Search Filters

Search Results

Found 194 results

510(k) Data Aggregation

    K Number
    K240506
    Date Cleared
    2024-10-22

    (244 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    KARL STORZ Endoscopy America, Inc

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The KARL STORZ Holders are manually operated surgical devices intended to hold endoscopes (such as arthroscopes, etc) and accessories during general and neurologic diagnostic and therapeutic procedures.

    Device Description

    The KARL STORZ Holder is a manually operated surgical instrument intended to hold scopes (such as endoscopes, arthroscopes, etc) in a desired position during diagnostic, therapeutic, and surgical procedures (including neurologic procedures). The holder is manufactured from Stainless Steel and functions by the user manually tightening or loosening a knob, which opens and closes around a surgical scope. The device is intended to attach to the several KARL STORZ holding systems, which are a Class 1, table-mounted, mechanical and electromechanical holding arms that have many general surgical functions in addition to scope holding; it can also be used for instrument holding, retraction, and positioning.

    AI/ML Overview

    The provided text is a 510(k) summary for the KARL STORZ Holders, which are manually operated surgical devices intended to hold endoscopes and accessories. This document primarily focuses on demonstrating substantial equivalence to a predicate device rather than providing extensive details about a clinical study with acceptance criteria for an AI/CAD-like device.

    Therefore, the requested information regarding acceptance criteria, study design, sample sizes, expert involvement, ground truth establishment, and MRMC studies for an AI/CAD device is not available in the provided text. The document describes a traditional medical device (a mechanical holder) and its non-clinical performance testing.

    However, based on the information that is available in the document, here's a summary of the non-clinical performance data and the demonstration of substantial equivalence:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not detail specific quantitative acceptance criteria or reported values in a table format for a clinical outcome beyond demonstrating functional equivalence. The performance data is described qualitatively as meeting the needs for substantial equivalence.

    Performance CategoryAcceptance Criteria (Implied for Substantial Equivalence)Reported Device Performance
    Holding PerformanceSufficient to securely hold intended devices (scopes, accessories) during procedures and maintain position.Testing was performed (Non-Clinical Performance Data: "Holding Performance testing"). Implied to be acceptable for intended use. Max loading weight of 3kg reported.
    Direction of View/Image StabilityMaintain stable direction of view/image during use.Testing was performed (Non-Clinical Performance Data: "Direction of view/Image stability testing"). Implied to be acceptable for intended use, as it contributes to substantial equivalence.
    Material of ConstructionBiocompatible and durable for surgical use.Stainless Steel, same as predicate device.
    Mode of OperationManual, user-friendly.Manual, same as predicate device.
    Cleaning & ReprocessingSterilizable and cleanable without degradation.Manual cleaning and Steam Sterilization, same as predicate device.

    2. Sample Size Used for the Test Set and Data Provenance:

    • The document describes non-clinical performance testing rather than a clinical test set involving human subjects or AI algorithm evaluation.
    • The exact sample size for the "Holding Performance testing" and "Direction of view/Image stability testing" is not specified.
    • Data Provenance: The testing was performed on the "subject device" (KARL STORZ Holders). The country of origin of the data is not explicitly stated beyond the submitter being KARL STORZ SE & Co. KG in Tuttlingen, Germany. The data would be considered prospective as it's testing of the new device for regulatory submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

    • This information is not applicable as the document describes non-clinical performance testing of a mechanical device, not the evaluation of an AI algorithm requiring expert ground truth for image interpretation.

    4. Adjudication Method for the Test Set:

    • This is not applicable for the same reason as point 3.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for evaluating the impact of AI on human reader performance, which is not the subject of this 510(k) submission for a mechanical endoscope holder.

    6. If a Standalone (algorithm only without human-in-the-loop performance) was done:

    • No, a standalone algorithm performance study was not done. This device is a mechanical surgical instrument, not an AI algorithm.

    7. The Type of Ground Truth Used:

    • The concept of "ground truth" as typically used for AI/CAD devices (e.g., expert consensus, pathology, outcome data) is not applicable. For this mechanical device, performance is evaluated against engineering specifications and functional requirements (e.g., ability to hold weight, maintain stability).

    8. The Sample Size for the Training Set:

    • This is not applicable. The device is a mechanical instrument, not an AI algorithm that requires a training set.

    9. How the Ground Truth for the Training Set was Established:

    • This is not applicable for the same reason as point 8.

    In summary, the provided FDA 510(k) summary (K240506) pertains to a mechanical endoscope holder, not an AI/CAD-enabled medical device. Therefore, the detailed requirements for acceptance criteria and study designs related to AI performance are not present in this document. The purpose of this 510(k) is to demonstrate substantial equivalence to a predicate mechanical device through non-clinical performance testing.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223885
    Date Cleared
    2023-09-14

    (261 days)

    Product Code
    Regulation Number
    884.1690
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Karl Storz Endoscopy America, Inc.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HOPKINS Telescopes when used with sheaths are intended to be used to permit viewing of the cervical and uterine cavity for the purpose of performing diagnostic and surgical procedures.

    Device Description

    The HOPKINS Telescopes are rigid telescopes that utilize the rod lens technology. At the distal end of the telescope's shaft is the lens and the other end of the shaft is attached to the eyepiece. Throughout the central lumen of the HOPKINS Telescopes, optical glass rods are used to transmit and magnify the image received from the lens. The HOPKINS Telescopes are available with 0°, 12° and 30° direction of view, 2mm, 2.9mm, 4mm diameter and 26cm, 30cm, 36cm working lengths.

    AI/ML Overview

    The provided text is a 510(k) Summary for the HOPKINS Telescopes, a medical device. This document focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study of its own performance against pre-defined acceptance criteria for a novel AI/software component.

    Therefore, the provided document does not contain the information required to answer your request regarding acceptance criteria and a study proving the device meets those criteria, specifically concerning AI/software performance.

    The document states:

    • "Clinical studies were not required to demonstrate substantial equivalence to the predicate device." (Page 9)
    • "The conclusions drawn from the nonclinical test demonstrate that the subject device is as safe and effective as the predicate device to support a substantial equivalence determination." (Page 9)

    This means the submission relies on bench testing and comparison of technological characteristics to a legally marketed predicate device, not on a clinical performance study with defined acceptance criteria and human readers (or AI algorithms) in the way your prompt describes.

    To directly answer your questions based only on the provided text, the answer is that this information is not present for the HOPKINS Telescopes device.

    If this were an AI/software device and the information was present, here's how I would answer each point:

    1. Table of acceptance criteria and reported device performance: This would be a table detailing metrics like sensitivity, specificity, AUC, or agreement rates, with specific target thresholds (acceptance criteria) and the actual performance achieved in the study.
    2. Sample size and data provenance: This would specify the number of cases/patients in the test set (distinct from training data), whether the data was collected retrospectively or prospectively, and from which countries or institutions.
    3. Number of experts and qualifications: This would state how many experts (e.g., radiologists, pathologists) were involved in establishing ground truth and their relevant experience (e.g., board-certified, years of experience, subspecialty).
    4. Adjudication method: This would describe how discordant readings or interpretations among experts were resolved (e.g., 2+1 means two initial readers, with a third if they disagree; 3+1 means three initial readers, with a fourth for adjudication if necessary).
    5. Multi-reader multi-case (MRMC) comparative effectiveness study: If conducted, this would provide details on how the AI system impacted human reader performance, including statistical metrics like the effect size (e.g., mean increase in AUC, sensitivity, or specificity when AI assistance was used compared to no AI assistance).
    6. Standalone performance: This would report the performance of the algorithm itself, without any human interaction, against the ground truth.
    7. Type of ground truth: This would specify the reference standard used (e.g., expert consensus, pathology results, follow-up clinical outcomes, surgical findings).
    8. Sample size for training set: This would state the number of unique cases/patients used to train the AI model.
    9. Ground truth for training set: This would explain the methodology used to label or establish the ground truth for the data utilized during the training phase of the algorithm.
    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Applicant Name (Manufacturer) :

    KARL STORZ Endoscopy America, Inc

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The KARL STORZ suction and irrigation system (consisting of handle and suction/irrigation tubes, cannulae or needles) is intended for use by qualified surgeons to provide suction and/or irrigation function to help flush or remove blood, fluids, and tissue debris from the operation site during general, gynecologic, urologic, and thoracic laparoscopic surgical procedures.

    Device Description

    The devices in the KARL STORZ Suction and Irrigation System are manually operated, reusable surgical devices consisting of handle, suction/irrigation tubes, cannulae or needles. The handle is intended to be used in combination with suction/irrigation cannulae and tubing and controls the suction and irrigation. Enabled by the handle with stop cock valve, the regulation of the irrigated quantity of fluids via the tubes, or rather the quantity of aspirated fluid (e.g., blood, fluids, and tissue debris) into or out of the patient is performed. Cannula and needle: Enabled by the pointed distal end, both puncture and injection/extraction cannulae and needles, perform puncturing of target organs or tissues and injecting fluids into operating site i.e. extracting blood or fluids from them. The prerequisite for suction is either a vacuum powered suction wall apparatus or a pump, which serve as impulse for the suction. The proximal end of the tube set is connected to the vacuum or pump. Either a pump is required as impulse for irrigation or the natural force of gravity, i.e., by positioning the flush bag at the appropriate height.

    AI/ML Overview

    The provided document is an FDA 510(k) Premarket Notification for medical devices, specifically KARL STORZ suction and irrigation systems. This type of submission focuses on demonstrating substantial equivalence to legally marketed predicate devices, primarily through non-clinical performance data and does not typically include studies involving artificial intelligence (AI) or machine learning (ML) components, nor does it present "acceptance criteria" and "device performance" in the context of an AI/ML model's diagnostic or predictive capabilities.

    Therefore, the requested information regarding AI/ML-related acceptance criteria and study details (such as sample sizes for test and training sets, expert adjudication methods, MRMC studies, standalone performance, ground truth establishment, etc.) cannot be extracted from this document, as it is not relevant to the type of device and submission described.

    The document focuses on:

    • Device Description: Manually operated, reusable surgical devices (handle, suction/irrigation tubes, cannulae, needles).
    • Intended Use: Suction and/or irrigation to flush or remove blood, fluids, and tissue debris during laparoscopic surgical procedures.
    • Basis for Substantial Equivalence: Comparison to predicate KARL STORZ devices (K945059 and K951190) based on technological characteristics and non-clinical performance data.
    • Non-Clinical Performance Data Provided:
      • Biocompatibility testing: Compliance with ISO 10993.
      • Reprocessing Validation: Compliance with standards like ANSI/AAMI/ISO 14937, 11138-3, 11607-1, 11737-1, 17664, 17665-1 regarding cleaning and sterilization.
      • Bench Top Performance Testing: Leak Test, Flow Incoming Test, Flow Outgoing Test.

    The document explicitly states: "Clinical testing was not required to demonstrate the substantial equivalence to the predicate device. Non-clinical bench testing was sufficient to assess safety and effectiveness and to establish the substantial equivalence of the modifications."

    In summary, this document does not contain information about an AI/ML device or its associated performance studies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223520
    Date Cleared
    2023-04-27

    (155 days)

    Product Code
    Regulation Number
    884.1690
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Karl Storz-Endoscopy-America, Inc.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The UNIDRIVE SIII System is intended for intrauterine use by trained gynecologists to hysteroscopically resect and remove tissue.

    The HOPKINS Telescope when used with the obturator is intended to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic and surgical procedures.

    Device Description

    The UNIDRIVE® SIII system is a motorized surgical device intended for intrauterine use to hysteroscopically resect and remove tissue. The UNIDRIVE SIII System includes the following components:
    UNIDRIVE SIII Control Unit (20701020-1) Footswitch (20016230) DrillCut-X II Shaver handpiece GYN (26702050) Shaver blades (26208SA, 26208SB) HOPKINS Telescope (26208AMA) Obturator (26208OC)

    The UNIDRIVE S III system has a motor control unit and can be used with the DrillCut-X II Shaver handpiece GYN which houses the shaver blades. The shaver blade upon being connected to the DrillCut-X II Shaver handpiece GYN is inserted into the working channel of the HOPKINS Telescopes. The shaver blades are activated via a one-pedal footswitch that is connected the UNIDRIVE SIII control unit. The shaver blades consist of two sheaths that fit into each other. The inner sheath oscillates within the outer sheath to enable tissue resection.

    AI/ML Overview

    The provided text is a 510(k) summary for the UNIDRIVE SIII System. It primarily focuses on demonstrating substantial equivalence to predicate devices through non-clinical testing rather than presenting a study with strict acceptance criteria of the type typically used for a novel AI/software performance claim.

    Therefore, the requested information specifically on acceptance criteria, reported performance, sample sizes, expert involvement, adjudication, MRMC studies, standalone performance, ground truth types, and training set details as they pertain to such a study is not available in the provided document. The document describes a comparison to predicate devices, stating that non-clinical testing was adequate to establish substantial equivalence without requiring clinical testing.

    However, I can extract the types of non-clinical tests performed, which could be considered as demonstrating the device meets established safety and performance standards relevant to its design.

    Regarding your specific questions, based on the provided text:

    1. A table of acceptance criteria and the reported device performance:
      The document does not provide a table of acceptance criteria with specific numerical targets and reported performance values in the context of a clinical or analytical performance study of the type you're asking for. Instead, it refers to successful demonstration of performance by comparison to predicate devices and adherence to recognized standards.

      • "The cutting performance of the subject device, UNIDRIVE SIII system has been successfully demonstrated by comparing to the primary predicate device TRUCLEAR Morcellator System and TRUCLEAR Morcellators cleared in K132015." This implies the UNIDRIVE SIII System performed comparably to the predicate for cutting, but specific metrics are not given.
      • "Additional comparative bench testing between the subject and secondary predicate device was conducted to demonstrate the optical performance of the HOPKINS Telescope that is part of the subject UNIDRIVE SIII system to meet its intended use." This indicates the optical performance was deemed sufficient but lacks specific criteria or quantitative results.
    2. Sample size used for the test set and the data provenance: Not applicable, as no clinical or analytical performance study (as implied by these questions) is detailed. The testing was non-clinical (bench testing).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth as typically defined for AI/software evaluation (e.g., expert consensus on images) is not relevant here as it was a non-clinical bench testing and comparative analysis.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. No MRMC study was conducted or mentioned. The device is not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable. This device is not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not applicable. For the non-clinical testing, "ground truth" would be the expected performance or measurements from the predicate devices and relevant engineering specifications.

    8. The sample size for the training set: Not applicable. Not an AI/ML device.

    9. How the ground truth for the training set was established: Not applicable.


    Summary of Non-Clinical Performance Testing (as described in the document):

    The device underwent various non-clinical tests to demonstrate substantial equivalence and adherence to recognized standards. These standards implicitly contain "acceptance criteria" for the device's functionality, safety, and performance.

    Non-Clinical Performance Data / Tests Performed:

    • Hysteroscopes and Gynecologic Laparoscopes Guidance: Adherence to FDA guidance for these types of devices.
    • ISO Endoscopic Standards:
      • ISO 8600-1:2015
      • ISO 8600-3:2019
      • ISO 8600-5:2020
    • Biocompatibility Summary (ISO 10993 series):
      • Cytotoxicity (ISO 10993-5)
      • Acute Systemic Toxicity (ISO 10993-11)
      • Intracutaneous Irritation (ISO 10993-10)
      • Maximization Sensitization (ISO 10993-10)
    • Electrical Safety and EMC (IEC 60601 series):
      • IEC 60601-1:2012 (3rd Edition)
      • IEC 60601-2-18:2009 (3RD Edition)
      • IEC 60601-1-2:2014 (4th Edition)
    • Reprocessing (Cleaning and Sterilization) (AAMI and ISO standards):
      • AAMI TIR12: 2010
      • AAMI TIR30: 2011
      • AAMI TIR39: 2009
      • ANSI/AAMI ST8: 2013
      • ANSI/AAMI ST77:2013
      • ANSI/AAMI ST79:2017
      • ANSI/AAMI ST81:2004/(R)2010
      • AAMI/ISO 14937:2009
      • ANSI/AAMI/ISO 17655-1:2006/2013
      • ANSI/AAMI/ISO 17655-2:2009
      • Reprocessing Medical Device in Health Care Settings: Validation Methods and Labeling
    • Software Verification and Validation Testing:
      • Guidance for the Content of Premarket Submissions for Software Contained in Medical Device
    • Cutting Performance: Successfully demonstrated by comparing to the primary predicate device.
    • Optical Performance: Demonstrated through comparative bench testing with the secondary predicate device for the HOPKINS Telescope.

    Conclusion stated: The nonclinical tests demonstrate that the subject device is as safe and effective as the primary and secondary predicate devices to support a substantial equivalence determination.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223923
    Date Cleared
    2023-03-30

    (90 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    KARL STORZ Endoscopy-America, Inc

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For telescopes with diameter ranging from 3mm-5mm
    The HOPKINS Telescopes are intended to provide visualization during laparoscopy and general surgery in adults and pediatrics.
    For telescopes with diameter ranging from 5.5mm- 11mm
    The HOPKINS Telescopes are intended to provide visualization during laparoscopy and general surgery in adults.

    Device Description

    The HOPKINS Telescopes are rigid telescopes that utilize the rod lens technology. At the distal end of the telescope's shaft is the lens and the other end of the shaft is attached to the eyepiece. Throughout the central lumen of the HOPKINS Telescopes, optical glass rods are used to transmit and magnify the image received from the lens. The HOPKINS Telescopes are available with 0°, 6°, 30° and 45° direction of view, diameters ranging from 3mm to 11mm and working lengths from 18cm- 50cm.

    AI/ML Overview

    The provided text describes the submission of a 510(k) summary for the HOPKINS Telescopes, comparing it to a predicate device. It details non-clinical performance data and references published literature for clinical performance, but it does not contain information about acceptance criteria or a specific study proving the device meets those criteria, especially in the context of an AI/ML device.

    The device described, "HOPKINS Telescopes," is a rigid endoscope that utilizes rod lens technology for visualization during surgery. There is no mention of AI or machine learning components. Therefore, many of the requested points, such as sample sizes for test/training sets, expert ground truth adjudication, MRMC studies, and standalone AI performance, are not applicable to the information provided in this document.

    However, based on the non-clinical performance data section, we can infer some "acceptance criteria" related to standards compliance and general performance.

    Here's an attempt to answer the questions based only on the provided text, recognizing that it does not pertain to AI/ML device performance:

    1. A table of acceptance criteria and the reported device performance

    Since this is not an AI/ML device, the "acceptance criteria" are related to compliance with recognized standards and successful bench testing. The document states:
    "the HOPKINS Telescopes has met all its design specification and is substantially equivalent to its predicate device."

    Criteria TypeAcceptance Criteria (Implied)Reported Device Performance
    Functional Standards (Endoscopy)Compliance with ISO 8600-1, ISO 8600-3, ISO 8600-5, ISO 8600-6Met these standards.
    BiocompatibilityCompliance with ISO 10993-5, -11, -10 (twice)Passed Cytotoxicity, Acute Systemic Toxicity, Intracutaneous Irritation, Maximization Sensitization tests.
    Electrical & Thermal SafetyCompliance with IEC 60601-2-18:2009 (3RD Edition)Met this standard.
    Reprocessing (Cleaning & Sterilization)Compliance with AAMI TIR12, TIR30, ST8, ST77, ST79, ST81, ISO 14937, ISO 17655-1Met these standards during reprocessing validation.
    General PerformanceSubstantial equivalence to predicate device in bench testingDemonstrated substantial equivalence; met all design specifications.

    2. Sample size used for the test set and the data provenance

    The document does not specify a "test set" in the context of data for an AI/ML model. The evaluation involves non-clinical bench testing and comparison to a predicate device. No specific sample sizes for these tests are mentioned, nor is there information about data provenance (country of origin, retrospective/prospective).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This is not applicable as the device is a physical endoscope, not an AI/ML diagnostic tool requiring expert ground truth for image interpretation.

    4. Adjudication method for the test set

    Not applicable.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable, as this is not an AI-assisted device.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable, as this is not an AI/ML algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical testing, the "ground truth" would be established by physical measurements, chemical analyses, and adherence to the specified performance parameters outlined in the referenced standards. For the pediatrics indication expansion, the ground truth is based on "published literature" supporting safety and effectiveness, which would implicitly rely on clinical data outcomes.

    8. The sample size for the training set

    Not applicable, as this is not an AI/ML device with a training set.

    9. How the ground truth for the training set was established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221893
    Date Cleared
    2023-03-17

    (261 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    KARL STORZ Endoscopy America, Inc

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The KARL STORZ bipolar resectoscopes with HF cable are intended for use by qualified surgeons for cutting, ablation, vaporization, and/or coagulation of tissue during various endoscopic urological procedures.

    The KARL STORZ bipolar resectoscopes with HF cable are intended for use by qualified surgeons for cutting, ablation, vaporization, and/or coagulation of tissue during various endoscopic electrosurgical procedures.

    The KARL STORZ bipolar resectoscopes with HF cable are intended for use by qualified surgeons for cutting, ablation, vaporization, and/or coagulation of tissue during various endoscopic urological and hysteroscopic electrosurgical procedures.

    Indications for Use in UROLOGY: The KARL STORZ bipolar resectoscopes with HF cable are intended for use by qualified surgeons for cutting, ablation, vaporization, and/or coagulation of tissue during various endoscopic urological procedures.

    Indications for Use in GYNECOLOGY: The KARL STORZ bipolar resectoscopes with HF cable are intended for use by qualified surgeons for cutting, ablation, vaporization, and/or coagulation of tissue during various endoscopic hysteroscopic electrosurgical procedures

    Indications for Use in UROLOGY and GYNECOLOGY: The KARL STORZ bipolar resectoscopes with HF cable are intended for use by qualified surgeons for cutting, ablation, vaporization, and/or coagulation of tissue during various endoscopic urological and hysteroscopic electrosurgical procedures.

    Device Description

    Resectoscopes are used for endoscopically controlled ablation of tissue. They are used for examination, diagnosis and/or therapy in conjunction with endoscopic accessories in medical disciplines such as Urology and Gynecology. A resectoscope is a combined cystoscope (urology) or hysteroscope (gynecology) and electrosurgical instruments and consists of a sheath, obturator, working element and HF electrode.

    AI/ML Overview

    This document describes the KARL STORZ Bipolar Resectoscopes with HF cable (K221893). It is important to note that this device is not an AI/ML powered device, therefore, many of the typical acceptance criteria and study details related to AI performance, such as sample sizes for test and training sets, expert qualifications, and adjudication methods, are not applicable.

    Here's an analysis of the provided information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a traditional medical device and not an AI/ML powered device, the "acceptance criteria" are based on compliance with recognized standards and design specifications, rather than quantitative performance metrics like sensitivity or specificity.

    Acceptance Criterion (Non-Clinical)Reported Device Performance
    Electrical Safety and EMCComplies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18
    BiocompatibilityComplies with ISO 10993 (specific parts not detailed, but generally covers biocompatibility)
    Performance Testing
    - Attachment/Detachment force of the electrodeVerified (Met design specifications)
    - System Interlocking TestVerified (Met design specifications)
    - Flow Test (comparative)Verified (Met design specifications)
    - Bending Force TestVerified (Met design specifications)
    - HF Tissue and HF Durability TestVerified (Met design specifications)
    Design SpecificationsAll design specifications were met through additional bench testing. (Specific quantitative specifications are not provided in this summary, but are internally verified by the manufacturer).
    Substantial Equivalence to Predicate DeviceThe conclusions drawn from the nonclinical tests demonstrate that the subject device performs as well as or better than devices currently marketed for the same intended use. (This is the overarching acceptance criterion for 510(k) clearance).

    2. Sample Size Used for the Test Set and Data Provenance

    Not applicable for a non-AI/ML medical device. Testing was primarily bench testing to verify engineering specifications and compliance with standards. There is no "test set" in the context of image or data analysis.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    Not applicable. Ground truth, in the context of AI, refers to expert-labeled data. For this device, "ground truth" would be established by engineering specifications and objective measurements against those specifications.

    4. Adjudication Method

    Not applicable. This is relevant for studies involving human interpretation or labeling of data, which is not the case here.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. This a traditional medical device, not an AI-assisted device, therefore, no such study was conducted to evaluate human reader improvement with AI.

    6. Standalone Performance Study

    Yes, in a sense. The non-clinical performance data (electrical safety, biocompatibility, and various performance tests) represent the "standalone" performance of the device without human interaction being a variable in the performance assessment itself (though human surgeons use the device). The tests verify the device's inherent functional characteristics and safety.

    7. Type of Ground Truth Used

    For non-clinical testing of a traditional medical device, the "ground truth" used would be:

    • Engineering Specifications: Pre-defined technical requirements for the device's dimensions, material properties, electrical characteristics, mechanical strength, and functional performance.
    • Recognized Standards: International (e.g., IEC, ISO) and national standards that set minimum safety and performance requirements for medical devices of this type.
    • Comparative Performance to Predicate Device: The performance of the legally marketed predicate device (K210651), which serves as a benchmark for substantial equivalence.

    8. Sample Size for the Training Set

    Not applicable. This device does not involve machine learning; therefore, there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no training set for a machine learning model, establishing ground truth for such a set is irrelevant.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221004
    Date Cleared
    2022-12-20

    (259 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    KARL STORZ Endoscopy America, Inc.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HOPKINS Telescopes are intended to provide visualization during laparoscopy and general surgery in adults and pediatrics.

    Device Description

    The HOPKINS Telescopes are rigid telescopes that utilize the rod lens technology. At the distal end of the telescope's shaft is the lens and the other end of the shaft is attached to the eyepiece. Throughout the central lumen of the HOPKINS Telescopes, optical glass rods are used to transmit and magnify the image received from the lens. The HOPKINS Telescopes are available with 0°, 25° and 45° direction of view, 5mm diameter and 29cm working lengths.

    AI/ML Overview

    The provided text is a 510(k) summary for the KARL STORZ HOPKINS Telescopes. This document is a premarket notification for a medical device to demonstrate substantial equivalence to a legally marketed predicate device. As such, it focuses on comparing the new device to an existing one based on technical characteristics, rather than establishing acceptance criteria and proving performance through a clinical study for an AI/ML-driven device.

    The document discusses the physical and optical characteristics of the endoscope, reprocessing methods, and non-clinical performance data (e.g., adherence to ISO standards, biocompatibility, thermal safety, reprocessing validation). It explicitly states: "There are no performance standards or special controls developed under Section 514 of the FD&C Act for endoscopes." This indicates that the regulatory pathway for this device does not require the kind of performance validation typically seen for AI/ML devices, which would involve acceptance criteria for metrics like sensitivity, specificity, or AUC, and detailed clinical study designs.

    Therefore, the document does not contain the information required to answer the prompt questions regarding acceptance criteria for an AI/ML device. The HOPKINS Telescopes are a traditional optical medical device, not an AI/ML product.

    To reiterate, the provided text does not contain any information about:

    1. Acceptance criteria or reported device performance for AI/ML metrics.
    2. Sample sizes for test sets or data provenance (as it's not an AI/ML study).
    3. Number and qualifications of experts for ground truth establishment.
    4. Adjudication methods.
    5. MRMC comparative effectiveness studies or effect sizes.
    6. Standalone algorithm performance studies.
    7. Type of ground truth (as it's not a diagnostic AI/ML device).
    8. Sample size for training sets (no AI/ML training is mentioned).
    9. How ground truth for a training set was established.
    Ask a Question

    Ask a specific question about this device

    K Number
    K222504
    Device Name
    H1
    Date Cleared
    2022-11-10

    (84 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Karl Storz Endoscopy America, Inc.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The KARL STORZ H1 camera head, in combination with an appropriately indicated camera control unit (CCU), light source, and monitor, and with an appropriately indicated endoscope, or microscope, is used for real-time visualization in diagnostic and surgical procedures.

    Device Description

    The camera head consists of an anodized aluminum, stainless steel enclosure containing a CMOS (Complementary metal-oxide-semiconductor) NTSC image sensor that converts light into electrons, the transistors in each pixel then amplify and move the charge using the more traditional wires forming conventional output signals. The camera head is intended to be attached to the proximal end of the endoscopes and is connected via cable to the compatible CCU for power and operational functions. The camera head is designed to be compatible for use with all standard KARL STORZ Endoscopes, Fiberscopes and Microscope for endoscopic observation in general endoscopic procedures.

    AI/ML Overview

    The provided text describes a 510(k) submission for the KARL STORZ H1 Camera Head, which is an endoscope camera. The document focuses on demonstrating substantial equivalence to a predicate device through non-clinical performance testing.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based solely on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document lists several performance tests conducted for the subject device to ensure it meets its design specifications and is substantially equivalent to the predicate device. However, it does not explicitly state numerical acceptance criteria for each test or a direct comparison of the subject device's performance against those criteria in a table format. It generally states that the device "met all its design specification" and "performs as well as the predicate device."

    Implied Acceptance Criteria Categories (from "Performance Testing" section):

    • Color Reproduction and Color Contrast Enhancement
    • Illumination
    • Detection Uniformity
    • Depth of field
    • Spatial Resolution
    • Dynamic Range & Detection Linearity
    • Distortion
    • Latency
    • Signal-to-Noise Ratio (SNR) & Sensitivity
    • Field of View

    Reported Device Performance:

    The document states, "The bench testing performed verified and validated that the H1 Camera Head has met all its design specification and is substantially equivalent to its predicate devices." It also concludes, "The conclusions drawn from the nonclinical tests demonstrate that the subject devices, the H1 Camera Head performs as well as the predicate device."

    Without specific numerical acceptance criteria and performance data, a detailed table cannot be created.

    2. Sample size used for the test set and the data provenance

    The document explicitly states: "Clinical testing was not required to demonstrate the substantial equivalence to the predicate devices. Non-clinical bench testing was sufficient to establish the substantial equivalence of the modifications."

    Therefore, there was no clinical test set in terms of human subjects or medical images. The testing was entirely non-clinical bench testing. The sample size for this non-clinical testing is not specified. The data provenance is also not applicable in the sense of country of origin for retrospective or prospective data, as it was bench testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    As no clinical testing was performed and the ground truth was established through non-clinical bench testing against design specifications, no human experts were used to establish ground truth in the traditional sense of medical image interpretation.

    4. Adjudication method for the test set

    Since no clinical test set involving human experts was used for ground truth establishment, no adjudication method (like 2+1 or 3+1) was employed.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC study was done. The device is a camera head, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not directly applicable. The device is a camera head, not an algorithm. The "standalone" performance was essentially the non-clinical bench testing of the device itself (hardware performance), which was done without human-in-the-loop in a clinical diagnostic context.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the non-clinical testing was based on design specifications and objective physical measurements during bench testing, ensuring the camera met its intended performance characteristics (e.g., spatial resolution measurements, color accuracy, brightness control functionality). This is not expert consensus, pathology, or outcomes data.

    8. The sample size for the training set

    This question is not applicable as the H1 Camera Head is a hardware device and not an AI/machine learning algorithm requiring a training set.

    9. How the ground truth for the training set was established

    This question is not applicable as the H1 Camera Head is a hardware device and not an AI/machine learning algorithm requiring a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213194
    Date Cleared
    2022-06-23

    (267 days)

    Product Code
    Regulation Number
    874.4720
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    Karl Storz Endoscopy America, Inc.

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The HD Mediastinoscope is intended to aid the surgeon in viewing the mediastinum and facilitate the introduction and removal of surgical instruments during ENT endoscopic surgical procedures.

    Device Description

    The HD Mediastinoscope is a reusable videoendoscope intended to be used in conjunction with the X-link module of the Image1 S Camera Control Unit. The HD Mediastinoscope is comprised of four main components: CMOS sensor at the distal end of the endoscope, an oval insertion portion (spatula), handle, and internal LED light source. The spatula can accommodate up to 2 x 5 mm instruments.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the HD Mediastinoscope:

    Due to the nature of the device (a physical surgical instrument rather than an AI/software device), the provided document does not contain information typically found in acceptance criteria and studies for AI/ML devices, such as sensitivity, specificity, or AUC.

    Based on the provided text, the "acceptance criteria" appear to be related to the device's technical specifications and safety/effectiveness in comparison to predicate devices, rather than specific performance metrics like those for an AI algorithm.

    Acceptance Criteria and Reported Device Performance

    The "acceptance criteria" are implied through the comparison with existing predicate and reference devices and compliance with recognized standards. The "performance" is demonstrated through non-clinical testing, particularly a cadaver test study.

    Acceptance Criteria (Implied)Reported Device Performance
    Substantial equivalence to predicate device (K954910)Demonstrated through technical characteristics, intended use, indications for use.
    Compliance with electrical safety and EMC standardsANSI/AAMI ES:60601-1:2005+A1:2012, IEC 60601-1-2:2014, IEC 62471:2006
    Compliance with biocompatibility standardsISO 10993-10:2010, ISO 10993-11:2006/(R) 2010
    Compliance with reprocessing validation standardsAAMI TIR 12:2010, AAMI TIR 30:2011, AAMI/ANSI/ISO 11737-1:2006/(R)2011, ASTM E1837-96:2014
    Change in spatula design does not raise new questions of safety and effectivenessDemonstrated through a cadaver test study.

    Details of the Study (Non-AI/ML Device)

    Since this is a traditional medical device (mediastinoscope) and not an AI/ML device, many of the requested points related to AI/ML studies (sample size for test set, data provenance, number of experts for ground truth, adjudication method, MRMC study, standalone performance, training set size, ground truth for training set) are not applicable or not specified in this 510(k) summary.

    However, based on the information provided:

    1. Sample size used for the test set and the data provenance:

      • The primary non-clinical performance test mentioned is a cadaver test study.
      • The sample size (number of cadavers or tests performed) is not specified in the document.
      • The data provenance (country of origin, retrospective/prospective) is not specified.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • The document mentions "a cadaver test study was conducted to demonstrate that the change in spatula design...". It does not specify the use of "experts" to establish a "ground truth" in the way an AI/ML study would (e.g., radiologists labeling images). Instead, the study likely involved technical assessment and observation by qualified personnel (e.g., surgeons, engineers).
      • Not explicitly stated in terms of "ground truth experts" for an AI/ML context.
    3. Adjudication method for the test set:

      • Not applicable / Not specified for this type of device and study.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, an MRMC study was not conducted. This is not an AI-assisted device.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This is a physical surgical instrument, not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • For the cadaver study, the "ground truth" would likely be the direct observation and technical assessment of the device's performance in a cadaveric model (e.g., ease of introduction, visualization, ability to facilitate instrument use, assessment of tissue damage, etc.) by the study investigators. This is not a "ground truth" in the sense of a diagnostic label for an AI.
    7. The sample size for the training set:

      • Not applicable. This device does not use a training set as it is not an AI/ML product.
    8. How the ground truth for the training set was established:

      • Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K221174
    Device Name
    Telepack +
    Date Cleared
    2022-05-13

    (18 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    KARL STORZ Endoscopy America, Inc

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TELE PACK + is an all-in-one Imaging System, which comprises a light source for illumination, Camera Control Unit (CCU) for use with compatible camera heads or video endoscopes for image processing, as well as a monitor for image display, intended for the visualization of endoscopic and microscopic procedures.

    Device Description

    The Telepack + is a portable and compact all-in-one imaging system that includes a 18.5 inch screen display, a camera control unit and internal LED light source, that is intended to be connected to a compatible device (camera head or videoendoscope) for the purpose of visualization and documentation of endoscopic and microscopic procedures as well as stroboscopy. The Telepack + includes a LED illumination light source to illuminate the intended area and a 18.5 inch monitor for display. It also allows the users to redefine the functions that take place when a button is pressed. The Telepack + is a non-patient contacting and require only wipe down as needed.

    AI/ML Overview

    The KARL STORZ TELEPACK+ is an all-in-one imaging system for endoscopic and microscopic procedures. This submission (K221174) is a modification of a previously cleared device (K193235). The key modifications include:

    • Swapping of two camera types (X-link and Office Link) via touchscreen.
    • Addition of capacitors to the Office Link front end.
    • Mechanical changes: added isolation to the microphone, an additional cover for the power supply, and a mechanical metal part as a protective earth.
    • Addition of video playback and image review functionality.
    • Additional cybersecurity features.
    • Additional camera and videoscope compatibility.

    The device's non-clinical performance data and bench testing were deemed sufficient to demonstrate substantial equivalence to its predicate device. This implies that the device meets its design specifications and is as safe and effective as the legally marketed predicate devices.

    Here's the breakdown of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria / Performance MetricReported Device Performance
    Electrical Safety and EMC
    IEC 60601-1 (General medical electrical equipment)Passed
    IEC 60601-1-2 (Electromagnetic compatibility)Passed
    IEC 60601-2-18 (Endoscopic equipment)Passed
    IEC 62471 (Photobiological safety)Passed
    Software Verification and Validation
    Guidance for the Content of Premarket SubmissionsMet (Level of concern: Moderate)
    Performance Testing
    Minimum IlluminationMet design specifications
    Spatial ResolutionMet design specifications
    Color PerformanceMet design specifications
    LatencyMet design specifications
    White BalanceMet design specifications
    AE (Automatic Exposure) Step ResponseMet design specifications
    Head Button FunctionalityMet design specifications
    Protection against electrical shockClass I protection, Type BF protection (stroboscopy & camera), Type CF protection (light)
    Moisture protectionDrip-water protection per IPX1

    2. Sample size used for the test set and the data provenance

    The document does not specify a separate "test set" and its sample size in the context of a clinical study, as no clinical performance data was required. The "test set" for demonstrating substantial equivalence primarily involved bench testing and electrical, EMC, and software validation tests. The data provenance is internal to KARL STORZ Endoscopy America, Inc., as these were laboratory and engineering tests conducted on the device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. For this type of device (an imaging system, not an AI diagnostic tool), "ground truth" established by medical experts for a test set is typically not germane to the substantial equivalence pathway, especially when no clinical data is required. The ground truth for engineering and performance measurements is established by standard measurement techniques and reference equipment.

    4. Adjudication method for the test set

    Not applicable. Since no clinical study or expert review for diagnostic accuracy was performed or required, an adjudication method for a test set is not relevant.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The TELEPACK+ is an imaging system and not an AI-assisted diagnostic device. Therefore, an MRMC study comparing human readers with and without AI assistance was not conducted.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. The TELEPACK+ is a medical imaging system, not a standalone algorithm. Its performance is evaluated on its ability to produce high-quality images and manage endoscopic/microscopic procedures, not on an algorithm's diagnostic capabilities.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the performance testing (e.g., minimum illumination, spatial resolution, color performance) was established through objective engineering measurements using calibrated equipment and standard test methods against the device's design specifications. For electrical safety and EMC, the "ground truth" is compliance with recognized international standards (e.g., IEC 60601 series).

    8. The sample size for the training set

    Not applicable. The TELEPACK+ relies on traditional imaging processing and control technologies, not machine learning or artificial intelligence that would require a "training set" in the context of algorithm development.

    9. How the ground truth for the training set was established

    Not applicable. As no training set for an AI algorithm was used, this question is not relevant.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 20