Search Results
Found 1 results
510(k) Data Aggregation
(132 days)
MASTEL PRECISION, INC.
The Mastel 1 Folder™ Implantation System is intended to fold and assist in inserting the SENSAR™, model AR40e intraocular lens, ONLY into the capsular bag.
The Mastel 1 Folder is an ophthalmic surgical device designed for use in inserting a foldable intraocular lens into the eye. The materials, basic scientific concepts, physical properties and intended use of the Mastel 1 Folder are similar to those of the predicate devices.
The Mastel 1 Folder Implantation System is comprised of a medical-grade titanium handpiece, a material commonly used in reusable surgical devices. The handpiece rod is designed to accept a single-use cartridge. The handpiece is reusable and is designed to withstand conventional in-office steam autoclaving. The handpiece is provided to the customer in a plastic case in a non-sterile state.
Acceptance Criteria and Study for Mastel 1 Folder™ Implantation System
1. Table of Acceptance Criteria and Reported Device Performance
This document summarizes the safety and effectiveness for a device that is a mechanical inserter; therefore, the "acceptance criteria" are related to maintaining the physical integrity and optical properties of the intraocular lenses (IOLs) during the insertion process. The reported device performance indicates successful delivery of IOLs without degradation of their critical parameters.
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Maintenance of IOL diopter | The results revealed that all Mastel 1 Folders provided acceptable results when delivering 130 intraocular lenses, ranging from 6.0 D to 23.0 D. |
Maintenance of IOL astigmatism | The results revealed that all Mastel 1 Folders provided acceptable results when delivering 130 intraocular lenses, ranging from 6.0 D to 23.0 D. |
Maintenance of IOL resolution | The results revealed that all Mastel 1 Folders provided acceptable results when delivering 130 intraocular lenses, ranging from 6.0 D to 23.0 D. |
Maintenance of IOL visual acuity characteristics | The results revealed that all Mastel 1 Folders provided acceptable results when delivering 130 intraocular lenses, ranging from 6.0 D to 23.0 D. |
Maintenance of IOL overall diameter | The results revealed that all Mastel 1 Folders provided acceptable results when delivering 130 intraocular lenses, ranging from 6.0 D to 23.0 D. |
Maintenance of IOL loop angles | The results revealed that all Mastel 1 Folders provided acceptable results when delivering 130 intraocular lenses, ranging from 6.0 D to 23.0 D. |
Maintenance of IOL sagitta | The results revealed that all Mastel 1 Folders provided acceptable results when delivering 130 intraocular lenses, ranging from 6.0 D to 23.0 D. |
No damage to the IOL during insertion | The results revealed that all Mastel 1 Folders provided acceptable results when delivering 130 intraocular lenses, ranging from 6.0 D to 23.0 D. (Implied by "acceptable results" across multiple IOL parameters, indicating no critical damage or functional impairment occurred.) |
Effective delivery of the IOL into the capsular bag (intended use) | "Mastel 1 Folders provided acceptable results when delivering 130 intraocular lenses" within the context of an "Implantation System" intended for insertion into the capsular bag. (Implied by successful delivery and maintenance of IOL properties.) |
2. Sample size used for the test set and the data provenance
- Sample Size for Test Set: 130 intraocular lenses (IOLs), ranging from 6.0 D to 23.0 D.
- Data Provenance: The study was an in vitro performance test. The country of origin of the data is not specified, but the submission is to the U.S. FDA, suggesting the testing was conducted to U.S. standards, likely in the U.S. or by a U.S. entity. The data is prospective in the sense that it was specifically generated for the submission to demonstrate device performance.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. The study was an in vitro performance test of a mechanical device's ability to handle and deliver intraocular lenses without damaging them or altering their critical physical and optical characteristics. The "ground truth" was established by objective measurements of the IOLs post-delivery against their known specifications/tolerances. No human expert interpretation of images or clinical outcomes was involved in establishing this in vitro performance ground truth.
4. Adjudication method for the test set
Not applicable. As described above, the "ground truth" was derived from direct physical and optical measurements of the IOLs. There was no need for expert adjudication because the assessment was objective and quantitative, comparing measured parameters against acceptable ranges.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a mechanical intraocular lens implantation system, not an AI-powered diagnostic or assistive tool for human readers. Therefore, an MRMC study comparing human reader performance with and without AI assistance is irrelevant to this device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, a standalone in vitro performance test was conducted. The device (Mastel 1 Folder Insertion System) was used to deliver 130 intraocular lenses, and the IOLs were subsequently evaluated for various physical and optical parameters. This test assesses the device's performance directly and independently of any human-in-the-loop clinical use evaluation.
7. The type of ground truth used
The ground truth used was based on objective physical and optical measurements of the intraocular lenses (IOLs) after being processed by the device. These measurements (diopter, astigmatism, resolution, visual acuity characteristics, overall diameter, loop angles, and sagitta) were compared against acceptable standards or specifications for IOLs, ensuring the device did not negatively impact their functional integrity.
8. The sample size for the training set
Not applicable. This device is a mechanical tool, not an artificial intelligence or machine learning system. Therefore, there is no "training set" in the context of algorithm development.
9. How the ground truth for the training set was established
Not applicable. As this is a mechanical device and not an AI/ML system, no training set or its associated ground truth establishment methods are relevant.
Ask a specific question about this device
Page 1 of 1