Search Results
Found 2 results
510(k) Data Aggregation
(272 days)
LUNEAU SAS
Measuring the refraction of the eye giving both lower and higher order aberrations
Measurement of the shape of the cornea
Retro-illumination imaging of the eye
Measuring the intraocular pressure without contacting the eye for glaucoma evaluation.
Photographing the eye and taking images of the eye to evaluate the thickness of the central cornea.
Full corneal thickness map.
Scheimpflug imaging.
Anterior chamber imaging.
Pupil image.
Image of the cornea relative to the iris.
The VX130 is a multifunctional ophthalmic diagnostic device.
The VX130 combined wavefront aberrometer, corneal topographer, cataract screening device, Scheimpflug pachymeter, and non-contact tonometer is a single platform that contains five different measurement units.
The wavefront aberrometer works on the Shack-Hartmann principle and is used as an advanced autorefractometer that measures both lower and higher order aberrations of the eye. Retro illumination is used to image ocular opacities. The corneal topographer uses a Placido disk to measure keratometry and the detailed shape of the cornea. A linear scanning Scheimpflug pachymeter measures the thickness of the cornea by illuminating it with a slit of light and photographing it using the Scheimpflug technique. Anterior and posterior corneal shape are also measured from the Scheimpflug images. An air puff non-contact tonometer is included for measurement of the intraocular pressure. The device is fully automated and a number of different measurements can be performed by a single command including alignment and focusing.
Here's an analysis of the acceptance criteria and study proving the device meets them, based on the provided text:
Device: VX130 Ophthalmic Diagnostic Device
Primary Function under review (for the comparison study): Pachymetry (corneal thickness measurement)
1. Table of Acceptance Criteria and Reported Device Performance
The document doesn't explicitly state formal "acceptance criteria" in terms of specific thresholds for performance metrics. Instead, the approach is one of demonstrating substantial equivalence to a predicate device (Galilei G4). The implicitly accepted performance is that the VX130's measurements are not significantly different from those of the predicate device, across various anterior segment parameters, and that it demonstrates sufficient repeatability.
Parameter | Acceptance Criteria (Implicit: No significant difference from predicate / Adequate repeatability) | Reported Device Performance (VX130 vs. Galilei G4 - Average Difference) | Reported Device Performance (VX130 Repeatability - Std Dev 1 / Std Dev 2) |
---|---|---|---|
K1 Anterior (D) | Should be comparable to predicate; good repeatability. | -0.27 | 0.173 / 0.132 |
K2 Anterior (D) | Should be comparable to predicate; good repeatability. | -0.29 | 0.095 / 0.237 |
Cyl Anterior (D) | Should be comparable to predicate; good repeatability. | -0.03 | 0.166 / 0.229 |
K1 Posterior (D) | Should be comparable to predicate; good repeatability. | -0.09 | 0.031 / 0.075 |
K2 Posterior (D) | Should be comparable to predicate; good repeatability. | -0.01 | 0.093 / 0.107 |
Cyl Posterior (D) | Should be comparable to predicate; good repeatability. | 0.08 | 0.103 / 0.112 |
CCT (Central Corneal Thickness, µm) | Should be comparable to predicate; good repeatability. | -0.43 | 2.89 / 3.68 |
Interpretation of Performance: The average differences between the VX130 and Galilei G4 are quite small (e.g., -0.43 µm for CCT), and the 95% confidence intervals (not explicitly provided for CCT, but -15.03 to 14.16 µm for CCT suggests the differences are within a clinically acceptable range for equivalence, especially considering the standard deviations). The repeatability standard deviations for the VX130 are also small, indicating consistent measurements.
2. Sample Size Used for the Test Set and Data Provenance
-
Comparison Study (VX130 vs. Galilei G4):
- Sample Size: Not explicitly stated, however, Table 2 (Repeatability of VX130) mentions "42 eyes were measured 3 times each". It is highly probable that the comparison study (Table 1) also used these 42 eyes, or a similar sample size, as it's part of the same "Tests for pachymetry" section. Assuming 42 subjects, likely n = 42 eyes.
- Data Provenance: Not specified; retrospective or prospective is not mentioned. Country of origin not mentioned.
-
Repeatability Study (Table 2 - Std dev 1):
- Sample Size: "a single eye was measured ten times".
-
Repeatability Study (Table 2 - Std dev 2):
- Sample Size: "42 eyes were measured 3 times each".
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications
The document does not mention the use of experts to establish ground truth for the test set. The study design is a comparison between two devices (VX130 and Galilei G4) and a repeatability assessment of the VX130 itself. The Galilei G4's measurements serve as the reference for comparison, not an expert-derived ground truth.
4. Adjudication Method for the Test Set
No adjudication method is mentioned as there were no human readers/interpreters involved in establishing the ground truth for the measurement comparison. The comparison is between automated device measurements.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No MRMC study was done. This study focuses on device performance for automated measurements, not on human reader performance with or without AI assistance.
6. Standalone Performance Study
Yes, a standalone performance study was done for the pachymetry function of the VX130 in two ways:
- Comparison to a predicate device: This evaluates the VX130's agreement with an established device (Galilei G4). The results are presented in Table 1, showing "small differences between the devices."
- Repeatability study: This demonstrates the consistency of the VX130's own measurements. The results are presented in Table 2, showing low standard deviations for repeated measurements.
7. Type of Ground Truth Used
The ground truth for the comparison study was the measurements obtained from a legally marketed predicate device (Galilei G4). This is a common approach for demonstrating substantial equivalence for quantitative diagnostic devices. For the repeatability study, the "ground truth" is inferred from the consistency of the device's own measurements over multiple trials.
8. Sample Size for the Training Set
The document does not mention a training set. This is not an AI/Machine Learning device that requires a training set in the conventional sense. The device's functionality is based on established optical and measurement principles (Shack-Hartmann, Placido disk, Scheimpflug), and its performance is evaluated against a predicate device and through repeatability.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no mention of a training set for an AI/ML algorithm.
Ask a specific question about this device
(217 days)
LUNEAU SAS
The VX120 is a multi-function diagnostic device combining wavefront aberometer, corneal topographer, retro-illuminator, tonometer and pachymeter, indicated for:
Measuring the refraction of the eye giving both lower and higher order aberrations
Measuring the shape of the comea
Retro-illumination imaging of the eye
Measuring the intraocular pressure without contacting the eye for glaucoma evaluation
Photographing the eye and taking images of the eye to evaluate the thickness of the cornea.
The VX120 is a multifunctional ophthalmic diagnostic device.
The VX120 combined wavefront aberrometer, corneal topographer, retro illumination device, Scheimpflug pachymeter, and non-contact tonometer is a single platform that contains five different measurement units.
The wavefront aberrometer works on the Shack-Hartmann principle and is used as an advanced autorefractometer that measures both lower and higher order aberrations of the refraction of the eye.
Retro illumination is used to image ocular opacities.
The corneal topographer uses a Placido disk to measure keratometry and the detailed shape of the cornea.
The Scheimpflug pachymeter measures the thickness of the central cornea by illuminating it with a slit of light and photographing it using the Scheimpflug technique. An air puff non-contact tonometer is included for measurement of the intraocular pressure.
The device is fully automated and a number of different measurements can be performed by a single command including alignment and focusing.
The provided text describes the 510(k) summary for the VX120 Ophthalmic Diagnostic Device, focusing on its tonometry and pachymetry functions for substantial equivalence to predicate devices. It outlines performance data including bench and clinical testing.
Here's an analysis of the acceptance criteria and the study that proves the device meets them, based only on the provided text:
Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly derived from the performance claims and the standards the device was tested against. The text explicitly states that the device meets the requirements of the specified standards.
Acceptance Criterion (Implicit/Explicit from Text) | Reported Device Performance |
---|---|
Tonometry Function: | |
Accuracy (against predicate/standard) | Equivalent to ±2mmHg or better (bench testing) |
Repeatability (standard deviation) | ±1.2 mmHg or better (bench testing) |
Compliance with ISO8612:2010 | Met all requirements of the standard, if eyes with astigmatism >3mm are excluded (Clinical evaluation) |
Compliance with ANSI Z80.10-2009 | Met all requirements of the standard, if eyes with astigmatism >3mm are excluded (Clinical evaluation) |
Pachymetry Function: | |
No significant statistical difference in CCT measurements compared to Pentacam | No significant statistical difference between measurements of CCT with VX120 and the Pentacam (Comparison study). |
General Device Performance: | |
Electrical safety | Complies with IEC60601-1:2006 |
EMC compatibility | Complies with IEC60601-1-2 |
Software verification and validation | Done according to IEC62304 |
Risk management | Evaluated according to ISO14971: 2009; all risks reduced to safe levels. |
Ophthalmic product testing | Evaluated in accordance with ISO15004-1:2009 and ISO15004-2:2007; met all requirements. |
Optical hazards | Evaluated in accordance with IEC60825-1: 2008; VX120 is laser class 1. |
Here's a breakdown of the study details based on the provided text:
-
Sample size used for the test set and the data provenance:
- Tonometry: The text mentions "clinical evaluation" and "bench testing." It does not specify the sample size for either.
- Pachymetry: The text mentions a "comparison study" with Pentacam. It does not specify the sample size for this study.
- Data Provenance: The text does not explicitly state the country of origin of the data or whether the studies were retrospective or prospective. The manufacturer, Luneau SAS, is based in France. The predicates are from the UK (Keeler) and Germany (Oculus), implying international standards and potentially international data, but this is not confirmed.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- The document describes performance testing against standards (ISO, ANSI) and predicate devices. It does not mention the use of human experts or human readers to establish ground truth for the test sets for tonometry or pachymetry performance. The "ground truth" for tonometry appears to be established by the performance of the predicate device and the specified accuracy and repeatability limits from the standards. For pachymetry, the ground truth is implicitly the measurements from the Pentacam predicate.
-
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable, as ground truth was not established by human experts requiring adjudication.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No MRMC or human-in-the-loop study comparing human readers with and without AI assistance is described. The device is referred to as a "diagnostic device," but the performance studies focus on its intrinsic measurement accuracy and comparison to predicate devices, not its assistance to human interpretation or diagnosis.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Yes, the performance data presented (bench testing, clinical evaluation against standards, and comparison studies for tonometry and pachymetry) appear to represent the standalone performance of the VX120 device/algorithm. The focus is on the device's output accuracy and consistency.
-
The type of ground truth used (expert concensus, pathology, outcomes data, etc):
- Tonometry: The ground truth for the tonometry function appears to be established by:
- Benchmarking against industry standards: ISO8612:2010 and ANSI Z80.10-2009 for clinical evaluation, and internal specifications for accuracy (±2mmHg) and repeatability (±1.2mmHg) from bench testing.
- Comparison to predicate device performance: The "Pulsair tonometer" is also stated to have an accuracy of ±2 mmHg.
- Pachymetry: The ground truth for the pachymetry function is established by comparison to a legally marketed predicate device, the Pentacam, noted as demonstrating "no significant statistical difference."
- Tonometry: The ground truth for the tonometry function appears to be established by:
-
The sample size for the training set:
- The document is a 510(k) summary, not a detailed technical report on the algorithm development. It does not mention a "training set" or specify its size. The device relies on established physical measurement principles (Shack-Hartmann, Placido disk, Scheimpflug, air puff tonometry) rather than machine learning that typically requires training data.
-
How the ground truth for the training set was established:
- Not applicable, as a "training set" is not mentioned or implied for this type of device and its described validation. The device's operation is based on fundamental physics and established optical/measurement techniques rather than trained algorithms in the sense of AI/machine learning.
Ask a specific question about this device
Page 1 of 1