Search Results
Found 2 results
510(k) Data Aggregation
(267 days)
Hunan Handlike Minimally Invasive Surgery Co., Ltd.
The Ultrasonic System is intended to be used to transect, dissect, and coagulate tissue. The instruments are indicated for use in open and endoscopic general surgical procedures where bleeding control and minimal thermal injury to tissue are desired. The instruments allow for the coagulation of vessels (veins and arteries) up to 3 mm in diameter.
Ultrasonic Surgical System is an ultrasonic dissection and coagulation system composed of four components: Ultrasonic Generator (HD-CSD01), Hand Piece (HD-CSH01), a disposable Ultrasonic Shears includes 4 shaft lengths (HD-CS-DJ45, HD-CS-DJ36, HD-CS-DJ22, HD-CS-DJ14) and foot switch (HRF-M52).
A disposable Ultrasonic Shears is a sterile component that has undergone EO sterilization, including an ultrasonic horn, jaws, main shaft, axis rotation, handle, operating handle, and control switch. The operating handle is used to open and close the jaws that grip the tissue. The Ultrasonic Surgical System can simultaneously cut and coagulate tissues by setting different output power levels. The high-power range allows for faster tissue cutting, while the low-power range allows for better tissue closure. The ultrasonic frequency current in the ultrasonic generator is transmitted to the hand piece, which converts electrical energy into mechanical energy for forward and backward vibration. Through the transmission and release of the blade, the end of the blade vibrates at a certain frequency. The heat generated by friction causes water vaporization in the tissue cells that come into contact with the blade, protein hydrogen bonds break, cells collapse and the tissue is cut open after solidification; When cutting blood vessels, the knife head contacts with Histone and generates heat through mechanical vibration, leading to the destruction of the collagen structure in the tissue, causing the protein to coagulate, and then sealing the blood vessels to stop bleeding. The disposable ultrasonic shears are EO sterilized, but the other parts are non-sterilized packaging, and the hand piece must be cleaned and sterilized by the user before being used for surgery.
This device is intended to be used in the hospital environment, such as surgical operation room and imaging intervention room.
The provided text is a 510(k) Summary for a medical device called "Ultrasonic Surgical System." It describes the device, its intended use, and compares it to predicate devices. However, this document does not contain details about specific acceptance criteria for performance tests or the detailed study results that would typically prove a device meets those criteria in a format directly applicable to the table requested.
The document mentions "Performance test and functional tests were conducted on the proposed device in accordance with product design specifications. Data generated from the test met the predetermined acceptance criteria." but does not list those criteria or the specific performance data against them.
It also includes results from animal testing (burst pressure, acute, and chronic animal studies) but these are comparative to a predicate device rather than against predefined, quantitative acceptance criteria for the subject device itself.
Therefore, many of the requested fields cannot be filled directly from the provided text.
Based on the information provided, here's what can be extracted:
1. Table of Acceptance Criteria and Reported Device Performance
This information is not explicitly detailed in the provided document. The document states that "Data generated from the test met the predetermined acceptance criteria" but does not list the specific criteria or the exact performance data against them for each test.
Acceptance Criteria (Not explicitly stated in the document) | Reported Device Performance (as per non-clinical tests summary) |
---|---|
Sterilization: (Presumed: Sterility Assurance Level [SAL] compliance, EO residuals within limits) | Ultrasonic Shears: Sterilized via ethylene oxide as per ISO 11135 (overkill method). Ethylene oxide residuals met ISO 10993-7 requirements. |
Biocompatibility: (Presumed: No toxicity, irritation, sensitization, acute systemic toxicity, pyrogenicity) | Ultrasonic Shears: Not toxic, irritating, or sensitizing; free of acute systemic toxicity and pyrogen, as per FDA guidance Use of International Standard ISO 10993-1. |
Electrical Safety & EMC: (Presumed: Compliance with relevant standards) | System: Complied with IEC 60601-1-2:2020 for electromagnetic compatibility, IEC 60601-1:2020 and IEC 60601-2-18 for electrical safety. |
Overall Performance & Functional Tests: (Presumed: Met product design specifications) | Device: "Data generated from the test met the predetermined acceptance criteria." (Specifics not provided). |
Burst Pressure Testing (Animal Study): (Presumed: Equivalent or superior burst pressure compared to predicate) | Animal Study: "Either there was no statistical difference of burst pressure between test group and control group or the average burst pressure of test group was higher than that of the control group where there was statistical difference." No error/note observed. |
Acute Animal Study: (Presumed: Equivalent times/thermal damage, no sealing failures) | Animal Study: Times and thermal damage range did not have statistical difference compared to predicate. No failure of sealing identified after first activation and after challenge. |
Chronic Animal Study: (Presumed: No abnormal phenomena, no adverse conditions at sealed site, acceptable tissue response) | Animal Study: Postoperative observation showed no abnormal phenomena. Gross examination showed no hematoma, bleeding, or other adverse conditions at the sealed site. Few tissues showed moderate inflammatory cell infiltration, most showed few inflammatory cells. Wide fibrosis and mild tissue necrosis at the test site. |
Packaging & Shelf Life: (Presumed: Package integrity throughout shelf life) | Packaging: Primary packaging validated as per ISO 11607. Shelf Life: Validated as per ASTM F1980. |
Software V&V: (Presumed: Compliance with FDA Guidance) | Software: In compliance with FDA Guidance-Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. |
Cleaning, Disinfection, Sterilization effectiveness (Hand Piece): (Presumed: Effectiveness validated as per ISO) | Hand Piece: Methods validated as per ISO 17664-1. |
Cybersecurity: (Presumed: No external interfaces requiring evaluation) | Device: No external interfaces, no need for cybersecurity evaluation per FDA guidance. |
2. Sample Size Used for the Test Set and Data Provenance
- The document specifies a "porcine model" for animal testing.
- It does not specify the sample size for the burst pressure, acute, or chronic animal studies (e.g., number of animals or number of vessels tested per animal/group).
- The data is prospective, generated from the animal study using the subject device and a predicate device.
- The country of origin of the data is not specified, but the applicant (Hunan Handlike Minimally Invasive Surgery Co., Ltd.) is from China.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- This information is not provided in the document. The animal studies would typically involve veterinarians and potentially pathologists, but their number and specific qualifications are not mentioned.
4. Adjudication Method for the Test Set
- This information is not provided in the document.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No MRMC study was done, as this is a surgical device, not an imaging/diagnostic AI device that would typically involve human readers interpreting cases. The comparative effectiveness assessment was against a predicate device in animal models.
6. Standalone (i.e., algorithm only without human-in-the-loop performance) Study
- Not applicable, as this is a physical surgical device, not a standalone algorithm. The device itself is evaluated.
7. The Type of Ground Truth Used
- For the animal studies, the "ground truth" would be established through direct observation during the procedures, histological examination of tissue samples, and measurements (e.g., burst pressure). These are effectively pathology and direct physiological measurements/observations.
8. The Sample Size for the Training Set
- Not applicable, as this is a physical surgical device, not an AI/ML algorithm that requires a training set in the conventional sense.
9. How the Ground Truth for the Training Set Was Established
- Not applicable, as this is a physical surgical device.
Ask a specific question about this device
(239 days)
Hunan Handlike Minimally Invasive Surgery Co., Ltd
Endoscopic Instruments are designed to cut, dissect, manipulate and/or cauterize various tissues during endoscopic/ laparoscopic, general surgical procedures.
Endoscopic Surgical Instrument is a kind of device used to cut, dissect, manipulate and/or cauterize various tissues during endoscopic/laparoscopic, general surgical procedures. The device includes Disposable Endoscopic Surgical Instrument and Reusable Endoscopic Surgical Instrument. Both of the disposable and reusable endoscopic instruments are provided sterile when they're released. After used for the first time, Reusable Endoscopic Instrument needs to be sterilized by the end users. The instruments are 5mm in diameter as well as 340mm in length.
Here's a breakdown of the acceptance criteria and study information based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state acceptance criteria in a quantitative format for specific performance metrics that would typically be found in an Acceptance Criteria table. Instead, it refers to compliance with international standards and internal design control procedures.
However, based on the non-clinical testing performed, we can infer some performance aspects:
Feature/Test | Acceptance Criteria (Inferred) | Reported Device Performance |
---|---|---|
Electrical Safety | Compliance with IEC 60601-1 | Testing in accordance with IEC 60601-1 (device passed) |
Electromagnetic Compatibility (EMC) | Compliance with IEC 60601-1-2 | Testing in accordance with IEC 60601-1-2 (device passed) |
Special Requirements for Electrosurgical Units | Compliance with IEC 60601-2-2 | Testing in accordance with IEC 60601-2-2 (device passed) |
Endoscopic Equipment (Electrosurgical) | Compliance with IEC 60601-2-18 | Testing in accordance with IEC 60601-2-18 (device passed) |
Appearance | Meets specification | All testing results show that the Endoscopic Surgical Instruments meet the specification and performance characteristics as identified in internal design control procedures. |
Basic Dimension | Meets specification | All testing results show that the Endoscopic Surgical Instruments meet the specification and performance characteristics as identified in internal design control procedures. |
Corrosion Resistance | Meets specification | All testing results show that the Endoscopic Surgical Instruments meet the specification and performance characteristics as identified in internal design control procedures. |
Transverse Holding Force | Meets specification | All testing results show that the Endoscopic Surgical Instruments meet the specification and performance characteristics as identified in internal design control procedures. |
Longitudinal Holding Force | Meets specification | All testing results show that the Endoscopic Surgical Instruments meet the specification and performance characteristics as identified in internal design control procedures. |
Ratchet Holding Force | Meets specification | All testing results show that the Endoscopic Surgical Instruments meet the specification and performance characteristics as identified in internal design control procedures. |
Cutting Efficacy | Meets specification | All testing results show that the Endoscopic Surgical Instruments meet the specification and performance characteristics as identified in internal design control procedures. |
Thermal Effects on Tissue | Acceptable thermal damage zone | The thermal effect has been performed on fresh pig's liver, kidney, and muscle tissue. The thermal damage zone is measured under magnification using histology after performing cauterizing. All testing results show the device meets internal specifications. |
Biocompatibility | Compliance with ISO 10993 series | The device meets the applicable requirements of the FDA Blue Book Memorandum G95-11 entitled Use of International Standards ISO 10993-1:2009 Biological Evaluation of Medical Devices-Part 1: Evaluation and Testing within a risk management process, including ISO 10993-5, -7, -10, -11, -12. |
Sterility (Reusable Instruments) | Sterility Assurance Level (SAL) of 10⁻⁶ | Both EO and Steam sterilization methods have been validated, and test reports show that the sterilization effect of the proposed device can achieve a SAL of 10⁻⁶. |
2. Sample Size Used for the Test Set and the Data Provenance
The document describes non-clinical laboratory testing (e.g., electrical safety, performance, biocompatibility). It does not involve a "test set" in the context of clinical data or AI model evaluation.
- Sample Size for performance tests: Not explicitly stated for each test, but implied to be sufficient for demonstrating compliance with specified standards and characteristics. For the thermal effects on tissue, it involved "fresh pig's liver, kidney, and muscle tissue," but the number of samples is not quantified.
- Data Provenance: The tests were conducted in a laboratory setting as part of the device's design control and regulatory submission. The document doesn't specify the country of origin for the non-clinical test data itself, beyond the manufacturer being based in China. The study is a prospective evaluation of the device against predefined performance standards and characteristics.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
Not applicable. This device is an electrosurgical instrument for cutting, dissecting, manipulating, and cauterizing tissues. There is no AI component or image analysis that would require expert adjudication to establish "ground truth" for a test set. The evaluation is based on objective measurements and compliance with technical standards.
4. Adjudication Method for the Test Set
Not applicable, as there is no "test set" in the context of expert review or clinical data.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If so, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance
Not applicable. This is a medical device (surgical instrument), not an AI-powered diagnostic or assistive tool for human readers.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done
Not applicable. This device is a manual surgical instrument that requires human operation. It does not have an autonomous algorithm component.
7. The Type of Ground Truth Used
The "ground truth" in this context is the objective, measurable performance of the physical device against established engineering standards and specifications. Examples include:
- Compliance with International Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2, IEC 60601-2-18, ISO 10993 series.
- Physical and Functional Specifications: Appearance, basic dimension, corrosion resistance, various holding forces (transverse, longitudinal, ratchet), cutting efficacy.
- Biological Outcome: Thermal damage zone measured via histology on tissue samples, sterility assurance level (SAL).
8. The Sample Size for the Training Set
Not applicable. This is a physical medical device, not an AI model requiring a training set.
9. How the Ground Truth for the Training Set Was Established
Not applicable. As there is no AI model, there is no training set or associated ground truth establishment process.
Ask a specific question about this device
Page 1 of 1