Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K232031
    Device Name
    LUNA 3D
    Date Cleared
    2024-03-29

    (266 days)

    Product Code
    Regulation Number
    892.5050
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LAP LUNA 3D system is used to support reproducible patient positioning and monitor patient surface motion during radiotherapy and radiosurgery treatments and during CT simulation for radiation therapy planning. LAP LUNA 3D may also be used to guide patients through breath-holds. The LAP LUNA 3D system can be used for all patients, undergoing radiotherapy and radiosurgery treatments and radiosurgery treatments and CT simulation for radiation therapy planning.

    Device Description

    The LUNA 3D optical camera system captures the current 3D patient skin surface with one or multiple camera pods. The software calculates the spacial deviations between the captured live surface and a reference surface within a selected region of interest using a registration algorithm. Reference surfaces may be generated with the optical camera system or may be imported using data received from Treatment Planning Systems (TPS). Based on the registration results the user can adjust the patient position for reproducible patient positioning relative to the treatment isocenter. During the imaging (simulation) and treatment delivery process, the system continuously calculates the deviations between live- and reference surface for patient motion monitoring to ensure that the patient position remains within pre-defined tolerances.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study information for the LUNA 3D device based on the provided FDA 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    The FDA 510(k) summary provides limited detailed acceptance criteria and mostly focuses on comparison to a predicate device. The performance claims primarily relate to accuracy.

    Acceptance Criteria / Performance MetricReported Device Performance (LUNA 3D)
    Positioning AccuracyTarget registration errors (as measured using calibration phantom)
    Ask a Question

    Ask a specific question about this device

    K Number
    K200907
    Date Cleared
    2020-06-03

    (58 days)

    Product Code
    Regulation Number
    892.5050
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The THALES 3D MR SCANNER system is designed to measure the characteristic radiation beam for an irradiation unit. The beam data is stored in the software program and can be exported from there for use during commissioning in a treatment planning system (TPS) or for quality assurance. The THALES 3D MR SCANNER system can be used with irradiation units with integrated magnetic resonance imaging and a magnetic field of less than or equal to 0.35 tesla.

    Device Description

    The THALES 3D MR SCANNER system consists of a water phantom with a carriage system, a manual control unit and a software program. A water basin with the corresponding pump, a control unit with an electrometer, and a motor controller are integrated in the carriage system. The software program is used to control the three axes of the water basin, perform measurements and analyse measured data.

    The software program has the following properties: Acquisition of measured data, analysis, and recording of photon beams with the water phantom. Generation and export of beam data libraries for a treatment planning system (TPS) for commissioning of irradiation units. The software program is equipped with a central database and/or a file system for central storage of all recorded data for different irradiation units within a large hospital network with different sites. The software program offers the option of defining different user levels to support the typical clinical workflows.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the THALES 3D MR Scanner:

    Important Note: The provided FDA 510(k) summary focuses on demonstrating substantial equivalence to a predicate device, not necessarily on a comprehensive efficacy study with predefined acceptance criteria for its performance metrics as one might see in a clinical trial. The "acceptance criteria" here are more about meeting safety, electromagnetic compatibility, and dosimeter standards, and showing that its operational parameters are comparable or superior without introducing new risks. There is no information about a comparative effectiveness study (MRMC) with human readers, nor a standalone algorithm performance study as would be applicable for AI/CADe devices.


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria (from predicate comparison/standards)Reported Device Performance (THALES 3D MR Scanner)
    Intended UseQA for radiation therapy systems, commissioning, and other occasions; use with integrated MRI (up to 1.5T for predicate)QA for radiation therapy systems, commissioning, and other occasions; use with integrated MRI with magnetic field ≤ 0.35 Tesla
    Scanning Range408mm * 355mm * 248mm (predicate)380mm * 380mm * 242mm
    Tank ShapeSquareSquare
    Wall MaterialPMMAPMMA
    Chamber Holder OrientationHorizontal and verticalHorizontal and vertical
    Motor Scanning ModeStep by stepStep by step
    Step Size0.1mm0.1mm
    Maximum Speed30mm/s (predicate)40mm/s
    Position Accuracy+/-0.1mm (predicate)+/- 0.25mm (at 0.35Tesla)
    Reproducibility+/-0.1mm (predicate)+/- 0.25mm (at 0.35Tesla)
    MRI CompatibilityUp to 1.5 Tesla (predicate)Up to 0.35 Tesla
    Electrometer Channels2 entries2 entries
    Electrometer Resolution10 fA (predicate)1 fA
    Electrical SafetyIEC 61010IEC 61010
    Dosimeter TestingIEC 60731IEC 60731
    EMCIEC 60601-1-2:2007 (predicate)IEC 60601-1-2:2015
    MRI SafetyN/A (implied by predicate's MRI compatibility)Tested in accordance with applicable ASTM standards

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not describe a "test set" in the context of an AI/ML algorithm or a specific patient-data based study. Instead, it details performance characteristics of a physical measurement device (water phantom) for radiation therapy. The performance testing refers to evaluating the device's physical and electrical characteristics against standards (IEC, ASTM). Therefore, there is no information about a sample size or data provenance related to patient data. The "test set" implies physical testing of the device itself.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    Not applicable. The ground truth for this device's performance is established by engineering and physics standards (IEC, ASTM) for safety, compatibility, and measurement accuracy, not by human expert consensus on clinical data.

    4. Adjudication Method for the Test Set

    Not applicable. There's no clinical data or expert review process described that would require an adjudication method.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. The document makes no mention of an MRMC study or any study involving human readers or comparative effectiveness in a clinical workflow with or without AI assistance. This device is a measurement tool, not an AI diagnostic or assistive tool in the traditional sense.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No. This is not an AI algorithm but a physical device with integrated software to control movement, acquire, analyze, and export measurement data. Its "performance" refers to how accurately it physically measures radiation beams, not its ability to interpret images or suggest diagnoses.

    7. The Type of Ground Truth Used

    The ground truth used for validating the device's performance comes from established engineering and medical device standards:

    • Electrical and mechanical safety: IEC 61010
    • EMC: IEC 60601-1-2
    • Dosimeter requirements: IEC 60731
    • MRI safety: Applicable ASTM standards
    • Performance metrics (e.g., position accuracy, electrometer resolution) are likely validated against manufacturer-defined specifications derived from engineering principles and comparison to the predicate device.

    8. The Sample Size for the Training Set

    Not applicable. This device is not an AI/ML model that requires a "training set" of data in the conventional sense. Its software controls hardware and processes raw measurement data based on established physics, not learned patterns from a "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173926
    Device Name
    DORADOnova MR3T
    Date Cleared
    2018-09-14

    (262 days)

    Product Code
    Regulation Number
    892.5050
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DORADOnova MR3T system assists in reproducible patient positioning during radiation therapy planning and is utilized during treatment planning workflow. The DORADOnova MR3T system projects visual laser beams which can be positioned at specified coordinates via a control system. The position coordinates are typically chosen by the operator or provided by an external system (e.g. treatment planning system). The DORADOnova MR3T system is allowed to be used in MR environment with a magnetic field, less or equal to 3 Tesla.

    Device Description

    The LAP moving laser positioning system contains laser devices and a control system. The laser system is provided in a configuration that enables the movement of two projection planes by using an appropriate composition of motorized and fixed laser devices. The installation of the system is realized in a combined bridge solution.

    AI/ML Overview

    This FDA 510(k) summary (K173926) describes the DORADOnova MR3T system, a medical device used for reproducible patient positioning during radiation therapy planning. The document primarily focuses on demonstrating substantial equivalence to predicate devices rather than providing a detailed clinical study with acceptance criteria for an AI/algorithm-driven device.

    Therefore, many of the requested elements for describing the acceptance criteria and the study that proves the device meets them (especially those related to AI/algorithm performance, ground truth establishment, expert adjudication, and MRMC studies) are not applicable to this specific device and document, as it is a laser positioning system, not an AI-powered diagnostic or therapeutic device.

    However, based on the provided text, I can extract the relevant performance specifications and how they were tested.


    Device Name: DORADOnova MR3T (LAP moving laser positioning system)

    Device Type: Medical charged-particle radiation therapy system (21 CFR 892.5050, Class II) - Specifically, a laser positioning system.


    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a physical positioning device and not an AI/algorithm, "acceptance criteria" are tied to its physical specifications and safety standards, largely demonstrated through bench testing and compliance with recognized standards.

    Feature / CriterionAcceptance Criteria (Stated or Implied)Reported Device Performance / Characteristics
    Intended UseMust assist in reproducible patient positioning during radiation therapy planning and be utilized during treatment planning workflow, by projecting visual laser beams positionable at specified coordinates. Must be usable in MR environment (≤ 3 Tesla).The DORADOnova MR3T system "assists in reproducible patient positioning during radiation therapy planning and is utilized during treatment planning workflow. The DORADOnova MR3T system projects visual laser beams which can be positioned at specified coordinates via a control system... The DORADOnova MR3T system is allowed to be used in MR environment with a magnetic field, less or equal to 3 Tesla." (Identical to intended use).
    Functional PrincipleMust use projection of visual light/laser lines on the patient's skin surface, with reflection aiding medical staff to determine patient's position."Projection of a visual laser lines on the patient's skin surface. Reflection of the laser lines aide's professional medical staff to determine the patient's position." (Similar to predicate/reference devices).
    Patient Reference Position InformationMust allow coordinates for 3 laser intersection points to be manually input or imported from a treatment planning system for reproducible patient positioning."The patient reference position information is the coordinates for 3 laser intersection points. They are either manually input into the device by the operator or imported from the treatment planning system." (Similar to predicate/reference devices).
    Positioning AccuracyMust be comparable to or better than predicate/reference devices. Reference device: +/- 0.1 mm. Predicate device: max mean: 0.27mm; max std dev: 0.65mm.+/- 0.1 mm (Matches reference device performance, superior to predicate device in terms of stated accuracy).
    Resolution / Line WidthMust be comparable to or better than predicate/reference devices. Reference device:
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1