standards for surgical gloves (ASTM D3577 and ISO 10993-1) and specific chemical assays (LAL test for endotoxins, protein content for extractable proteins). It's a materials science and manufacturing quality ground truth, not a clinical diagnostic or pathology ground truth.

    8. The sample size for the training set

    Not applicable. This is a physical medical device, not an AI model that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. As there's no AI model or training set, this question is not relevant.

    Ask a Question

    Ask a specific question about this device

    K Number
    K071465
    Date Cleared
    2007-09-11

    (105 days)

    Product Code
    Regulation Number
    878.4460
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    MOLNLYCKE HEALTHCARE, US LLC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Biogel Orthopedic Surgeon's Glove is a disposable device made of natural rubber latex that is intended to be worn on the hands, usually in a surgical setting, to provide a barrier against potentially infectious material and other contaminants.

    Device Description

    The Biogel Orthodepic Surgical Glove is a sterile powder free, polymer coated latex surgical glove. The glove contains 50 micrograms or less of total water extractable protein per gram.

    AI/ML Overview

    The provided document is a 510(k) summary and FDA clearance letter for a medical device: the Biogel Orthopedic Surgical Glove. This type of submission is for demonstrating substantial equivalence to a predicate device, not for proving novel clinical effectiveness with extensive studies. As such, many of the requested elements (like sample sizes for test sets, expert qualifications, adjudication methods, MRMC studies, or training set details) are not relevant or typically included in such filings for this class of device.

    Here's a breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    CharacteristicAcceptance Criteria (Standard)Reported Device Performance
    DimensionsMeets ASTM D3577Meets ASTM D3577
    Physical PropertiesMeets ASTM D3577Meets ASTM D3577
    Freedom from HolesMeets ASTM D3577Meets ASTM D3577
    BiocompatibilityMeets ISO 10993-1Meets ISO 10993-1
    LAL Test (Endotoxin)
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1