Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K992559
    Manufacturer
    Date Cleared
    1999-09-20

    (52 days)

    Product Code
    Regulation Number
    870.4350
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Applicant Name (Manufacturer) :

    JOSTRA BENTLEY

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Hollow Fibre Membrane Oxygenator HMO 1010 is intended for use in an extracorporeal perfusion circuit to oxygenate blood and remove carbon dioxide and to temper blood during short duration cardiopulmonary bypass procedures lasting 6 hours or less.

    Device Description

    The Hollow Fibre Membrane Oxygenator Quadrox HMO 1010 is a blood gas exchanger with integrated heat exchanger. In open heart surgery it is used in an extracorporeal perfusion circuit first to oxygenate blood and remove carbondioxide second to regulate the blood temperature during short duration cardiopulmonary bypass procedures lasting 6 hours or less.

    The oxygenation system is principally based on microporous polypropylene hollow fiber membranes. The heat exchanger is made of thight polyethylene fibers. The QUADROX consists of two membrane compartments. In the first chamber of the oxygenator, sheets of heat exchanger fibers and sheets of microporous oxygenation fibers are arranged crosswise. In the second chamber only sheets of the oxygenation fibers are arranged.

    Blood enters the housing via the inlet connector and is distributed into a pre-chamber. Then the blood streams through the membrane package. In the first section it is tempered and oxygenated. In the second part only oxygenation and removal of carbondioxide takes place.

    Positioned and integrated at the top of the oxygenator is a de-airing membrane. This membrane is a hydrophobic membrane allowing only gaseous substances to pass through the membrane but fluids are held back. The de-airing membrane allows easier priming, deairing and elimination of air throughout the whole procedure. To prime the oxygenator the Luer cap has to be removed. It should be kept open during perfusion to eliminate air continuously.

    AI/ML Overview

    Acceptance Criteria and Device Performance for Jostra Medizintechnik AG Hollow Fibre Membrane Oxygenator Quadrox HMO 1010

    This document outlines the acceptance criteria and performance data for the Hollow Fibre Membrane Oxygenator Quadrox HMO 1010, based on the provided 510(k) summary. The device was found to be substantially equivalent to the Affinity Hollow Fiber Oxygenator by Avecor Cardiovascular, Inc.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by the substantial equivalence determination to the predicate device. The performance data for the Quadrox HMO 1010 are compared to the predicate device's specifications.

    ParameterAcceptance Criteria (Predicate Device: Affinity Hollow Fiber Oxygenator)Reported Device Performance (Quadrox HMO 1010)
    Membrane TypeMicroporous Polypropylene Hollow FibersMicroporous Polypropylene Hollow Fibre
    Membrane Surface Area$2.5 m^2$$1.8 m^2$
    Static Priming Volume270 ml250 ml
    Recommended Blood Flow Rate1-7 liters/minute0.5-7 liters/minute
    Maximum Water Side Pressure30 psi14 psi
    Material of Heat ExchangerStainless steelPolyethylene
    Arterial Outlet Port3/8"3/8"
    Venous Inlet Port3/8"3/8"
    Arterial Sample PortFemale Luer PortLuer-Port
    Recirculation Port1/4"1/4"
    Gas Inlet Port1/4"1/4"
    Gas Outlet Port3/8" non-barbed1/4"
    Water Ports½" quick disconnects½" Quick-Connect Fittings (Hansen)
    Method of SterilizationunknownEthylene Oxide
    UseSingle-use deviceSingle-use device
    BiocompatibilityBiocompatible and non-toxicBiocompatible and non-toxic
    Sterility Assurance Level (SAL)10-610-6
    EtO Residuals (Ethylene Oxide)
    Ask a Question

    Ask a specific question about this device

    K Number
    K984338
    Manufacturer
    Date Cleared
    1999-08-19

    (259 days)

    Product Code
    Regulation Number
    870.2100
    Reference & Predicate Devices
    Why did this record match?
    Applicant Name (Manufacturer) :

    JOSTRA BENTLEY

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Jostra HL-20 Twin Pump is intended for use as an extra corporeal circulation device for perfusion appropriate to cardiopulmonary bypass procedures of six hours or less. It is only intended for use with the Jostra HL-20 Heart Lung Machine.

    Device Description

    Not Found

    AI/ML Overview

    The provided text describes a 510(k) submission for the Jostra HL-20 Twin Pump, aiming to demonstrate its substantial equivalence to a predicate device, the Jostra HL-20 Single Head Roller Pump. The submission focuses on functional and design similarities, rather than a clinical study evaluating diagnostic performance. Therefore, many of the requested categories for AI/diagnostic device studies (like sample size, ground truth, expert qualifications, MRMC studies, standalone performance) are not applicable or cannot be extracted directly from this document.

    However, I can extract the acceptance criteria and performance related to the engineering and safety standards that the device adheres to, as this is the basis of the equivalence claim.

    Acceptance Criteria and Reported Device Performance

    Acceptance CriterionReported Device Performance (Reference to Standard)
    General SafetyComplies with EN93/42: EU Council Directive 93/42/EEC concerning Medical Devices
    Constructional SafetyComplies with IEC 601-1: IEC 601-1, 2nd Edition: 1988 (identical with DIN VDE 0750 Part 1)
    Electromagnetic Compatibility (EMC)Complies with EU 89/336: EMC Directive 89/336/EEC 1993 (identical with DIN VDE 0750, Part 1-2)
    Functional Performance"performs within its specifications and also within acceptable limits for performance standards" (Implied: this covers the specific operational aspects of the pump, which are not detailed in the provided text but are indicated to be satisfactory.)

    Study Details (as inferable from the document, recognizing this is not a diagnostic AI study):

    1. Sample size used for the test set and the data provenance: Not applicable. This is not a study involving a "test set" of patient data for diagnostic performance. The "performance data" mentioned in F and G likely refers to engineering tests and measurements on the device itself to verify it meets the specified standards.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for diagnostic purposes is not established here. The "ground truth" in this context is adherence to established engineering and safety standards (e.g., IEC, EU Directives).
    3. Adjudication method: Not applicable.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done: No. This is not a diagnostic device or an AI system that interacts with human readers in a diagnostic context.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: No. This is a mechanical/electronic medical device, not an algorithm.
    6. The type of ground truth used: Not applicable in the context of diagnostic performance. The "ground truth" for the device's substantial equivalence claim is its conformity to established international safety and performance standards (e.g., EN93/42, IEC 601-1, EU 89/336).
    7. The sample size for the training set: Not applicable. This device does not use machine learning or require a training set.
    8. How the ground truth for the training set was established: Not applicable.

    Summary of the "Study" (Demonstration of Equivalence):

    The "study" or demonstration of equivalence for the Jostra HL-20 Twin Pump involves:

    • Comparison to a Predicate Device: The new device (Twin Pump) is compared to a legally marketed predicate device (Single Head Roller Pump) regarding indications for use, technological characteristics (design, materials), and safety/effectiveness concerns.
    • Adherence to Standards: Performance data (Attachment #5, not provided here) was submitted to demonstrate that the Twin Pump performs within its specifications and meets the acceptable limits of several key performance and safety standards:
      • EU Council Directive 93/42/EEC (General Medical Devices)
      • IEC 601-1, 2nd Edition: 1988 (Safety of Medical Electrical Equipment, Part 1: General Requirements)
      • EMC Directive 89/336/EEC 1993 (Electromagnetic Compatibility)
    • Conclusion: The provided data demonstrated that the HL-20 Twin Pump model is "similar in materials, design and function to that of the predicate device" and that its performance falls within acceptable limits relative to established safety and performance standards. This leads to the FDA's determination of substantial equivalence.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1