Search Results
Found 2 results
510(k) Data Aggregation
(270 days)
International Marketing Specialists Inc.
Tiger Reusable Sharps Containers and accessories are intended to be used in healthcare facilities including nursing stations, medication carts, laboratories, dental offices, emergency rooms, treatment rooms, emergency vehicles, veterinarian offices and other small quantity waste generators for the safe disposal, storage and transportation of hazardous sharps.
Tiger Reusable Sharps Containers are of injection-molded polypropylene plastic, designed for reusable by healthcare professionals. No part of the container is intended to come in contact with patients. The containers are designed to be puncture-resistant, leak-resistant on the sides and bottom, impact-resistant, closable, and stable.
The base is made from a high-strength material to support the capacity of the container. The recommended fill level is engraved onto the plastic and corresponds to the product identification label's level line.
The device in question is the Tiger Reusable Sharps Container.
Here's a breakdown of the acceptance criteria and study information:
1. Table of Acceptance Criteria and Reported Device Performance
Test | Standard | Acceptance Criteria | Reported Device Performance (Results) |
---|---|---|---|
Puncture Resistance of Materials Used in Containers for Discarded Medical Needles and Other Sharps | ASTM F 2132 -01 (Reapproved 2008)e1 | The force needed to penetrate test specimens shall be a minimum of 20 N or greater. | Pass |
DOT Drop Test | 49 CFR 178.603 | There is no damage to the outer packaging likely to adversely affect safety during transport, there is no leakage of the filling substance from the inner packaging. | Pass |
DOT Stacking Test | 49 CFR 178.606 | No test sample may show any deterioration or distortion, which could adversely affect safety or container strength during transport causing instability in stacks of packages. | Pass |
DOT Vibration standard | 49 CFR 178.608 | A packaging passes the vibration test if there is no rupture or leakage from any of the packages. No test sample should show any deterioration which could adversely affect transportation safety or any distortion liable to reduce packaging strength. | Pass |
Container stability | ISO 23907, ISO 23907-2 | The container shall not topple over when tested. | Pass |
Aperture and closure | ISO 23907, ISO 23907-2 | It shall be possible to place sharps into the sharps container without using a second hand to manipulate the aperture. The aperture of containers intended to be placed in public access areas should be designed to restrict hand entry and removal of contents from the container. The aperture should be designed to prevent the risk of overfilling. | Pass |
Resistance to penetration | ISO 23907, ISO 23907-2, ASTM F2132-01 | The force needed to penetrate test specimens shall be a minimum of 20 N or greater. | Pass |
Resistance to damage or leakage after dropping | ISO 23907, ISO 23907-2 | There shall be no evidence of leakage and no breach of the sharps containment area. Minimum five minutes after every topple: - There shall be no evidence that the performance or function of the container has been compromised. - The container's temporary closure shall remain intact. | Pass |
Resistance to spillage by toppling | ISO 23907, ISO 23907-2 | There shall be no evidence of leakage and no breach of the sharps containment area. Minimum five minutes after every topple: - There shall be no evidence that the performance or function of the container has been compromised. - The container's temporary closure shall remain intact. | Pass |
Accessory strength test | ISO 23907, ISO 23907-2 | The strength of Wall Enclosures and Metal Wall brackets shall be tested by filling their associated sharps container with water. The filled container shall be placed into one of each of the respective wall-mounted accessories. After 48 hours, the containers shall be removed, and the accessories shall be inspected for any loss of integrity. There shall be no evidence of sagging, breakage, liquid leakage, or changes in performance for the locking mechanisms. | Pass |
Fill line | ISO 23907, ISO 23907-2 | The fill line indicator shall be determined by the design of the container, considering the risk of sharps extending above the fill line. It shall be at a level no greater than 85 % of the total capacity of the container. | Pass |
Lifespan simulation testing | ISO 23907 First edition 2012-09-01, ASTM F2132-01 (reapproved 2008), ISO 23907-2 First edition 2019-11 | To verify the performance of the devices conforms to the applicable performance standards requirement after Lifespan Simulation Testing. | Pass |
Cytotoxicity Tests | ISO10993-5 | 8.6 Evaluation Criteria (Details not provided in the document, but implies a pass/fail determination based on standard cytotoxicity evaluation). | Based on the conditions of the test, the device was found to be non-cytotoxic. |
Skin Sensitization Test | ISO10993-10 | 9.4 Evaluation of results (Details not provided in the document, but implies a pass/fail determination based on standard skin sensitization evaluation). | Based on the conditions of the test, the device was found to be non-sensitizing. |
Skin Irritation Test | ISO10993-10 | 9.4 Evaluation of results (Details not provided in the document, but implies a pass/fail determination based on standard skin irritation evaluation). | Based on the conditions of the test, the device was found to be non-irritating. |
Chemical Residues Test | ISO/DIS 15883-5 | 5.3 Quantitative Study of Extracts (Details not provided in the document, but implies evaluation of chemical residues). | Pass |
Cleaning Validation | FDA Guidance, AAMI TIR 30, ISO/DIS 15883-5 | 1 Evaluation pass/fail criteria for the residual protein and hemoglobin (blood) by Spectrophotometric method (Method 1) | |
2 Evaluation pass/fail criteria for residual TOC by total organic carbon analyzer (Method 2) | Pass | ||
Disinfection Validation | FDA Guidance, AAMI TIR 30, ISO/DIS 15883-5 | The disinfection process should be deemed effective if all three of the following are met: | |
The concentration of the challenge suspension meets 1.1.2 c) requirements. | |||
The positive control 1.2.2 e) shows microbial growth of each challenge organism. | |||
After processing, no challenge organism was recovered at any site. | Pass |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
The document does not explicitly state the specific sample sizes used for each individual test or the data provenance (country of origin, retrospective/prospective). It generally refers to "test specimens" or "test samples" for each performance test. The performance testing was done to demonstrate compliance with recognized international and national standards.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
This information is not applicable as the tests performed are physical, chemical, and biological performance tests for a medical device (sharps container), not diagnostic image analysis or clinical studies requiring expert ground truth establishment. The "ground truth" for these tests is defined by the technical specifications and acceptance criteria of the referenced standards.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
This information is not applicable for the type of performance testing conducted. Adjudication methods are typically used in clinical trials or studies involving human assessment or interpretation, not for objective engineering or laboratory tests.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not applicable. The device is a physical sharps container and not an AI-powered diagnostic or assistive tool. Therefore, MRMC studies and AI assistance are irrelevant to its evaluation.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
This information is not applicable. The device does not involve an algorithm or AI.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The ground truth used for these performance tests is based on objective measurements and predefined thresholds established by the relevant national and international standards (e.g., ISO, ASTM, DOT CFR). For biological tests, the ground truth is based on the absence of cytotoxicity, sensitization, irritation, or successful disinfection/cleaning as per the specified methods and criteria. No expert consensus, pathology, or outcomes data were used to establish the ground truth for these device performance tests.
8. The sample size for the training set
This information is not applicable. The device is not an AI/machine learning model and therefore does not have a "training set."
9. How the ground truth for the training set was established
This information is not applicable for the same reason as above; there is no training set for this type of device.
Ask a specific question about this device
(114 days)
International Marketing Specialists Inc.
Tiger Sharps Containers are intended to provide a receptacle for used, contaminated medical sharps and act as an enclosure during transport to ultimate disposal.
The Containers are single-use, disposable, non-sterile containers intended to be used for health-care purposes for safe disposal of hazardous sharps such as hypodermic needles, syringes, lancets and blood needles. The target population is for qualified personnel in health care facilities in which medical sharps may be used. All device models are not for use in areas with unsupervised patient access.
All device models only be used with appropriate mounting accessories.
Tiger Sharps Containers constructed of injection molded polypropylene plastic. They are designed for a single-use (disposable) by healthcare professionals. No part of the container is intended to come in contact with patients. The containers are designed to be puncture resistant, leak resistant on the sides and bottom, impact resistant, closable and stable.
Lids and closures are uncolored translucent material allowing for a visual determination of fill level. The base is made from a high strength material to support the capacity of the container. The recommended fill level is engraved onto the plastic and corresponds to the level line on the product identification label.
The provided text describes the regulatory clearance of "Tiger Sharps Containers" and includes information about its performance testing. Here's a breakdown of the requested information:
1. A table of acceptance criteria and the reported device performance
Acceptance Criteria (Test) | Device Performance (Result) | Standard/Method |
---|---|---|
Puncture Resistance | Pass | ASTM F2132-01(2008)e1 |
Resistance to Damage/Leaking after Drop | Pass | ISO 23907-2012 |
Handle Strength | Pass | ISO 23907-2012 |
Container Stability | Pass | ISO 23907-2012 |
Stacking | Pass | 49 CFR 178.606 |
Vibration | Pass | 49 CFR 178.608 |
Drop test | Pass | 49 CFR 178.603 |
Biocompatibility | Demonstrated compliance | Related international standards |
Sterility | Not applicable (device is non-sterile) | N/A |
Product Specifications (e.g., dimensions, capacity, materials) | Demonstrated compliance | Related international standards |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
The document does not specify the sample size used for each performance test. It only states that "Performance testing was provided in support of the substantial equivalence determination." The provenance of the data (country of origin, retrospective/prospective) is also not mentioned.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
This information is not applicable as the device is a sharps container and the performance tests described are physical and mechanical tests, not clinical evaluations requiring expert interpretation of medical data. Therefore, there is no "ground truth" established by medical experts in this context.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
This information is not applicable for the same reasons as #3. The performance tests are objective measurements against established standards.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This information is not applicable. The device is a physical sharps container, not an AI-powered diagnostic or assistive tool. Therefore, an MRMC study comparing human readers with and without AI assistance is irrelevant.
6. If a standalone (i.e., algorithm only without human-in-the loop performance) was done
This information is not applicable as the device is a physical sharps container and does not involve an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for the performance tests appears to be defined by the consensus standards (e.g., ASTM F2132-01(2008)e1, ISO 23907-2012, 49 CFR) themselves. These standards specify the test methods and the acceptable criteria for "passing" a test, rather than relying on expert clinical consensus, pathology, or outcomes data in a medical sense.
8. The sample size for the training set
This information is not applicable. The device is a physical product that undergoes engineering performance testing, not a machine learning model that requires a training set.
9. How the ground truth for the training set was established
This information is not applicable as there is no training set for a machine learning model.
Ask a specific question about this device
Page 1 of 1