510(k) Data Aggregation
K Number
K141848Device Name
IOGYN SYSTEM
Manufacturer
Date Cleared
2014-08-29
(51 days)
Product Code
Regulation Number
884.1710Why did this record match?
Applicant Name (Manufacturer) :
IOGYN, INC.
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
The IOGYN System is intended to distend the uterus by filling it with saline to facilitate viewing with a hysteroscope during diagnostic and operative hysteroscopy and provide fluid management through the closed-loop recirculation of filtered distension fluid. It is also intended for cutting and coagulation of uterine tissue such as intrauterine polyps and myomas using a bipolar resecting device.
Device Description
The IOGYN System is comprised of the following:
- IOGYN Controller with Integrated Fluid Management .
- Footswitch O
- Fluid Management Accessories o
- IOGYN Resecting Device ●
The IOGYN System provides an integrated control system with bipolar radiofrequency outputs (cut and coagulation) and fluid management through the use of two integrated peristaltic pumps. The Resecting Device is a bipolar radiofrequency device configured for the resection and aspiration of uterine pathology. Fluid infusion and aspiration of the uterine cavity are controlled by the IOGYN Controller's peristaltic pumps, in conjunction with the Fluid Management Accessories; these components form a closed-loop recirculating system. These integrated peristaltic pumps are operated by a software pressure control algorithm that measures and controls intra-uterine cavity pressure by varying saline infusion and aspiration rates in response to pressure changes during aspiration and tissue resection using the Resecting Device. The IOGYN Controller Graphical User Interface (GUI) has fluid control settings which allow the user to toggle infusion ON/OFF and to set the cavity pressure from 30-125mmHg.
The IOGYN System uses a closed-loop fluid management system with a single 3liter saline bag that continuously recirculates filtered distension fluid throughout the procedure. The fluid absorption is limited by the volume within the 3-liter saline bag minus the dead volume within the system, which limits the deliverable volume to less than 2.5L. The 2.5L of recirculated fluid is passed through two levels of filtration - the first level of filtration removes resected tissue and bulk particles; the second level of filtration removes cellular materials, hemoglobin, plasma proteins, cytokines, coagulation factors, bacteria and viruses. These two levels of filtration generate optically clear, sterile, filtered distension fluid which is returned to the 3 liter saline bag and recirculated.
Ask a Question
Ask a specific question about this device
K Number
DEN130040Device Name
IOGYN SYSTEM
Manufacturer
Date Cleared
2014-03-28
(210 days)
Product Code
Regulation Number
884.1710Why did this record match?
Applicant Name (Manufacturer) :
IOGYN, INC.
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
The IOGYN System is intended to distend the uterus by filling it with saline to facilitate viewing with a hysteroscope during diagnostic and operative hysteroscopy and provide fluid management through the closed loop recirculation of filtered distension fluid. It is also intended for cutting and coagulation of uterine tissue such as intrauterine polyps and myomas using a bipolar resecting device.
Device Description
The IOGYN System is comprised of the following: IOGYN Controller with Integrated Fluid Management, Footswitch, Fluid Management Accessories, IOGYN Resecting Device. The IOGYN System provides an integrated control system with bipolar radiofrequency outputs (cut and coagulation) and fluid management through the use of two integrated peristaltic pumps. The Resecting Device is a bipolar radiofrequency device configured for the resection and aspiration of uterine pathology. Fluid infusion and aspiration of the uterine cavity are controlled by the IOGYN Controller's peristaltic pumps, in conjunction with the Fluid Management Accessories; these components form a closed loop recirculating system. These integrated peristaltic pumps are operated by a software pressure control algorithm that measures and controls intrauterine cavity pressure by varying saline infusion and aspiration rates in response to pressure changes during aspiration and tissue resection using the Resecting Device. The IOGYN Controller Graphical User Interface (GUI) has fluid control settings which allow the user to toggle infusion ON/OFF and to set the cavity pressure from 30-125mmHg. The IOGYN System uses a closed loop fluid management system with a single 3-liter saline bag that continuously recirculates filtered distension fluid throughout the procedure. The fluid absorption is limited by the volume within the 3-liter saline bag minus the dead volume within the system, which limits the deliverable volume to less than 2.5L. The 2.5L of recirculated fluid is passed through a tissue catch and then a molecular filter to generate optically clear, sterile, filtered distension fluid which is returned to the 3 liter saline bag and recirculated.
Ask a Question
Ask a specific question about this device
K Number
K123330Device Name
IOGYN ENDOSCOPE
Manufacturer
Date Cleared
2013-03-19
(144 days)
Product Code
Regulation Number
884.1690Why did this record match?
Applicant Name (Manufacturer) :
IOGYN, INC.
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
The IOGYN Endoscope is intended to provide the physician with a means for endoscopic diagnostic and therapeutic surgical procedures. It is indicated for use in diagnostic examination and therapeutic surgical procedures of, but not limited to, urology and gynecology.
Device Description
The IOGYN Endoscope is a rigid multi-channel endoscope provided with a rod lens system to transmit light and images, a working channel, and two fluid channels. The body consists of an outer and an inner tube of surgical steel. The light fibers, rod-lens optical system, and two fluid channels are sandwiched between these tubes. The inner tube serves as a working channel to provide access for resection/coagulation instruments. The fluid channels are intended to move fluid. At the proximal end, the endoscope is provided with a light post that is compatible with standard endoscopic light sources, and with an eyepiece intended for visualization. The IOGYN Endoscope is dimensionally compatible with the IOGYN System.