Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K232328
    Device Name
    Insignis™ Syringe Infusion System
    Manufacturer
    Innovative Health Sciences, LLC
    Date Cleared
    2024-03-01

    (211 days)

    Product Code
    FPA, FRN, PKP
    Regulation Number
    880.5440
    Why did this record match?
    Applicant Name (Manufacturer) :

    Innovative Health Sciences, LLC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
    Intended Use
    The Insignis™ Syringe Infusion System is specifically indicated for the subcutaneous infusion of the following human plasma-derived immunoglobulins when used according to the FDA approved biologic labeling: Cuvitru®, Immune Globulin Infusion (Human) 20% (manufactured by Takeda®); Hizentra®, Immune Globulin Subcutaneous (Human) 20% Liquid (manufactured by CSL Behring®); and Xembify®, Immune Globulin Subcutaneous (Human) 20% Liquid (manufactured by Grifols®) in the home, hospital, or ambulatory settings when administered according to the approved biologic or drug product labeling. The system is intended for use by clinicians and lay users. It is intended for use in adults and in adolescents ages 13-21. For adolescents, it is to be operated under the supervision of an adult after instruction is provided by a healthcare provider. The Insignis TM Syringe Infusion System with the Insignis Syringe Driver and Intravenous Controller, is specifically indicated for the intravenous infusion of fluids such as 0.9% Sodium Chloride (normal saline) and dextrose solutions, and the following antibiotics when used according to the FDA approved drug product labeling: vancomycin, ertapenem, meropenem, oxacillin, and tobramycin.
    Device Description
    The Insignis™ Syringe Infusion System is a portable mechanical infusion system that operates without electricity. The Insignis system is intended to deliver indicated fluids subcutaneously or intravenously, depending on the prescribed medicine type, route of administration, and dose. The Insignis system incorporates into its design, a means to maintain a constant nominal pressure (13.5 psi), a controlled flowrate within a restrictive tubing circuit and needles to ensure the desired flowrate of medication to the patient is achieved. The spring-loaded Insignis™ Syringe Driver a constant mechanical force to the plunger of a BD 50 mL syringe (ref. #309653), that dispenses fluid through a restricted flowrate tubing and / or needle set. The Insignis system can infuse the rated BD syringe volume of 0 - 50 mL. Through a selection of IHS products total flowrates of 8 mL/hr -250 mL/hr are available: application specific. - A. For subcutaneous immunoglobulin (SCIg) administrations, the Insignis system includes the OneSett™ Subcutaneous Administration Set, which consists of a selectable rate flow controller connected in-line to an administration set comprised of 1 - 4 needle set configurations (6mm, 9mm, 12 mm, and 14mm lengths). The OneSett is dedicated to deliver 20% concentrate SCIg drugs at specific flowrates which meet the drug manufacturer's package inserts recommended flowrate limitations. The selectable rate flow controller controls the flowrate delivered to each subcutaneous needle site. - B. For intravenous administrations, the Insignis system includes the Intravenous Controller, which consists of a selectable rate flow controller connected in-line to a tubing set. The Intravenous Controller is dedicated to delivering select antibiotics, and saline and dextrose solutions. Again, the selectable rate flow controller allows for selectable and controllable flowrates. For administering the prescribed medications, the Insignis system requires that the user fill the syringe with the prescribed medications, connect the infusion set to the syringe, eliminate air from the fluid line, secure the syringe into the Insignis™ Syringe Driver, connect to the infusion set to the patient, and load the syringe driver to begin the infusion. The system is comprised of the Insignis™ Syringe Driver, a syringe, and either intravenous or subcutaneous infusion sets. To elaborate: - a. The Insignis™ Syringe Driver provides the constant driving force to dispense fluids. - b. The syringe is the reservoir that holds the fluids (manufactured by Becton & Dickinson). - c. The infusion sets (intravenous or subcutaneous) form the fluid path and delivers the fluid to the patient at controlled flowrates.
    Ask a Question
    K Number
    K202279
    Device Name
    Insignis Subcutaneous Needle Sets
    Manufacturer
    Innovative Health Sciences, LLC
    Date Cleared
    2021-08-13

    (367 days)

    Product Code
    FPA
    Regulation Number
    880.5440
    Why did this record match?
    Applicant Name (Manufacturer) :

    Innovative Health Sciences, LLC

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
    Intended Use
    Insignis® Subcutaneous Needle Sets are intended for the delivery of medication to the subcutaneous tissue for use not to exceed 24 hours.
    Device Description
    The Insignis® Subcutaneous Needle Sets are sterile, single use, packaged subcutaneous needle sets. Each Subcutaneous Needle Set has a luer lock at one end and a 90-degree 26G (0.46mm) needle mounted to a closing wing-stabilizer (or butterfly wings) at the distal end of one or more needles. The needles are available in four different lengths: 6mm, 9mm, 12mm, and 14mm. Each set has a slide clamp to stop the flow immediately as well as a butterfly wing that can be closed around the needle upon completion of the infusion. The Insignis® Subcutaneous Needle Sets is available as a single set, as well as a two, three and four needles sets. The Insignis® Subcutaneous Needle Sets is designed to be used for less than 24 hours
    Ask a Question