Search Results
Found 2 results
510(k) Data Aggregation
(279 days)
The Medcomp® Vessel Dilators are designed for percutaneous entry into a vessel in order to enlarge the opening of the vessel for the placement of a catheter in a vein or artery.
Medcomp®s Vessel Dilators are designed for percutaneous entry into a vessel in order to enlarge the opening of the vessel for the placement of a catheter into a vein or artery. The Vessel Dilators range in size from 4F to 24F with lengths of either 15cm (6.0 inches) or 20cm (8.0 inches), are either single or dual taper, and can use guidewire sizes .025", .035", and .038". The Medcomp® Vessel dilators are offered in two different tubing materials (polyethylene or polypropylene) attached to an HDPE hub. Vessel dilators are supplied sterile, for use in an aseptic technique and can only be used once.
The provided text is a 510(k) summary for the Medcomp Vessel Dilator, a medical device for percutaneous entry into a vessel to enlarge the opening for catheter placement. This document focuses on demonstrating substantial equivalence to predicate devices rather than proving the device meets specific performance acceptance criteria through a clinical study involving AI.
Therefore, most of the requested information regarding acceptance criteria, device performance, sample sizes, expert involvement, adjudication methods, MRMC studies, standalone performance, and training set details for an AI-powered device is not applicable or present in this document.
However, I can extract information related to the bench/performance data/non-clinical testing that demonstrates the device's substantial equivalence to its predicate and reference devices.
Here's the relevant information based on the provided text:
1. A table of acceptance criteria and the reported device performance
The document does not present specific acceptance criteria in a table format with corresponding reported performance values for an AI device. Instead, it states that "The proposed device, Vessel Dilators, meets the performance criteria of design verification as specified by ISO standards, guidance documents, and internal test protocols." It also mentions "The results of performance testing, in conjunction with the substantial equivalence claims, effectively demonstrate the proposed device, Medcomp® Vessel Dilator, is equivalent to the reference device, Cook® Dilator, preamendment."
The performance data mentioned includes:
- Compliance with ISO 11070:2014 (Sterile single-use intravascular introducers, dilators and guidewires)
- Compliance with ISO 10555-1:2013 (Sterile, single-use intravascular catheters - part 1: general requirements) - described as "Bench Testing"
- Compliance with AAMI / ANSI / ISO 10993-1:2009/(R) 2013 (Biological evaluation of medical devices -- Part 1: evaluation and testing within a risk management process) - described as "Biocompatibility"
- Compliance with ISO 11135:2014 (Sterilization of health-care products - ethylene oxide - requirements for the development, validation and routine control of a sterilization process for medical devices) - described as "Sterility"
- Compliance with AAMI / ANSI / ISO 10993-7:2008(R) 2012 (Biological evaluation of medical devices - part 7: ethylene oxide sterilization residuals) - described as "Sterility"
- Compliance with ISO 14971:2007 (Medical devices - application of risk management to medical devices)
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
This information is not provided. The document describes non-clinical performance testing and biocompatibility studies, not a test set related to AI performance.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not applicable as the document is not about an AI device or a study requiring expert ground truth for image interpretation.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not applicable. The document is about a physical medical device (vessel dilator) and its substantial equivalence, not an AI-powered image analysis tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This information is not applicable.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For the non-clinical performance testing, the "ground truth" would be the established specifications and requirements outlined in the referenced ISO standards and internal test protocols. For biocompatibility, it's the biological response against predefined safety limits. This isn't "ground truth" in the context of an AI diagnostic device.
8. The sample size for the training set
This information is not applicable.
9. How the ground truth for the training set was established
This information is not applicable.
Ask a specific question about this device
(127 days)
The Dignity® Dual Port is a power injectable implantable infusion port that is indicated for patient therapies requiring repeated access to the vascular system. The Dignity® Dual Port can be used for infusion of medications, I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples.
When used with a power injectable needle, the Dignity® Dual Port is indicated for power injection of contrast media. For power injection of contrast media, the maximum recommended infusion rate is 5 ml/s with a 19 or 20 gauge non-coring power injectable needle. The maximum recommended infusion rate is 2 ml/s with a 22 gauge non-coring power injectable needle.
The Dignity® Dual power injectable implantable infusion port is an implantable access device designed to provide repeated access to the vascular system. Port access is performed by percutaneous needle insertion using a non-coring needle. Power injection is performed using a power injectable needle only. The Dignity® Dual Port device consists of two primary components: an injection port with a self-sealing septa and a radiopaque catheter. The Dignity® Dual Ports can be identified subcutaneously by feeling the top of the septum and the top rim of the port housing. Power Injectable Implantable Infusion Ports can be identified by the letters "CT" under radiographic imaging.
The Dignity® Dual Port will be marketed in four kit configurations, which are listed below:
- 9.5F Standard Port Kit (Catalog number MRDP95ADN)
- 9.5F Port Kit with Micro-Stick (Catalog number MIDP95ADN)
- 9.5F Standard Port Kit with Silicone Filled Suture Holes (Catalog number MRDP95ADS)
- 9.5F Port kit with Silicone Filled Suture Holes and with Micro-Stick (Catalog number MIDP95ADS)
The Dignity® Dual Port is comprised of a polysulfone cap (with silicone filled suture holes, or open suture holes), two silicone septa, and a polysulfone base assembly with a titanium tube that provides a channel from the stem to the distal reservoir. The reservoir is plastic (polysulfone). The Dignity® Dual catheter locking assembly (makrolon/pellethane) locks the chronoflex lumen to the plastic stem of the Dignity® Dual Port.
The provided text describes the "Dignity® Dual Port," a power injectable implantable infusion port. The document is a 510(k) summary submitted to the FDA, demonstrating substantial equivalence to pre-existing devices. Therefore, the "study" referred to is a demonstration of equivalence through functional and material comparisons, rather than a traditional clinical study with human patients and outcomes data.
Here's the breakdown of the acceptance criteria and the study that proves the device meets those criteria, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (from Predicate/Reference Devices & Standards) | Reported Device Performance (Dignity® Dual Port - K153238) |
|---|---|
| Design: Double lumen, dual reservoir, power injection | Equivalent: Double lumen, dual reservoir, power injection |
| Dimensions: Base fits within 29mm x 39mm; Internal Volume 0.6 ml each reservoir; Depth Marking every 1cm (Predicate: K090512) | Equivalent: Base fits within 41mm x 12mm; Internal Volume: 0.7ml and 0.6 ml; Depth Marking every 1cm (Note: Dimensions are different, but deemed "Equivalent" in the context of the 510(k) summary, likely indicating functional equivalence despite size variation). |
| Indications for Use: Infusion of medications, I.V. fluids, parenteral nutrition solutions, blood products, withdrawal of blood samples; Power injection of contrast media up to 5 ml/s. (Predicate: K090512) | Equivalent: Infusion of medications, I.V. fluids, parenteral nutrition solutions, blood products, withdrawal of blood samples; Power injection of contrast media: 5 ml/s (19 or 20 gauge non-coring needle), 2 ml/s (22 gauge non-coring needle). (Additional detail for 22 gauge, but generally equivalent) |
| Catheter Size: 9.5F Double Lumen ChronoFlex Polyurethane Catheter (Predicate: K090512) | Equivalent: 9.5F Double Lumen ChoronFlex Polyurethane Catheter |
| Sterilization: ETO (Predicate: K090512) | Equivalent: ETO, ANSI/AAMI/ISO 11135-1:2007 SAL 10-6 |
| Power Injectable Rate: 5 ml/s (Predicate: K090512) | Equivalent: 5 ml/s |
| Materials: Plastic, Silicone, Polyurethane (Predicate: K090512) | Equivalent to Reference Devices (K070003, K120281): Thermoplastic Polymer, Silicone, Polyurethane, Titanium, Polycarbonate (Note: Uses different and additional materials compared to the predicate, but is deemed equivalent based on comparison to reference devices which use similar materials and have established biocompatibility.) |
| MRI Performance Testing: MRI Conditional; Power Injection (19Ga 4.9ml/s, 20Ga 4.9ml/s, 22Ga 1.9ml/s); Infusion Testing (1700ml/hr); Catheter Lock Disengagement (10 lbs); Needle Insertion (19Ga 2.88lb, 22Ga 2.05 lbs); Gravity Flow (1680 ml/hr). (Predicate: K090512) | Equivalent: MRI Conditional; Power Injection (19Ga 4.9ml/s, 20Ga 4.9ml/s, 22Ga 1.9ml/s); Infusion Testing: 1600ml/hr; Catheter Lock Disengagement: 6.9 lb; Needle Insertion: 19Ga 3.8lb, 22Ga 3.6 lbs; Gravity Flow: 1760 ml/hr. (Some values differ but are deemed equivalent by the submission.) |
| Biocompatibility: Meet ISO 10993-1 requirements for permanent implant with tissue and both indirect/direct blood contact (via reference devices K070003, K120281) | Met: Materials are equivalent to cleared reference devices (K070003, K120281) which performed biocompatibility per ISO 10993-1. Summary of prior testing confirms compliance. |
2. Sample Size Used for the Test Set and Data Provenance
This document describes a 510(k) submission, which relies on demonstrating substantial equivalence to a legally marketed predicate device rather than conducting a de novo clinical trial with a distinct human "test set."
- Test Set: No independent human "test set" in the context of a clinical trial is described. The "testing" involves bench testing, material comparisons, and a comparison matrix against a predicate and reference devices.
- Data Provenance: The data provenance is primarily from bench/performance data and comparison to previously FDA-cleared devices (predicate K090512, reference K070003, K120281). This is not data from human subjects or from a specific country of origin in the way a clinical trial would generate it. All data would be generated in a lab setting by Medcomp® or its contracted testing facilities.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
- Ground Truth Establishment: Not applicable in the context of this 510(k) summary. The "ground truth" for the current device's acceptability is established by demonstrating its equivalence to predicate devices that have already been determined safe and effective by the FDA. The performance standards are derived from established industry standards (e.g., ISO, ASTM) and the performance characteristics of the predicate device.
- Experts: No external experts are described as establishing "ground truth" for the test set. The submission itself is prepared by Medcomp®'s regulatory associate, Ms. Courtney Nix, and reviewed internally by Medcomp® and externally by the FDA (specifically the Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices).
4. Adjudication Method for the Test Set
- Adjudication Method: Not applicable. There is no "test set" of patient data requiring adjudication. The assessment is a comparison of design, materials, and performance data against established standards and predicate device specifications.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size
- MRMC Study: No, an MRMC comparative effectiveness study was not done. This type of study typically applies to diagnostic imaging devices where human readers interpret results with and without AI assistance. The Dignity® Dual Port is an implantable infusion port, a physical medical device, not an imaging or diagnostic AI system.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
- Standalone Performance: Not applicable. The device is a physical implantable infusion port, not an algorithm or AI system.
7. The Type of Ground Truth Used
- Type of Ground Truth: The "ground truth" for the submission is the established safety and effectiveness of the legally marketed predicate device (C.R. Bard PowerPort©, K090512) and reference devices (Medcomp® Pro-Fuse®, K070003; Medcomp® Dignity® Power Injectable Titanium Port, K120281), as interpreted through FDA regulations and consensus standards (ISO, ASTM). The equivalence is demonstrated through comparative analysis of specifications, materials, and bench test results against these established benchmarks.
8. The Sample Size for the Training Set
- Training Set: Not applicable. The device is a physical medical device, not an AI or machine learning model that requires a training set.
9. How the Ground Truth for the Training Set Was Established
- Ground Truth for Training Set: Not applicable, as there is no training set for this type of device submission.
Ask a specific question about this device
Page 1 of 1