510(k) Data Aggregation

    K Number
    K231957
    Device Name
    SafeBreak® Vascular
    Manufacturer
    Date Cleared
    2023-09-29

    (88 days)

    Product Code
    Regulation Number
    880.5220
    Why did this record match?
    Applicant Name (Manufacturer) :

    Lineus Medical

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
    Intended Use
    SafeBreak® Vascular is intended to separate when excessive tension is exerted across intravenous (IV) and intraosseous (IO) administration sets. When SafeBreak® Vascular separates, fluid flow is stopped from the infusion pump and blood flow is stopped from the patient's IV or IO catheter. SafeBreak® Vascular is intended to aid in reduction of IV and IO mechanical complications requiring IV and IO replacement. SafeBreak® Vascular is intended to be used on peripheral IV catheters, midlines, peripherally inserted central catheter (PICCs), central venous catheters (CVCs), IV ports and port needles, and IOs in adults and pediatric populations two (2) weeks of age and older receiving intermittent or continuous infusions with an electronic pump.
    Device Description
    SafeBreak® Vascular is placed in-line with an intravenous or intraosseous catheter and an intravascular administration set, including any administration set accessories. It separates into two parts when a specified force is applied. The device has been shown to reduce the risk of IV catheter failure(s) requiring IV catheter replacement. When SafeBreak® Vascular separates, fluid flow is stopped from the infusion pump and blood flow is stopped from the patient's IV or IO catheter. SafeBreak® Vascular is provided sterilized by Ethylene Oxide. The product is single use only and is not designed for reprocessing or re-sterilization by the user.
    Ask a Question

    Ask a specific question about this device

    K Number
    K223486
    Device Name
    SafeBreak® Vascular
    Manufacturer
    Date Cleared
    2023-05-19

    (179 days)

    Product Code
    Regulation Number
    880.5220
    Why did this record match?
    Applicant Name (Manufacturer) :

    Lineus Medical

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
    Intended Use
    SafeBreak® Vascular is intended to separate when excessive tension is exerted across a peripheral IV administration set. When SafeBreak® Vascular separates, fluid flow is stopped from the infusion pump and blood flow is stopped from the patient's IV catheter. SafeBreak® Vascular is intended to aid in reduction of peripheral IV mechanical complications requiring IV replacement. SafeBreak® Vascular is intended to be used on peripheral IV catheters in adults and pediations greater than one (1) year of age receiving intermittent or continuous infusions with an electronic pump.
    Device Description
    SafeBreak® Vascular is placed in-line with an intravenous catheter and an intravascular administration set, including any administration set accessories. It separates into two parts when a specified force is applied. The device has been shown to reduce the risk of IV catheter failure(s) requiring IV catheter replacement. When SafeBreak® Vascular separates, fluid flow is stopped from the infusion pump and blood flow is stopped from the patient's IV catheter. SafeBreak® Vascular is provided sterilized by Ethylene Oxide. The product is single use only and is not designed for reprocessing or re-sterilization by the user. The purpose of this 510(k) submission is to expand the indications for use of the product to include pediatric populations greater than one (1) year of age.
    Ask a Question

    Ask a specific question about this device

    K Number
    K222791
    Device Name
    SafeBreak Vascular
    Date Cleared
    2022-10-16

    (31 days)

    Product Code
    Regulation Number
    880.5220
    Why did this record match?
    Applicant Name (Manufacturer) :

    Lineus Medical, LLC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
    Intended Use
    SafeBreak® Vascular is intended to separate when is exerted across a peripheral IV administration set. When SafeBreak@ Vascular separates, fluid flow is stopped from the infusion pump and blood flow is stopped from the patient's IV catheter. SafeBreak® Vascular is intended to aid in reduction of peripheral IV mechanical complications requiring IV replacement. SafeBreak® Vascular is intended to be used on peripheral IV catheters in adults and adolescent populations eighteen (18) years of age and older receiving intermittent or continuous infusions with an electronic pump.
    Device Description
    SafeBreak® Vascular is placed in-line with an intravenous catheter and an intravascular administration set, including any administration set accessories. It separates into two parts when a specified force is applied. The device has been shown to reduce the risk of IV catheter failure(s) requiring IV catheter replacement. When SafeBreak® Vascular separates, fluid flow is stopped from the infusion pump and blood flow is stopped from the patient's IV catheter. SafeBreak® Vascular is provided sterilized by Ethylene Oxide. The product is single use only and is not designed for reprocessing or re-sterilization by the user. The purpose of this special 510(k) is to expand the separation force tolerance of SafeBreak® Vascular to 1-5 lbf.
    Ask a Question

    Ask a specific question about this device

    K Number
    DEN190043
    Device Name
    SafeBreak Vascular
    Date Cleared
    2021-05-27

    (617 days)

    Product Code
    Regulation Number
    880.5220
    Why did this record match?
    Applicant Name (Manufacturer) :

    d/b/a Lineus Medical

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
    Intended Use
    SafeBreak® Vascular is intended to separate when excessive tension is exerted across a peripheral IV administration set. When SafeBreak® Vascular separates, fluid flow is stopped from the infusion pump and blood flow is stopped from the patient's IV catheter. SafeBreak® Vascular is intended to aid in reduction of peripheral IV mechanical complications requiring IV replacement. SafeBreak® Vascular is intended to be used on peripheral IV catheters in adults and adolescent populations eighteen (18) years of age and older receiving intermittent or continuous infusions with an electronic pump.
    Device Description
    SafeBreak® Vascular is used with peripheral intravenous (IV) catheters and installed for in-line use with an IV administration set. It connects between the IV administration tubing and the patient's peripheral IV access device. When excessive force is applied to the IV administration set, SafeBreak® Vascular separates at a force lower than the force required to dislodge the peripheral IV catheter or disrupt the IV securement dressing (4 ± 1 1bf). When SafeBreak® Vascular separates, fluid flow is stopped from the infusion pump, and blood flow is stopped from the patient's IV catheter. SafeBreak® Vascular may be used for a maximum of seven days.