K Number
K212644Device Name
Aura Storm
Manufacturer
Invictus Lighting
Date Cleared
2022-05-26
(279 days)
Product Code
FRA
Regulation Number
880.6500Why did this record match?
Applicant Name (Manufacturer) :
Invictus Lighting
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Aura Storm air purifier is a device intended for medical purposes that is used to capture and destroy bacteria and viruses in the air through the multi-stage filtration system and exposure to ultraviolet radiation.
The Aura Storm air purifier has been demonstrated to destroy the following bacteria: Staphylococcus albicans, Staphylococcus aureus, and Escherichia Coli, and virus: A/PR8/34 HIN1 virus entrained on the filter of the subject device under the following exposure conditions:
Average Maximum log reduction / entrainment time (minutes) at Fan Speed 4. Room Temperature test: Log 4 (99.99%) / 60 minutes.
Average Maximum log reduction / entrainment time (minutes) at Fan Speed 1. Room Temperature test: Log 4 (99.99%) / 120 minutes.
Device Description
The Aura Storm is a multi-stage UV-C air purifier that delivers air purification for rooms up to 2700 square feet (*52ft x 52ft). The Storm inactivates and eliminates bacteria and viruses in the Aura Storm may be used for medical purposes which can include hospitals, medical facilities, medical clinics, nursing facilities, and dental facilities. The Aura Storm is a standalone device that is controlled via buttons on the top of the unit.
The Aura Storm is a dual-sided system with each side consisting of four-stage is a prefilter. The second stage is a HEPA H13+ filter. The third stage is a TiO2 cold photo-catalyst filter. Finally, the UV-C sterilization lamp emits a 253.7mm designed to activate the photo-catalyst filter and eliminate bacteria and viruses. The anion generator is used strictly to increase the size of the particulate molecules to increase the ability to capture and eliminate the bacteria and viruses. This designed system delivers a third-party tested 99.99% (4-log) sterilization in 60 minutes at the highest fan speed. The Aura storm was also tested at the lowest fan speed and delivered a 99.99% (4-log) sterilization in 120 minutes. The air flow diagram has been submitted in the 510(k) submission.
There is an internal fan which draws in air from both sides of the unit, filters it through the multi-stage system and outputs the filtered air through the top of the unit features a capacitive touch button control to operate the On/Off, fan speed (4 speeds), Auto mode, UV lamp, Anion generator, Child lock, timer, and filter reset. There is also a particulate matter sensor and color bar to indicate the level of clean air.