Search Results
Found 3 results
510(k) Data Aggregation
K Number
K242872Device Name
iDart Lyme IgM ImmunoBlot Kit
Manufacturer
Date Cleared
2025-06-12
(262 days)
Product Code
Regulation Number
866.3830Why did this record match?
Applicant Name (Manufacturer) :
ID-FISH Technology, Inc
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
The iDart™ Lyme IgM ImmunoBlot Kit is an immunoblot assay intended for the in vitro qualitative detection of IgM antibodies to Borrelia burgdorferi in human serum. The iDart™ Lyme IgM ImmunoBlot Kit is intended to detect antibodies to Lyme Screen Antigen (LSA) and multiple other B. burgdorferi antigens following a modified two-tier test methodology. Positive results from the iDart™ Lyme IgM ImmunoBlot Kit are supportive evidence for the presence of antibodies and exposure to B. burgdorferi. Negative results do not preclude infection with B. burgdorferi. iDart™ Lyme IgM ImmunoBlot Kit is intended to aid in the diagnosis of Lyme disease and the test kit should only be used on samples from patients with clinical history, signs and symptoms consistent with Lyme disease. The iDart™ Lyme IgM Immunoblot Kit is not intended as a screen for asymptomatic patients.
Test results are to be used in conjunction with information obtained from the patient's clinical evaluation and other diagnostic procedures.
Device Description
The iDart™ Lyme IgM ImmunoBlot tests are line immunoblot assays. Antigenic proteins specific for Borrelia species that cause Lyme Disease are produced by recombinant DNA technology in Escherichia coli. The purified proteins are then applied as discrete lines on a nitrocellulose membrane along with two control proteins. The iDart™ Lyme IgM ImmunoBlot Kit contains IgM ImmunoBlot strips and the proteins are applied in the following order: C1 (IgG/IgM – conjugate control), C2 (Protein L – calibrator/serum control), P93, P41 (2 antigen bands), P39 (2 antigen bands), P23 (9 antigen bands), P31 (9 antigen bands), P34, C10 and LSA (a chimeric VlsE peptide termed the Lyme Screen Antigen).
Ask a Question
Ask a specific question about this device
K Number
K233367Device Name
iDart Lyme IgG ImmunoBlot Kit
Manufacturer
Date Cleared
2024-08-12
(315 days)
Product Code
Regulation Number
866.3830Why did this record match?
Applicant Name (Manufacturer) :
ID-FISH Technology, Inc
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
The iDart™ Lyme IgG ImmunoBlot Kit is an immunoblot assay intended for the in vitro qualitative detection of IgG antibodies to Borrelia burgdorferi in human serum. The iDart Lyme IgG ImmunoBlot Kit is intended to detect antibodies to LSA and multiple other B. burgdorferi antigens following a modified two-tier test methodology. Positive results from the iDart Lyme IgG ImmunoBlot Kit are supportive evidence for the presence of antibodies and exposure to B. burgdorferi. Negative results do not preclude infection with B. burgdorferi. iDart™ Lyme IgG ImmunoBlot Kit is intended to aid in the diagnosis of Lyme disease and the test kit should only be used on samples from patients with clinical history, signs and symptoms consistent with Lyme disease. The iDart Lyme IgG Immunoblot Kit is not intended as a screen for asymptomatic patients.
Test results are to be used in conjunction with information obtained from the patient's clinical evaluation and other diagnostic procedures.
For in vitro diagnostic use only
For professional use only
For prescription use only
Device Description
The iDart™ Lyme IgG ImmunoBlot tests are line immunoblot assays. Antigenic proteins specific for Borrelia species that cause Lyme Disease are produced by recombinant DNA technology in Escherichia coli. The purified proteins are then applied as discrete lines on a nitrocellulose membrane along with two control proteins.
The iDart™ Lyme IgG ImmunoBlot Kit contains IgG ImmunoBlot strips and the proteins are applied in the following order: C1 (lgG/lgM - conjugate control), C2 (Protein L - calibrator/serum control), P93, P41, P39, P23, P31, P66, P58, P45, P34, P30, P28, P18 and LSA (a chimeric VISE peptide termed the Lyme Screen Antigen).
Ask a Question
Ask a specific question about this device
K Number
DEN160025Device Name
ID-FISH Plasmodium Genus Test Kit, ID-FISH Plasmodium falciparum and P. vivax Combo Test Kit
Manufacturer
Date Cleared
2017-08-18
(417 days)
Product Code
Regulation Number
866.3367Why did this record match?
Applicant Name (Manufacturer) :
ID-FISH TECHNOLOGY, INC
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
ID-FISH Plasmodium Genus Test Kit (PlasG) and ID-FISH Plasmodium falciparum and P. vivax Combo Test Kit (PlasFV) are intended for in vitro diagnostic use in the clinical laboratory for detection of Plasmodium species in human venous whole blood (EDTA) samples from patients suspected of Plasmodium infection. The test kits are intended to aid in the diagnosis of malaria and to aid in the differential diagnosis of P. falciparum and P. vivax infection. The test kits should be used only on samples from patients with a clinical history, signs and symptoms consistent with malaria, and are not intended as a screen for asymptomatic patients.
The ID-FISH Plasmodium Genus Test Kit is a qualitative test for detection of malaria parasites in blood smears. Positive results should be supplemented with the Plasmodium species specific test kit, ID-FISH Plasmodium falciparum and P. vivax Combo Test Kit for identification and differentiation of Plasmodium falciparum and Plasmodium vivax. The results of these test kits should be used in conjunction with other diagnostic test results. Clinical performance has not been established for P. ovale, P. malariae, or P. knowlesi.
Device Description
The ID-FISH Plasmodium Genus Test Kit (PlasG) and ID-FISH Plasmodium falciparum and P. vivax Combo Test Kit (PlasFV) are fluorescence in situ hybridization (FISH) assays to detect Plasmodium spp. or P. falciparum or P. vivax parasites in thin film blood smears prepared from EDTA venous whole blood samples.
Ask a Question
Ask a specific question about this device
Page 1 of 1